Perioperative Cardiac Output Monitoring Utilizing Non-pulse Contour Methods

Authors: Yohei Fujimoto, Koichi Suehiro, Akira Mukai, Kiyonobu Nishikawa

Published in: Current Anesthesiology Reports | Issue 4/2017

Login to get access

Abstract

Purpose of Review

The purpose of this review was to provide an overview of the recent concepts regarding cardiac output measurement devices that utilize non-pulse contour methods, especially in an intraoperative setting. The techniques include inert gas rebreathing method, partial CO2 rebreathing method, impedance cardiography, and its derivative technologies such as electrical velocimetry and bioreactance, transesophageal echocardiography/Doppler, and transthoracic echocardiography/Doppler. We focused on the invasiveness of the devices and their underlying technology.

Recent Findings

Although various types of cardiac output monitoring devices are available, none of them may be considered as an ideal device in terms of accuracy, trending ability of cardiac output changes, and reproducibility of measurements. There are increasing types of devices applicable for intraoperative use, yet only few data are available regarding the trending ability of cardiac output changes and reproducibility of the measurements. Therefore, the empirical application of these devices for various surgical patients may be done under the consideration of their invasiveness and their underlying technology, and it may provide us with more data over time.

Summary

The non-pulse contour-derived cardiac output measurement devices are classified according to their underlying principles, which closely reflect their advantages and disadvantages in the perioperative setting.
Appendix
Available only for authorised users
Literature
5.
go back to reference Smart FW, Husserl FE. Complications of flow-directed balloon-tipped catheters. Chest. 1990;97(1):227–8.PubMedCrossRef Smart FW, Husserl FE. Complications of flow-directed balloon-tipped catheters. Chest. 1990;97(1):227–8.PubMedCrossRef
7.
go back to reference Practice guidelines for pulmonary artery catheterization: an updated report by the American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization. Anesthesiology. 2003;99(4):988–1014. Practice guidelines for pulmonary artery catheterization: an updated report by the American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization. Anesthesiology. 2003;99(4):988–1014.
9.
go back to reference Lennon M, Zaw NN, Popping DM, Wenk M. Procedural complications of central venous catheter insertion. Minerva Anestesiol. 2012;78(11):1234–40.PubMed Lennon M, Zaw NN, Popping DM, Wenk M. Procedural complications of central venous catheter insertion. Minerva Anestesiol. 2012;78(11):1234–40.PubMed
11.
go back to reference Janik JE, Conlon SJ, Janik JS. Percutaneous central access in patients younger than 5 years: size does matter. J Pediatr Surg. 2004;39(8):1252–6.PubMedCrossRef Janik JE, Conlon SJ, Janik JS. Percutaneous central access in patients younger than 5 years: size does matter. J Pediatr Surg. 2004;39(8):1252–6.PubMedCrossRef
12.
go back to reference Heidemann L, Nathani N, Sagana R, Chopra V, Hueng M. A contemporary assessment of mechanical complication rates and trainee perceptions of central venous catheter insertion. J Hosp Med. 2017;12(8):646–51. 10.12788/jhm.2784.PubMedCrossRef Heidemann L, Nathani N, Sagana R, Chopra V, Hueng M. A contemporary assessment of mechanical complication rates and trainee perceptions of central venous catheter insertion. J Hosp Med. 2017;12(8):646–51. 10.​12788/​jhm.​2784.PubMedCrossRef
21.
