Open Access 01-12-2021 | Aortic Valve Replacement | Research article

Application of three-dimensional transesophageal echocardiography in preoperative evaluation of transcatheter aortic valve replacement

Authors: Peng Ding, Chennian Xu, Yang Liu, Xin Meng, Ping Jin, Jiayou Tang, Lanlan Li, Yanyan Ma, Jian Yang

Published in: BMC Cardiovascular Disorders | Issue 1/2021

Login to get access

Abstract

Background

Our goal was to determine the accuracy of 3-dimensional transesophageal echocardiography (3D-TEE) compared with that of computed tomography (CT) in the preoperative evaluation for transcatheter aortic valve replacement (TAVR) when the errors caused by inconsistent software and method have been eliminated and the representativeness of the sample has been improved. We also investigated the influence of aortic root calcification on the accuracy of 3D-TEE in aortic annulus evaluations.

Methods

Part I: 45 of 233 patients who underwent TAVR in the department of cardiovascular surgery at the Xijing hospital from January 2016 to August 2019 were studied retrospectively. Materialise Mimics software and the multiplanar reconstruction method were used for evaluation, based on 3D-TEE and CT. The annulus area-derived diameter, the annulus perimeter-derived diameter (Dp), the annulus mean diameter, the left ventricular outflow tract Dp diameter, the sinotubular junction (STJ) diameter-Dp, and the aortic sinus diameter were compared and analyzed. Part II: 31 of 233 patients whose 3D-TEE and CT data were well preserved and in the required format were included. HU450 and HU850 were used as indicators to measure the severity of calcification. The Spearman rank correlation and Linear regression were used to analyze the correlation between aortic root calcification and the accuracy of 3D-TEE in aortic annulus measurement.

Results

The measurement results based on 3D-TEE were significantly lower than those obtained using CT (P < 0.05), except for the STJ diameter-Dp in diastole (P = 0.11). The correlation coefficient of the two groups was 0.699–0.954 (P < 0.01), which also indicated a significant correlation between the two groups. A Bland–Altman plot showed that the ordinate values were mostly within the 95% consistency limit; the consistency of the two groups was good. By establishing the linear regression equation, the two groups can be inferred from each other. The Spearman rank correlation analysis and the Linear regression analysis showed that the influence of aortic calcification on the accuracy of the 3D-TEE annulus evaluation was limited.

Conclusions

Although an evaluation based on 3D-TEE underestimated the results, we can deduce CT results from 3D-TEE because the two methods exhibit considerable correlation and consistency.

Trial registration

Name: Surgery and Transcatheter Intervention for Structural Heart Diseases. Number: NCT02917980. URL: https://​clinicaltrials.​gov/​ct2/​results?​term=​NCT02917980.
Footnotes
1
Mean diameter = \(\frac{{{\text{maximum diameter}} + {\text{minimal diameter}}}}{2}.\)
 
