Published in: Radiation Oncology 1/2016

Open Access 01-12-2016 | Research

Accelerated partial breast irradiation using robotic radiotherapy: a dosimetric comparison with tomotherapy and three-dimensional conformal radiotherapy

Authors: Erwann Rault, Thomas Lacornerie, Hong-Phuong Dang, Frederik Crop, Eric Lartigau, Nick Reynaert, David Pasquier

Published in: Radiation Oncology | Issue 1/2016

Login to get access

Abstract

Background

Accelerated partial breast irradiation (APBI) is a new breast treatment modality aiming to reduce treatment time using hypo fractionation. Compared to conventional whole breast irradiation that takes 5 to 6 weeks, APBI is reported to induce worse cosmetic outcomes both when using three-dimensional conformal radiotherapy (3D-CRT) and intensity-modulated radiotherapy (IMRT). These late normal tissue effects may be attributed to the dose volume effect because a large portion of the non-target breast tissue volume (NTBTV) receives a high dose. In the context of APBI, non-coplanar beams could spare the NTBTV more efficiently. This study evaluates the dosimetric benefit of using the Cyberknife (CK) for APBI in comparison to IMRT (Tomotherapy) and three dimensional conformal radiotherapy (3D-CRT).

Methods

The possibility of using surgical clips, implanted during surgery, to track target movements is investigated first. A phantom of a female thorax was designed in-house using the measurements of 20 patients. Surgical clips of different sizes were inserted inside the breast. A treatment plan was delivered to the mobile and immobile phantom. The motion compensation accuracy was evaluated using three radiochromic films inserted inside the breast. Three dimensional conformal radiotherapy (3D-CRT), Tomotherapy (TOMO) and CK treatment plans were calculated for 10 consecutive patients who received APBI in Lille. To ensure a fair comparison of the three techniques, margins applied to the CTV were set to 10 mm. However, a second CK plan was prepared using 3 mm margins to evaluate the benefits of motion compensation.

Results

Only the larger clips (VITALITEC Medium-Large) could be tracked inside the larger breast (all gamma indices below 1 for 1 % of the maximum dose and 1 mm). All techniques meet the guidelines defined in the NSABP/RTOG and SHARE protocols. As the applied dose volume constraints are very strong, insignificant dosimetric differences exist between techniques regarding the PTV coverage and the sparing of the lung and heart. However, the CK may be used to reduce high doses received by the NTBTV more efficiently.

