Dual mobility cups provide biomechanical advantages in situations at risk for dislocation: a finite element analysis

Authors: Alexandre Terrier, Adeliya Latypova, Maika Guillemin, Valérie Parvex, Olivier Guyen

Published in: International Orthopaedics | Issue 3/2017

Login to get access

Abstract

Purpose

Constrained devices, standard implants with large heads, and dual mobility systems have become popular options to manage instability after total hip arthroplasty (THA). Clinical results with these options have shown variable success rates and significant higher rates of aseptic loosening and mechanical failures with constrained implants. Literature suggests potential advantages of dual mobility, however little is known about its biomechanics. We present a comparative biomechanical study of a standard implant, a constrained implant, and a dual mobility system.

Methods

A finite element analysis was developed to assess and compare these acetabular options with regard to the range of motion (ROM) to impingement, the angle of dislocation, the resistive torque, the volume of polyethylene (PE) with a stress above 80% of the elastic limit, and the interfacial cup/bone stress.

Results

Dual mobility implants provided the greatest ROM to impingement and allowed delaying subluxation and dislocation when compared to standard and constrained implants. Dual mobility also demonstrated the lowest resistive torque at subluxation while the constrained implant provided the greatest one. The lowest critical PE volume was observed with the dual mobility implant, and the highest stress at the interfaces was observed with the constrained implant.

Conclusion

This study highlights the biomechanical advantages of dual mobility systems over constrained and standard implants, and is supported by the clinical results reported. Therefore, the use of dual mobility systems in situations at risk for instability should be advocated and constrained implants should be restricted to salvage situations.
Literature
2.
go back to reference Berry DJ (2001) Unstable total hip arthroplasty: detailed overview. Instr Course Lect 50:265–274PubMed Berry DJ (2001) Unstable total hip arthroplasty: detailed overview. Instr Course Lect 50:265–274PubMed
3.
go back to reference Bartz RL, Noble PC, Kadakia NR, Tullos HS (2000) The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am 82:1300–1307CrossRefPubMed Bartz RL, Noble PC, Kadakia NR, Tullos HS (2000) The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am 82:1300–1307CrossRefPubMed
6.
go back to reference Goetz DD, Capello WN, Callaghan JJ, et al (1998) Salvage of total hip instability with a constrained acetabular component. Clin Orthop 171–181 Goetz DD, Capello WN, Callaghan JJ, et al (1998) Salvage of total hip instability with a constrained acetabular component. Clin Orthop 171–181
8.
15.
go back to reference van Heumen M, Heesterbeek PJC, Swierstra BA et al (2015) Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after 1–5 years. J Orthop Traumatol 16:15–20. doi:10.1007/s10195-014-0318-7 CrossRefPubMed van Heumen M, Heesterbeek PJC, Swierstra BA et al (2015) Dual mobility acetabular component in revision total hip arthroplasty for persistent dislocation: no dislocations in 50 hips after 1–5 years. J Orthop Traumatol 16:15–20. doi:10.​1007/​s10195-014-0318-7 CrossRefPubMed
19.
go back to reference Wu G, Siegler S, Allard P et al (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part 1: ankle, hip, and spine. J Biomech 35:543–548CrossRefPubMed Wu G, Siegler S, Allard P et al (2002) ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part 1: ankle, hip, and spine. J Biomech 35:543–548CrossRefPubMed
21.
go back to reference Teoh SH, Chan WH, Thampuran R (2002) An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. J Biomech 35:323–30CrossRefPubMed Teoh SH, Chan WH, Thampuran R (2002) An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. J Biomech 35:323–30CrossRefPubMed
22.
23.
go back to reference Saikko VO (1996) A three-axis hip joint simulator for wear and friction studies on total hip prostheses. Proc Inst Mech Eng H 210:175–85CrossRefPubMed Saikko VO (1996) A three-axis hip joint simulator for wear and friction studies on total hip prostheses. Proc Inst Mech Eng H 210:175–85CrossRefPubMed
25.
go back to reference Shrader MW, Parvizi J, Lewallen DG (2003) The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am 85:2179–2183CrossRefPubMed Shrader MW, Parvizi J, Lewallen DG (2003) The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am 85:2179–2183CrossRefPubMed
27.
go back to reference Muratoglu OK, Bragdon CR, O’Connor D et al (2001) Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty 16:24–30. doi:10.1054/arth.2001.28376 CrossRefPubMed Muratoglu OK, Bragdon CR, O’Connor D et al (2001) Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J Arthroplasty 16:24–30. doi:10.​1054/​arth.​2001.​28376 CrossRefPubMed
Metadata
Title
Dual mobility cups provide biomechanical advantages in situations at risk for dislocation: a finite element analysis
Authors
Alexandre Terrier
Adeliya Latypova
Maika Guillemin
Valérie Parvex
Olivier Guyen
Publication date
01-03-2017
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 3/2017
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-016-3368-z

Other articles of this Issue 3/2017

International Orthopaedics 3/2017 Go to the issue