Published in: BMC Musculoskeletal Disorders 1/2010

Open Access 01-12-2010 | Research article

Biomechanical investigation of an alternative concept to angular stable plating using conventional fixation hardware

Authors: Markus Windolf, Kajetan Klos, Dirk Wähnert, Bas van der Pol, Roman Radtke, Karsten Schwieger, Roland P Jakob

Published in: BMC Musculoskeletal Disorders | Issue 1/2010

Login to get access

Abstract

Background

Angle-stable locking plates have improved the surgical management of fractures. However, locking implants are costly and removal can be difficult. The aim of this in vitro study was to evaluate the biomechanical performance of a newly proposed crossed-screw concept ("Fence") utilizing conventional (non-locked) implants in comparison to conventional LC-DCP (limited contact dynamic compression plate) and LCP (locking compression plate) stabilization, in a human cadaveric diaphyseal gap model.

Methods

In eight pairs of human cadaveric femora, one femur per pair was randomly assigned to receive a Fence construct with either elevated or non-elevated plate, while the contralateral femur received either an LCP or LC-DCP instrumentation. Fracture gap motion and fatigue performance under cyclic loading was evaluated successively in axial compression and in torsion. Results were statistically compared in a pairwise setting.

Results

The elevated Fence constructs allowed significantly higher gap motion compared to the LCP instrumentations (axial compression: p ≤ 0.011, torsion p ≤ 0.015) but revealed similar performance under cyclic loading (p = 0.43). The Fence instrumentation with established bone-plate contact revealed larger fracture gap motion under axial compression compared to the conventional LC-DCP osteosynthesis (p ≤ 0.017). However, all contact Fence specimens survived the cyclic test, whereas all LC-DCP constructs failed early during torsion testing (p < 0.001). All failures occurred due to breakage of the screw heads.

