Use of dual-mobility cup in revision hip arthroplasty reduces the risk for further dislocation: analysis of seven hundred and ninety one first-time revisions performed due to dislocation, reported to the Swedish Hip Arthroplasty Register

Authors: Maziar Mohaddes, Peter Cnudde, Ola Rolfson, Alexander Wall, Johan Kärrholm

Published in: International Orthopaedics | Issue 3/2017

Login to get access

Abstract

Purpose

Dislocation after total hip arthroplasty (THA) is a common reason for revision. The last decade fostered a significant increase in the use of dual-mobility cups (DMCs). Here we report our study on the short-term survival rate of a cemented DMC reported to the Swedish Hip Arthroplasty Register (SHAR) compared with other cemented designs used in first-time revision due to dislocation.

Methods

During 2005–2015, 984 first-time revisions for dislocation were reported to SHAR. In 436 of these cases a cemented dual articular cup was used. During the same time period, 355 revisions performed with a standard cemented cup (femoral head size 28–36 mm) were reported to the SHAR. Patients receiving a DMC were slightly older (75 years, p = 0.005). Re-revision for all reasons was used as primary endpoint. We also anlaysed risk for re-revision of the acetabular component and re-revision due to dislocation. Kaplan–Meier implant survival and a Cox regression analyses adjusted for age and gender were performed.

Results

Implant survival at 4 years for all reasons (91% ± 3.7% vs 86% ± 4.1%, p = 0.02), and especially for re-operation because of dislocation, favours the DMC group (96% ± 3.0% vs 92% ± 3.3%, p = 0.001).

