Published in: Clinical Orthopaedics and Related Research® 12/2010

01-12-2010 | Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation

Authors: Moussa Hamadouche, MD, PhD, David J. Biau, MD, Denis Huten, MD, Thierry Musset, MD, François Gaucher, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2010

Login to get access

Abstract

Background

The treatment of recurrent dislocation after total hip arthroplasty remains challenging. Dual mobility sockets have been associated with a low rate of dislocation but it is not known whether they are useful for treating recurrent dislocation.

Questions/purposes

We therefore asked whether a cemented dual mobility socket would (1) restore hip stability following recurrent dislocation; (2) provide a pain-free and mobile hip; and (3) show durable radiographic fixation.

Methods

We retrospectively reviewed 51 patients treated with a cemented dual mobility socket for recurrent dislocation after total hip arthroplasty between August 2002 and June 2005. The mean age at the time of the index procedure of was 71.3 years. Of the 51 patients, 47 have had complete clinical and radiographic evaluation data at a mean followup of 51.4 months (range, 25–76.3 months).

Results

The cemented dual mobility socket restored complete stability of the hip in 45 of the 47 patients (96%). The mean Merle d’Aubigné hip score was 16 ± 2 at the latest followup. Radiographic analysis revealed no or radiolucent lines less than 1 mm thick located in a single acetabular zone in 43 of 47 hips (91.5%). The cumulative survival rate of the acetabular component at 72 months using revision for dislocation and/or mechanical failure as the end point was 96% ± 4% (95% confidence interval, 90%–100%).

Conclusions

A cemented dual mobility socket was able to restore hip stability in 96% of recurrent dislocating hips. However, longer-term followup is needed to ensure that dislocation and loosening rates will not increase.