Conclusions

Even though accentuated fracture gap motion became obvious, the "Fence" technique is considered an alternative to cost-intensive locking-head devices. The concept can be of interest in cases were angle-stable implants are unavailable and can lead to new strategies in implant design.
Appendix
Available only for authorised users
Literature
1.
go back to reference Perren SM: Backgrounds of the technology of internal fixators. Injury. 2003, 34 (Suppl 2): B1-B3. 10.1016/j.injury.2003.09.018.CrossRefPubMed Perren SM: Backgrounds of the technology of internal fixators. Injury. 2003, 34 (Suppl 2): B1-B3. 10.1016/j.injury.2003.09.018.CrossRefPubMed
2.
go back to reference Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ: Biomechanics of locked plates and screws. J Orthop Trauma. 2004, 18: 488-493. 10.1097/00005131-200409000-00003.CrossRefPubMed Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ: Biomechanics of locked plates and screws. J Orthop Trauma. 2004, 18: 488-493. 10.1097/00005131-200409000-00003.CrossRefPubMed
3.
go back to reference Wagner M: General principles for the clinical use of the LCP. Injury. 2003, 34 (Suppl 2): B31-B42. 10.1016/j.injury.2003.09.023.CrossRefPubMed Wagner M: General principles for the clinical use of the LCP. Injury. 2003, 34 (Suppl 2): B31-B42. 10.1016/j.injury.2003.09.023.CrossRefPubMed
4.
go back to reference Fitzpatrick DC, Doornink J, Madey SM, Bottlang M: Relative stability of conventional and locked plating fixation in a model of the osteoporotic femoral diaphysis. Clin Biomech (Bristol, Avon). 2009, 24: 203-209. 10.1016/j.clinbiomech.2008.11.002.CrossRef Fitzpatrick DC, Doornink J, Madey SM, Bottlang M: Relative stability of conventional and locked plating fixation in a model of the osteoporotic femoral diaphysis. Clin Biomech (Bristol, Avon). 2009, 24: 203-209. 10.1016/j.clinbiomech.2008.11.002.CrossRef
5.
go back to reference Fulkerson E, Egol KA, Kubiak EN, Liporace F, Kummer FJ, Koval KJ: Fixation of diaphyseal fractures with a segmental defect: a biomechanical comparison of locked and conventional plating techniques. J Trauma. 2006, 60: 830-835. 10.1097/01.ta.0000195462.53525.0c.CrossRefPubMed Fulkerson E, Egol KA, Kubiak EN, Liporace F, Kummer FJ, Koval KJ: Fixation of diaphyseal fractures with a segmental defect: a biomechanical comparison of locked and conventional plating techniques. J Trauma. 2006, 60: 830-835. 10.1097/01.ta.0000195462.53525.0c.CrossRefPubMed
6.
go back to reference Miller DL, Goswami T: A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing. Clin Biomech (Bristol, Avon). 2007, 22: 1049-1062. 10.1016/j.clinbiomech.2007.08.004.CrossRef Miller DL, Goswami T: A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing. Clin Biomech (Bristol, Avon). 2007, 22: 1049-1062. 10.1016/j.clinbiomech.2007.08.004.CrossRef
7.
go back to reference Haidukewych GJ, Ricci W: Locked plating in orthopaedic trauma: a clinical update. J Am Acad Orthop Surg. 2008, 16: 347-355.CrossRefPubMed Haidukewych GJ, Ricci W: Locked plating in orthopaedic trauma: a clinical update. J Am Acad Orthop Surg. 2008, 16: 347-355.CrossRefPubMed
8.
go back to reference Dewo P, Sharma PK, Tas van der HF, Houwen van der EB, Timmer M, Magetsari R, Busscher HJ, van Horn JR, Verkerke GJ: Surface properties of Indonesian-made narrow dynamic compression plates. Med J Malaysia. 2008, 63 (Suppl A): 21-22.PubMed Dewo P, Sharma PK, Tas van der HF, Houwen van der EB, Timmer M, Magetsari R, Busscher HJ, van Horn JR, Verkerke GJ: Surface properties of Indonesian-made narrow dynamic compression plates. Med J Malaysia. 2008, 63 (Suppl A): 21-22.PubMed
9.
go back to reference Angelini AJ, Livani B, Flierl MA, Morgan SJ, Belangero WD: Less invasive percutaneous wave plating of simple femur shaft fractures: A prospective series. Injury. 2010 Angelini AJ, Livani B, Flierl MA, Morgan SJ, Belangero WD: Less invasive percutaneous wave plating of simple femur shaft fractures: A prospective series. Injury. 2010
10.
go back to reference Ahmad M, Nanda R, Bajwa AS, Candal-Couto J, Green S, Hui AC: Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability?. Injury. 2007, 38: 358-364. 10.1016/j.injury.2006.08.058.CrossRefPubMed Ahmad M, Nanda R, Bajwa AS, Candal-Couto J, Green S, Hui AC: Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability?. Injury. 2007, 38: 358-364. 10.1016/j.injury.2006.08.058.CrossRefPubMed
11.
go back to reference DeTora M, Kraus K: Mechanical testing of 3.5 mm locking and non-locking bone plates. Vet Comp Orthop Traumatol. 2008, 21: 318-322.PubMed DeTora M, Kraus K: Mechanical testing of 3.5 mm locking and non-locking bone plates. Vet Comp Orthop Traumatol. 2008, 21: 318-322.PubMed
12.
go back to reference Gardner MJ, Griffith MH, Demetrakopoulos D, Brophy RH, Grose A, Helfet DL, Lorich DG: Hybrid locked plating of osteoporotic fractures of the humerus. J Bone Joint Surg Am. 2006, 88: 1962-1967. 10.2106/JBJS.E.00893.CrossRefPubMed Gardner MJ, Griffith MH, Demetrakopoulos D, Brophy RH, Grose A, Helfet DL, Lorich DG: Hybrid locked plating of osteoporotic fractures of the humerus. J Bone Joint Surg Am. 2006, 88: 1962-1967. 10.2106/JBJS.E.00893.CrossRefPubMed
13.
go back to reference O'Toole RV, Andersen RC, Vesnovsky O, Alexander M, Topoleski LD, Nascone JW, Sciadini MF, Turen C, Eglseder WA: Are locking screws advantageous with plate fixation of humeral shaft fractures? A biomechanical analysis of synthetic and cadaveric bone. J Orthop Trauma. 2008, 22: 709-715. 10.1097/BOT.0b013e31818df8cb.CrossRefPubMed O'Toole RV, Andersen RC, Vesnovsky O, Alexander M, Topoleski LD, Nascone JW, Sciadini MF, Turen C, Eglseder WA: Are locking screws advantageous with plate fixation of humeral shaft fractures? A biomechanical analysis of synthetic and cadaveric bone. J Orthop Trauma. 2008, 22: 709-715. 10.1097/BOT.0b013e31818df8cb.CrossRefPubMed
14.
go back to reference Uhl JM, Seguin B, Kapatkin AS, Schulz KS, Garcia TC, Stover SM: Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models. Vet Surg. 2008, 37: 663-673. 10.1111/j.1532-950X.2008.00433.x.CrossRefPubMed Uhl JM, Seguin B, Kapatkin AS, Schulz KS, Garcia TC, Stover SM: Mechanical comparison of 3.5 mm broad dynamic compression plate, broad limited-contact dynamic compression plate, and narrow locking compression plate systems using interfragmentary gap models. Vet Surg. 2008, 37: 663-673. 10.1111/j.1532-950X.2008.00433.x.CrossRefPubMed
15.
go back to reference Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN: Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg. 2003, 123: 74-81.PubMed Lill H, Hepp P, Korner J, Kassi JP, Verheyden AP, Josten C, Duda GN: Proximal humeral fractures: how stiff should an implant be? A comparative mechanical study with new implants in human specimens. Arch Orthop Trauma Surg. 2003, 123: 74-81.PubMed
Metadata
Title
Biomechanical investigation of an alternative concept to angular stable plating using conventional fixation hardware
Authors
Markus Windolf
Kajetan Klos
Dirk Wähnert
Bas van der Pol
Roman Radtke
Karsten Schwieger
Roland P Jakob
Publication date
01-12-2010
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2010
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-11-95

Other articles of this Issue 1/2010

BMC Musculoskeletal Disorders 1/2010 Go to the issue