Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans

Authors: Murugesan Kathirvel, Shanmuga Subramanian, Alessandro Clivio, Gandhi Arun, Antonella Fogliata, Giorgia Nicolini, Vellaiyan Subramani, Shanmugam Thirumalai Swamy, Eugenio Vanetti, Luca Cozzi

Published in: Radiation Oncology | Issue 1/2013

Login to get access

Abstract

Background

The accuracy of the two dose calculation engines available for RapidArc planning (both released for clinical use) is investigated in comparison to the COMPASS data.

Methods

Two dose calculation algorithms (Acuros-XB and Anisotropic Analytic Algorithm (AAA)) were used to calculate RA plans and compared to calculations with the Collapsed Cone Convolution algorithm (CC) from the COMPASS system (IBA Dosimetry). CC calculations, performed on patient data, are based on experimental fluence measurements with a 2D array of ion chambers mounted on the linac head. The study was conducted on clinical cases treated with RA. Five cases for each of the following groups were included: Brain, Head and Neck, Thorax, Pelvis and stereotactic body radiation therapy for hypo-fractionated treatments with small fields. COMPASS measurements were performed with the iMatrixx-2D array. RapidArc plans were optimized for delivery using 6MV photons from a Clinac-iX (Varian, Palo Alto, USA).
Accuracy of the RA calculation was appraised by means of: 1) comparison of Dose Volume histograms (DVH) metrics; 2) analysis of differential dose distributions and determination of mean dose differences per organ; 3) 3D gamma analysis with distance-to-agreement and dose difference thresholds set to 3%/3 mm or 2%/2 mm for targets, organs at risks and for the volumes encompassed by the 50 and 10% isodoses.

Results

For almost all parameters, the better agreement was between Acuros-XB and COMPASS independently from the anatomical site and fractionation. The same result was obtained from the mean dose difference per organ with Acuros-CC average differences below 0.5% while for AAA-CC data, average deviations exceeded 0.5% and in the case of the pelvis 1%. Relevance of observed differences determined with the 3D gamma analysis resulted in a pass rate exceeding 99.5% for Acuros-CC and exceeding 97.5% for AAA-CC.

