Use of screw locking elements improves radiological and biomechanical results of femoral osteotomies

Authors: Gerardo L Garcés, Oliver Rodríguez, Enrique Rodríguez Grau-Bassas, Syra Roiz, Alejandro Yánez, Alberto Cuadrado

Published in: BMC Musculoskeletal Disorders | Issue 1/2014

Login to get access

Abstract

Background

Dynamic compression plate (DCP) constructs provide inadequate fixation in cases of poor bone quality and early weight-bearing. Screw locking elements (SLE) are flat locking nuts placed at the end of the screw to prevent screw stripping from the bone, improving fixation stability. The purpose of this work was to compare biomechanical and radiological evaluations of femoral ovine osteotomies fixed using DCP constructs with and without SLE.

Method

A dyaphyseal femoral osteotomy was performed in sixteen adult sheep and fixed with a DCP and cortical screws. Half of the animals were operated on with a SLE on each side of the osteotomy and the rest without the addition of SLE. Four animals of each group were euthanized after 8 weeks, and the remaining after 16 weeks. Both femora of each animal were radiographed and mechanically tested in torsion.

Results

Radiologically femoral malalignment or screw loosening was observed in six out of the eight animals operated on without SLE. In contrast, all animals subjected to the operation with SLE showed complete radiological consolidation of the osteotomy. Seven of these eight animals showed normal femoral alignment and no osteosynthesis failure. Stiffness of the bones fixed with SLE was among 145% and 177% the value of their contralateral non-operated femurs (all animals of this group showed greater stiffness on the operated bone than its contralateral non-operated femur). However, stiffness of the bones operated on without SLE was among 58% and 87% the value of the stiffness of their contralateral non-operated bone (all animals of this group showed greater stiffness on the non-operated bone than the osteotomized ones).

