Published in: Radiation Oncology 1/2014

Open Access 01-12-2014 | Research

The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT

Authors: Jung-in Kim, So-Yeon Park, Hak Jae Kim, Jin Ho Kim, Sung-Joon Ye, Jong Min Park

Published in: Radiation Oncology | Issue 1/2014

Login to get access

Abstract

Background

To investigate the sensitivity of various gamma criteria used in the gamma-index method for patient-specific volumetric modulated arc therapy (VMAT) quality assurance (QA) for stereotactic body radiation therapy (SBRT) using a flattening filter free (FFF) photon beam.

Methods

Three types of intentional misalignments were introduced to original high-definition multi-leaf collimator (HD-MLC) plans. The first type, referred to Class Out, involved the opening of each bank of leaves. The second type, Class In, involved the closing of each bank of leaves. The third type, Class Shift, involved the shifting of each bank of leaves towards the ground. Patient-specific QAs for the original and the modified plans were performed with MapCHECK2 and EBT2 films. The sensitivity of the gamma-index method using criteria of 1%/1 mm, 1.5%/1.5 mm, 1%/2 mm, 2%/1 mm and 2%/2 mm was investigated with absolute passing rates according to the magnitudes of MLCs misalignments. In addition, the changes in dose-volumetric indicators due to the magnitudes of MLC misalignments were investigated. The correlations between passing rates and the changes in dose-volumetric indicators were also investigated using Spearman’s rank correlation coefficient (γ).

Results

The criterion of 2%/1 mm was able to detect Class Out and Class In MLC misalignments of 0.5 mm and Class Shift misalignments of 1 mm. The widely adopted clinical criterion of 2%/2 mm was not able to detect 0.5 mm MLC errors of the Class Out or Class In types, and also unable to detect 3 mm Class Shift errors. No correlations were observed between dose-volumetric changes and gamma passing rates (γ < 0.8).

