Comparison of monitoring performance of Bioreactance vs. pulse contour during lung recruitment maneuvers

Authors: Pierre Squara, Dominique Rotcajg, Dominique Denjean, Philippe Estagnasie, Alain Brusset

Published in: Critical Care | Issue 4/2009

Login to get access

Abstract

Introduction

This study was designed to test the hypothesis of equivalence in cardiac output (CO) and stroke volume (SV) monitoring capabilities of two devices: non invasive transthoracic bioreactance (NICOM), and a pulse contour analysis (PICCO PC) coupled to transpulmonary thermodilution (PICCO TD).

Methods

We included consecutive patients of a single ICU following cardiac surgery. Continuous minute-by-minute hemodynamic variables obtained from NICOM and PICCO PC were recorded and compared in 20 patients at baseline, during a lung recruitment maneuver (20 cmH2O of PEEP) and following withdrawal of PEEP. PICCO TD measurements were also determined. We evaluated the accuracy of these two technologies at baseline using PICCO TD as reference and we estimated the precision by the fluctuation around the mean value (2SD/mean). Then, we assessed time response, amplitude response and reliability for detecting expected decreases when PEEP was applied. Type I and type II errors were analyzed.

Results

CO values (PICCO TD) ranged from 1.6 to 8.0 L.min-1. At baseline, CO values were comparable for NICOM, PICCO PC and PICCO TD: 5.0 ± 1.2, 4.7 ± 1.4 and 4.6 ± 1.3 L.min.-1, respectively (NS). Limits of agreements with PICCO TD were 1.52 L.min.-1 for NICOM and 1.77 L.min.-1 for PICCO PC, NS. The 95% statistical power gives an equivalence with a threshold of 0.52 L.min.-1 for NICOM vs. PICCO PC. The CO precision was 6 ± 3% and 6 ± 5% for NICOM and PICCO PC, respectively, NS. When PEEP was applied, CO was reduced by 33 ± 12%, 31 ± 14% and 32 ± 13%, for NICOM, PICCO PC and PICCO TD, respectively (NS). Time response was 3.2 ± 0.7 minute for NICOM vs. 2 ± 0.5 minute for PICCO PC (NS). SV results were comparable to those for CO.

