Published in: Abdominal Radiology 6/2015

01-08-2015 | Review Article

Whole-body FDG PET-MR oncologic imaging: pitfalls in clinical interpretation related to inaccurate MR-based attenuation correction

Authors: Ulrike Attenberger, Ciprian Catana, Hersh Chandarana, Onofrio A Catalano, Kent Friedman, Stefan A Schonberg, James Thrall, Marco Salvatore, Bruce R. Rosen, Alexander R. Guimaraes

Published in: Abdominal Radiology | Issue 6/2015

Login to get access

Abstract

Simultaneous data collection for positron emission tomography and magnetic resonance imaging (PET/MR) is now a reality. While the full benefits of concurrently acquiring PET and MR data and the potential added clinical value are still being evaluated, initial studies have identified several important potential pitfalls in the interpretation of fluorodeoxyglucose (FDG) PET/MRI in oncologic whole-body imaging, the majority of which being related to the errors in the attenuation maps created from the MR data. The purpose of this article was to present such pitfalls and artifacts using case examples, describe their etiology, and discuss strategies to overcome them. Using a case-based approach, we will illustrate artifacts related to (1) Inaccurate bone tissue segmentation; (2) Inaccurate air cavities segmentation; (3) Motion-induced misregistration; (4) RF coils in the PET field of view; (5) B0 field inhomogeneity; (6) B1 field inhomogeneity; (7) Metallic implants; (8) MR contrast agents.
Literature
2.
go back to reference Antoch G, Vogt FM, Freudenberg LS, et al. (2003) Whole-body dual-modality PET/CT and whole-body MRI for tumor staging in oncology. JAMA 290:3199–3206PubMedCrossRef Antoch G, Vogt FM, Freudenberg LS, et al. (2003) Whole-body dual-modality PET/CT and whole-body MRI for tumor staging in oncology. JAMA 290:3199–3206PubMedCrossRef
3.
go back to reference Fonti R, Salvatore B, Quarantelli M, et al. (2008) 18F-FDG PET/CT, 99mTc-MIBI, and MRI in evaluation of patients with multiple myeloma. J Nucl Med 49:195–200PubMedCrossRef Fonti R, Salvatore B, Quarantelli M, et al. (2008) 18F-FDG PET/CT, 99mTc-MIBI, and MRI in evaluation of patients with multiple myeloma. J Nucl Med 49:195–200PubMedCrossRef
4.
go back to reference Lauenstein TC, Goehde SC, Herborn CU, et al. (2004) Whole-body MR imaging: evaluation of patients for metastases. Radiology 233:139–148PubMedCrossRef Lauenstein TC, Goehde SC, Herborn CU, et al. (2004) Whole-body MR imaging: evaluation of patients for metastases. Radiology 233:139–148PubMedCrossRef
5.
go back to reference Martinez-Moller A, Souvatzoglou M, Delso G, et al. (2009) Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 50:520–526PubMedCrossRef Martinez-Moller A, Souvatzoglou M, Delso G, et al. (2009) Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med 50:520–526PubMedCrossRef
6.
go back to reference Schmidt GP, Haug AR, Schoenberg SO, Reiser MF (2006) Whole-body MRI and PET-CT in the management of cancer patients. Eur Radiol 16:1216–1225PubMedCrossRef Schmidt GP, Haug AR, Schoenberg SO, Reiser MF (2006) Whole-body MRI and PET-CT in the management of cancer patients. Eur Radiol 16:1216–1225PubMedCrossRef
7.
go back to reference Seemann MD (2005) Whole-body PET/MRI: the future in oncological imaging. Technol Cancer Res Treat 4:577–582PubMedCrossRef Seemann MD (2005) Whole-body PET/MRI: the future in oncological imaging. Technol Cancer Res Treat 4:577–582PubMedCrossRef
8.
go back to reference Catana C, Wu Y, Judenhofer MS, et al. (2006) Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 47:1968–1976PubMed Catana C, Wu Y, Judenhofer MS, et al. (2006) Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 47:1968–1976PubMed
9.