Conclusions

Gamma criterion of 2%/1 mm was found to be suitable as a tolerance level with passing rates of 90% and 80% for patient-specific VMAT QA for SBRT when using MapCHECK2 and EBT2 film, respectively.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX: Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys 2003, 30: 2089-2115. 10.1118/1.1591194CrossRefPubMed Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, Xia P, Xiao Y, Xing L, Yu CX: Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys 2003, 30: 2089-2115. 10.1118/1.1591194CrossRefPubMed
2.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed
3.
go back to reference Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y: IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36: 5359-5373. 10.1118/1.3238104CrossRefPubMed Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y: IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36: 5359-5373. 10.1118/1.3238104CrossRefPubMed
4.
go back to reference Both S, Alecu IM, Stan AR, Alecu M, Ciura A, Hansen JM, Alecu R: A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy. J Appl Clinl Med Phys 2007, 8: 1-8. Both S, Alecu IM, Stan AR, Alecu M, Ciura A, Hansen JM, Alecu R: A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy. J Appl Clinl Med Phys 2007, 8: 1-8.
5.
go back to reference Davidson MT, Blake SJ, Batchelar DL, Cheung P, Mah K: Assessing the role of volumetric modulated arc therapy (VMAT) relative to IMRT and helical tomotherapy in the management of localized, locally advanced, and post-operative prostate cancer. Int J Radiat Oncol Biol Phys 2011, 80: 1550-1558. 10.1016/j.ijrobp.2010.10.024CrossRefPubMed Davidson MT, Blake SJ, Batchelar DL, Cheung P, Mah K: Assessing the role of volumetric modulated arc therapy (VMAT) relative to IMRT and helical tomotherapy in the management of localized, locally advanced, and post-operative prostate cancer. Int J Radiat Oncol Biol Phys 2011, 80: 1550-1558. 10.1016/j.ijrobp.2010.10.024CrossRefPubMed
6.
go back to reference Quan EM, Li X, Li Y, Wang X, Kudchadker RJ, Johnson JL, Kuban DA, Lee AK, Zhang X: A comprehensive comparison of IMRT and VMAT plan quality for prostate cancer treatment. Int J Radiat Oncol Biol Phys 2012, 83: 1169-1178. 10.1016/j.ijrobp.2011.09.015PubMedCentralCrossRefPubMed Quan EM, Li X, Li Y, Wang X, Kudchadker RJ, Johnson JL, Kuban DA, Lee AK, Zhang X: A comprehensive comparison of IMRT and VMAT plan quality for prostate cancer treatment. Int J Radiat Oncol Biol Phys 2012, 83: 1169-1178. 10.1016/j.ijrobp.2011.09.015PubMedCentralCrossRefPubMed
7.
go back to reference Sung W, Park JM, Choi CH: The effect of photon energy on intensity-modulated radiation therapy (IMRT) plans for prostate cancer. Radiat Oncol J 2012, 30: 27-35. 10.3857/roj.2012.30.1.27PubMedCentralCrossRefPubMed Sung W, Park JM, Choi CH: The effect of photon energy on intensity-modulated radiation therapy (IMRT) plans for prostate cancer. Radiat Oncol J 2012, 30: 27-35. 10.3857/roj.2012.30.1.27PubMedCentralCrossRefPubMed
8.
go back to reference Joo YK, Sung YP, Doo HL: Comparison of treatment plans with multileaf collimators of different leaf width. K J Med Phys 2004, 15: 173-178. Joo YK, Sung YP, Doo HL: Comparison of treatment plans with multileaf collimators of different leaf width. K J Med Phys 2004, 15: 173-178.
9.
go back to reference Wagner D, Vorwerk H: Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array. Radiat oncology 2011, 6: 21. 10.1186/1748-717X-6-21CrossRef Wagner D, Vorwerk H: Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array. Radiat oncology 2011, 6: 21. 10.1186/1748-717X-6-21CrossRef
10.
go back to reference Nelms BE, Zhen H, Tome WA: Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys 2011, 38: 1037-1044. 10.1118/1.3544657PubMedCentralCrossRefPubMed Nelms BE, Zhen H, Tome WA: Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys 2011, 38: 1037-1044. 10.1118/1.3544657PubMedCentralCrossRefPubMed
11.
go back to reference Zhen H, Nelms BE, Tome WA: Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. Med Phys 2011, 38: 5477-5489. 10.1118/1.3633904CrossRefPubMed Zhen H, Nelms BE, Tome WA: Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. Med Phys 2011, 38: 5477-5489. 10.1118/1.3633904CrossRefPubMed
12.
go back to reference Yan G, Liu C, Simon TA, Peng LC, Fox C, Li JG: On the sensitivity of patient-specific IMRT QA to MLC positioning errors. J Appl Clinl Med Phys 2009, 10: 2915. Yan G, Liu C, Simon TA, Peng LC, Fox C, Li JG: On the sensitivity of patient-specific IMRT QA to MLC positioning errors. J Appl Clinl Med Phys 2009, 10: 2915.
13.
go back to reference Heilemann G, Poppe B, Laub W: On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys 2013, 40: 031702. 10.1118/1.4789580CrossRefPubMed Heilemann G, Poppe B, Laub W: On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys 2013, 40: 031702. 10.1118/1.4789580CrossRefPubMed
14.
go back to reference Fredh A, Scherman JB, Fog LS, Munck af Rosenschold P: Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys 2013, 40: 031716. 10.1118/1.4788645CrossRefPubMed Fredh A, Scherman JB, Fog LS, Munck af Rosenschold P: Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys 2013, 40: 031716. 10.1118/1.4788645CrossRefPubMed
15.
go back to reference Oliver M, Gagne I, Bush K, Zavgorodni S, Ansbacher W, Beckham W: Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol 2010, 97: 554-560. 10.1016/j.radonc.2010.06.013CrossRefPubMed Oliver M, Gagne I, Bush K, Zavgorodni S, Ansbacher W, Beckham W: Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapy. Radiother Oncol 2010, 97: 554-560. 10.1016/j.radonc.2010.06.013CrossRefPubMed
