Abstract

Background

Failure to restore femoral offset in metal on polyethylene total hip arthroplasty (THA) causes polyethylene wear and aseptic loosening. To our knowledge, no study to date has investigated the relationship between femoral offset and wear in metal-on-metal (MOM) THA.

Aims

In this study, we investigated the relationship between femoral offset and wear by measuring circulating metal ion levels in MOM THA.

Methods

In this retrospective study, we identified patients who had undergone MOM THA with the ASR XL system (DePuy International Ltd., Leeds, UK). Femoral offset was measured using anteroposterior radiographs, and circulating metal ion levels (cobalt and chromium) were recorded.

Results

In total, 95 patients were included (68 males and 27 females). The mean age at the time of surgery was 64.9. Mean time from surgery to blood sampling was 15.4 months. No statistically significant relationship was found between femoral offset and cobalt (p = 0.313) or chromium (p = 0.401) ions.

Conclusion

It is known that failure to restore femoral offset during THA can lead to high rates of wear in metal-on-polyethylene articulations. In our study, no statistically significant relationship was found between femoral offset and serum cobalt or chromium ions. This study adds to the information available to surgeons regarding factors that increase wear in metal-on-metal total hip arthroplasty.
Literature
1.
go back to reference Sieber HP, Rieker CB, Köttig P (1999) Analysis of 118 second-generation metal-on-metal retrieved hip implants. J Bone Joint Surg (Br) 81:46–50CrossRef Sieber HP, Rieker CB, Köttig P (1999) Analysis of 118 second-generation metal-on-metal retrieved hip implants. J Bone Joint Surg (Br) 81:46–50CrossRef
2.
go back to reference Cuckler JM (2005) The rationale for metal-on-metal total hip arthroplasty. Clin Orthop Relat Res 441:132–136CrossRefPubMed Cuckler JM (2005) The rationale for metal-on-metal total hip arthroplasty. Clin Orthop Relat Res 441:132–136CrossRefPubMed
3.
go back to reference Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C (2002) In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am 84–A:1836–1841CrossRef Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C (2002) In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am 84–A:1836–1841CrossRef
5.
12.
go back to reference Hart AJ, Ilo K, Underwood R, Cann P, Henckel J, Lewis A et al (2011) The relationship between the angle of version and rate of wear of retrieved metal-on-metal resurfacings. Bone Joint J 93–B:315–320 Hart AJ, Ilo K, Underwood R, Cann P, Henckel J, Lewis A et al (2011) The relationship between the angle of version and rate of wear of retrieved metal-on-metal resurfacings. Bone Joint J 93–B:315–320
13.
go back to reference De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. Bone Joint J 90–B:1291–1297 De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. Bone Joint J 90–B:1291–1297
14.
go back to reference Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P et al (2009) Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and birmingham hip resurfacing arthroplasties. J Bone Joint Surg (Br) 9191:1287–1295. https://doi.org/10.1302/0301-620X.91B10 CrossRef Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P et al (2009) Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and birmingham hip resurfacing arthroplasties. J Bone Joint Surg (Br) 9191:1287–1295. https://​doi.​org/​10.​1302/​0301-620X.​91B10 CrossRef
17.
go back to reference Devane PA, Robinson EJ, Bourne RB, Rorabeck CH, Nayak NN, Horne JG (1997) Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J Bone Joint Surg Am 79:682–689CrossRefPubMed Devane PA, Robinson EJ, Bourne RB, Rorabeck CH, Nayak NN, Horne JG (1997) Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J Bone Joint Surg Am 79:682–689CrossRefPubMed
19.
go back to reference McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME (1995) Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. Bone Joint J 77–B:865–869 McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME (1995) Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. Bone Joint J 77–B:865–869
20.
go back to reference Fackler CD, Poss R (1980) Dislocation in total hip arthroplasties. Clin Orthop Relat Res 151:169–178 Fackler CD, Poss R (1980) Dislocation in total hip arthroplasties. Clin Orthop Relat Res 151:169–178
21.
go back to reference Devane PA, Horne JG (1999) Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res 369:59–72CrossRef Devane PA, Horne JG (1999) Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res 369:59–72CrossRef
23.
go back to reference Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRef Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRef
28.
go back to reference Liebs TR, Nasser L, Herzberg W, Rüther W, Hassenpflug J (2014) The influence of femoral offset on health-related quality of life after total hip replacement. Bone Joint J 96–B:36–42CrossRefPubMed Liebs TR, Nasser L, Herzberg W, Rüther W, Hassenpflug J (2014) The influence of femoral offset on health-related quality of life after total hip replacement. Bone Joint J 96–B:36–42CrossRefPubMed
31.
go back to reference Knight SR, Aujla R, Biswas SP (2011) Total Hip Arthroplasty - over 100 years of operative history. Orthop Rev (Pavia) 3:2–4CrossRef Knight SR, Aujla R, Biswas SP (2011) Total Hip Arthroplasty - over 100 years of operative history. Orthop Rev (Pavia) 3:2–4CrossRef
32.
go back to reference Dobbs H (1980) Survivorship of total hip replacements. J Bone Joint Surg (Br) 62-B(2):168–173CrossRef Dobbs H (1980) Survivorship of total hip replacements. J Bone Joint Surg (Br) 62-B(2):168–173CrossRef
33.
go back to reference Vendittoli P-A, Lavigne M, Girard J, Roy AG (2006) A randomised study comparing resection of acetabular bone at resurfacing and total hip replacement. Bone Joint J 88–B:997–1002 Vendittoli P-A, Lavigne M, Girard J, Roy AG (2006) A randomised study comparing resection of acetabular bone at resurfacing and total hip replacement. Bone Joint J 88–B:997–1002
35.
go back to reference Hallab N, Merritt K, Jacobs JJ (2001) Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am 83:428–436CrossRefPubMed Hallab N, Merritt K, Jacobs JJ (2001) Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am 83:428–436CrossRefPubMed
36.
go back to reference MacDonald SJ (2004) Metal-on-metal total hip arthroplasty: the concerns. Clin Orthop Relat Res 429:86–93CrossRef MacDonald SJ (2004) Metal-on-metal total hip arthroplasty: the concerns. Clin Orthop Relat Res 429:86–93CrossRef
38.
go back to reference Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP et al (2011) Adverse reaction to metal debris following hip resurfacing. Bone Joint J 93–B:164–171 Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP et al (2011) Adverse reaction to metal debris following hip resurfacing. Bone Joint J 93–B:164–171
Metadata
Title
Femoral offset found not to affect metal ion levels in metal-on-metal total hip arthroplasty
Authors
Eoghan Pomeroy
Fergal Macnamara
Evelyn Murphy
Niall McGoldrick
Anant Mahapatra
Nasir Awan
Publication date
01-02-2019
Publisher
Springer London
Published in
Irish Journal of Medical Science (1971 -) / Issue 1/2019
Print ISSN: 0021-1265
Electronic ISSN: 1863-4362
DOI
https://doi.org/10.1007/s11845-018-1808-z

Other articles of this Issue 1/2019

Irish Journal of Medical Science (1971 -) 1/2019 Go to the issue