go back to reference Judenhofer MS, Wehrl HF, Newport DF, et al. (2008) Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med 14:459–465PubMedCrossRef Judenhofer MS, Wehrl HF, Newport DF, et al. (2008) Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med 14:459–465PubMedCrossRef
10.
go back to reference Schlemmer HP, Pichler BJ, Schmand M, et al. (2008) Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology 248:1028–1035PubMedCrossRef Schlemmer HP, Pichler BJ, Schmand M, et al. (2008) Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology 248:1028–1035PubMedCrossRef
11.
go back to reference Zaidi H (2007) Is MR-guided attenuation correction a viable option for dual-modality PET/MR imaging? Radiology 244:639–642PubMedCrossRef Zaidi H (2007) Is MR-guided attenuation correction a viable option for dual-modality PET/MR imaging? Radiology 244:639–642PubMedCrossRef
12.
go back to reference Catana C, van der Kouwe A, Benner T, et al. (2010) Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype. J Nucl Med 51:1431–1438PubMedCentralPubMedCrossRef Catana C, van der Kouwe A, Benner T, et al. (2010) Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype. J Nucl Med 51:1431–1438PubMedCentralPubMedCrossRef
13.
go back to reference Eiber M, Martinez-Moller A, Souvatzoglou M, et al. (2011) Value of a Dixon-based MR/PET attenuation correction sequence for the localization and evaluation of PET-positive lesions. Eur J Nucl Med Mol Imaging 38:1691–1701PubMedCrossRef Eiber M, Martinez-Moller A, Souvatzoglou M, et al. (2011) Value of a Dixon-based MR/PET attenuation correction sequence for the localization and evaluation of PET-positive lesions. Eur J Nucl Med Mol Imaging 38:1691–1701PubMedCrossRef
14.
go back to reference Paulus DH, Braun H, Aklan B, Quick HH (2012) Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils. Med Phys 39:4306–4315PubMedCrossRef Paulus DH, Braun H, Aklan B, Quick HH (2012) Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils. Med Phys 39:4306–4315PubMedCrossRef
15.
go back to reference Schulz V, Torres-Espallardo I, Renisch S, et al. (2011) Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging 38:138–152PubMedCrossRef Schulz V, Torres-Espallardo I, Renisch S, et al. (2011) Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging 38:138–152PubMedCrossRef
16.
go back to reference Steinberg J, Jia G, Sammet S, et al. (2010) Three-region MRI-based whole-body attenuation correction for automated PET reconstruction. Nucl Med Biol 37:227–235PubMedCrossRef Steinberg J, Jia G, Sammet S, et al. (2010) Three-region MRI-based whole-body attenuation correction for automated PET reconstruction. Nucl Med Biol 37:227–235PubMedCrossRef
17.
go back to reference Berker Y, Franke J, Salomon A, et al. (2012) MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence. J Nucl Med 53:796–804PubMedCrossRef Berker Y, Franke J, Salomon A, et al. (2012) MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence. J Nucl Med 53:796–804PubMedCrossRef
18.
go back to reference Keereman V, Holen RV, Mollet P, Vandenberghe S (2011) The effect of errors in segmented attenuation maps on PET quantification. Med Phys 38:6010–6019PubMedCrossRef Keereman V, Holen RV, Mollet P, Vandenberghe S (2011) The effect of errors in segmented attenuation maps on PET quantification. Med Phys 38:6010–6019PubMedCrossRef
19.
go back to reference Akbarzadeh A, Ay MR, Ahmadian A, Riahi Alam N, Zaidi H (2012) MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation. Ann Nucl Med Akbarzadeh A, Ay MR, Ahmadian A, Riahi Alam N, Zaidi H (2012) MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation. Ann Nucl Med
20.
go back to reference Weiger M, Brunner DO, Dietrich BE, Muller CF, Pruessmann KP (2013) ZTE imaging in humans. Magn Reson Med 70:328–332PubMedCrossRef Weiger M, Brunner DO, Dietrich BE, Muller CF, Pruessmann KP (2013) ZTE imaging in humans. Magn Reson Med 70:328–332PubMedCrossRef
21.
