Published in: Journal of Nuclear Cardiology 2/2018

Open Access 01-04-2018 | Original Article

Value of automatic patient motion detection and correction in myocardial perfusion imaging using a CZT-based SPECT camera

Authors: Joris D. van Dijk, MSc, Jorn A. van Dalen, PhD, Mohamed Mouden, MD, PhD, Jan Paul Ottervanger, MD, PhD, Siert Knollema, MD, PhD, MSc, Cornelis H. Slump, PhD, Pieter L. Jager, MD, PhD

Published in: Journal of Nuclear Cardiology | Issue 2/2018

Login to get access

Abstract

Background

Correction of motion has become feasible on cadmium-zinc-telluride (CZT)-based SPECT cameras during myocardial perfusion imaging (MPI). Our aim was to quantify the motion and to determine the value of automatic correction using commercially available software.

Methods and Results

We retrospectively included 83 consecutive patients who underwent stress-rest MPI CZT-SPECT and invasive fractional flow reserve (FFR) measurement. Eight-minute stress acquisitions were reformatted into 1.0- and 20-second bins to detect respiratory motion (RM) and patient motion (PM), respectively. RM and PM were quantified and scans were automatically corrected. Total perfusion deficit (TPD) and SPECT interpretation—normal, equivocal, or abnormal—were compared between the noncorrected and corrected scans. Scans with a changed SPECT interpretation were compared with FFR, the reference standard. Average RM was 2.5 ± 0.4 mm and maximal PM was 4.5 ± 1.3 mm. RM correction influenced the diagnostic outcomes in two patients based on TPD changes ≥7% and in nine patients based on changed visual interpretation. In only four of these patients, the changed SPECT interpretation corresponded with FFR measurements. Correction for PM did not influence the diagnostic outcomes.