Level of Evidence

Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop. 2005;91:627–636.PubMed Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop. 2005;91:627–636.PubMed
2.
go back to reference Anderson MJ, Murray WR, Skinner HB. Constrained acetabular components. J Arthroplasty. 1994;9:17–23.CrossRefPubMed Anderson MJ, Murray WR, Skinner HB. Constrained acetabular components. J Arthroplasty. 1994;9:17–23.CrossRefPubMed
3.
go back to reference Aubriot J, Lesimple P, Leclercq S. Study of Bousquet’s non-cemented acetabular implant in 100 hybrid total hip prostheses (Charnley type cemented femoral component). Average 5-year follow-up [in French]. Acta Orthop Belg. 1993;59(Suppl 1):267–271.PubMed Aubriot J, Lesimple P, Leclercq S. Study of Bousquet’s non-cemented acetabular implant in 100 hybrid total hip prostheses (Charnley type cemented femoral component). Average 5-year follow-up [in French]. Acta Orthop Belg. 1993;59(Suppl 1):267–271.PubMed
4.
go back to reference Barrack RL, Mulroy RD, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992;74:385–389.PubMed Barrack RL, Mulroy RD, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992;74:385–389.PubMed
5.
go back to reference Berend KR, Lombardi AV, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty. 2005;20(Suppl 3):93–102.CrossRefPubMed Berend KR, Lombardi AV, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty. 2005;20(Suppl 3):93–102.CrossRefPubMed
6.
go back to reference Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed
7.
go back to reference Bousquet G, Gazielly DF, Giradin P, Debiesse JL, Relave M, Israeli A. The ceramic coated cementless total hip arthroplasty. Basic concepts and surgical technique. J Orthop Surg Tech. 1985;1:15–28. Bousquet G, Gazielly DF, Giradin P, Debiesse JL, Relave M, Israeli A. The ceramic coated cementless total hip arthroplasty. Basic concepts and surgical technique. J Orthop Surg Tech. 1985;1:15–28.
8.
go back to reference Cates HE, Faris PM, Keating EM, Ritter MA. Polyethylene wear in cemented metal-backed acetabular cups. J Bone Joint Surg Br. 1993;75:249–253.PubMed Cates HE, Faris PM, Keating EM, Ritter MA. Polyethylene wear in cemented metal-backed acetabular cups. J Bone Joint Surg Br. 1993;75:249–253.PubMed
9.
go back to reference Chandler RW, Dorr LD, Perry J. The functional cost of dislocation following total hip arthroplasty. Clin Orthop Relat Res. 1982;168:168–172.PubMed Chandler RW, Dorr LD, Perry J. The functional cost of dislocation following total hip arthroplasty. Clin Orthop Relat Res. 1982;168:168–172.PubMed
10.
go back to reference Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.CrossRefPubMed Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.CrossRefPubMed
11.
go back to reference Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed
12.
go back to reference De Lee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32. De Lee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.
13.
go back to reference Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG. High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplasty. 2005;20(Suppl 3):103–107.CrossRefPubMed Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG. High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplasty. 2005;20(Suppl 3):103–107.CrossRefPubMed
14.
go back to reference Farizon F, de LR, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup with dual mobility. A twelve-year follow-up study. Int Orthop. 1998;22:219–224.CrossRefPubMed Farizon F, de LR, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup with dual mobility. A twelve-year follow-up study. Int Orthop. 1998;22:219–224.CrossRefPubMed
15.
16.
go back to reference Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A concise follow-up of a previous report. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A concise follow-up of a previous report. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed
17.
go back to reference Goetz DD, Capello WN, Callaghan JJ, Brown TD, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg Am. 1998;80:502–509.PubMed Goetz DD, Capello WN, Callaghan JJ, Brown TD, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg Am. 1998;80:502–509.PubMed
18.
go back to reference Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
19.
go back to reference Guyen O, Chen QS, Bejui-Hugues J, Berry DJ, An KN. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res. 2007;455:202–208.CrossRefPubMed Guyen O, Chen QS, Bejui-Hugues J, Berry DJ, An KN. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res. 2007;455:202–208.CrossRefPubMed
20.
go back to reference Guyen O, Lewallen DG, Cabanela ME. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am. 2008;90:1553–1560.CrossRefPubMed Guyen O, Lewallen DG, Cabanela ME. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am. 2008;90:1553–1560.CrossRefPubMed
21.
go back to reference Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465–472.CrossRefPubMed Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465–472.CrossRefPubMed
22.
go back to reference Guyen O, Pibarot V, Vaz G, Chevillotte C, Carret JP, Bejui-Hugues J. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplast. 2007;22:849–58. Guyen O, Pibarot V, Vaz G, Chevillotte C, Carret JP, Bejui-Hugues J. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplast. 2007;22:849–58.
23.
go back to reference Harris WH, Penenberg BL. Further follow-up on socket fixation using a metal-backed acetabular component for total hip replacement. A minimum ten-year follow-up study. J Bone Joint Surg Am. 1987;69:1140–1143.PubMed Harris WH, Penenberg BL. Further follow-up on socket fixation using a metal-backed acetabular component for total hip replacement. A minimum ten-year follow-up study. J Bone Joint Surg Am. 1987;69:1140–1143.PubMed
24.
go back to reference Kaper BP, Bernini PM. Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg Am. 1998;80:561–565.PubMed Kaper BP, Bernini PM. Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg Am. 1998;80:561–565.PubMed
25.
go back to reference Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.CrossRefPubMed Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.CrossRefPubMed
26.
go back to reference Khan RJ, Fick D, Alakeson R, Haebich S, de Cruz M, Nivbrant B, Wood D. A constrained acetabular component for recurrent dislocation. J Bone Joint Surg Br. 2006;88:870–876.CrossRefPubMed Khan RJ, Fick D, Alakeson R, Haebich S, de Cruz M, Nivbrant B, Wood D. A constrained acetabular component for recurrent dislocation. J Bone Joint Surg Br. 2006;88:870–876.CrossRefPubMed
27.
go back to reference Krein SW, Chao EY. Biomechanics of bipolar hip endoprostheses. J Orthop Res. 1984;4:356–368.CrossRef Krein SW, Chao EY. Biomechanics of bipolar hip endoprostheses. J Orthop Res. 1984;4:356–368.CrossRef
28.
go back to reference Langlais F, Ropars M, Gaucher F, Musset T, Chaix O. Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res. 2008;466:389–395.CrossRefPubMed Langlais F, Ropars M, Gaucher F, Musset T, Chaix O. Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res. 2008;466:389–395.CrossRefPubMed