Literature
1.
go back to reference Baumgartner H, Falk V, Bax JJ, De Bonis M, Hamm C, Holm PJ, et al. 2017 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. 2017;38:2739–91.CrossRef Baumgartner H, Falk V, Bax JJ, De Bonis M, Hamm C, Holm PJ, et al. 2017 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. 2017;38:2739–91.CrossRef
2.
go back to reference Leon MB, Smith CR, Mack MJ, Makkar RR, Svensson LG, Kodali SK, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. 2016;374:1609–20.CrossRef Leon MB, Smith CR, Mack MJ, Makkar RR, Svensson LG, Kodali SK, et al. Transcatheter or surgical aortic-valve replacement in intermediate-risk patients. N Engl J Med. 2016;374:1609–20.CrossRef
3.
go back to reference Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. 2019;380:1695–705.CrossRef Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med. 2019;380:1695–705.CrossRef
4.
go back to reference Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med. 2019;380:1706–15.CrossRef Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med. 2019;380:1706–15.CrossRef
5.
go back to reference Hakim D, Ghimire G, Alli OO, Singh S, Sasse MF, Booker OJ, et al. Large-field intravascular ultrasound for annular sizing and predicting paravalvular regurgitation during TAVR: comparisons with multidetector computed tomography and transoesophageal echocardiography. Eur Heart J Cardiovasc Imaging. 2017;18:1404–13.CrossRef Hakim D, Ghimire G, Alli OO, Singh S, Sasse MF, Booker OJ, et al. Large-field intravascular ultrasound for annular sizing and predicting paravalvular regurgitation during TAVR: comparisons with multidetector computed tomography and transoesophageal echocardiography. Eur Heart J Cardiovasc Imaging. 2017;18:1404–13.CrossRef
6.
go back to reference Vaquerizo B, Spaziano M, Alali J, Mylote D, Theriault-Lauzier P, Alfagih R, et al. Three-dimensional echocardiography vs. computed tomography for transcatheter aortic valve replacement sizing. Eur Heart J Cardiovasc Imaging. 2016;17:15–23.PubMed Vaquerizo B, Spaziano M, Alali J, Mylote D, Theriault-Lauzier P, Alfagih R, et al. Three-dimensional echocardiography vs. computed tomography for transcatheter aortic valve replacement sizing. Eur Heart J Cardiovasc Imaging. 2016;17:15–23.PubMed
7.
go back to reference Wystub N, Baz L, Mobius-Winkler S, Porner TC, Goebel B, Hamadanchi A, et al. Aortic annulus measurement with computed tomography angiography reduces aortic regurgitation after transfemoral aortic valve replacement compared to 3-D echocardiography: a single-centre experience. Clin Res Cardiol. 2019;108:1266–75.CrossRef Wystub N, Baz L, Mobius-Winkler S, Porner TC, Goebel B, Hamadanchi A, et al. Aortic annulus measurement with computed tomography angiography reduces aortic regurgitation after transfemoral aortic valve replacement compared to 3-D echocardiography: a single-centre experience. Clin Res Cardiol. 2019;108:1266–75.CrossRef
8.
go back to reference Abdelghani M, Cavalcante R, Miyazaki Y, de Winter RJ, Tijssen JG, Sarmento-Leite R, et al. Transcatheter aortic valve implantation for mixed versus pure stenotic aortic valve disease. EuroIntervention. 2017;13:1157–65.CrossRef Abdelghani M, Cavalcante R, Miyazaki Y, de Winter RJ, Tijssen JG, Sarmento-Leite R, et al. Transcatheter aortic valve implantation for mixed versus pure stenotic aortic valve disease. EuroIntervention. 2017;13:1157–65.CrossRef
9.
go back to reference Schafer U, Schirmer J, Niklas S, Harmel E, Deuschl F, Conradi L. First-in-human implantation of a novel transfemoral selfexpanding transcatheter heart valve to treat pure aortic regurgitation. EuroIntervention. 2017;13:1296–9.CrossRef Schafer U, Schirmer J, Niklas S, Harmel E, Deuschl F, Conradi L. First-in-human implantation of a novel transfemoral selfexpanding transcatheter heart valve to treat pure aortic regurgitation. EuroIntervention. 2017;13:1296–9.CrossRef
10.
go back to reference Wernly B, Eder S, Navarese EP, Kretzschmar D, Franz M, Alushi B, et al. Transcatheter aortic valve replacement for pure aortic valve regurgitation: “on-label” versus “off-label” use of TAVR devices. Clin Res Cardiol. 2019;108:921–30.CrossRef Wernly B, Eder S, Navarese EP, Kretzschmar D, Franz M, Alushi B, et al. Transcatheter aortic valve replacement for pure aortic valve regurgitation: “on-label” versus “off-label” use of TAVR devices. Clin Res Cardiol. 2019;108:921–30.CrossRef
11.