go back to reference Haryadi DG, Orr JA, Kuck K, McJames S, Westenskow DR. Partial CO2 rebreathing indirect Fick technique for non-invasive measurement of cardiac output. J Clin Monit Comput. 2000;16(5–6):361–74.PubMedCrossRef Haryadi DG, Orr JA, Kuck K, McJames S, Westenskow DR. Partial CO2 rebreathing indirect Fick technique for non-invasive measurement of cardiac output. J Clin Monit Comput. 2000;16(5–6):361–74.PubMedCrossRef
23.
go back to reference Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing technique for cardiac output monitoring in critically ill patients. Acta Anaesthesiol Scand. 2002;46(2):152–9.PubMedCrossRef Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing technique for cardiac output monitoring in critically ill patients. Acta Anaesthesiol Scand. 2002;46(2):152–9.PubMedCrossRef
25.
go back to reference Binder JC, Parkin WG. Non-invasive cardiac output determination: comparison of a new partial-rebreathing technique with thermodilution. Anaesth Intensive Care. 2001;29(1):19–23.PubMed Binder JC, Parkin WG. Non-invasive cardiac output determination: comparison of a new partial-rebreathing technique with thermodilution. Anaesth Intensive Care. 2001;29(1):19–23.PubMed
27.
go back to reference Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18(5):563–72.PubMedCrossRef Botero M, Kirby D, Lobato EB, Staples ED, Gravenstein N. Measurement of cardiac output before and after cardiopulmonary bypass: comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing. J Cardiothorac Vasc Anesth. 2004;18(5):563–72.PubMedCrossRef
28.
go back to reference Nilsson LB, Eldrup N, Berthelsen PG. Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output. Acta Anaesthesiol Scand. 2001;45(6):680–5.PubMedCrossRef Nilsson LB, Eldrup N, Berthelsen PG. Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output. Acta Anaesthesiol Scand. 2001;45(6):680–5.PubMedCrossRef
30.
go back to reference Valiatti JL, Amaral JL. Comparison between cardiac output values measured by thermodilution and partial carbon dioxide rebreathing in patients with acute lung injury. Sao Paulo Medical Journal = Revista paulista de Medicina. 2004;122(6):233–8.PubMedCrossRef Valiatti JL, Amaral JL. Comparison between cardiac output values measured by thermodilution and partial carbon dioxide rebreathing in patients with acute lung injury. Sao Paulo Medical Journal = Revista paulista de Medicina. 2004;122(6):233–8.PubMedCrossRef
32.
go back to reference Maxwell RA, Gibson JB, Slade JB, Fabian TC, Proctor KG. Noninvasive cardiac output by partial CO2 rebreathing after severe chest trauma. J Trauma. 2001;51(5):849–53.PubMedCrossRef Maxwell RA, Gibson JB, Slade JB, Fabian TC, Proctor KG. Noninvasive cardiac output by partial CO2 rebreathing after severe chest trauma. J Trauma. 2001;51(5):849–53.PubMedCrossRef
33.
go back to reference Tachibana K, Imanaka H, Takeuchi M, Takauchi Y, Miyano H, Nishimura M. Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present. Anesthesiology. 2003;98(4):830–7.PubMedCrossRef Tachibana K, Imanaka H, Takeuchi M, Takauchi Y, Miyano H, Nishimura M. Noninvasive cardiac output measurement using partial carbon dioxide rebreathing is less accurate at settings of reduced minute ventilation and when spontaneous breathing is present. Anesthesiology. 2003;98(4):830–7.PubMedCrossRef
36.
go back to reference van Heerden PV, Baker S, Lim SI, Weidman C, Bulsara M. Clinical evaluation of the non-invasive cardiac output (NICO) monitor in the intensive care unit. Anaesth Intensive Care. 2000;28(4):427–30.PubMed van Heerden PV, Baker S, Lim SI, Weidman C, Bulsara M. Clinical evaluation of the non-invasive cardiac output (NICO) monitor in the intensive care unit. Anaesth Intensive Care. 2000;28(4):427–30.PubMed
37.