Discussion

Our findings indicate that use of a cemented DMC reduces the short- to mid-term risk of a second revision in first-time revisions compared with classic cup designs. Longer follow-up is needed to establish any long-term clinical advantages when DMCs are used in revisions performed due to dislocation.
Appendix
Available only for authorised users
Literature
1.
go back to reference Khatod M, Cafri G, Namba RS, Inacio MC, Paxton EW (2014) Risk factors for total hip arthroplasty aseptic revision. J Arthroplasty 29(7):1412–7CrossRefPubMed Khatod M, Cafri G, Namba RS, Inacio MC, Paxton EW (2014) Risk factors for total hip arthroplasty aseptic revision. J Arthroplasty 29(7):1412–7CrossRefPubMed
2.
go back to reference Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ (2009) The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am 91(1):128–33CrossRefPubMed Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ (2009) The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am 91(1):128–33CrossRefPubMed
3.
go back to reference Australian Orthopaedic Association National Joint Replacement Registry. Annual Report. Adelaide:AOA; 2015. 2015 [cited 2016 31 October 2016]. Australian Orthopaedic Association National Joint Replacement Registry. Annual Report. Adelaide:AOA; 2015. 2015 [cited 2016 31 October 2016].
4.
go back to reference Garellick, G., Kärrholm, J., Lindahl, H., Malchau, H., Rogmark, C., Rolfson, O. The Swedish Hip Arthroplasty Register, Annual Report 2014. 2015 [cited 2016 31 October 2016]. Garellick, G., Kärrholm, J., Lindahl, H., Malchau, H., Rogmark, C., Rolfson, O. The Swedish Hip Arthroplasty Register, Annual Report 2014. 2015 [cited 2016 31 October 2016].
5.
go back to reference Andersen AV, Kjersgaard AG, Solgaard S (2013) Trilogy-constrained acetabular component for recurrent dislocation. ISRN Orthop, 2013: 629201 Andersen AV, Kjersgaard AG, Solgaard S (2013) Trilogy-constrained acetabular component for recurrent dislocation. ISRN Orthop, 2013: 629201
6.
go back to reference Guyen O, Lewallen DG, Cabanela ME (2008) Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am 90(7):1553–60CrossRefPubMed Guyen O, Lewallen DG, Cabanela ME (2008) Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am 90(7):1553–60CrossRefPubMed
7.
go back to reference Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL (2005) The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty 20(7 Suppl 3):93–102CrossRefPubMed Berend KR, Lombardi AV Jr, Mallory TH, Adams JB, Russell JH, Groseth KL (2005) The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty 20(7 Suppl 3):93–102CrossRefPubMed
8.
go back to reference Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ (2012) Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 94(18):1655–61CrossRefPubMed Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM, Jacobs JJ (2012) Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 94(18):1655–61CrossRefPubMed
9.
go back to reference Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK (2009) Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty 24(1):158 e15–9CrossRef Duffy GP, Wannomae KK, Rowell SL, Muratoglu OK (2009) Fracture of a cross-linked polyethylene liner due to impingement. J Arthroplasty 24(1):158 e15–9CrossRef
10.
go back to reference Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty 20(1):11–9CrossRefPubMed Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty 20(1):11–9CrossRefPubMed
11.
go back to reference Berry DJ, von Knoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am 87(11):2456–63PubMed Berry DJ, von Knoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am 87(11):2456–63PubMed
12.
go back to reference Jameson SS, Lees D, James P, Serrano-Pedraza I, Partington PF, Muller SD, Meek RM, Reed MR (2011) Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J Bone Joint Surg (Br) 93(7):876–80CrossRef Jameson SS, Lees D, James P, Serrano-Pedraza I, Partington PF, Muller SD, Meek RM, Reed MR (2011) Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. J Bone Joint Surg (Br) 93(7):876–80CrossRef
13.
go back to reference Garbuz DS, Masri BA, Duncan CP, Greidanus NV, Bohm ER, Petrak MJ, Della Valle CJ, Gross AE (2012) The Frank Stinchfield Award: Dislocation in revision THA: do large heads (36 and 40 mm) result in reduced dislocation rates in a randomized clinical trial? Clin Orthop Relat Res 470(2):351–6CrossRefPubMed Garbuz DS, Masri BA, Duncan CP, Greidanus NV, Bohm ER, Petrak MJ, Della Valle CJ, Gross AE (2012) The Frank Stinchfield Award: Dislocation in revision THA: do large heads (36 and 40 mm) result in reduced dislocation rates in a randomized clinical trial? Clin Orthop Relat Res 470(2):351–6CrossRefPubMed
14.
go back to reference Farizon F, de Lavison R, Azoulai JJ, Bousquet G (1998) Results with a cementless alumina-coated cup with dual mobility - A twelve-year follow-up study. Int Orthop 22(4):219–224CrossRefPubMedPubMedCentral Farizon F, de Lavison R, Azoulai JJ, Bousquet G (1998) Results with a cementless alumina-coated cup with dual mobility - A twelve-year follow-up study. Int Orthop 22(4):219–224CrossRefPubMedPubMedCentral
15.
go back to reference Langlais FL, Gaucher F, Musset T, Chaix O (2008) Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res 466:389–395CrossRefPubMedPubMedCentral Langlais FL, Gaucher F, Musset T, Chaix O (2008) Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res 466:389–395CrossRefPubMedPubMedCentral
16.
go back to reference Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabriona A, Mertla P (2011) Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: A 59 cases series with a mean 8 years’ follow-up. Elsevier, 97: 8–13 Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabriona A, Mertla P (2011) Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: A 59 cases series with a mean 8 years’ follow-up. Elsevier, 97: 8–13
17.
go back to reference Hamadouche M, Biau DJ, Huten D, Musset T, Gaucher F (2010) The use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 468:3248–3254CrossRefPubMedPubMedCentral Hamadouche M, Biau DJ, Huten D, Musset T, Gaucher F (2010) The use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 468:3248–3254CrossRefPubMedPubMedCentral
18.