go back to reference Tsuneyoshi H, Komiya T, Shimamoto T. Accuracy of aortic annulus diameter measurement: comparison of multi-detector CT, two- and three-dimensional echocardiography. J Card Surg. 2016;31:18–22.CrossRef Tsuneyoshi H, Komiya T, Shimamoto T. Accuracy of aortic annulus diameter measurement: comparison of multi-detector CT, two- and three-dimensional echocardiography. J Card Surg. 2016;31:18–22.CrossRef
12.
go back to reference Genereux P, Kodali SK, Green P, Paradis JM, Daneault B, Rene G, et al. Incidence and effect of acute kidney injury after transcatheter aortic valve replacement using the new valve academic research consortium criteria. Am J Cardiol. 2013;111(1):100–5.CrossRef Genereux P, Kodali SK, Green P, Paradis JM, Daneault B, Rene G, et al. Incidence and effect of acute kidney injury after transcatheter aortic valve replacement using the new valve academic research consortium criteria. Am J Cardiol. 2013;111(1):100–5.CrossRef
13.
go back to reference Willson AB, Webb JG, Freeman M, Wood DA, Gurvitch R, Thompson CR, et al. Computed tomography-based sizing recommendations for transcatheter aortic valve replacement with balloon-expandable valves: comparison with transesophageal echocardiography and rationale for implementation in a prospective trial. J Cardiovasc Comput Tomogr. 2012;6(6):406–14.CrossRef Willson AB, Webb JG, Freeman M, Wood DA, Gurvitch R, Thompson CR, et al. Computed tomography-based sizing recommendations for transcatheter aortic valve replacement with balloon-expandable valves: comparison with transesophageal echocardiography and rationale for implementation in a prospective trial. J Cardiovasc Comput Tomogr. 2012;6(6):406–14.CrossRef
14.
go back to reference Ng AC, Delgado V, van der Kley F, Shanks M, van de Veire NR, Bertini M, et al. Comparison of aortic root dimensions and geometries before and after transcatheter aortic valve implantation by 2- and 3-dimensional transesophageal echocardiography and multislice computed tomography. Circ Cardiovasc Imaging. 2010;3:94–102.CrossRef Ng AC, Delgado V, van der Kley F, Shanks M, van de Veire NR, Bertini M, et al. Comparison of aortic root dimensions and geometries before and after transcatheter aortic valve implantation by 2- and 3-dimensional transesophageal echocardiography and multislice computed tomography. Circ Cardiovasc Imaging. 2010;3:94–102.CrossRef
15.
go back to reference Khalique OK, Kodali SK, Paradis JM, Nazif TM, Williams MR, Einstein AJ, et al. Aortic annular sizing using a novel 3-dimensional echocardiographic method: use and comparison with cardiac computed tomography. Circ Cardiovasc Imaging. 2014;7:155–63.CrossRef Khalique OK, Kodali SK, Paradis JM, Nazif TM, Williams MR, Einstein AJ, et al. Aortic annular sizing using a novel 3-dimensional echocardiographic method: use and comparison with cardiac computed tomography. Circ Cardiovasc Imaging. 2014;7:155–63.CrossRef
16.
go back to reference Mediratta A, Addetia K, Medvedofsky D, Schneider RJ, Kruse E, Shah AP, et al. 3D echocardiographic analysis of aortic annulus for transcatheter aortic valve replacement using novel aortic valve quantification software: comparison with computed tomography. Echocardiography (Mount Kisco, NY). 2017;34:690–9.CrossRef Mediratta A, Addetia K, Medvedofsky D, Schneider RJ, Kruse E, Shah AP, et al. 3D echocardiographic analysis of aortic annulus for transcatheter aortic valve replacement using novel aortic valve quantification software: comparison with computed tomography. Echocardiography (Mount Kisco, NY). 2017;34:690–9.CrossRef
17.
go back to reference Prihadi EA, van Rosendael PJ, Vollema EM, Bax JJ, Delgado V, Ajmone MN. Feasibility, accuracy, and reproducibility of aortic annular and root sizing for transcatheter aortic valve replacement using novel automated three-dimensional echocardiographic software: comparison with multi-detector row computed tomography. J Am Soc Echocardiog. 2018;31:505-514.e3.CrossRef Prihadi EA, van Rosendael PJ, Vollema EM, Bax JJ, Delgado V, Ajmone MN. Feasibility, accuracy, and reproducibility of aortic annular and root sizing for transcatheter aortic valve replacement using novel automated three-dimensional echocardiographic software: comparison with multi-detector row computed tomography. J Am Soc Echocardiog. 2018;31:505-514.e3.CrossRef
18.