go back to reference Sobanski P, Sinkiewicz W, Kubica J, Blazejewski J, Bujak R. The reliability of noninvasive cardiac output measurement using the inert gas rebreathing method in patients with advanced heart failure. Cardiol J. 2008;15(1):63–70.PubMed Sobanski P, Sinkiewicz W, Kubica J, Blazejewski J, Bujak R. The reliability of noninvasive cardiac output measurement using the inert gas rebreathing method in patients with advanced heart failure. Cardiol J. 2008;15(1):63–70.PubMed
41.
47.
go back to reference •• Perak AM, Opotowsky AR, Walsh BK, Esch JJ, DiNardo JA, Kussman BD, et al. Noninvasive cardiac output estimation by inert gas rebreathing in mechanically ventilated pediatric patients. J Pediatr. 2016;177:184–90.e3. https://doi.org/10.1016/j.jpeds.2016.07.007. Accuracy and precision data of inert gas rebreathing methods in pediatric mechanically ventilated patients. PubMedCrossRef •• Perak AM, Opotowsky AR, Walsh BK, Esch JJ, DiNardo JA, Kussman BD, et al. Noninvasive cardiac output estimation by inert gas rebreathing in mechanically ventilated pediatric patients. J Pediatr. 2016;177:184–90.e3. https://​doi.​org/​10.​1016/​j.​jpeds.​2016.​07.​007. Accuracy and precision data of inert gas rebreathing methods in pediatric mechanically ventilated patients. PubMedCrossRef
49.
50.
go back to reference Haruna M, Kumon K, Yahagi N, Watanabe Y, Ishida Y, Kobayashi N, et al. Blood volume measurement at the bedside using ICG pulse spectrophotometry. Anesthesiology. 1998;89(6):1322–8.PubMedCrossRef Haruna M, Kumon K, Yahagi N, Watanabe Y, Ishida Y, Kobayashi N, et al. Blood volume measurement at the bedside using ICG pulse spectrophotometry. Anesthesiology. 1998;89(6):1322–8.PubMedCrossRef
52.
go back to reference Imai T, Takahashi K, Fukura H, Morishita Y. Measurement of cardiac output by pulse dye densitometry using indocyanine green: a comparison with the thermodilution method. Anesthesiology. 1997;87(4):816–22.PubMedCrossRef Imai T, Takahashi K, Fukura H, Morishita Y. Measurement of cardiac output by pulse dye densitometry using indocyanine green: a comparison with the thermodilution method. Anesthesiology. 1997;87(4):816–22.PubMedCrossRef
55.
go back to reference Taguchi N, Nakagawa S, Miyasaka K, Fuse M, Aoyagi T. Cardiac output measurement by pulse dye densitometry using three wavelengths. Pediatr Crit Care Med. 2004;5(4):343–50.PubMedCrossRef Taguchi N, Nakagawa S, Miyasaka K, Fuse M, Aoyagi T. Cardiac output measurement by pulse dye densitometry using three wavelengths. Pediatr Crit Care Med. 2004;5(4):343–50.PubMedCrossRef
62.
go back to reference • Vrancken SL, van Heijst AF, Hopman JC, Liem KD, van der Hoeven JG, de Boode WP. Hemodynamic volumetry using transpulmonary ultrasound dilution (TPUD) technology in a neonatal animal model. J Clin Monit Comput. 2015;29(5):643–52. https://doi.org/10.1007/s10877-014-9647-6. Providing explanation of transpulmonary ultrasound dilution technique. PubMedCrossRef • Vrancken SL, van Heijst AF, Hopman JC, Liem KD, van der Hoeven JG, de Boode WP. Hemodynamic volumetry using transpulmonary ultrasound dilution (TPUD) technology in a neonatal animal model. J Clin Monit Comput. 2015;29(5):643–52. https://​doi.​org/​10.​1007/​s10877-014-9647-6. Providing explanation of transpulmonary ultrasound dilution technique. PubMedCrossRef
71.
go back to reference •• Leclercq T, Lilot M, Schulz T, Meyer A, Farhat F, Fellahi JL. Endotracheal bioimpedance cardiography improves immediate postoperative outcome: a case-control study in off-pump coronary surgery. J Clin Monit Comput. 2017; https://doi.org/10.1007/s10877-017-9996-z. The authors have revealed that utilization of endotracheal bioimpedance cardiography improved postoperative outcome. •• Leclercq T, Lilot M, Schulz T, Meyer A, Farhat F, Fellahi JL. Endotracheal bioimpedance cardiography improves immediate postoperative outcome: a case-control study in off-pump coronary surgery. J Clin Monit Comput. 2017; https://​doi.​org/​10.​1007/​s10877-017-9996-z. The authors have revealed that utilization of endotracheal bioimpedance cardiography improved postoperative outcome.