Conclusions

Use of SLE avoided loosening of the system and stimulated stronger osteotomy consolidation. Clinical application of this improved system may thus be a feasible and cost-effective alternative to other more rigid and expensive bone fixation techniques.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bottlang M, Doornink J, Fitzpatrick DC, Madey SM: Far cortical locking can reduce stiffness of locked plating constructs while retaining construct strength. J Bone Joint Surg Am. 2009, 91: 1985-1994. 10.2106/JBJS.H.01038.CrossRefPubMedPubMedCentral Bottlang M, Doornink J, Fitzpatrick DC, Madey SM: Far cortical locking can reduce stiffness of locked plating constructs while retaining construct strength. J Bone Joint Surg Am. 2009, 91: 1985-1994. 10.2106/JBJS.H.01038.CrossRefPubMedPubMedCentral
3.
go back to reference Bottlang M, Lesser M, Koerber J, Doornink J, von Rechenberg B, Augat P, Fitzpatrick DC, Madey SM, Marsh JL: Far cortical locking can improve healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010, 92: 1652-1660. 10.2106/JBJS.I.01111.CrossRefPubMedPubMedCentral Bottlang M, Lesser M, Koerber J, Doornink J, von Rechenberg B, Augat P, Fitzpatrick DC, Madey SM, Marsh JL: Far cortical locking can improve healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010, 92: 1652-1660. 10.2106/JBJS.I.01111.CrossRefPubMedPubMedCentral
4.
go back to reference Gardner MJ, Nork SE, Huber P, Krieg JC: Less rigid stable fracture fixation in osteoporotic bone using locked plates with near cortical slots. Injury. 2010, 41: 652-656. 10.1016/j.injury.2010.02.022.CrossRefPubMed Gardner MJ, Nork SE, Huber P, Krieg JC: Less rigid stable fracture fixation in osteoporotic bone using locked plates with near cortical slots. Injury. 2010, 41: 652-656. 10.1016/j.injury.2010.02.022.CrossRefPubMed
5.
go back to reference Bottlang M, Doornink J, Lujan TJ, Fitzpatrick DC, Marsh JL, Augat P, von Rechenberg B, Lesser M, Madey SM: Effects of construct stiffness on healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010, 92 (Suppl 2): 12-22.CrossRefPubMedPubMedCentral Bottlang M, Doornink J, Lujan TJ, Fitzpatrick DC, Marsh JL, Augat P, von Rechenberg B, Lesser M, Madey SM: Effects of construct stiffness on healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010, 92 (Suppl 2): 12-22.CrossRefPubMedPubMedCentral
6.
go back to reference Bottlang M, Doornink J, Byrd GD, Fitzpatrick D, Madey S: A nonlocking end screw can decrease fracture risk caused by locked plating in the osteoporotic diaphysis. J Bone Joint Surg Am. 2009, 91: 620-627. 10.2106/JBJS.H.00408.CrossRefPubMed Bottlang M, Doornink J, Byrd GD, Fitzpatrick D, Madey S: A nonlocking end screw can decrease fracture risk caused by locked plating in the osteoporotic diaphysis. J Bone Joint Surg Am. 2009, 91: 620-627. 10.2106/JBJS.H.00408.CrossRefPubMed
7.
go back to reference Claes LE, Heigele CA: Magnitudes of local stress and strain along bony surfaces predict the course and type of fracture healing. J Biomech. 1999, 32: 255-266. 10.1016/S0021-9290(98)00153-5.CrossRefPubMed Claes LE, Heigele CA: Magnitudes of local stress and strain along bony surfaces predict the course and type of fracture healing. J Biomech. 1999, 32: 255-266. 10.1016/S0021-9290(98)00153-5.CrossRefPubMed
8.
go back to reference Gardner MJ, Evans JM, Dunbar RP: Failure of fracture plate fixation. J Am Acad Orthop Surg. 2009, 17: 647-657.CrossRefPubMed Gardner MJ, Evans JM, Dunbar RP: Failure of fracture plate fixation. J Am Acad Orthop Surg. 2009, 17: 647-657.CrossRefPubMed
9.
go back to reference Tan SLE, Balogh ZJ: Indications and limitations of locked plating. Injury. 2009, 40: 683-691. 10.1016/j.injury.2009.01.003.CrossRefPubMed Tan SLE, Balogh ZJ: Indications and limitations of locked plating. Injury. 2009, 40: 683-691. 10.1016/j.injury.2009.01.003.CrossRefPubMed
10.
go back to reference Angelini AJ, Livani B, Flierl MA, Morgan SJ, Belangero WD: Less invasive percutaneous wave plating of simple femur shaft fractures: A prospective series. Injury. 2010, 41: 624-628. 10.1016/j.injury.2010.01.101.CrossRefPubMed Angelini AJ, Livani B, Flierl MA, Morgan SJ, Belangero WD: Less invasive percutaneous wave plating of simple femur shaft fractures: A prospective series. Injury. 2010, 41: 624-628. 10.1016/j.injury.2010.01.101.CrossRefPubMed
11.
go back to reference Dewo P, Sharma PK, van der Tas HF, van der Houwen EB, Timmer M, Magetsari R, Busscher HJ, Van Horn JR, Verkerke GJ: Surface properties of Indonesian-made narrow dynamic compression plates. Med J Malaysia. 2008, 63 (Suppl A): 21-22.PubMed Dewo P, Sharma PK, van der Tas HF, van der Houwen EB, Timmer M, Magetsari R, Busscher HJ, Van Horn JR, Verkerke GJ: Surface properties of Indonesian-made narrow dynamic compression plates. Med J Malaysia. 2008, 63 (Suppl A): 21-22.PubMed