Conclusions

This study demonstrated that i) a good agreement exists between COMPASS-CC calculations based on measured fluences with respect to dose distributions obtained with both Acuros-XB and AAA algorithms; ii) 3D dose distributions reconstructed from actual delivery coincide very precisely with the planned data; iii) a slight preference in favor of Acuros-XB was observed suggesting the preference for this algorithm in clinical applications.
Appendix
Available only for authorised users
Literature
1.
go back to reference Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35: 310-317. 10.1118/1.2818738CrossRefPubMed Otto K: Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35: 310-317. 10.1118/1.2818738CrossRefPubMed
2.
go back to reference Korreman S, Medin J, Kjaer-Kristoffersen F: Dosimetric verification of RapidArc treatment delivery. Acta Oncol 2009, 48: 185-191. 10.1080/02841860802287116CrossRefPubMed Korreman S, Medin J, Kjaer-Kristoffersen F: Dosimetric verification of RapidArc treatment delivery. Acta Oncol 2009, 48: 185-191. 10.1080/02841860802287116CrossRefPubMed
3.
go back to reference Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J, Cozzi L: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008, 3: 24. 10.1186/1748-717X-3-24CrossRefPubMedPubMedCentral Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J, Cozzi L: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008, 3: 24. 10.1186/1748-717X-3-24CrossRefPubMedPubMedCentral
4.
go back to reference Schreibmann E, Dhabaan A, Elder E, Fox T: Patient specific quality assurance method for VMAT treatment delivery. Med Phys 2009, 36: 4530-4535. 10.1118/1.3213085CrossRefPubMed Schreibmann E, Dhabaan A, Elder E, Fox T: Patient specific quality assurance method for VMAT treatment delivery. Med Phys 2009, 36: 4530-4535. 10.1118/1.3213085CrossRefPubMed
5.
go back to reference Teke T, Bergman A, Kwa W, Gill B, Duzenli C, Popescu A: Monte Carlo based patient specific RapidArc QA using Linac log files. Med Phys 2010, 37: 116-123. 10.1118/1.3266821CrossRefPubMed Teke T, Bergman A, Kwa W, Gill B, Duzenli C, Popescu A: Monte Carlo based patient specific RapidArc QA using Linac log files. Med Phys 2010, 37: 116-123. 10.1118/1.3266821CrossRefPubMed
6.
go back to reference Qian J, Lee L, Liu W, Chu K, Mok E, Luxton G, Le Q, Xing L: Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone beam CT and dynamic log files. Phys Med Biol 2010, 55: 3597-3610. 10.1088/0031-9155/55/13/002CrossRefPubMed Qian J, Lee L, Liu W, Chu K, Mok E, Luxton G, Le Q, Xing L: Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone beam CT and dynamic log files. Phys Med Biol 2010, 55: 3597-3610. 10.1088/0031-9155/55/13/002CrossRefPubMed
7.
go back to reference Han Z, Ng S, Bhagwat M, Lyatskaya Y, Zygmanski P: Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions. Med Phys 2010, 37: 3704-3714. 10.1118/1.3455707CrossRefPubMed Han Z, Ng S, Bhagwat M, Lyatskaya Y, Zygmanski P: Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions. Med Phys 2010, 37: 3704-3714. 10.1118/1.3455707CrossRefPubMed
8.
go back to reference Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe S, Papanikolau N: Consistency and reproducibility of the VMAT plan delivery using three independent validation methods. J Appl Clin Med Phys 2010, 12: 3373.PubMed Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe S, Papanikolau N: Consistency and reproducibility of the VMAT plan delivery using three independent validation methods. J Appl Clin Med Phys 2010, 12: 3373.PubMed
9.
go back to reference Bakhtiari M, Kumaraswamy L, Bailey D, de Boer S, Malhotra H, Podgorsak M: Using an EPID for patient specific VMAT quality assurance. Med Phys 2011, 38: 1366-1373. 10.1118/1.3552925CrossRefPubMed Bakhtiari M, Kumaraswamy L, Bailey D, de Boer S, Malhotra H, Podgorsak M: Using an EPID for patient specific VMAT quality assurance. Med Phys 2011, 38: 1366-1373. 10.1118/1.3552925CrossRefPubMed
10.
go back to reference Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe S, Papanikolau N: Comparison of four commercial devices for RapidArc and sliding window IMRT QA. J Appl Clin Med Phys 2011, 12: 3367.PubMed Chandraraj V, Stathakis S, Manickam R, Esquivel C, Supe S, Papanikolau N: Comparison of four commercial devices for RapidArc and sliding window IMRT QA. J Appl Clin Med Phys 2011, 12: 3367.PubMed
11.
go back to reference Gloi A, Buchana R, Zuge C, Goettler A: RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys 2011, 12: 3251.PubMed Gloi A, Buchana R, Zuge C, Goettler A: RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys 2011, 12: 3251.PubMed
12.