Conclusion

Respiratory motion and patient motion were small. Motion correction did not appear to improve the diagnostic outcome and, hence, the added value seems limited in MPI using CZT-based SPECT cameras.
Appendix
Available only for authorised users
Literature
1.
go back to reference Underwood SR, Anagnostopoulos C, Cerqueira MD, Ell PJ, Flint EJ, Harbinson M, et al. Myocardial perfusion scintigraphy: The evidence. Eur J Nucl Med Mol Imaging. 2004;31:261-91.CrossRefPubMed Underwood SR, Anagnostopoulos C, Cerqueira MD, Ell PJ, Flint EJ, Harbinson M, et al. Myocardial perfusion scintigraphy: The evidence. Eur J Nucl Med Mol Imaging. 2004;31:261-91.CrossRefPubMed
2.
go back to reference Depuey EG, Mahmarian JJ, Miller TD, Einstein AJ, Hansen CL, Holly TA, et al. Patient-centered imaging. J Nucl Cardiol. 2012;19:185-215.CrossRefPubMed Depuey EG, Mahmarian JJ, Miller TD, Einstein AJ, Hansen CL, Holly TA, et al. Patient-centered imaging. J Nucl Cardiol. 2012;19:185-215.CrossRefPubMed
3.
go back to reference Heller GV, Bateman TM, Johnson LL, Cullom SJ, Case JA, Galt JR, et al. Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol. 2004;11:273-81.CrossRefPubMed Heller GV, Bateman TM, Johnson LL, Cullom SJ, Case JA, Galt JR, et al. Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol. 2004;11:273-81.CrossRefPubMed
4.
go back to reference van Dijk JD, Jager PL, Mouden M, Slump CH, Ottervanger JP, de Boer J, et al. Development and validation of a patient-tailored dose regime in myocardial perfusion imaging using CZT-SPECT. J Nucl Cardiol. 2014;21:1158-67.CrossRefPubMed van Dijk JD, Jager PL, Mouden M, Slump CH, Ottervanger JP, de Boer J, et al. Development and validation of a patient-tailored dose regime in myocardial perfusion imaging using CZT-SPECT. J Nucl Cardiol. 2014;21:1158-67.CrossRefPubMed
5.
go back to reference Agarwal V, DePuey EG. Myocardial perfusion SPECT horizontal motion artifact. J Nucl Cardiol. 2014;21:1260-5.CrossRefPubMed Agarwal V, DePuey EG. Myocardial perfusion SPECT horizontal motion artifact. J Nucl Cardiol. 2014;21:1260-5.CrossRefPubMed
6.
go back to reference Buechel RR, Pazhenkottil AP, Herzog BA, Husmann L, Nkoulou RN, Burger IA, et al. Real-time breath-hold triggering of myocardial perfusion imaging with a novel cadmium-zinc-telluride detector gamma camera. Eur J Nucl Med Mol Imaging. 2010;37:1903-8.CrossRefPubMed Buechel RR, Pazhenkottil AP, Herzog BA, Husmann L, Nkoulou RN, Burger IA, et al. Real-time breath-hold triggering of myocardial perfusion imaging with a novel cadmium-zinc-telluride detector gamma camera. Eur J Nucl Med Mol Imaging. 2010;37:1903-8.CrossRefPubMed
7.
go back to reference Cooper JA, Neumann PH, Mccandless BK. Effect of patient motion on tomographic myocardial perfusion imaging. J Nucl Med. 1992;33:1566-72.PubMed Cooper JA, Neumann PH, Mccandless BK. Effect of patient motion on tomographic myocardial perfusion imaging. J Nucl Med. 1992;33:1566-72.PubMed
8.
go back to reference Ko CL, Wu YW, Cheng MF, Yen RF, Wu WC, Tzen KY. Data-driven respiratory motion tracking and compensation in CZT cameras: A comprehensive analysis of phantom and human images. J Nucl Cardiol. 2014;22:308-18.CrossRefPubMed Ko CL, Wu YW, Cheng MF, Yen RF, Wu WC, Tzen KY. Data-driven respiratory motion tracking and compensation in CZT cameras: A comprehensive analysis of phantom and human images. J Nucl Cardiol. 2014;22:308-18.CrossRefPubMed
9.
go back to reference Matsumoto N, Berman DS, Kavanagh PB, Gerlach J, Hayes SW, Lewin HC, et al. Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT. J Nucl Med. 2001;42:687-94.PubMed Matsumoto N, Berman DS, Kavanagh PB, Gerlach J, Hayes SW, Lewin HC, et al. Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT. J Nucl Med. 2001;42:687-94.PubMed
10.
go back to reference Eisner R, Churchwell A, Noever T, Nowak D, Cloninger K, Dunn D, et al. Quantitative analysis of the tomographic thallium-201 myocardial bullseye display: Critical role of correcting for patient motion. J Nucl Med. 1988;29:91-7.PubMed Eisner R, Churchwell A, Noever T, Nowak D, Cloninger K, Dunn D, et al. Quantitative analysis of the tomographic thallium-201 myocardial bullseye display: Critical role of correcting for patient motion. J Nucl Med. 1988;29:91-7.PubMed
11.
go back to reference Bai C, Maddahi J, Kindem J, Conwell R, Gurley M, Old R. Development and evaluation of a new fully automatic motion detection and correction technique in cardiac SPECT imaging. J Nucl Cardiol. 2009;16:580-9.CrossRefPubMedPubMedCentral Bai C, Maddahi J, Kindem J, Conwell R, Gurley M, Old R. Development and evaluation of a new fully automatic motion detection and correction technique in cardiac SPECT imaging. J Nucl Cardiol. 2009;16:580-9.CrossRefPubMedPubMedCentral