72.
go back to reference Taniguchi Y, Emoto N, Miyagawa K, Nakayama K, Kinutani H, Tanaka H, et al. Noninvasive and simple assessment of cardiac output and pulmonary vascular resistance with whole-body impedance cardiography is useful for monitoring patients with pulmonary hypertension. Circ J. 2013;77(9):2383–9.PubMedCrossRef Taniguchi Y, Emoto N, Miyagawa K, Nakayama K, Kinutani H, Tanaka H, et al. Noninvasive and simple assessment of cardiac output and pulmonary vascular resistance with whole-body impedance cardiography is useful for monitoring patients with pulmonary hypertension. Circ J. 2013;77(9):2383–9.PubMedCrossRef
73.
go back to reference Paredes OL, Shite J, Shinke T, Watanabe S, Otake H, Matsumoto D, et al. Impedance cardiography for cardiac output estimation: reliability of wrist-to-ankle electrode configuration. Circ J. 2006;70(9):1164–8.PubMedCrossRef Paredes OL, Shite J, Shinke T, Watanabe S, Otake H, Matsumoto D, et al. Impedance cardiography for cardiac output estimation: reliability of wrist-to-ankle electrode configuration. Circ J. 2006;70(9):1164–8.PubMedCrossRef
74.
go back to reference Matsuda Y, Kawate H, Shimada S, Matsuzaki C, Nagata H, Adachi M, et al. Perioperative sequential monitoring of hemodynamic parameters in patients with pheochromocytoma using the Non-Invasive Cardiac System (NICaS). Endocr J. 2014;61(6):571–5.PubMedCrossRef Matsuda Y, Kawate H, Shimada S, Matsuzaki C, Nagata H, Adachi M, et al. Perioperative sequential monitoring of hemodynamic parameters in patients with pheochromocytoma using the Non-Invasive Cardiac System (NICaS). Endocr J. 2014;61(6):571–5.PubMedCrossRef
75.
go back to reference Ram M, Lavie A, Lev S, Blecher Y, Amikam U, Shulman Y, et al. Cardiac hemodynamics before, during and after elective cesarean section under spinal anesthesia in low-risk women. J Perinatol: official journal of the California Perinatal Association. 2017; https://doi.org/10.1038/jp.2017.53. Ram M, Lavie A, Lev S, Blecher Y, Amikam U, Shulman Y, et al. Cardiac hemodynamics before, during and after elective cesarean section under spinal anesthesia in low-risk women. J Perinatol: official journal of the California Perinatal Association. 2017; https://​doi.​org/​10.​1038/​jp.​2017.​53.
94.
go back to reference •• Bernstein DP, Henry IC, Lemmens HJ, Chaltas JL, DeMaria AN, Moon JB, et al. Validation of stroke volume and cardiac output by electrical interrogation of the brachial artery in normals: assessment of strengths, limitations, and sources of error. J Clin Monit Comput. 2015;29(6):789–800. https://doi.org/10.1007/s10877-015-9668-9. The authors have investigated a brand new technology, the transbrachial electrical bioimpedance velocimetry method. PubMedPubMedCentralCrossRef •• Bernstein DP, Henry IC, Lemmens HJ, Chaltas JL, DeMaria AN, Moon JB, et al. Validation of stroke volume and cardiac output by electrical interrogation of the brachial artery in normals: assessment of strengths, limitations, and sources of error. J Clin Monit Comput. 2015;29(6):789–800. https://​doi.​org/​10.​1007/​s10877-015-9668-9. The authors have investigated a brand new technology, the transbrachial electrical bioimpedance velocimetry method. PubMedPubMedCentralCrossRef
98.
go back to reference •• Suehiro K, Joosten A, Murphy LS, Desebbe O, Alexander B, Kim SH, et al. Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis. J Clin Monit Comput. 2016;30(5):603–20. https://doi.org/10.1007/s10877-015-9757-9. The literature provides thorough information on agreement of cardiac output monitoring devices in pediatric patients. PubMedCrossRef •• Suehiro K, Joosten A, Murphy LS, Desebbe O, Alexander B, Kim SH, et al. Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis. J Clin Monit Comput. 2016;30(5):603–20. https://​doi.​org/​10.​1007/​s10877-015-9757-9. The literature provides thorough information on agreement of cardiac output monitoring devices in pediatric patients. PubMedCrossRef