12.
go back to reference Windolf M, Klos K, Wahnert D, van der Pol B, Radtke R, Schwieger K, Jakob RP: Biomechanical investigation of an alternative concept to angular stable plating using conventional fixation hardware. BMC Musculoskelet Disord. 2010, 11: 95-10.1186/1471-2474-11-95.CrossRefPubMedPubMedCentral Windolf M, Klos K, Wahnert D, van der Pol B, Radtke R, Schwieger K, Jakob RP: Biomechanical investigation of an alternative concept to angular stable plating using conventional fixation hardware. BMC Musculoskelet Disord. 2010, 11: 95-10.1186/1471-2474-11-95.CrossRefPubMedPubMedCentral
13.
go back to reference Yánez A, Carta JA, Garcés G: Biomechanical evaluation of a new system to improve screw fixation in osteoporotic bones. Med Engin Phys. 2010, 32: 532-534. 10.1016/j.medengphy.2010.02.014.CrossRef Yánez A, Carta JA, Garcés G: Biomechanical evaluation of a new system to improve screw fixation in osteoporotic bones. Med Engin Phys. 2010, 32: 532-534. 10.1016/j.medengphy.2010.02.014.CrossRef
14.
go back to reference Yánez A, Garcés GL, Carta JA, Cuadrado A: A new system to improve screw fixation to bones. Journal of Medical Devices ASME. 2011, 5: 044501-10.1115/1.4005227.CrossRef Yánez A, Garcés GL, Carta JA, Cuadrado A: A new system to improve screw fixation to bones. Journal of Medical Devices ASME. 2011, 5: 044501-10.1115/1.4005227.CrossRef
15.
go back to reference Yánez A, Cuadrado A, Carta JA, Garcés GL: Screw locking elements: A means to modify the flexibility of osteoporotic fracture fixation with DCPs without compromising system strength or stability. Med Eng Phys. 2012, 34: 717-724. 10.1016/j.medengphy.2011.09.015.CrossRefPubMed Yánez A, Cuadrado A, Carta JA, Garcés GL: Screw locking elements: A means to modify the flexibility of osteoporotic fracture fixation with DCPs without compromising system strength or stability. Med Eng Phys. 2012, 34: 717-724. 10.1016/j.medengphy.2011.09.015.CrossRefPubMed
16.
go back to reference Epari DR, Kassi JP, Schell H, Duda GN: Timely fracture-healing requires optimization of axial fixation stability. J Bone Joint Surg Am. 2007, 89: 1575-1585. 10.2106/JBJS.F.00247.CrossRefPubMed Epari DR, Kassi JP, Schell H, Duda GN: Timely fracture-healing requires optimization of axial fixation stability. J Bone Joint Surg Am. 2007, 89: 1575-1585. 10.2106/JBJS.F.00247.CrossRefPubMed
17.
go back to reference Jagodzinski M, Kretteck C: Effect of mechanical stability on fracture healing an update. Injury. 2007, 38 (Suppl1): S3-S10.CrossRefPubMed Jagodzinski M, Kretteck C: Effect of mechanical stability on fracture healing an update. Injury. 2007, 38 (Suppl1): S3-S10.CrossRefPubMed
18.
go back to reference Korner J, Lill H, Müller LP, Hessmann M, Kopf K, Goldhahn J, Gonschorek O, Josten C, Rommens P: Distal humerus fractures in elderly patients: results after open reduction and internal fixation. Osteoporos Int. 2005, 16 (Suppl 2): S73-79.CrossRefPubMed Korner J, Lill H, Müller LP, Hessmann M, Kopf K, Goldhahn J, Gonschorek O, Josten C, Rommens P: Distal humerus fractures in elderly patients: results after open reduction and internal fixation. Osteoporos Int. 2005, 16 (Suppl 2): S73-79.CrossRefPubMed
19.
go back to reference Giannoudis P, Tzioupis C, Almalki T, Buckley R: Fracture healing in osteoporotic fractures: is it really different? A basic science perspective. Injury. 2007, 38 (Suppl 1): S90-99.CrossRefPubMed Giannoudis P, Tzioupis C, Almalki T, Buckley R: Fracture healing in osteoporotic fractures: is it really different? A basic science perspective. Injury. 2007, 38 (Suppl 1): S90-99.CrossRefPubMed
20.
go back to reference Pesce V, Speciale D, Sammarco G, Patella S, Spinarelli A, Patella V: Surgical approach to bone healing in osteoporosis. Clin Cases Miner Bone Metab. 2009, 6: 131-135.PubMed Pesce V, Speciale D, Sammarco G, Patella S, Spinarelli A, Patella V: Surgical approach to bone healing in osteoporosis. Clin Cases Miner Bone Metab. 2009, 6: 131-135.PubMed
21.
go back to reference Clavert P, Ducrot G, Sirveaux F, Fabre T, Mansat P: Outcomes of distal humerus fractures in patients above 65 years of age treated by plate fixation. Orthop Traumatol Surg Res. 2013, 99: 771-777. 10.1016/j.otsr.2013.08.001.CrossRefPubMed Clavert P, Ducrot G, Sirveaux F, Fabre T, Mansat P: Outcomes of distal humerus fractures in patients above 65 years of age treated by plate fixation. Orthop Traumatol Surg Res. 2013, 99: 771-777. 10.1016/j.otsr.2013.08.001.CrossRefPubMed