go back to reference Kato N, Shibayama K, Noguchi M, Makihara Y, Okumura H, Obunai K, et al. Superiority of novel automated assessment of aortic annulus by intraoperative three-dimensional transesophageal echocardiography in patients with severe aortic stenosis: Comparison with conventional cross-sectional assessment. J Cardiol. 2018;72:321–7.CrossRef Kato N, Shibayama K, Noguchi M, Makihara Y, Okumura H, Obunai K, et al. Superiority of novel automated assessment of aortic annulus by intraoperative three-dimensional transesophageal echocardiography in patients with severe aortic stenosis: Comparison with conventional cross-sectional assessment. J Cardiol. 2018;72:321–7.CrossRef
19.
go back to reference Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Baron-Esquivias G, Baumgartner H, et al. Guidelines on the management of valvular heart disease (version 2012): the Joint Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur J Cardio-Thorac. 2012;42:S1-44.CrossRef Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Baron-Esquivias G, Baumgartner H, et al. Guidelines on the management of valvular heart disease (version 2012): the Joint Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur J Cardio-Thorac. 2012;42:S1-44.CrossRef
20.
go back to reference Grube E, Laborde JC, Gerckens U, Felderhoff T, Sauren B, Buellesfeld L, et al. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study. Circulation. 2006;114:1616–24.CrossRef Grube E, Laborde JC, Gerckens U, Felderhoff T, Sauren B, Buellesfeld L, et al. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study. Circulation. 2006;114:1616–24.CrossRef
21.
go back to reference Pasupati S, Devlin G, Davis M, Kejriwal N, Jogia P, Liew TV, et al. Transcatheter solution for pure aortic insufficiency. JACC Cardiovasc Imaging. 2014;7:315–8.CrossRef Pasupati S, Devlin G, Davis M, Kejriwal N, Jogia P, Liew TV, et al. Transcatheter solution for pure aortic insufficiency. JACC Cardiovasc Imaging. 2014;7:315–8.CrossRef
22.
go back to reference Seiffert M, Bader R, Kappert U, Rastan A, Krapf S, Bleiziffer S, et al. Initial German experience with transapical implantation of a second-generation transcatheter heart valve for the treatment of aortic regurgitation. JACC Cardiovasc Interv. 2014;7:1168–74.CrossRef Seiffert M, Bader R, Kappert U, Rastan A, Krapf S, Bleiziffer S, et al. Initial German experience with transapical implantation of a second-generation transcatheter heart valve for the treatment of aortic regurgitation. JACC Cardiovasc Interv. 2014;7:1168–74.CrossRef
23.
go back to reference Wei L, Liu H, Zhu L, Yang Y, Zheng J, Guo K, et al. A new transcatheter aortic valve replacement system for predominant aortic regurgitation implantation of the J-Valve and early outcome. JACC Cardiovasc Interv. 2015;8:1831–41.CrossRef Wei L, Liu H, Zhu L, Yang Y, Zheng J, Guo K, et al. A new transcatheter aortic valve replacement system for predominant aortic regurgitation implantation of the J-Valve and early outcome. JACC Cardiovasc Interv. 2015;8:1831–41.CrossRef
24.
go back to reference Wendt D, Kahlert P, Pasa S, El-Chilali K, Al-Rashid F, Tsagakis K, et al. Transapical transcatheter aortic valve for severe aortic regurgitation: expanding the limits. JACC Cardiovasc Interv. 2014;7:1159–67.CrossRef Wendt D, Kahlert P, Pasa S, El-Chilali K, Al-Rashid F, Tsagakis K, et al. Transapical transcatheter aortic valve for severe aortic regurgitation: expanding the limits. JACC Cardiovasc Interv. 2014;7:1159–67.CrossRef
25.
go back to reference Cerillo AG, Griese D, Berti S. Successful percutaneous implantation of symetis ACURATE neo transcatheter aortic bioprosthesis for the treatment of pure aortic regurgitation. Catheter Cardiovasc Interv. 2016;88:319–23.CrossRef Cerillo AG, Griese D, Berti S. Successful percutaneous implantation of symetis ACURATE neo transcatheter aortic bioprosthesis for the treatment of pure aortic regurgitation. Catheter Cardiovasc Interv. 2016;88:319–23.CrossRef
Metadata
Title
Application of three-dimensional transesophageal echocardiography in preoperative evaluation of transcatheter aortic valve replacement
Authors
Peng Ding
Chennian Xu
Yang Liu
Xin Meng
Ping Jin
Jiayou Tang
Lanlan Li
Yanyan Ma
Jian Yang
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Cardiovascular Disorders / Issue 1/2021
Electronic ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-021-02101-7

Other articles of this Issue 1/2021

BMC Cardiovascular Disorders 1/2021 Go to the issue