go back to reference Catana C, Benner T, van der Kouwe A, et al. (2011) MRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner. J Nucl Med 52:154–161PubMedCentralPubMedCrossRef Catana C, Benner T, van der Kouwe A, et al. (2011) MRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner. J Nucl Med 52:154–161PubMedCentralPubMedCrossRef
22.
go back to reference Beyer T, Antoch G, Blodgett T, et al. (2003) Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur J Nucl Med Mol Imaging 30:588–596PubMedCrossRef Beyer T, Antoch G, Blodgett T, et al. (2003) Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur J Nucl Med Mol Imaging 30:588–596PubMedCrossRef
23.
go back to reference Martinez-Moller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG (2007) Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 48:188–193PubMed Martinez-Moller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG (2007) Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med 48:188–193PubMed
24.
go back to reference Marshall HR, Prato FS, Deans L, et al. (2012) Variable lung density consideration in attenuation correction of whole-body PET/MRI. J Nucl Med 53:977–984PubMedCrossRef Marshall HR, Prato FS, Deans L, et al. (2012) Variable lung density consideration in attenuation correction of whole-body PET/MRI. J Nucl Med 53:977–984PubMedCrossRef
25.
go back to reference Togao O, Tsuji R, Ohno Y, Dimitrov I, Takahashi M (2010) Ultrashort echo time (UTE) MRI of the lung: assessment of tissue density in the lung parenchyma. Magn Reson Med 64:1491–1498PubMedCrossRef Togao O, Tsuji R, Ohno Y, Dimitrov I, Takahashi M (2010) Ultrashort echo time (UTE) MRI of the lung: assessment of tissue density in the lung parenchyma. Magn Reson Med 64:1491–1498PubMedCrossRef
26.
29.
go back to reference Nehmeh SA, Erdi YE, Ling CC, et al. (2002) Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med 43:876–881PubMed Nehmeh SA, Erdi YE, Ling CC, et al. (2002) Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med 43:876–881PubMed
30.
go back to reference Lamare F, Ledesma Carbayo MJ, Cresson T, et al. (2007) List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations. Phys Med Biol 52:5187–5204PubMedCrossRef Lamare F, Ledesma Carbayo MJ, Cresson T, et al. (2007) List-mode-based reconstruction for respiratory motion correction in PET using non-rigid body transformations. Phys Med Biol 52:5187–5204PubMedCrossRef
31.
go back to reference Guerin B, Cho S, Chun SY, et al. (2011) Nonrigid PET motion compensation in the lower abdomen using simultaneous tagged-MRI and PET imaging. Med Phys 38:3025–3038PubMedCentralPubMedCrossRef Guerin B, Cho S, Chun SY, et al. (2011) Nonrigid PET motion compensation in the lower abdomen using simultaneous tagged-MRI and PET imaging. Med Phys 38:3025–3038PubMedCentralPubMedCrossRef
32.
33.
go back to reference Qiao F, Pan T, Clark JW Jr, Mawlawi OR (2006) A motion-incorporated reconstruction method for gated PET studies. Phys Med Biol 51:3769–3783PubMedCrossRef Qiao F, Pan T, Clark JW Jr, Mawlawi OR (2006) A motion-incorporated reconstruction method for gated PET studies. Phys Med Biol 51:3769–3783PubMedCrossRef
34.
go back to reference Lamare F, Cresson T, Savean J, et al. (2007) Respiratory motion correction for PET oncology applications using affine transformation of list mode data. Phys Med Biol 52:121–140PubMedCrossRef Lamare F, Cresson T, Savean J, et al. (2007) Respiratory motion correction for PET oncology applications using affine transformation of list mode data. Phys Med Biol 52:121–140PubMedCrossRef
35.
go back to reference Rakheja R, DeMello L, Chandarana H, et al. (2013) Comparison of the accuracy of PET/CT and PET/MRI spatial registration of multiple metastatic lesions. Am J Roentgenol 201:1120–1123CrossRef Rakheja R, DeMello L, Chandarana H, et al. (2013) Comparison of the accuracy of PET/CT and PET/MRI spatial registration of multiple metastatic lesions. Am J Roentgenol 201:1120–1123CrossRef
36.
