Published in: Radiation Oncology 1/2017

Open Access 01-12-2017 | Research

Comprehensive target geometric errors and margin assessment in stereotactic partial breast irradiation

Authors: Xin Zhen, Bo Zhao, Zhuoyu Wang, Robert Timmerman, Ann Spangler, Nathan Kim, Asal Rahimi, Xuejun Gu

Published in: Radiation Oncology | Issue 1/2017

Login to get access

Abstract

Background

Recently developed stereotactic partial breast irradiation (S-PBI) allows delivery of a high biologically potent dose to the target while sparing adjacent critical organs and normal tissue. With S-PBI tumoricidal doses, accurate and precise dose delivery is critical to achieve high treatment quality. This study is to investigate both rigid and non-rigid components of target geometric error and their corresponding margins in S-PBI and identify correlated clinical factors.

Methods

Forty-three early-stage breast cancer patients with implanted gold fiducial markers were enrolled in the study. Fiducial positions recorded on the orthogonal kV images on a Cyberknife system during treatment were used to estimate intra-fraction errors and composite errors (including intra-fraction errors and residual errors after patient setup). Both rigid and non-rigid components of intra-fraction and composite errors were analyzed and used to estimate rigid and non-rigid margins, respectively. Univariate and multivariate linear regressions were conducted to evaluate correlations between clinical factors and errors.

Results

For the study group, the intra-fraction rigid and non-rigid errors are 2.0 ± 0.6 mm and 0.3 ± 0.2 mm, respectively. The composite rigid and non-rigid errors are 2.3 ± 0.5 mm and 1.3 ± 0.8 mm, respectively. The rigid margins in the left-right, anterior-posterior, and superior-inferior directions are estimated as 2.1, 2.4, and 2.3 mm, respectively. The estimated non-rigid margin, assumed to be isotropic, is 1.7 mm. The outer breast quadrants are more susceptible to composite errors occurrence than the inner breast quadrants. The target to chest wall distance is the clinical factor correlated with target geometric errors.