Conclusions

Although limited to 20 patients, this study has enough power to show comparable CO and SV monitoring capabilities of Bioreactance and pulse contour analysis calibrated by transpulmonary thermodilution.
Appendix
Available only for authorised users
Literature
1.
go back to reference Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D: Catheterization of the heart in man with used of a flow directed balloon-tipped catheter. N Engl J Med 1970, 283: 447-451.CrossRefPubMed Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D: Catheterization of the heart in man with used of a flow directed balloon-tipped catheter. N Engl J Med 1970, 283: 447-451.CrossRefPubMed
2.
go back to reference Ganz W, Donoso R, Marcus H, Forrester J, Swan H: A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 1971, 27: 392-396. 10.1016/0002-9149(71)90436-XCrossRefPubMed Ganz W, Donoso R, Marcus H, Forrester J, Swan H: A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 1971, 27: 392-396. 10.1016/0002-9149(71)90436-XCrossRefPubMed
3.
go back to reference Biais M, Nouette-Gaulain K, Cottenceau V, Revel P, Sztark F: Uncalibrated pulse contour-derived stroke volume variation predicts fluid responsiveness in mechanically ventilated patients undergoing liver transplantation. Br J Anaesth 2008, 101: 761-768. 10.1093/bja/aen277CrossRefPubMed Biais M, Nouette-Gaulain K, Cottenceau V, Revel P, Sztark F: Uncalibrated pulse contour-derived stroke volume variation predicts fluid responsiveness in mechanically ventilated patients undergoing liver transplantation. Br J Anaesth 2008, 101: 761-768. 10.1093/bja/aen277CrossRefPubMed
4.
go back to reference Lamia B, Ochagavia A, Monnet X, Chemla D, Richard C, Teboul JL: Echocardiographic prediction of volume responsiveness in critically ill patients with spontaneously breathing activity. Intensive Care Med 2007, 33: 1125-1132. 10.1007/s00134-007-0646-7CrossRefPubMed Lamia B, Ochagavia A, Monnet X, Chemla D, Richard C, Teboul JL: Echocardiographic prediction of volume responsiveness in critically ill patients with spontaneously breathing activity. Intensive Care Med 2007, 33: 1125-1132. 10.1007/s00134-007-0646-7CrossRefPubMed
5.
go back to reference Hofer C, Senn A, Weibel L, Zollinger A: Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac and PiCCO plus system. Crit Care 2008, 12: R82. 10.1186/cc6933PubMedCentralCrossRefPubMed Hofer C, Senn A, Weibel L, Zollinger A: Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac and PiCCO plus system. Crit Care 2008, 12: R82. 10.1186/cc6933PubMedCentralCrossRefPubMed
6.
go back to reference Squara P, Cecconi C, Singer M, Rhodes A, Chiche J: Tracking changes in cardiac output; Methodological considerations for the validation of monitoring devices. Intensive Care Med 2009, in press. Squara P, Cecconi C, Singer M, Rhodes A, Chiche J: Tracking changes in cardiac output; Methodological considerations for the validation of monitoring devices. Intensive Care Med 2009, in press.
7.
go back to reference Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds R: Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output. Crit Care 2009, 13: 201. 10.1186/cc7129PubMedCentralCrossRefPubMed Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds R: Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output. Crit Care 2009, 13: 201. 10.1186/cc7129PubMedCentralCrossRefPubMed
8.
go back to reference Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C: Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 2007, 33: 1191-1194. 10.1007/s00134-007-0640-0CrossRefPubMed Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C: Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 2007, 33: 1191-1194. 10.1007/s00134-007-0640-0CrossRefPubMed
9.
go back to reference Keren H, Burkhoff D, Squara P: Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol 2007, 293: H583-589. 10.1152/ajpheart.00195.2007CrossRefPubMed Keren H, Burkhoff D, Squara P: Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol 2007, 293: H583-589. 10.1152/ajpheart.00195.2007CrossRefPubMed
10.
go back to reference Raval N, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D: Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 2008, 22: 113-119.CrossRefPubMed Raval N, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D: Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput 2008, 22: 113-119.CrossRefPubMed
11.
go back to reference Marqué S, Cariou A, Chiche J, Squara P: Non Invasive Cardiac Output Monitoring (NICOM) Compared to Minimally Invasive Monitoring (VIGILEO). Crit Care 2009, 13: R73. 10.1186/cc7884PubMedCentralCrossRefPubMed Marqué S, Cariou A, Chiche J, Squara P: Non Invasive Cardiac Output Monitoring (NICOM) Compared to Minimally Invasive Monitoring (VIGILEO). Crit Care 2009, 13: R73. 10.1186/cc7884PubMedCentralCrossRefPubMed