go back to reference Antoch G, Bockisch A (2009) Combined PET/MRI: a new dimension in whole-body oncology imaging? Eur J Nucl Med Mol Imaging 36(Suppl 1):S113–S120PubMedCrossRef Antoch G, Bockisch A (2009) Combined PET/MRI: a new dimension in whole-body oncology imaging? Eur J Nucl Med Mol Imaging 36(Suppl 1):S113–S120PubMedCrossRef
37.
go back to reference Tsoumpas C, Mackewn JE, Halsted P, et al. (2010) Simultaneous PET-MR acquisition and MR-derived motion fields for correction of non-rigid motion in PET. Ann Nucl Med 24:745–750PubMedCrossRef Tsoumpas C, Mackewn JE, Halsted P, et al. (2010) Simultaneous PET-MR acquisition and MR-derived motion fields for correction of non-rigid motion in PET. Ann Nucl Med 24:745–750PubMedCrossRef
38.
go back to reference Ledesma-Carbayo MJ, Derbyshire JA, Sampath S, et al. (2008) Unsupervised estimation of myocardial displacement from tagged MR sequences using nonrigid registration. Magn Reson Med 59:181–189PubMedCrossRef Ledesma-Carbayo MJ, Derbyshire JA, Sampath S, et al. (2008) Unsupervised estimation of myocardial displacement from tagged MR sequences using nonrigid registration. Magn Reson Med 59:181–189PubMedCrossRef
39.
go back to reference Osman NF, Prince JL (2004) Regenerating MR tagged images using harmonic phase (HARP) methods. IEEE Trans BioMed Eng 51:1428–1433PubMedCrossRef Osman NF, Prince JL (2004) Regenerating MR tagged images using harmonic phase (HARP) methods. IEEE Trans BioMed Eng 51:1428–1433PubMedCrossRef
40.
go back to reference Prince JL, McVeigh ER (1992) Motion estimation from tagged MR image sequences. IEEE Trans Med Imaging 11:238–249PubMedCrossRef Prince JL, McVeigh ER (1992) Motion estimation from tagged MR image sequences. IEEE Trans Med Imaging 11:238–249PubMedCrossRef
41.
go back to reference von Siebenthal M, Szekely G, Gamper U, et al. (2007) 4D MR imaging of respiratory organ motion and its variability. Phys Med Biol 52:1547–1564CrossRef von Siebenthal M, Szekely G, Gamper U, et al. (2007) 4D MR imaging of respiratory organ motion and its variability. Phys Med Biol 52:1547–1564CrossRef
42.
go back to reference Wang Y, Rossman PJ, Grimm RC, Riederer SJ, Ehman RL (1996) Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. Radiology 198:55–60PubMedCrossRef Wang Y, Rossman PJ, Grimm RC, Riederer SJ, Ehman RL (1996) Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. Radiology 198:55–60PubMedCrossRef
43.
go back to reference Bryant DJ, Payne JA, Firmin DN, Longmore DB (1984) Measurement of flow with NMR imaging using a gradient pulse and phase difference technique. J Comput Assist Tomogr 8:588–593PubMedCrossRef Bryant DJ, Payne JA, Firmin DN, Longmore DB (1984) Measurement of flow with NMR imaging using a gradient pulse and phase difference technique. J Comput Assist Tomogr 8:588–593PubMedCrossRef
44.
go back to reference Nayler GL, Firmin DN, Longmore DB (1986) Blood flow imaging by cine magnetic resonance. J Comput Assist Tomogr 10:715–722PubMedCrossRef Nayler GL, Firmin DN, Longmore DB (1986) Blood flow imaging by cine magnetic resonance. J Comput Assist Tomogr 10:715–722PubMedCrossRef
45.
go back to reference Axel L, Dougherty L (1989) Heart wall motion: improved method of spatial modulation of magnetization for MR imaging. Radiology 172:349–350PubMedCrossRef Axel L, Dougherty L (1989) Heart wall motion: improved method of spatial modulation of magnetization for MR imaging. Radiology 172:349–350PubMedCrossRef
46.