22.
go back to reference Cornell C: Internal fracture fixation in patients with osteoporosis. J Am Acad Orthop Surg. 2003, 11: 109-119.CrossRefPubMed Cornell C: Internal fracture fixation in patients with osteoporosis. J Am Acad Orthop Surg. 2003, 11: 109-119.CrossRefPubMed
23.
go back to reference Yamaji T, Ando K, Wolf S, Augat P, Claes L: The effect of micromovement on callus formation. J Orthop Sci. 2001, 6: 571-575. 10.1007/s007760100014.CrossRefPubMed Yamaji T, Ando K, Wolf S, Augat P, Claes L: The effect of micromovement on callus formation. J Orthop Sci. 2001, 6: 571-575. 10.1007/s007760100014.CrossRefPubMed
24.
go back to reference Cuadrado A, Yánez A, Carta JA, Garcés GL: Suitability of DCPs with screw locking elements to allow sufficient interfragmentary motion to promote secondary bone healing of osteoporotic fractures. Med Eng Phys. 2013, 35: 852-859. 10.1016/j.medengphy.2012.08.018.CrossRefPubMed Cuadrado A, Yánez A, Carta JA, Garcés GL: Suitability of DCPs with screw locking elements to allow sufficient interfragmentary motion to promote secondary bone healing of osteoporotic fractures. Med Eng Phys. 2013, 35: 852-859. 10.1016/j.medengphy.2012.08.018.CrossRefPubMed
25.
go back to reference Auer JA, Goodship A, Arnoczky S, Pearce S, Price J, Claes L, von Rechenberg B, Hofmann-Amtenbrinck M, Schneider E, Müller-Terpitz R, Thiele F, Rippe KP, Grainger DW: Refining animal models in fracture research: seeking consensus in optimising both animal welfare and scientific validity for appropriate biomedical use. BMC Musculoskelet Disord. 2007, 8: 72-10.1186/1471-2474-8-72.CrossRefPubMedPubMedCentral Auer JA, Goodship A, Arnoczky S, Pearce S, Price J, Claes L, von Rechenberg B, Hofmann-Amtenbrinck M, Schneider E, Müller-Terpitz R, Thiele F, Rippe KP, Grainger DW: Refining animal models in fracture research: seeking consensus in optimising both animal welfare and scientific validity for appropriate biomedical use. BMC Musculoskelet Disord. 2007, 8: 72-10.1186/1471-2474-8-72.CrossRefPubMedPubMedCentral
26.
go back to reference Zeiter S, Montavon P, Schneider E, Glen N: Plate stabilization with bone rivets: an alternative method for internal fixation of fractures. J Orthop Trauma. 2004, 18: 279-285. 10.1097/00005131-200405000-00003.CrossRefPubMed Zeiter S, Montavon P, Schneider E, Glen N: Plate stabilization with bone rivets: an alternative method for internal fixation of fractures. J Orthop Trauma. 2004, 18: 279-285. 10.1097/00005131-200405000-00003.CrossRefPubMed
27.
go back to reference Axelrad W, Einhorn T: Use of clinical assessment tools in the evaluation of fracture healing. Injury. 2011, 42: 301-305. 10.1016/j.injury.2010.11.043.CrossRefPubMed Axelrad W, Einhorn T: Use of clinical assessment tools in the evaluation of fracture healing. Injury. 2011, 42: 301-305. 10.1016/j.injury.2010.11.043.CrossRefPubMed
28.
go back to reference Corrales LA, Morshed S, Bhandari M, Miclau T: Variability in the assessment of fracture-healing in orthopaedic trauma studies. J Bone Joint Surg Am. 2008, 90: 1862-1868. 10.2106/JBJS.G.01580.CrossRefPubMedPubMedCentral Corrales LA, Morshed S, Bhandari M, Miclau T: Variability in the assessment of fracture-healing in orthopaedic trauma studies. J Bone Joint Surg Am. 2008, 90: 1862-1868. 10.2106/JBJS.G.01580.CrossRefPubMedPubMedCentral
29.
go back to reference Kooistra B, Sprague S, Bhandari M, Schemitsch E: Outcomes assessment in fracture healing trials: a primer. J Orthop Trauma. 2010, 24: S71-S75.CrossRefPubMed Kooistra B, Sprague S, Bhandari M, Schemitsch E: Outcomes assessment in fracture healing trials: a primer. J Orthop Trauma. 2010, 24: S71-S75.CrossRefPubMed
30.
31.
go back to reference Simon J, Dennis M, Kummer F, Koval K: Schuhli augmentation of plate and screw fixation for humeral shaft fractures: a laboratory study. J Orthop Trauma. 1999, 13: 196-199. 10.1097/00005131-199903000-00007.CrossRefPubMed Simon J, Dennis M, Kummer F, Koval K: Schuhli augmentation of plate and screw fixation for humeral shaft fractures: a laboratory study. J Orthop Trauma. 1999, 13: 196-199. 10.1097/00005131-199903000-00007.CrossRefPubMed
Metadata
Title
Use of screw locking elements improves radiological and biomechanical results of femoral osteotomies
Authors
Gerardo L Garcés
Oliver Rodríguez
Enrique Rodríguez Grau-Bassas
Syra Roiz
Alejandro Yánez
Alberto Cuadrado
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2014
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-15-387

Other articles of this Issue 1/2014

BMC Musculoskeletal Disorders 1/2014 Go to the issue