12.
go back to reference de Wilde R, Breukers R, Berg P, Jansen J: Monitoring cardiac output using the femoral and radial arterial pressure waveform. Anaesthesia 2006, 61: 743-746. 10.1111/j.1365-2044.2006.04712.xCrossRefPubMed de Wilde R, Breukers R, Berg P, Jansen J: Monitoring cardiac output using the femoral and radial arterial pressure waveform. Anaesthesia 2006, 61: 743-746. 10.1111/j.1365-2044.2006.04712.xCrossRefPubMed
13.
go back to reference Squara P: Bioreactance, A new method for non invasive cardiac output monitoring. In Year book of Intensive Care And Emergency Medicine. Edited by: Vincent J. Paris: Springer; 2008:619-630.CrossRef Squara P: Bioreactance, A new method for non invasive cardiac output monitoring. In Year book of Intensive Care And Emergency Medicine. Edited by: Vincent J. Paris: Springer; 2008:619-630.CrossRef
14.
go back to reference Claxton B, Morgan P, McKeague H, Mulpur A, Berridge J: Alveolar recruitment strategy improves arterial oxygenation after cardiopulmonary bypass. Anaesthesia 2003, 58: 111-116. 10.1046/j.1365-2044.2003.02892.xCrossRefPubMed Claxton B, Morgan P, McKeague H, Mulpur A, Berridge J: Alveolar recruitment strategy improves arterial oxygenation after cardiopulmonary bypass. Anaesthesia 2003, 58: 111-116. 10.1046/j.1365-2044.2003.02892.xCrossRefPubMed
15.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed
16.
go back to reference Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999, 15: 85-91. 10.1023/A:1009982611386CrossRefPubMed Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999, 15: 85-91. 10.1023/A:1009982611386CrossRefPubMed
17.
go back to reference Altman D, Bland J: Absence of evidence is not evidence of absence. Aust Vet J 1996, 74: 311. 10.1111/j.1751-0813.1996.tb13786.xCrossRefPubMed Altman D, Bland J: Absence of evidence is not evidence of absence. Aust Vet J 1996, 74: 311. 10.1111/j.1751-0813.1996.tb13786.xCrossRefPubMed
18.
go back to reference Sagan C: The Demon-Haunted World: Science as a Candle in the Dark. New York: Random House Publishing Group; 1995. Sagan C: The Demon-Haunted World: Science as a Candle in the Dark. New York: Random House Publishing Group; 1995.
19.
go back to reference Henning R: Effects of positive end-expiratory pressure on the right ventricle. J Appl Physiol 1986, 61: 819-826.PubMed Henning R: Effects of positive end-expiratory pressure on the right ventricle. J Appl Physiol 1986, 61: 819-826.PubMed
20.
go back to reference Lim S, Adams A, Simonson D, Dries D, Broccard A, Hotchkiss J, Marini J: Transient hemodynamic effects of recruitment maneuvers in three experimental models of acute lung injury. Crit Care Med 2004, 32: 2378-2384. 10.1097/01.CCM.0000147444.58070.72CrossRefPubMed Lim S, Adams A, Simonson D, Dries D, Broccard A, Hotchkiss J, Marini J: Transient hemodynamic effects of recruitment maneuvers in three experimental models of acute lung injury. Crit Care Med 2004, 32: 2378-2384. 10.1097/01.CCM.0000147444.58070.72CrossRefPubMed
21.
go back to reference Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 2003, 50: 707-711. 10.1007/BF03018714CrossRefPubMed Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P: Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 2003, 50: 707-711. 10.1007/BF03018714CrossRefPubMed
22.
go back to reference Combes A, Berneau JB, Luyt CE, Trouillet JL: Estimation of left ventricular systolic function by single transpulmonary thermodilution. Intensive Care Med 2004, 30: 1377-1383.PubMed Combes A, Berneau JB, Luyt CE, Trouillet JL: Estimation of left ventricular systolic function by single transpulmonary thermodilution. Intensive Care Med 2004, 30: 1377-1383.PubMed
23.
go back to reference Hüter L, Schwarzkopf K, Preussler N, Schubert H, Schreiber T: The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiothorac Vasc Anesth 2007, 21: 659-663. 10.1053/j.jvca.2007.01.005CrossRefPubMed Hüter L, Schwarzkopf K, Preussler N, Schubert H, Schreiber T: The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model. J Cardiothorac Vasc Anesth 2007, 21: 659-663. 10.1053/j.jvca.2007.01.005CrossRefPubMed
Metadata
Title
Comparison of monitoring performance of Bioreactance vs. pulse contour during lung recruitment maneuvers
Authors
Pierre Squara
Dominique Rotcajg
Dominique Denjean
Philippe Estagnasie
Alain Brusset
Publication date
01-08-2009
Publisher
BioMed Central
Published in
Critical Care / Issue 4/2009
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc7981

Other articles of this Issue 4/2009

Critical Care 4/2009 Go to the issue