go back to reference Axel L, Dougherty L (1989) MR imaging of motion with spatial modulation of magnetization. Radiology 171:841–845PubMedCrossRef Axel L, Dougherty L (1989) MR imaging of motion with spatial modulation of magnetization. Radiology 171:841–845PubMedCrossRef
47.
go back to reference Crum WR, Berry E, Ridgway JP, et al. (1998) Frequency-domain simulation of MR tagging. J Magn Reson Imaging 8:1040–1050PubMedCrossRef Crum WR, Berry E, Ridgway JP, et al. (1998) Frequency-domain simulation of MR tagging. J Magn Reson Imaging 8:1040–1050PubMedCrossRef
48.
go back to reference Fischer SE, McKinnon GC, Maier SE, Boesiger P (1993) Improved myocardial tagging contrast. Magn Reson Med 30:191–200PubMedCrossRef Fischer SE, McKinnon GC, Maier SE, Boesiger P (1993) Improved myocardial tagging contrast. Magn Reson Med 30:191–200PubMedCrossRef
49.
go back to reference Osman NF, McVeigh ER, Prince JL (2000) Imaging heart motion using harmonic phase MRI. IEEE Trans Med Imaging 19:186–202PubMedCrossRef Osman NF, McVeigh ER, Prince JL (2000) Imaging heart motion using harmonic phase MRI. IEEE Trans Med Imaging 19:186–202PubMedCrossRef
50.
go back to reference Grimm R, Furst S, Souvatzoglou M, et al. (2014) Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI. Med Image Anal 19:110–120PubMedCrossRef Grimm R, Furst S, Souvatzoglou M, et al. (2014) Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI. Med Image Anal 19:110–120PubMedCrossRef
51.
go back to reference Zhang B, Pal D, Hu Z, Ojha N, Guo T, Muswick G, Chi-hua T, Kaste J (2009) Attenuation correction for MR table and coils for a sequential PET/MR system. In: Proceedings of the 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC), pp 3303–3306 Zhang B, Pal D, Hu Z, Ojha N, Guo T, Muswick G, Chi-hua T, Kaste J (2009) Attenuation correction for MR table and coils for a sequential PET/MR system. In: Proceedings of the 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC), pp 3303–3306
52.
go back to reference Tellmann L, Quick HH, Bockisch A, Herzog H, Beyer T (2011) The effect of MR surface coils on PET quantification in whole-body PET/MR: results from a pseudo-PET/MR phantom study. Med Phys 38:2795–2805PubMedCrossRef Tellmann L, Quick HH, Bockisch A, Herzog H, Beyer T (2011) The effect of MR surface coils on PET quantification in whole-body PET/MR: results from a pseudo-PET/MR phantom study. Med Phys 38:2795–2805PubMedCrossRef
53.
go back to reference Blumhagen JO LR, Fenchel M, Kampmeier J, Scheffler K (2011) MR-based field-of-fiew extension: compensation of field imperfection. 19th Annual Meeting of ISMRM, p 2693 Blumhagen JO LR, Fenchel M, Kampmeier J, Scheffler K (2011) MR-based field-of-fiew extension: compensation of field imperfection. 19th Annual Meeting of ISMRM, p 2693
54.
go back to reference Bakker CJ, Moerland MA, Bhagwandien R, Beersma R (1992) Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imaging. Magn Reson Imaging 10:597–608PubMedCrossRef Bakker CJ, Moerland MA, Bhagwandien R, Beersma R (1992) Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imaging. Magn Reson Imaging 10:597–608PubMedCrossRef
55.
go back to reference Blumhagen JOLR, Fenchel M, Scheffler K, Quick HH (2012) MR-based FOV extension in whole-body MR/PET using continuous table move. Proc Intl Soc Mag Reson Med (ISMRM) 20:p2720 Blumhagen JOLR, Fenchel M, Scheffler K, Quick HH (2012) MR-based FOV extension in whole-body MR/PET using continuous table move. Proc Intl Soc Mag Reson Med (ISMRM) 20:p2720
56.