16.
go back to reference Kathirvel M, Subramanian S, Clivio A, Arun G, Fogliata A, Nicolini G, Subramani V, Thirumalai Swamy S, Vanetti E, Thirumalai Swamy S, Vanetti E, Cozzi L: Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans. Radiat oncology 2013, 8: 140. 10.1186/1748-717X-8-140CrossRef Kathirvel M, Subramanian S, Clivio A, Arun G, Fogliata A, Nicolini G, Subramani V, Thirumalai Swamy S, Vanetti E, Thirumalai Swamy S, Vanetti E, Cozzi L: Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans. Radiat oncology 2013, 8: 140. 10.1186/1748-717X-8-140CrossRef
17.
go back to reference LoSasso T, Chui CS, Ling CC: Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys 1998, 25: 1919-1927. 10.1118/1.598381CrossRefPubMed LoSasso T, Chui CS, Ling CC: Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys 1998, 25: 1919-1927. 10.1118/1.598381CrossRefPubMed
18.
go back to reference Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, Keall P, Lovelock M, Meeks S, Papiez L, Purdie T, Sadagopan R, Schell MC, Salter B, Schlesinger DJ, Shiu AS, Solberg T, Song DY, Stieber V, Timmerman R, Tomé WA, Verellen D, Wang L, Yin FF: Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys 2010, 37: 4078-4101. 10.1118/1.3438081CrossRefPubMed Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, Keall P, Lovelock M, Meeks S, Papiez L, Purdie T, Sadagopan R, Schell MC, Salter B, Schlesinger DJ, Shiu AS, Solberg T, Song DY, Stieber V, Timmerman R, Tomé WA, Verellen D, Wang L, Yin FF: Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys 2010, 37: 4078-4101. 10.1118/1.3438081CrossRefPubMed
19.
go back to reference Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, Rogers DW: AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999, 26: 1847-1870. 10.1118/1.598691CrossRefPubMed Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, Rogers DW: AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999, 26: 1847-1870. 10.1118/1.598691CrossRefPubMed
20.
go back to reference Iftimia I, Cirino ET, Xiong L, Mower HW: Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clinl Med Phys 2010, 11: 3164. Iftimia I, Cirino ET, Xiong L, Mower HW: Quality assurance methodology for Varian RapidArc treatment plans. J Appl Clinl Med Phys 2010, 11: 3164.
21.
go back to reference Gloi AM, Buchana RE, Zuge CL, Goettler AM: RapidArc quality assurance through MapCHECK. J Appl Clinl Med Phys 2011, 12: 3251. Gloi AM, Buchana RE, Zuge CL, Goettler AM: RapidArc quality assurance through MapCHECK. J Appl Clinl Med Phys 2011, 12: 3251.
22.
go back to reference Li H, Dong L, Zhang L, Yang JN, Gillin MT, Zhu XR: Toward a better understanding of the gamma index: Investigation of parameters with a surface-based distance method. Med Phys 2011, 38: 6730-6741. 10.1118/1.3659707PubMedCentralCrossRefPubMed Li H, Dong L, Zhang L, Yang JN, Gillin MT, Zhu XR: Toward a better understanding of the gamma index: Investigation of parameters with a surface-based distance method. Med Phys 2011, 38: 6730-6741. 10.1118/1.3659707PubMedCentralCrossRefPubMed
23.
go back to reference Timmerman R, Heinzerling J, Abdulrahman R, Choy H, Meyer JL: Stereotactic body radiation therapy for thoracic cancers: recommendations for patient selection, setup and therapy. Front Radiat Ther Oncol 2011, 43: 395-411.CrossRefPubMed Timmerman R, Heinzerling J, Abdulrahman R, Choy H, Meyer JL: Stereotactic body radiation therapy for thoracic cancers: recommendations for patient selection, setup and therapy. Front Radiat Ther Oncol 2011, 43: 395-411.CrossRefPubMed
24.
go back to reference Poppe B, Blechschmidt A, Djouguela A, Kollhoff R, Rubach A, Willborn KC, Harder D: Two-dimensional ionization chamber arrays for IMRT plan verification. Med Phys 2006, 33: 1005-1015. 10.1118/1.2179167CrossRefPubMed Poppe B, Blechschmidt A, Djouguela A, Kollhoff R, Rubach A, Willborn KC, Harder D: Two-dimensional ionization chamber arrays for IMRT plan verification. Med Phys 2006, 33: 1005-1015. 10.1118/1.2179167CrossRefPubMed
25.
go back to reference Reinhardt S, Hillbrand M, Wilkens JJ, Assmann W: Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. Med Phys 2012, 39: 5257-5262. 10.1118/1.4737890CrossRefPubMed Reinhardt S, Hillbrand M, Wilkens JJ, Assmann W: Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. Med Phys 2012, 39: 5257-5262. 10.1118/1.4737890CrossRefPubMed
26.
go back to reference Li JG, Yan G, Liu C: Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clinl Med Phys 2009, 10: 2942. Li JG, Yan G, Liu C: Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clinl Med Phys 2009, 10: 2942.
27.
go back to reference Kruse JJ: On the insensitivity of single field planar dosimetry to IMRT inaccuracies. Med Phys 2010, 37: 2516-2524. 10.1118/1.3425781CrossRefPubMed Kruse JJ: On the insensitivity of single field planar dosimetry to IMRT inaccuracies. Med Phys 2010, 37: 2516-2524. 10.1118/1.3425781CrossRefPubMed
28.
go back to reference Rangel A, Palte G, Dunscombe P: The sensitivity of patient specific IMRT QC to systematic MLC leaf bank offset errors. Med Phys 2010, 37: 3862-3867. 10.1118/1.3453576CrossRefPubMed Rangel A, Palte G, Dunscombe P: The sensitivity of patient specific IMRT QC to systematic MLC leaf bank offset errors. Med Phys 2010, 37: 3862-3867. 10.1118/1.3453576CrossRefPubMed
Metadata
Title
The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT
Authors
Jung-in Kim
So-Yeon Park
Hak Jae Kim
Jin Ho Kim
Sung-Joon Ye
Jong Min Park
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2014
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-9-167

Other articles of this Issue 1/2014

Radiation Oncology 1/2014 Go to the issue