go back to reference Guyen O, Pibarot V, Vaz G, Chevillotte C, Béjui-Hugues J (2009) Use of a Dual Mobility Socket to Manage Total Hip Arthroplasty Instability. Clin Orthop Relat Res 467:465–472CrossRefPubMed Guyen O, Pibarot V, Vaz G, Chevillotte C, Béjui-Hugues J (2009) Use of a Dual Mobility Socket to Manage Total Hip Arthroplasty Instability. Clin Orthop Relat Res 467:465–472CrossRefPubMed
19.
go back to reference Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R (2014) A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess wether a dual mobility cup has a decreased dislocation risk. Int Orthop 38(6):1125–1129CrossRefPubMedPubMedCentral Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R (2014) A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess wether a dual mobility cup has a decreased dislocation risk. Int Orthop 38(6):1125–1129CrossRefPubMedPubMedCentral
20.
go back to reference Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabrion A, Mertl P (2011) Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 8 years’ follow-up. Orthop Traumatol Surg Res 97(1):8–13CrossRefPubMed Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabrion A, Mertl P (2011) Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 8 years’ follow-up. Orthop Traumatol Surg Res 97(1):8–13CrossRefPubMed
21.
go back to reference van Heumen M, Heesterbeek PJ, Swierstra BA, Van Hellemondt GG, Goosen JH (2015) Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after 1–5 years. J Orthop Traumatol 16(1):15–20CrossRefPubMed van Heumen M, Heesterbeek PJ, Swierstra BA, Van Hellemondt GG, Goosen JH (2015) Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after 1–5 years. J Orthop Traumatol 16(1):15–20CrossRefPubMed
22.
go back to reference Jakobsen T, Kappel A, Hansen F, Krarup N (2014) The dislocating hip replacement - revision with a dual mobility cup in 56 consecutive patients. Open Orthop J 8:268–71CrossRefPubMedPubMedCentral Jakobsen T, Kappel A, Hansen F, Krarup N (2014) The dislocating hip replacement - revision with a dual mobility cup in 56 consecutive patients. Open Orthop J 8:268–71CrossRefPubMedPubMedCentral
23.
go back to reference Philippot R, Adam P, Reckhaus M, Delangle F, Verdot F-X, Curvale G, Farizon F (2009) Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res 95(6):407–413CrossRefPubMed Philippot R, Adam P, Reckhaus M, Delangle F, Verdot F-X, Curvale G, Farizon F (2009) Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res 95(6):407–413CrossRefPubMed
24.
go back to reference Prudhon JL, Steffann F, Ferreira A, Verdier R, Aslanian T, Caton J (2014) Cementless dual-mobility cup in total hip arthroplasty revision. Int Orthop 38(12):2463–2468CrossRefPubMed Prudhon JL, Steffann F, Ferreira A, Verdier R, Aslanian T, Caton J (2014) Cementless dual-mobility cup in total hip arthroplasty revision. Int Orthop 38(12):2463–2468CrossRefPubMed
25.
go back to reference Hailer NP, Weiss RJ, Stark A, Kärrholm J (2012) Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocation: 228 patients from the Swedish Hip Arthroplasty Register. Acta Orthop 83(6):566–571CrossRefPubMedPubMedCentral Hailer NP, Weiss RJ, Stark A, Kärrholm J (2012) Dual-mobility cups for revision due to instability are associated with a low rate of re-revisions due to dislocation: 228 patients from the Swedish Hip Arthroplasty Register. Acta Orthop 83(6):566–571CrossRefPubMedPubMedCentral
26.
go back to reference Leclercq S, El Blidi S, Aubriot J (1995) Traitement de la luxation récidivante de prothèse totale de hanche par le cotyle de Bousquet. A propos de 13 cas. Rev Chir Orthop Reparatrice Appar Mot 81(5):389–394PubMed Leclercq S, El Blidi S, Aubriot J (1995) Traitement de la luxation récidivante de prothèse totale de hanche par le cotyle de Bousquet. A propos de 13 cas. Rev Chir Orthop Reparatrice Appar Mot 81(5):389–394PubMed
27.
go back to reference Soderman P, Malchau H, Herberts P, Zügner R, Regnér H, Garellick G (2001) Outcome after total hip arthroplasty: Part II. Disease-specific follow-up and the Swedish National Total Hip Arthroplasty Register. Acta Orthop 72(2):113–119CrossRef Soderman P, Malchau H, Herberts P, Zügner R, Regnér H, Garellick G (2001) Outcome after total hip arthroplasty: Part II. Disease-specific follow-up and the Swedish National Total Hip Arthroplasty Register. Acta Orthop 72(2):113–119CrossRef
28.
go back to reference Timperley AJ, Biau D, Chew D, Whitehouse SL (2016) Dislocation after total hip replacement: there is no such thing as a safe zone for socket placement with the posterior approach. Hip Int Timperley AJ, Biau D, Chew D, Whitehouse SL (2016) Dislocation after total hip replacement: there is no such thing as a safe zone for socket placement with the posterior approach. Hip Int
30.
go back to reference Ritter MA (1995) The Cemented Acetabular Component of a Total Hip Replacement: All Polyethylene Versus Metal Backing. Clin Orthop Relat Res 311:69–75 Ritter MA (1995) The Cemented Acetabular Component of a Total Hip Replacement: All Polyethylene Versus Metal Backing. Clin Orthop Relat Res 311:69–75
31.
go back to reference Peraldi P, Vandenbussche E, Augereau B (1996) Bad clinical results of cemented caps with metal-backed acetabular components. 124 cases with 21 months follow-up. Rev Chir Orthop Reparatrice Appar Mot 83(6):561–565 Peraldi P, Vandenbussche E, Augereau B (1996) Bad clinical results of cemented caps with metal-backed acetabular components. 124 cases with 21 months follow-up. Rev Chir Orthop Reparatrice Appar Mot 83(6):561–565
32.
go back to reference Saragaglia D, Ruatti S, Refaie R (2013) Relevance of a press-fit dual mobility cup to deal with recurrent dislocation of conventional total hip arthroplasty: a 29-case series. Eur J Orthop Surg Traumatol 23(4):431–436CrossRefPubMed Saragaglia D, Ruatti S, Refaie R (2013) Relevance of a press-fit dual mobility cup to deal with recurrent dislocation of conventional total hip arthroplasty: a 29-case series. Eur J Orthop Surg Traumatol 23(4):431–436CrossRefPubMed
Metadata
Title
Use of dual-mobility cup in revision hip arthroplasty reduces the risk for further dislocation: analysis of seven hundred and ninety one first-time revisions performed due to dislocation, reported to the Swedish Hip Arthroplasty Register
Authors
Maziar Mohaddes
Peter Cnudde
Ola Rolfson
Alexander Wall
Johan Kärrholm
Publication date
01-03-2017
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 3/2017
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-016-3381-2

Other articles of this Issue 3/2017

International Orthopaedics 3/2017 Go to the issue