102.
105.
go back to reference Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, et al. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Critical Care (London). 2014;18(6):603. https://doi.org/10.1186/s13054-014-0603-0.CrossRef Blohm ME, Obrecht D, Hartwich J, Mueller GC, Kersten JF, Weil J, et al. Impedance cardiography (electrical velocimetry) and transthoracic echocardiography for non-invasive cardiac output monitoring in pediatric intensive care patients: a prospective single-center observational study. Critical Care (London). 2014;18(6):603. https://​doi.​org/​10.​1186/​s13054-014-0603-0.CrossRef
107.
go back to reference Torigoe T, Sato S, Nagayama Y, Sato T, Yamazaki H. Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants. J Perinatol: official journal of the California Perinatal Association. 2015;35(7):485–9. https://doi.org/10.1038/jp.2014.245.CrossRef Torigoe T, Sato S, Nagayama Y, Sato T, Yamazaki H. Influence of patent ductus arteriosus and ventilators on electrical velocimetry for measuring cardiac output in very-low/low birth weight infants. J Perinatol: official journal of the California Perinatal Association. 2015;35(7):485–9. https://​doi.​org/​10.​1038/​jp.​2014.​245.CrossRef
109.
go back to reference Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, et al. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.PubMedCrossRef Nidorf SM, Picard MH, Triulzi MO, Thomas JD, Newell J, King ME, et al. New perspectives in the assessment of cardiac chamber dimensions during development and adulthood. J Am Coll Cardiol. 1992;19(5):983–8.PubMedCrossRef
111.
go back to reference • McNamara H, Barclay P, Sharma V. Accuracy and precision of the ultrasound cardiac output monitor (USCOM 1A) in pregnancy: comparison with three-dimensional transthoracic echocardiography. Br J Anaesth. 2014;113(4):669–76. https://doi.org/10.1093/bja/aeu162. The authors proved the adequate accuracy of USCOM in pregnant women. PubMedCrossRef • McNamara H, Barclay P, Sharma V. Accuracy and precision of the ultrasound cardiac output monitor (USCOM 1A) in pregnancy: comparison with three-dimensional transthoracic echocardiography. Br J Anaesth. 2014;113(4):669–76. https://​doi.​org/​10.​1093/​bja/​aeu162. The authors proved the adequate accuracy of USCOM in pregnant women. PubMedCrossRef
112.
119.
go back to reference • Moller-Sorensen H, Cordtz J, Ostergaard M, Nilsson JC, Hansen KL. Transesophageal Doppler reliably tracks changes in cardiac output in comparison with intermittent pulmonary artery thermodilution in cardiac surgery patients. J Clin Monit Comput. 2017;31(1):135–42. https://doi.org/10.1007/s10877-015-9806-4. The authors properly validate precision of the method. PubMedCrossRef • Moller-Sorensen H, Cordtz J, Ostergaard M, Nilsson JC, Hansen KL. Transesophageal Doppler reliably tracks changes in cardiac output in comparison with intermittent pulmonary artery thermodilution in cardiac surgery patients. J Clin Monit Comput. 2017;31(1):135–42. https://​doi.​org/​10.​1007/​s10877-015-9806-4. The authors properly validate precision of the method. PubMedCrossRef
122.
126.
go back to reference Permpikul C, Leelayuthachai T. Non-invasive estimated continuous cardiac output (escCO) during severe sepsis and septic shock resuscitation. J Med Assoc Thailand = Chotmaihet thangphaet. 2014;97(Suppl 3):S184–8. Permpikul C, Leelayuthachai T. Non-invasive estimated continuous cardiac output (escCO) during severe sepsis and septic shock resuscitation. J Med Assoc Thailand = Chotmaihet thangphaet. 2014;97(Suppl 3):S184–8.
