Single-isocenter stereotactic non-coplanar arc treatment of 200 patients with brain metastases: multileaf collimator size and setup uncertainties

Authors: Lucie Calmels, Susan Blak Nyrup Biancardo, Patrik Sibolt, Susanne Nørring Bekke, Ulf Bjelkengren, Eva Wilken, Poul Geertsen, David Sjöström, Claus F. Behrens

Published in: Strahlentherapie und Onkologie | Issue 5/2022

Login to get access

Abstract

Purpose

The purpose of this study was to evaluate our 2 years’ experience with single-isocenter, non-coplanar, volumetric modulated arc therapy (VMAT) for brain metastasis (BM) stereotactic radiosurgery (SRS).

Methods

A total of 202 patients treated with the VMAT SRS solution were analyzed retrospectively. Plan quality was assessed for 5 mm (120) and 2.5 mm (high-definition, HD) central leaf width multileaf collimators (MLCs). For BMs at varying distances from the plan isocenter, the geometric offset from the ideal position for two image-guided radiotherapy workflows was calculated. In the workflow with ExacTrac (BrainLAB, München, Germany; W‑ET), patient positioning errors were corrected at each couch rotation. In the workflow without ExacTrac (W-noET), only the initial patient setup correction was considered. The dose variation due to rotational errors was simulated for multiple-BM plans with the HD-MLC.

Results

Plan conformity and quality assurance were equivalent for plans delivered with the two MLCs while the HD-MLC plans provided better healthy brain tissue (BmP) sparing. 95% of the BMs had residual intrafractional setup errors ≤ 2 mm for W‑ET and 68% for W‑noET. For small BM (≤1 cc) situated >3 cm from the plan isocenter, the dose received by 95% of the BM decreased in median (interquartile range) by 6.3% (2.8–8.8%) for a 1-degree rotational error.

Conclusion

This study indicates that the HD-MLC is advantageous compared to the 120-MLC for sparing healthy brain tissue. When a 2-mm margin is applied, W‑noET is sufficient to ensure coverage of BM situated ≤ 3 cm of the plan isocenter, while for BM further away, W‑ET is recommended.
Literature
Metadata
Title
Single-isocenter stereotactic non-coplanar arc treatment of 200 patients with brain metastases: multileaf collimator size and setup uncertainties
Authors
Lucie Calmels
Susan Blak Nyrup Biancardo
Patrik Sibolt
Susanne Nørring Bekke
Ulf Bjelkengren
Eva Wilken
Poul Geertsen
David Sjöström
Claus F. Behrens
Publication date
01-05-2022
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 5/2022
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-021-01846-6

Other articles of this Issue 5/2022

Strahlentherapie und Onkologie 5/2022 Go to the issue