go back to reference Delso G, Martinez-Moller A, Bundschuh RA, Nekolla SG, Ziegler SI (2010) The effect of limited MR field of view in MR/PET attenuation correction. Med Phys 37:2804–2812PubMedCrossRef Delso G, Martinez-Moller A, Bundschuh RA, Nekolla SG, Ziegler SI (2010) The effect of limited MR field of view in MR/PET attenuation correction. Med Phys 37:2804–2812PubMedCrossRef
57.
go back to reference Delso G, Martinez-Moller A, Bundschuh RA, et al. (2010) Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner. Phys Med Biol 55:4361–4374PubMedCrossRef Delso G, Martinez-Moller A, Bundschuh RA, et al. (2010) Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner. Phys Med Biol 55:4361–4374PubMedCrossRef
58.
go back to reference Beyer T, Bockisch A, Kuhl H, Martinez MJ (2006) Whole-body 18F-FDG PET/CT in the presence of truncation artifacts. J Nucl Med 47:91–99PubMed Beyer T, Bockisch A, Kuhl H, Martinez MJ (2006) Whole-body 18F-FDG PET/CT in the presence of truncation artifacts. J Nucl Med 47:91–99PubMed
59.
go back to reference Nuyts J, Bal G, Kehren F, et al. (2013) Completion of a truncated attenuation image from the attenuated PET emission data. IEEE Trans Med Imaging 32:237–246PubMedCrossRef Nuyts J, Bal G, Kehren F, et al. (2013) Completion of a truncated attenuation image from the attenuated PET emission data. IEEE Trans Med Imaging 32:237–246PubMedCrossRef
60.
go back to reference Baldwin LN, Wachowicz K, Thomas SD, Rivest R, Fallone BG (2007) Characterization, prediction, and correction of geometric distortion in 3 T MR images. Med Phys 34:388–399PubMedCrossRef Baldwin LN, Wachowicz K, Thomas SD, Rivest R, Fallone BG (2007) Characterization, prediction, and correction of geometric distortion in 3 T MR images. Med Phys 34:388–399PubMedCrossRef
61.
go back to reference Baldwin LN, Wachowicz K, Fallone BG (2009) A two-step scheme for distortion rectification of magnetic resonance images. Med Phys 36:3917–3926PubMedCrossRef Baldwin LN, Wachowicz K, Fallone BG (2009) A two-step scheme for distortion rectification of magnetic resonance images. Med Phys 36:3917–3926PubMedCrossRef
62.
go back to reference Chang H, Fitzpatrick JM (1992) A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities. IEEE Trans Med Imaging 11:319–329PubMedCrossRef Chang H, Fitzpatrick JM (1992) A technique for accurate magnetic resonance imaging in the presence of field inhomogeneities. IEEE Trans Med Imaging 11:319–329PubMedCrossRef
63.
go back to reference Doran SJ, Charles-Edwards L, Reinsberg SA, Leach MO (2005) A complete distortion correction for MR images: I. Gradient warp correction. Phys Med Biol 50:1343–1361PubMedCrossRef Doran SJ, Charles-Edwards L, Reinsberg SA, Leach MO (2005) A complete distortion correction for MR images: I. Gradient warp correction. Phys Med Biol 50:1343–1361PubMedCrossRef
64.
go back to reference Kannengiesser SA, Wang Y, Haacke EM (1999) Geometric distortion correction in gradient-echo imaging by use of dynamic time warping. Magn Reson Med 42:585–590PubMedCrossRef Kannengiesser SA, Wang Y, Haacke EM (1999) Geometric distortion correction in gradient-echo imaging by use of dynamic time warping. Magn Reson Med 42:585–590PubMedCrossRef
65.
go back to reference Langlois S, Desvignes M, Constans JM, Revenu M (1999) MRI geometric distortion: a simple approach to correcting the effects of non-linear gradient fields. J Magn Reson Imaging 9:821–831PubMedCrossRef Langlois S, Desvignes M, Constans JM, Revenu M (1999) MRI geometric distortion: a simple approach to correcting the effects of non-linear gradient fields. J Magn Reson Imaging 9:821–831PubMedCrossRef
66.