29.
go back to reference Lautridou C, Lebel B, Burdin G, Vielpeau C. Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:731–739.PubMed Lautridou C, Lebel B, Burdin G, Vielpeau C. Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:731–739.PubMed
30.
go back to reference Leclercq S, El Bidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis. Apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed Leclercq S, El Bidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis. Apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed
31.
go back to reference Lecuire F, Benareau I, Rubini J, Basso M. [Intra-prosthetic dislocation of the Bousquet dual mobility socket] [in French].Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255. Lecuire F, Benareau I, Rubini J, Basso M. [Intra-prosthetic dislocation of the Bousquet dual mobility socket] [in French].Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255.
32.
go back to reference Lombardi AV, Mallory TH, Kraus TJ, Vaughn BK. Preliminary report on the S-ROM constraining acetabular insert: a retrospective clinical experience. Orthopedics. 1991;14:297–303.PubMed Lombardi AV, Mallory TH, Kraus TJ, Vaughn BK. Preliminary report on the S-ROM constraining acetabular insert: a retrospective clinical experience. Orthopedics. 1991;14:297–303.PubMed
33.
go back to reference Mantel N, Haenzel W. Statistical aspects of the analysis of data from retrospective studies of disease. J Natl Cancer Inst. 1959;22:719–748.PubMed Mantel N, Haenzel W. Statistical aspects of the analysis of data from retrospective studies of disease. J Natl Cancer Inst. 1959;22:719–748.PubMed
34.
go back to reference McCarthy JC, Lee JA. Constrained acetabular components in complex revision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:210–215.CrossRefPubMed McCarthy JC, Lee JA. Constrained acetabular components in complex revision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:210–215.CrossRefPubMed
35.
go back to reference Merle d’aubigné R. Numerical classification of the function of the hip [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1970;56:481–486. Merle d’aubigné R. Numerical classification of the function of the hip [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1970;56:481–486.
36.
go back to reference Murray DW, Carr AJ, Bulstrode C. Survival analysis of joint replacements. J Bone Joint Surg Br. 1993;75:697–704.PubMed Murray DW, Carr AJ, Bulstrode C. Survival analysis of joint replacements. J Bone Joint Surg Br. 1993;75:697–704.PubMed
37.
go back to reference Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed
38.
go back to reference Philippot R, Adam P, Reckhaus M, Delangle F, Verdot FX, Curvale G, Farizon F. Prevention of dislocation in total hip revision surgery using a dual mobility design [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2009;95:407–413. Philippot R, Adam P, Reckhaus M, Delangle F, Verdot FX, Curvale G, Farizon F. Prevention of dislocation in total hip revision surgery using a dual mobility design [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2009;95:407–413.
39.
go back to reference Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop. 2009;33:927–932.CrossRefPubMed Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop. 2009;33:927–932.CrossRefPubMed
40.
go back to reference Philippot R, Farizon F, Camilleri JP, Boyer B, Derhi G, Bonnan J, Fessy MH, Lecuire F. Survival of dual mobility socket with a mean 17 years follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:43–48.PubMed Philippot R, Farizon F, Camilleri JP, Boyer B, Derhi G, Bonnan J, Fessy MH, Lecuire F. Survival of dual mobility socket with a mean 17 years follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:43–48.PubMed
41.
go back to reference Ritter MA. The cemented acetabular component of a total hip replacement. All polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75. Ritter MA. The cemented acetabular component of a total hip replacement. All polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75.
42.
go back to reference Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552, vii. Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552, vii.
43.
go back to reference Sanchez-Sotelo J, Haidukewych GJ, Boberg C. Hospital cost of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2006;88:290–294.CrossRefPubMed Sanchez-Sotelo J, Haidukewych GJ, Boberg C. Hospital cost of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2006;88:290–294.CrossRefPubMed
44.
go back to reference Shapiro G, Weiland D, Markel D, Padgett D, Sculco T, Pellicci P. The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.CrossRefPubMed Shapiro G, Weiland D, Markel D, Padgett D, Sculco T, Pellicci P. The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.CrossRefPubMed
45.
go back to reference Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed
46.
go back to reference Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res. 2004;420:122–129.CrossRefPubMed Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res. 2004;420:122–129.CrossRefPubMed
47.
go back to reference Tufescu TV, Dust W. Failure of a new constrained acetabular insert: a report of 2 cases. J Arthroplasty. 2004;19:238–239. CrossRefPubMed Tufescu TV, Dust W. Failure of a new constrained acetabular insert: a report of 2 cases. J Arthroplasty. 2004;19:238–239. CrossRefPubMed
48.
go back to reference Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed
49.
go back to reference Yun AG, Padgett D, Pellicci P, Dorr LD. Constrained acetabular liners: mechanisms of failure. J Arthroplasty. 2005;20:536–541.CrossRefPubMed Yun AG, Padgett D, Pellicci P, Dorr LD. Constrained acetabular liners: mechanisms of failure. J Arthroplasty. 2005;20:536–541.CrossRefPubMed
Metadata
Title
The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation
Authors
Moussa Hamadouche, MD, PhD
David J. Biau, MD
Denis Huten, MD
Thierry Musset, MD
François Gaucher, MD
Publication date
01-12-2010
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2010
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1404-7

Other articles of this Issue 12/2010

Clinical Orthopaedics and Related Research® 12/2010 Go to the issue

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Papers Presented at the 2009 Closed Meeting of the International Hip Society: Editorial Comment

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Cemented Hip Designs are a Reasonable Option in Young Patients

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Highly Crosslinked Polyethylene Reduces Wear in Total Hip Arthroplasty at 5 Years

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Cemented Total Hip Arthroplasty With Subtrochanteric Osteotomy in Dysplastic Hips

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Incidence and Factors Associated with Squeaking in Alumina-on-Alumina THA

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Reduced Femoral Component Subsidence with Improved Impaction Grafting at Revision Hip Arthroplasty