127.
go back to reference • Thonnerieux M, Alexander B, Binet C, Obadia JF, Bastien O, Desebbe O. The ability of esCCO and ECOM monitors to measure trends in cardiac output during alveolar recruitment maneuver after cardiac surgery: a comparison with the pulmonary thermodilution method. Anesth Analg. 2015;121(2):383–91. https://doi.org/10.1213/ane.0000000000000753. The authors properly handled the precision and trending ability. PubMedCrossRef • Thonnerieux M, Alexander B, Binet C, Obadia JF, Bastien O, Desebbe O. The ability of esCCO and ECOM monitors to measure trends in cardiac output during alveolar recruitment maneuver after cardiac surgery: a comparison with the pulmonary thermodilution method. Anesth Analg. 2015;121(2):383–91. https://​doi.​org/​10.​1213/​ane.​0000000000000753​. The authors properly handled the precision and trending ability. PubMedCrossRef
131.
go back to reference Magliocca A, Rezoagli E, Anderson TA, Burns SM, Ichinose F, Chitilian HV. Cardiac output measurements based on the pulse wave transit time and thoracic impedance exhibit limited agreement with thermodilution method during orthotopic liver transplantation. Anesth Analg. 2017; https://doi.org/10.1213/ane.0000000000002171. Magliocca A, Rezoagli E, Anderson TA, Burns SM, Ichinose F, Chitilian HV. Cardiac output measurements based on the pulse wave transit time and thoracic impedance exhibit limited agreement with thermodilution method during orthotopic liver transplantation. Anesth Analg. 2017; https://​doi.​org/​10.​1213/​ane.​0000000000002171​.
132.
go back to reference Terada T, Oiwa A, Maemura Y, Robert S, Kessoku S, Ochiai R. Comparison of the ability of two continuous cardiac output monitors to measure trends in cardiac output: estimated continuous cardiac output measured by modified pulse wave transit time and an arterial pulse contour-based cardiac output device. J Clin Monit Comput. 2016;30(5):621–7. https://doi.org/10.1007/s10877-015-9772-x.PubMedCrossRef Terada T, Oiwa A, Maemura Y, Robert S, Kessoku S, Ochiai R. Comparison of the ability of two continuous cardiac output monitors to measure trends in cardiac output: estimated continuous cardiac output measured by modified pulse wave transit time and an arterial pulse contour-based cardiac output device. J Clin Monit Comput. 2016;30(5):621–7. https://​doi.​org/​10.​1007/​s10877-015-9772-x.PubMedCrossRef
134.
go back to reference Raissuni Z, Zores F, Henriet O, Dallest S, Roul G. Can we obtain a noninvasive and continuous estimation of cardiac output? Comparison between three noninvasive methods. Int Heart J. 2013;54(6):395–400.PubMedCrossRef Raissuni Z, Zores F, Henriet O, Dallest S, Roul G. Can we obtain a noninvasive and continuous estimation of cardiac output? Comparison between three noninvasive methods. Int Heart J. 2013;54(6):395–400.PubMedCrossRef
Metadata
Title
Perioperative Cardiac Output Monitoring Utilizing Non-pulse Contour Methods
Authors
Yohei Fujimoto
Koichi Suehiro
Akira Mukai
Kiyonobu Nishikawa
Publication date
01-12-2017
Publisher
Springer US
Published in
Current Anesthesiology Reports / Issue 4/2017
Electronic ISSN: 2167-6275
DOI
https://doi.org/10.1007/s40140-017-0240-7

Other articles of this Issue 4/2017

Current Anesthesiology Reports 4/2017 Go to the issue

Patient Safety in Anesthesia (A Barbeito, Section Editor)

Improvement Science in Anaesthesia

Geriatric Anesthesia (S Akhtar, Section Editor)

Fluid Management in the Elderly

Patient Safety in Anesthesia (A Barbeito, Section Editor)

Transitions of Care in the Perioperative Period

Advances in Monitoring for Anesthesia (LAH Critchley, Section Editor)

Cardiac Output Monitoring: Validation Studies–how Results Should be Presented