Conclusions

This is the first comprehensive analysis of breast target geometric rigid and non-rigid errors in S-PBI. Upon the estimation, the non-rigid margin is comparable to rigid margin, and therefore should be included in planning target volume as it cannot be accounted for by the Cyberknife system. Treatment margins selection also need to consider the impact of relevant clinical factor.
Appendix
Available only for authorised users
Literature
1.
go back to reference Rault E, Lacornerie T, Dang HP, Crop F, Lartigau E, Reynaert N, Pasquier D. Accelerated partial breast irradiation using robotic radiotherapy: a dosimetric comparison with tomotherapy and three-dimensional conformal radiotherapy. Radiat Oncol. 2016;11:29.CrossRefPubMedPubMedCentral Rault E, Lacornerie T, Dang HP, Crop F, Lartigau E, Reynaert N, Pasquier D. Accelerated partial breast irradiation using robotic radiotherapy: a dosimetric comparison with tomotherapy and three-dimensional conformal radiotherapy. Radiat Oncol. 2016;11:29.CrossRefPubMedPubMedCentral
2.
go back to reference Marta GN, Macedo CR, Carvalho HA, Hanna SA, JLF d S, Riera R. Accelerated partial irradiation for breast cancer: systematic review and meta-analysis of 8653 women in eight randomized trials. Radiother Oncol. 2015;114:42–9.CrossRefPubMed Marta GN, Macedo CR, Carvalho HA, Hanna SA, JLF d S, Riera R. Accelerated partial irradiation for breast cancer: systematic review and meta-analysis of 8653 women in eight randomized trials. Radiother Oncol. 2015;114:42–9.CrossRefPubMed
3.
go back to reference Vicini FA, Baglan KL, Kestin LL, Mitchell C, Chen PY, Frazier RC, Edmundson G, Goldstein NS, Benitez P, Huang RR, Martinez A. Accelerated treatment of breast cancer. J Clin Oncol. 2001;19:1993–2001.CrossRefPubMed Vicini FA, Baglan KL, Kestin LL, Mitchell C, Chen PY, Frazier RC, Edmundson G, Goldstein NS, Benitez P, Huang RR, Martinez A. Accelerated treatment of breast cancer. J Clin Oncol. 2001;19:1993–2001.CrossRefPubMed
4.
go back to reference Keisch M, Vicini F, Kuske RR, Hebert M, White J, Quiet C, Arthur D, Scroggins T, Streeter O. Initial clinical experience with the MammoSite breast brachytherapy applicator in women with early-stage breast cancer treated with breast-conserving therapy. Int J Radiat Oncol Biol Phys. 2003;55:289–93.CrossRefPubMed Keisch M, Vicini F, Kuske RR, Hebert M, White J, Quiet C, Arthur D, Scroggins T, Streeter O. Initial clinical experience with the MammoSite breast brachytherapy applicator in women with early-stage breast cancer treated with breast-conserving therapy. Int J Radiat Oncol Biol Phys. 2003;55:289–93.CrossRefPubMed
5.
go back to reference Olivotto IA, Whelan TJ, Parpia S, Kim DH, Berrang T, Truong PT, Kong I, Cochrane B, Nichol A, Roy I, et al. Interim cosmetic and toxicity results from RAPID: a randomized trial of accelerated partial breast irradiation using three-dimensional conformal external beam radiation therapy. J Clin Oncol. 2013;31:4038–45.CrossRefPubMed Olivotto IA, Whelan TJ, Parpia S, Kim DH, Berrang T, Truong PT, Kong I, Cochrane B, Nichol A, Roy I, et al. Interim cosmetic and toxicity results from RAPID: a randomized trial of accelerated partial breast irradiation using three-dimensional conformal external beam radiation therapy. J Clin Oncol. 2013;31:4038–45.CrossRefPubMed
6.
go back to reference Bentzen SM, Yarnold JR. Reports of unexpected late side effects of accelerated partial breast irradiation--radiobiological considerations. Int J Radiat Oncol Biol Phys. 2010;77:969–73.CrossRefPubMedPubMedCentral Bentzen SM, Yarnold JR. Reports of unexpected late side effects of accelerated partial breast irradiation--radiobiological considerations. Int J Radiat Oncol Biol Phys. 2010;77:969–73.CrossRefPubMedPubMedCentral
7.
go back to reference Obayomi-Davies O, Kole TP, Oppong B, Rudra S, Makariou EV, Campbell LD, Anjum HM, Collins SP, Unger K, Willey S, et al. Stereotactic accelerated partial breast irradiation for early-stage breast cancer: rationale, feasibility, and early experience using the CyberKnife radiosurgery delivery platform. Front Oncol. 2016;6:129.CrossRefPubMedPubMedCentral Obayomi-Davies O, Kole TP, Oppong B, Rudra S, Makariou EV, Campbell LD, Anjum HM, Collins SP, Unger K, Willey S, et al. Stereotactic accelerated partial breast irradiation for early-stage breast cancer: rationale, feasibility, and early experience using the CyberKnife radiosurgery delivery platform. Front Oncol. 2016;6:129.CrossRefPubMedPubMedCentral