go back to reference Syamkumar SA, Padmanabhan S, Sukumar P, Nagarajan V: Characterization of responses of 2d array seven29 detector and its combined use with Octavius phantom for the patient-specific quality assurance in RapidArc treatment delivery. Med Dosim 2011, 37: 53-60.CrossRefPubMed Syamkumar SA, Padmanabhan S, Sukumar P, Nagarajan V: Characterization of responses of 2d array seven29 detector and its combined use with Octavius phantom for the patient-specific quality assurance in RapidArc treatment delivery. Med Dosim 2011, 37: 53-60.CrossRefPubMed
13.
go back to reference Fakir H, Gaede S, Mulligan M, Chen J: Development of a novel ArcCheck insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities. Med Phys 2012, 39: 4203-4208. 10.1118/1.4728222CrossRefPubMed Fakir H, Gaede S, Mulligan M, Chen J: Development of a novel ArcCheck insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities. Med Phys 2012, 39: 4203-4208. 10.1118/1.4728222CrossRefPubMed
14.
go back to reference Bush K, Townson R, Zavgorodni S: Monte Carlo simulation of RapidArc radiotherapy delivery. Phys Med Biol 2008, 53: N359-N371. 10.1088/0031-9155/53/19/N01CrossRefPubMed Bush K, Townson R, Zavgorodni S: Monte Carlo simulation of RapidArc radiotherapy delivery. Phys Med Biol 2008, 53: N359-N371. 10.1088/0031-9155/53/19/N01CrossRefPubMed
15.
go back to reference Gagne I, Ansbacher W, Zavgorodni S, Popescu C, Beckham W: A Monte Carlo evaluation of RapidArc dose calculations for oropharynx radiotherapy. Phys Med Biol 2008, 53: 7167-7185. 10.1088/0031-9155/53/24/011CrossRefPubMed Gagne I, Ansbacher W, Zavgorodni S, Popescu C, Beckham W: A Monte Carlo evaluation of RapidArc dose calculations for oropharynx radiotherapy. Phys Med Biol 2008, 53: 7167-7185. 10.1088/0031-9155/53/24/011CrossRefPubMed
16.
go back to reference van Elmpt W, Petit S, De Ruysscher D, Lambin P, Dekker A: 3D dose delivery verification using repeated cone beam imaging and EPID dosimetry for stereotactic body radiotherapy of non small cell lung cancer. Radiother Oncol 2010, 94: 188-194. 10.1016/j.radonc.2009.12.024CrossRefPubMed van Elmpt W, Petit S, De Ruysscher D, Lambin P, Dekker A: 3D dose delivery verification using repeated cone beam imaging and EPID dosimetry for stereotactic body radiotherapy of non small cell lung cancer. Radiother Oncol 2010, 94: 188-194. 10.1016/j.radonc.2009.12.024CrossRefPubMed
17.
go back to reference Korevaar E, Wauben D, van der Hulst P, Lagendijk J, van’t Veld A: Clinical introduction of a linac head mounted 2D detector array based quality assurance system in head and neck IMRT. Radiother Oncol 2011, 100: 446-453. 10.1016/j.radonc.2011.09.007CrossRefPubMed Korevaar E, Wauben D, van der Hulst P, Lagendijk J, van’t Veld A: Clinical introduction of a linac head mounted 2D detector array based quality assurance system in head and neck IMRT. Radiother Oncol 2011, 100: 446-453. 10.1016/j.radonc.2011.09.007CrossRefPubMed
18.
go back to reference Godart J, Korevaar E, Visser R, Wauben D, van’t Veld A: Reconstruction of high resolution 3D dose from matrix measurements: error detection capability of the COMPASS correction kernel method. Phys Med Biol 2011, 56: 5029-5043. 10.1088/0031-9155/56/15/023CrossRefPubMed Godart J, Korevaar E, Visser R, Wauben D, van’t Veld A: Reconstruction of high resolution 3D dose from matrix measurements: error detection capability of the COMPASS correction kernel method. Phys Med Biol 2011, 56: 5029-5043. 10.1088/0031-9155/56/15/023CrossRefPubMed
19.
go back to reference Nakaguchi Y, Araki F, Maruyama M, Saiga S: Dose verification of IMRT by use of a COMPASS transmission detector. Radiol Phys Technol 2012, 5: 63-70. 10.1007/s12194-011-0137-yCrossRefPubMed Nakaguchi Y, Araki F, Maruyama M, Saiga S: Dose verification of IMRT by use of a COMPASS transmission detector. Radiol Phys Technol 2012, 5: 63-70. 10.1007/s12194-011-0137-yCrossRefPubMed
20.
go back to reference Boggula R, Jahnke L, Wertz H, Lohr F, Wenz F: Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans. Int J Radiat Oncol Biol Phys 2011, 81: 1168-1175. 10.1016/j.ijrobp.2010.09.010CrossRefPubMed Boggula R, Jahnke L, Wertz H, Lohr F, Wenz F: Patient-specific 3D pretreatment and potential 3D online dose verification of Monte Carlo-calculated IMRT prostate treatment plans. Int J Radiat Oncol Biol Phys 2011, 81: 1168-1175. 10.1016/j.ijrobp.2010.09.010CrossRefPubMed
21.
go back to reference Boggula R, Lorenz F, Müller L, Birkner M, Wertz H, Stieler F, Steil V, Lohr F, Wenz F: Experimental validation of a commercial 3D dose verification system for intensity-modulated arc therapies. Phys Med Biol 2010, 55: 5619-5633. 10.1088/0031-9155/55/19/001CrossRefPubMed Boggula R, Lorenz F, Müller L, Birkner M, Wertz H, Stieler F, Steil V, Lohr F, Wenz F: Experimental validation of a commercial 3D dose verification system for intensity-modulated arc therapies. Phys Med Biol 2010, 55: 5619-5633. 10.1088/0031-9155/55/19/001CrossRefPubMed