go back to reference Reinsberg SA, Doran SJ, Charles-Edwards EM, Leach MO (2005) A complete distortion correction for MR images: II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 50:2651–2661PubMedCrossRef Reinsberg SA, Doran SJ, Charles-Edwards EM, Leach MO (2005) A complete distortion correction for MR images: II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 50:2651–2661PubMedCrossRef
67.
go back to reference Kuhl CK, Traber F, Schild HH (2008) Whole-body high-field-strength (3.0-T) MR Imaging in Clinical Practice. Part I. Technical considerations and clinical applications. Radiology 246:675–696PubMedCrossRef Kuhl CK, Traber F, Schild HH (2008) Whole-body high-field-strength (3.0-T) MR Imaging in Clinical Practice. Part I. Technical considerations and clinical applications. Radiology 246:675–696PubMedCrossRef
68.
go back to reference Dietrich O, Reiser MF, Schoenberg SO (2008) Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol 65:29–35PubMedCrossRef Dietrich O, Reiser MF, Schoenberg SO (2008) Artifacts in 3-T MRI: physical background and reduction strategies. Eur J Radiol 65:29–35PubMedCrossRef
69.
go back to reference Vaughan JT, Adriany G, Snyder CJ, et al. (2004) Efficient high-frequency body coil for high-field MRI. Magn Reson Med 52:851–859PubMedCrossRef Vaughan JT, Adriany G, Snyder CJ, et al. (2004) Efficient high-frequency body coil for high-field MRI. Magn Reson Med 52:851–859PubMedCrossRef
70.
go back to reference Keller SH, Holm S, Hansen AE, et al. (2013) Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. MAGMA 26:173–181PubMedCrossRef Keller SH, Holm S, Hansen AE, et al. (2013) Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. MAGMA 26:173–181PubMedCrossRef
71.
go back to reference Ai T, Padua A, Goerner F, et al. (2012) SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Invest Radiol 47:267–276PubMedCrossRef Ai T, Padua A, Goerner F, et al. (2012) SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Invest Radiol 47:267–276PubMedCrossRef
72.
go back to reference Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW (2012) Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology 265:204–214PubMedCrossRef Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW (2012) Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology 265:204–214PubMedCrossRef
73.
go back to reference Ladefoged CN, Andersen FL, Keller SH, Lofgren J, Hansen AE, Holm S, Hojgaard L, Beyer T (2013) PET/MR imaging of the pelvis in the presence of endoprostheses: reducing image artifacts and increasing accuracy through inpainting. Eur J Nucl Med Mol Imaging Ladefoged CN, Andersen FL, Keller SH, Lofgren J, Hansen AE, Holm S, Hojgaard L, Beyer T (2013) PET/MR imaging of the pelvis in the presence of endoprostheses: reducing image artifacts and increasing accuracy through inpainting. Eur J Nucl Med Mol Imaging
74.
go back to reference Lois C, Bezrukov I, Schmidt H, et al. (2012) Effect of MR contrast agents on quantitative accuracy of PET in combined whole-body PET/MR imaging. Eur J Nucl Med Mol Imaging 39:1756–1766PubMedCrossRef Lois C, Bezrukov I, Schmidt H, et al. (2012) Effect of MR contrast agents on quantitative accuracy of PET in combined whole-body PET/MR imaging. Eur J Nucl Med Mol Imaging 39:1756–1766PubMedCrossRef
Metadata
Title
Whole-body FDG PET-MR oncologic imaging: pitfalls in clinical interpretation related to inaccurate MR-based attenuation correction
Authors
Ulrike Attenberger
Ciprian Catana
Hersh Chandarana
Onofrio A Catalano
Kent Friedman
Stefan A Schonberg
James Thrall
Marco Salvatore
Bruce R. Rosen
Alexander R. Guimaraes
Publication date
01-08-2015
Publisher
Springer US
Published in
Abdominal Radiology / Issue 6/2015
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-015-0455-3

Other articles of this Issue 6/2015

Abdominal Radiology 6/2015 Go to the issue