8.
go back to reference Vermeulen SS, Haas JA. CyberKnife stereotactic body radiotherapy and CyberKnife accelerated partial breast irradiation for the treatment of early breast cancer. Transl Cancer Res. 2014;3:295–302. Vermeulen SS, Haas JA. CyberKnife stereotactic body radiotherapy and CyberKnife accelerated partial breast irradiation for the treatment of early breast cancer. Transl Cancer Res. 2014;3:295–302.
9.
go back to reference Rahimi A, Thomas K, Spangler A, Rao R, Leitch M, Wooldridge R, Rivers A, Seiler S, Albuquerque K, Stevenson S, et al. Preliminary results of a phase 1 dose-escalation trial for early-stage breast cancer using 5-fraction stereotactic body radiation therapy for partial-breast irradiation. Int J Radiat Oncol Biol Phys. 2017;98:196–205.e192.CrossRefPubMed Rahimi A, Thomas K, Spangler A, Rao R, Leitch M, Wooldridge R, Rivers A, Seiler S, Albuquerque K, Stevenson S, et al. Preliminary results of a phase 1 dose-escalation trial for early-stage breast cancer using 5-fraction stereotactic body radiation therapy for partial-breast irradiation. Int J Radiat Oncol Biol Phys. 2017;98:196–205.e192.CrossRefPubMed
10.
go back to reference Yue NJ, Goyal S, Zhou J, Khan AJ, Haffty BG. Intrafractional target motions and uncertainties of treatment setup reference systems in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;79:1549–56.CrossRefPubMed Yue NJ, Goyal S, Zhou J, Khan AJ, Haffty BG. Intrafractional target motions and uncertainties of treatment setup reference systems in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;79:1549–56.CrossRefPubMed
11.
go back to reference Park CK, Pritz J, Zhang GG, Forster KM, Harris EE. Validating fiducial markers for image-guided radiation therapy for accelerated partial breast irradiation in early-stage breast cancer. Int J Radiat Oncol Biol Phys. 2012;82:e425–31.CrossRefPubMed Park CK, Pritz J, Zhang GG, Forster KM, Harris EE. Validating fiducial markers for image-guided radiation therapy for accelerated partial breast irradiation in early-stage breast cancer. Int J Radiat Oncol Biol Phys. 2012;82:e425–31.CrossRefPubMed
12.
go back to reference Trovo M, Polesel J, Biasutti C, Sartor G, Roncadin M, Trovo GM. Fiducial markers for image-guided partial breast irradiation. Radiol Med. 2013;118:1212–9.CrossRefPubMed Trovo M, Polesel J, Biasutti C, Sartor G, Roncadin M, Trovo GM. Fiducial markers for image-guided partial breast irradiation. Radiol Med. 2013;118:1212–9.CrossRefPubMed
13.
go back to reference Yue NJ, Goyal S, Kim LH, Khan A, Haffty BG. Patterns of intrafractional motion and uncertainties of treatment setup reference systems in accelerated partial breast irradiation for right- and left-sided breast cancer. Pract Radiat Oncol. 2014;4:6–12.CrossRefPubMed Yue NJ, Goyal S, Kim LH, Khan A, Haffty BG. Patterns of intrafractional motion and uncertainties of treatment setup reference systems in accelerated partial breast irradiation for right- and left-sided breast cancer. Pract Radiat Oncol. 2014;4:6–12.CrossRefPubMed
14.
go back to reference Yamashita H, Okuma K, Tada K, Shiraishi K, Takahashi W, Shibata-Mobayashi S, Sakumi A, Saotome N, Haga A, Onoe T, et al. Four-dimensional measurement of the displacement of internal fiducial and skin markers during 320-multislice computed tomography scanning of breast cancer. Int J Radiat Oncol Biol Phys. 2012;84:331–5.CrossRefPubMed Yamashita H, Okuma K, Tada K, Shiraishi K, Takahashi W, Shibata-Mobayashi S, Sakumi A, Saotome N, Haga A, Onoe T, et al. Four-dimensional measurement of the displacement of internal fiducial and skin markers during 320-multislice computed tomography scanning of breast cancer. Int J Radiat Oncol Biol Phys. 2012;84:331–5.CrossRefPubMed
15.
go back to reference Seiler S, Rahimi A, Choudhery S, Garwood D, Spangler A, Cherian S, Goudreau S. Ultrasound-guided placement of gold Fiducial markers for stereotactic partial-breast irradiation. AJR Am J Roentgenol. 2016;207:685–8.CrossRefPubMed Seiler S, Rahimi A, Choudhery S, Garwood D, Spangler A, Cherian S, Goudreau S. Ultrasound-guided placement of gold Fiducial markers for stereotactic partial-breast irradiation. AJR Am J Roentgenol. 2016;207:685–8.CrossRefPubMed