Conclusions

Robotic stereotactic radiotherapy may be used for APBI to more efficiently spare the NTBTV and improve cosmetic results of APBI.
Literature
1.
go back to reference Ferlay J, Soerjomataram I, Ervik M, Dikshit R, Eser S, Mathers C, Rebelo M, Parkin DM, Forman D, Bray F. Globocan 2012 v1, Cancer incidence and mortality worldwide: IARC CancerBase No.11. ISBN-13 978-9-28-322447-1. Ferlay J, Soerjomataram I, Ervik M, Dikshit R, Eser S, Mathers C, Rebelo M, Parkin DM, Forman D, Bray F. Globocan 2012 v1, Cancer incidence and mortality worldwide: IARC CancerBase No.11. ISBN-13 978-9-28-322447-1.
2.
go back to reference Belkacemi Y, Bourgier C, Kramar A, Auzac G, Dumas I, Lacornerie T, Mége JP, Mijonnet S, Lemonnier J, Lartigau E. SHARE: a French multicenter phase III trial comparing accelerated partial irradiation versus standard or hypofractionated whole breast irradiation in breast cancer patients at low risk of local recurrence. Clin Adv Hematol Oncol. 2013;11(2):76–83. Belkacemi Y, Bourgier C, Kramar A, Auzac G, Dumas I, Lacornerie T, Mége JP, Mijonnet S, Lemonnier J, Lartigau E. SHARE: a French multicenter phase III trial comparing accelerated partial irradiation versus standard or hypofractionated whole breast irradiation in breast cancer patients at low risk of local recurrence. Clin Adv Hematol Oncol. 2013;11(2):76–83.
3.
go back to reference Lehman M, Hickey BE, Francis DP, See AM. Partial breast irradiation for early breast cancer. Cochrane Database Syst Rev. 2014;18:6. Lehman M, Hickey BE, Francis DP, See AM. Partial breast irradiation for early breast cancer. Cochrane Database Syst Rev. 2014;18:6.
4.
go back to reference Hepel JT, Tokita M, Macausland SG, Evans SB, Hiatt JR, Price LL, Dipetrillo T, Wazer DE. Toxicity of three-dimensional conformal radiotherapy for accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2009;75(5):1290–6. Hepel JT, Tokita M, Macausland SG, Evans SB, Hiatt JR, Price LL, Dipetrillo T, Wazer DE. Toxicity of three-dimensional conformal radiotherapy for accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2009;75(5):1290–6.
5.
go back to reference Jagsi R, Ben-David MA, Moran JM, Marsh RB, Griffith KA, Hayman JA, Pierce LJ. Unacceptable cosmesis in a protocol investigating intensity-modulated radiotherapy with active breathing control for accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2010;76:71–8. Jagsi R, Ben-David MA, Moran JM, Marsh RB, Griffith KA, Hayman JA, Pierce LJ. Unacceptable cosmesis in a protocol investigating intensity-modulated radiotherapy with active breathing control for accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2010;76:71–8.
6.
go back to reference Olivotto I, Whelan T, Parpia S, Kim DH, Berrang T, Truong PT, et al. Interim cosmetic and toxicity results from RAPID: A randomized trial of accelerated partial breast irradiation using three-dimensional conformal external beam radiation therapy. J Clin Oncol. 2013;31:4038–45.CrossRefPubMed Olivotto I, Whelan T, Parpia S, Kim DH, Berrang T, Truong PT, et al. Interim cosmetic and toxicity results from RAPID: A randomized trial of accelerated partial breast irradiation using three-dimensional conformal external beam radiation therapy. J Clin Oncol. 2013;31:4038–45.CrossRefPubMed
7.
go back to reference Bentzen S, Yarnold J. Reports of unexpected late side effects of accelerated partial breast irradiation—radiobiological considerations. Int J Radiat Oncol Biol Phys. 2010;77(4):969–73.PubMedCentralCrossRefPubMed Bentzen S, Yarnold J. Reports of unexpected late side effects of accelerated partial breast irradiation—radiobiological considerations. Int J Radiat Oncol Biol Phys. 2010;77(4):969–73.PubMedCentralCrossRefPubMed
8.
go back to reference Heron DE, Mintz A. CyberKnife robotic stereotactic radiosurgery. Technol Cancer Res Treat. 2010;9(5):431–2.CrossRefPubMed Heron DE, Mintz A. CyberKnife robotic stereotactic radiosurgery. Technol Cancer Res Treat. 2010;9(5):431–2.CrossRefPubMed
9.
go back to reference Kilby W, Dooley JR, Kuduvalli G, Sayeh S, Maurer CR. The CyberKnife(R) Robotic Radiosurgery System in 2010. Technol Cancer Res Treat. 2010;9(5):433–52.CrossRefPubMed Kilby W, Dooley JR, Kuduvalli G, Sayeh S, Maurer CR. The CyberKnife(R) Robotic Radiosurgery System in 2010. Technol Cancer Res Treat. 2010;9(5):433–52.CrossRefPubMed
10.
go back to reference Vermeulen S, Cotrutz C, Morris A, Meier R, Buchanan C, Dawson P, Porter B. Accelerated Partial Breast Irradiation: Using the Cyberknife as the radiation delivery platform in the treatment of early breast cancer. Front Oncol. 2001; 1(43). doi: doi:10.3389/fonc.2011.00043. eCollection 2011. Vermeulen S, Cotrutz C, Morris A, Meier R, Buchanan C, Dawson P, Porter B. Accelerated Partial Breast Irradiation: Using the Cyberknife as the radiation delivery platform in the treatment of early breast cancer. Front Oncol. 2001; 1(43). doi: doi:10.​3389/​fonc.​2011.​00043. eCollection 2011.
11.
go back to reference Xu Q, Chen Y, Grimm J, Fan J, An L, Xue J, Pahlajani N, Lacouture T. Dosimetric investigation of accelerated partial breast irradiation (APBI) using CyberKnife. Med Phys. 2012;39(11):6621–8. Xu Q, Chen Y, Grimm J, Fan J, An L, Xue J, Pahlajani N, Lacouture T. Dosimetric investigation of accelerated partial breast irradiation (APBI) using CyberKnife. Med Phys. 2012;39(11):6621–8.
12.
go back to reference Mackie TR. Accurate convolution/superposition for multi-resolution dose calculation using cumulative tabulated kernels. Phys Med Biol. 2005;50:655–80.CrossRefPubMed Mackie TR. Accurate convolution/superposition for multi-resolution dose calculation using cumulative tabulated kernels. Phys Med Biol. 2005;50:655–80.CrossRefPubMed