22.
go back to reference Vanetti E, Nicolini G, Nord J, Peltola J, Clivio A, Fogliata A, Cozzi L: On the role of the optimization algorithm of RapidArc(®) volumetric modulated arc therapy on plan quality and efficiency. Med Phys 2011, 38: 5844. 10.1118/1.3641866CrossRefPubMed Vanetti E, Nicolini G, Nord J, Peltola J, Clivio A, Fogliata A, Cozzi L: On the role of the optimization algorithm of RapidArc(®) volumetric modulated arc therapy on plan quality and efficiency. Med Phys 2011, 38: 5844. 10.1118/1.3641866CrossRefPubMed
23.
go back to reference Ulmer W, Pyyry J, Kaissl W: A 3D photon superposition convolution algorithm and its foundation on results of Monte Carlo calculations. Phys Med Biol 2005, 50: 1767-1790. 10.1088/0031-9155/50/8/010CrossRefPubMed Ulmer W, Pyyry J, Kaissl W: A 3D photon superposition convolution algorithm and its foundation on results of Monte Carlo calculations. Phys Med Biol 2005, 50: 1767-1790. 10.1088/0031-9155/50/8/010CrossRefPubMed
24.
go back to reference Vassiliev O, Wareing T, McGhee J, Failla G, Salehpour M, Mourtada F: Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beams. Phys Med Biol 2010, 55: 581-598. 10.1088/0031-9155/55/3/002CrossRefPubMed Vassiliev O, Wareing T, McGhee J, Failla G, Salehpour M, Mourtada F: Validation of a new grid-based Boltzmann equation solver for dose calculation in radiotherapy with photon beams. Phys Med Biol 2010, 55: 581-598. 10.1088/0031-9155/55/3/002CrossRefPubMed
25.
go back to reference ICRU report 83: Prescribing, recording and reporting Intensity Modulated Photon Beam Therapy (IMRT) (ICRU report 83). Washington, DC: International Commission on Radiation Units and Measurements; 2010. ICRU report 83: Prescribing, recording and reporting Intensity Modulated Photon Beam Therapy (IMRT) (ICRU report 83). Washington, DC: International Commission on Radiation Units and Measurements; 2010.
26.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA: A technique for quantitative evaluation of dose distributions. Med Phys 2008, 25: 656-661.CrossRef Low DA, Harms WB, Mutic S, Purdy JA: A technique for quantitative evaluation of dose distributions. Med Phys 2008, 25: 656-661.CrossRef
27.
go back to reference Fogliata A, Nicolini G, Vanetti E, Clivio A, Winkler P, Cozzi L: The impact of photon dose calculation algorithms on expected dose distributions in lungs under different respiratory phases. Phys Med Biol 2008, 53: 2375-2390. 10.1088/0031-9155/53/9/011CrossRefPubMed Fogliata A, Nicolini G, Vanetti E, Clivio A, Winkler P, Cozzi L: The impact of photon dose calculation algorithms on expected dose distributions in lungs under different respiratory phases. Phys Med Biol 2008, 53: 2375-2390. 10.1088/0031-9155/53/9/011CrossRefPubMed
28.
go back to reference Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment. Radiat Oncol 2011, 6: 103. 10.1186/1748-717X-6-103CrossRefPubMedPubMedCentral Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment. Radiat Oncol 2011, 6: 103. 10.1186/1748-717X-6-103CrossRefPubMedPubMedCentral
29.
go back to reference Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: Critical appraisal of Acuros XB and Anisotropic Analytical Algorithm dose calculation in advanced non-small cell lung cancer treatments. Int J Radiat Oncol Biol Phys 2012, 83: 1587-1595. 10.1016/j.ijrobp.2011.10.078CrossRefPubMed Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L: Critical appraisal of Acuros XB and Anisotropic Analytical Algorithm dose calculation in advanced non-small cell lung cancer treatments. Int J Radiat Oncol Biol Phys 2012, 83: 1587-1595. 10.1016/j.ijrobp.2011.10.078CrossRefPubMed
30.
go back to reference Kan M, Leung L, Yu P: Dosimetric impact of using the Acuros XB algorithm for intensity modulated radiation therapy and RapidArc planning in nasopharyngeal carcinomas. Int J Radiat Oncol Biol Phys 2012, 85: e73-e80.CrossRefPubMed Kan M, Leung L, Yu P: Dosimetric impact of using the Acuros XB algorithm for intensity modulated radiation therapy and RapidArc planning in nasopharyngeal carcinomas. Int J Radiat Oncol Biol Phys 2012, 85: e73-e80.CrossRefPubMed
Metadata
Title
Critical appraisal of the accuracy of Acuros-XB and Anisotropic Analytical Algorithm compared to measurement and calculations with the compass system in the delivery of RapidArc clinical plans
Authors
Murugesan Kathirvel
Shanmuga Subramanian
Alessandro Clivio
Gandhi Arun
Antonella Fogliata
Giorgia Nicolini
Vellaiyan Subramani
Shanmugam Thirumalai Swamy
Eugenio Vanetti
Luca Cozzi
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2013
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-8-140

Other articles of this Issue 1/2013

Radiation Oncology 1/2013 Go to the issue