16.
go back to reference Kilby W, Dooley JR, Kuduvalli G, Sayeh S, Maurer CR Jr. The CyberKnife robotic Radiosurgery system in 2010. Technol Cancer Res Treat. 2010;9:433–52.CrossRefPubMed Kilby W, Dooley JR, Kuduvalli G, Sayeh S, Maurer CR Jr. The CyberKnife robotic Radiosurgery system in 2010. Technol Cancer Res Treat. 2010;9:433–52.CrossRefPubMed
17.
go back to reference Horn BKP. Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am A. 1987;4:629–42.CrossRef Horn BKP. Closed-form solution of absolute orientation using unit quaternions. J Opt Soc Am A. 1987;4:629–42.CrossRef
18.
go back to reference Zhang Q, Xiong W, Chan MF, Song Y, Burman C. Rotation effects on the target-volume margin determination. Physica Medica. 2015;31:80–4.CrossRefPubMed Zhang Q, Xiong W, Chan MF, Song Y, Burman C. Rotation effects on the target-volume margin determination. Physica Medica. 2015;31:80–4.CrossRefPubMed
19.
go back to reference R Core Team. R: a language and environment for statistical computing. 3.3.0. Vienna: R Foundation for Statistical Computing; 2016. R Core Team. R: a language and environment for statistical computing. 3.3.0. Vienna: R Foundation for Statistical Computing; 2016.
20.
go back to reference Martyn P, Alexey S, Matt D. Bayesian Graphical Models using MCMC. 4–6 ed; 2016. Martyn P, Alexey S, Matt D. Bayesian Graphical Models using MCMC. 4–6 ed; 2016.
21.
go back to reference Acharya S, Fischer-Valuck BW, Mazur TR, Curcuru A, Sona K, Kashani R, Green O, Ochoa L, Mutic S, Zoberi I, et al. Magnetic resonance image guided radiation therapy for external beam accelerated partial-breast irradiation: evaluation of delivered dose and intrafractional cavity motion. Int J Radiat Oncol Biol Phys. 2016;96:785–92.CrossRefPubMed Acharya S, Fischer-Valuck BW, Mazur TR, Curcuru A, Sona K, Kashani R, Green O, Ochoa L, Mutic S, Zoberi I, et al. Magnetic resonance image guided radiation therapy for external beam accelerated partial-breast irradiation: evaluation of delivered dose and intrafractional cavity motion. Int J Radiat Oncol Biol Phys. 2016;96:785–92.CrossRefPubMed
22.
go back to reference van Heijst TC, Philippens ME, Charaghvandi RK, den Hartogh MD, Lagendijk JJ, van den Bongard HJ, van Asselen B. Quantification of intra-fraction motion in breast radiotherapy using supine magnetic resonance imaging. Phys Med Biol. 2016;61:1352–70.CrossRefPubMed van Heijst TC, Philippens ME, Charaghvandi RK, den Hartogh MD, Lagendijk JJ, van den Bongard HJ, van Asselen B. Quantification of intra-fraction motion in breast radiotherapy using supine magnetic resonance imaging. Phys Med Biol. 2016;61:1352–70.CrossRefPubMed
23.
go back to reference Leonard CE, Tallhamer M, Johnson T, Hunter K, Howell K, Kercher J, Widener J, Kaske T, Paul D, Sedlacek S, Carter DL. Clinical experience with image-guided radiotherapy in an accelerated partial breast intensity-modulated radiotherapy protocol. Int J Radiat Oncol Biol Phys. 2010;76:528–34.CrossRefPubMed Leonard CE, Tallhamer M, Johnson T, Hunter K, Howell K, Kercher J, Widener J, Kaske T, Paul D, Sedlacek S, Carter DL. Clinical experience with image-guided radiotherapy in an accelerated partial breast intensity-modulated radiotherapy protocol. Int J Radiat Oncol Biol Phys. 2010;76:528–34.CrossRefPubMed
24.
go back to reference Kim LH, Wong J, Yan D. On-line localization of the lumpectomy cavity using surgical clips. Int J Radiat Oncol Biol Phys. 2007;69:1305–9.CrossRefPubMed Kim LH, Wong J, Yan D. On-line localization of the lumpectomy cavity using surgical clips. Int J Radiat Oncol Biol Phys. 2007;69:1305–9.CrossRefPubMed
Metadata
Title
Comprehensive target geometric errors and margin assessment in stereotactic partial breast irradiation
Authors
Xin Zhen
Bo Zhao
Zhuoyu Wang
Robert Timmerman
Ann Spangler
Nathan Kim
Asal Rahimi
Xuejun Gu
Publication date
01-12-2017
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2017
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-017-0889-6

Other articles of this Issue 1/2017

Radiation Oncology 1/2017 Go to the issue