13.
go back to reference ICRU report 83 Prescribing, recording and reporting photon beam intensity modulated radiation therapy (IMRT). J ICRU Vol. 10 No. 1 (2010). ICRU report 83 Prescribing, recording and reporting photon beam intensity modulated radiation therapy (IMRT). J ICRU Vol. 10 No. 1 (2010).
14.
go back to reference Weiss E, Wijesooriya K, Dill SV, Keall PJ. Tumor and normal tissue motion in the thorax during respiration: Analysis of volumetric and positional variations using 4D CT. Int J Radiat Oncol Biol Phys. 2007;67(1):296–307.CrossRefPubMed Weiss E, Wijesooriya K, Dill SV, Keall PJ. Tumor and normal tissue motion in the thorax during respiration: Analysis of volumetric and positional variations using 4D CT. Int J Radiat Oncol Biol Phys. 2007;67(1):296–307.CrossRefPubMed
15.
go back to reference Qi XS, White J, Rabinovitch R, Merrell K, Sood A, Bauer A, et al. Respiratory organ motion and dosimetric impact on breast and nodal irradiation. Int J Radiat Oncol Biol Phys. 2010;78(2):609–17.CrossRefPubMed Qi XS, White J, Rabinovitch R, Merrell K, Sood A, Bauer A, et al. Respiratory organ motion and dosimetric impact on breast and nodal irradiation. Int J Radiat Oncol Biol Phys. 2010;78(2):609–17.CrossRefPubMed
16.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656–61.CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656–61.CrossRefPubMed
17.
go back to reference Julian TB, Costantino JP, Vicini FA, White JR, Winter KA, Arthur DW, Kuske RR, Rabinovitch R, Curran WJ, Wolmark N. Early Toxicity Results with 3-D Conformal External Beam Therapy (CEBT) from the NSABP B-39/RTOG 0413 Accelerated Partial Breast Irradiation (APBI) Trial. Int J Radiat Oncol Biol Phys. 2011;81(2):S7. abstract. Julian TB, Costantino JP, Vicini FA, White JR, Winter KA, Arthur DW, Kuske RR, Rabinovitch R, Curran WJ, Wolmark N. Early Toxicity Results with 3-D Conformal External Beam Therapy (CEBT) from the NSABP B-39/RTOG 0413 Accelerated Partial Breast Irradiation (APBI) Trial. Int J Radiat Oncol Biol Phys. 2011;81(2):S7. abstract.
18.
go back to reference Leonard KL, Hepel JT, Hiatt JR, Dipetrillo TA, Price LL, Wazer DE. The effect of dose-volume parameters and interfraction interval on cosmetic outcome and toxicity after 3-dimensional conformal accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2013;85(3):623–9.CrossRefPubMed Leonard KL, Hepel JT, Hiatt JR, Dipetrillo TA, Price LL, Wazer DE. The effect of dose-volume parameters and interfraction interval on cosmetic outcome and toxicity after 3-dimensional conformal accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2013;85(3):623–9.CrossRefPubMed
19.
go back to reference Mellon EA, Sreeraman R, Gebhardt BJ, Mierzejewski A, Correa CR. Impact of radiation treatment parameters and adjuvant systemic therapy on cosmetic outcomes after accelerated partial breast irradiation using 3-dimensional conformal radiation therapy technique. Pract Radiat Oncol. 2014;4(3):e159–66.CrossRefPubMed Mellon EA, Sreeraman R, Gebhardt BJ, Mierzejewski A, Correa CR. Impact of radiation treatment parameters and adjuvant systemic therapy on cosmetic outcomes after accelerated partial breast irradiation using 3-dimensional conformal radiation therapy technique. Pract Radiat Oncol. 2014;4(3):e159–66.CrossRefPubMed
20.
go back to reference Peterson D, Truong PT, Parpia S, Olivotto IA, Berrang T, Kim DH, et al. Predictors of adverse cosmetic outcome in the RAPID tria: an exploratory analysis. Int J Radiat Oncol Biol Phys. 2015;91(5):968–76.CrossRefPubMed Peterson D, Truong PT, Parpia S, Olivotto IA, Berrang T, Kim DH, et al. Predictors of adverse cosmetic outcome in the RAPID tria: an exploratory analysis. Int J Radiat Oncol Biol Phys. 2015;91(5):968–76.CrossRefPubMed
21.
go back to reference Patel RR, Becker SJ, Das RK, Mackie TR. A dosimetric comparison of accelerated partial breast irradiation techniques: multicatheter interstitial brachitherapy, three-dimensional conformal radiotherapy, and supine versus prone helical tomotherapy. Int J Radiat Oncol Biol Phys. 2007;68(3):935–42.CrossRefPubMed Patel RR, Becker SJ, Das RK, Mackie TR. A dosimetric comparison of accelerated partial breast irradiation techniques: multicatheter interstitial brachitherapy, three-dimensional conformal radiotherapy, and supine versus prone helical tomotherapy. Int J Radiat Oncol Biol Phys. 2007;68(3):935–42.CrossRefPubMed
22.
go back to reference Moon SH, Shin KH, Kim TH, Yoon M, Park S, Lee DH, Kim JW, Kim DW, Park SY, Cho KH. Dosimetric comparison of four different external beam partial breast irradiation techniques: Three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, helical tomotherapy, and proton beam therapy. Radiat Oncol. 2009;90:66–73. Moon SH, Shin KH, Kim TH, Yoon M, Park S, Lee DH, Kim JW, Kim DW, Park SY, Cho KH. Dosimetric comparison of four different external beam partial breast irradiation techniques: Three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, helical tomotherapy, and proton beam therapy. Radiat Oncol. 2009;90:66–73.
Metadata
Title
Accelerated partial breast irradiation using robotic radiotherapy: a dosimetric comparison with tomotherapy and three-dimensional conformal radiotherapy
Authors
Erwann Rault
Thomas Lacornerie
Hong-Phuong Dang
Frederik Crop
Eric Lartigau
Nick Reynaert
David Pasquier
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2016
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-016-0607-9

Other articles of this Issue 1/2016

Radiation Oncology 1/2016 Go to the issue