12.
go back to reference Massardo T, Jaimovich R, Faure R, Muñoz M, Alay R, Gatica H. Motion correction and myocardial perfusion SPECT using manufacturer provided software. Does it affect image interpretation? Eur J Nucl Med Mol Imaging. 2010;37:758-64.CrossRefPubMed Massardo T, Jaimovich R, Faure R, Muñoz M, Alay R, Gatica H. Motion correction and myocardial perfusion SPECT using manufacturer provided software. Does it affect image interpretation? Eur J Nucl Med Mol Imaging. 2010;37:758-64.CrossRefPubMed
13.
go back to reference Redgate S, Barber DC, Fenner JW, Al-Mohammad A, Taylor JC, Hanney MB, Tindale WB. A study to quantify the effect of patient motion and develop methods to detect and correct for motion during myocardial perfusion imaging on a CZT solid-state dedicated cardiac camera. J Nucl Cardiol. 2016;23:514-26.CrossRefPubMed Redgate S, Barber DC, Fenner JW, Al-Mohammad A, Taylor JC, Hanney MB, Tindale WB. A study to quantify the effect of patient motion and develop methods to detect and correct for motion during myocardial perfusion imaging on a CZT solid-state dedicated cardiac camera. J Nucl Cardiol. 2016;23:514-26.CrossRefPubMed
14.
go back to reference Mouden M, Ottervanger JP, Knollema S, Timmer JR, Reiffers S, Oostdijk AH, et al. Myocardial perfusion imaging with a cadmium zinc telluride-based gamma camera versus invasive fractional flow reserve. Eur J Nucl Med Mol Imaging. 2013;41:956-62.CrossRefPubMed Mouden M, Ottervanger JP, Knollema S, Timmer JR, Reiffers S, Oostdijk AH, et al. Myocardial perfusion imaging with a cadmium zinc telluride-based gamma camera versus invasive fractional flow reserve. Eur J Nucl Med Mol Imaging. 2013;41:956-62.CrossRefPubMed
15.
go back to reference Mouden M, Timmer JR, Ottervanger JP, Reiffers S, Oostdijk AH, Knollema S, Jager PL. Impact of a new ultrafast CZT SPECT camera for myocardial perfusion imaging: Fewer equivocal results and lower radiation dose. Eur J Nucl Med Mol Imaging. 2012;39:1048-55.CrossRefPubMed Mouden M, Timmer JR, Ottervanger JP, Reiffers S, Oostdijk AH, Knollema S, Jager PL. Impact of a new ultrafast CZT SPECT camera for myocardial perfusion imaging: Fewer equivocal results and lower radiation dose. Eur J Nucl Med Mol Imaging. 2012;39:1048-55.CrossRefPubMed
16.
go back to reference Koopman D, van Dalen JA, Slump C, Lots D, Timmer JR, Jager PL. Impact of image processing in the detection of ischemia using CZT-SPECT/CT. Nucl Med Commun. 2015;36:60-8.CrossRefPubMed Koopman D, van Dalen JA, Slump C, Lots D, Timmer JR, Jager PL. Impact of image processing in the detection of ischemia using CZT-SPECT/CT. Nucl Med Commun. 2015;36:60-8.CrossRefPubMed
17.
go back to reference Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging. 2010;37:773-8.CrossRefPubMed Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging. 2010;37:773-8.CrossRefPubMed
18.
go back to reference Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol. 2009;16:927-34.CrossRefPubMedPubMedCentral Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol. 2009;16:927-34.CrossRefPubMedPubMedCentral
19.
go back to reference Duvall WL, Croft LB, Godiwala T, Ginsberg E, George T, Henzlova MJ. Reduced isotope dose with rapid SPECT MPI imaging: Initial experience with a CZT SPECT camera. J Nucl Cardiol. 2010;17:1009-14.CrossRefPubMed Duvall WL, Croft LB, Godiwala T, Ginsberg E, George T, Henzlova MJ. Reduced isotope dose with rapid SPECT MPI imaging: Initial experience with a CZT SPECT camera. J Nucl Cardiol. 2010;17:1009-14.CrossRefPubMed
20.
go back to reference Berman DS, Kang X, Gransar H, Gerlach J, Friedman JD, Hayes SW, et al. Quantitative assessment of myocardial perfusion abnormality on SPECT myocardial perfusion imaging is more reproducible than expert visual analysis. J Nucl Cardiol. 2009;16:45-53.CrossRefPubMedPubMedCentral Berman DS, Kang X, Gransar H, Gerlach J, Friedman JD, Hayes SW, et al. Quantitative assessment of myocardial perfusion abnormality on SPECT myocardial perfusion imaging is more reproducible than expert visual analysis. J Nucl Cardiol. 2009;16:45-53.CrossRefPubMedPubMedCentral
21.
go back to reference Tonino PAL, De Bruyne B, Pijls NHJ, Siebert U, Ikeno F, van Veer M, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med. 2009;360:213-24.CrossRefPubMed Tonino PAL, De Bruyne B, Pijls NHJ, Siebert U, Ikeno F, van Veer M, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med. 2009;360:213-24.CrossRefPubMed
22.
go back to reference Iskandrian AE, Hage FG, Shaw LJ, Mahmarian JJ, Berman DS. Serial myocardial perfusion imaging. JACC Cardiovasc Imaging. 2014;7:79-96.CrossRefPubMed Iskandrian AE, Hage FG, Shaw LJ, Mahmarian JJ, Berman DS. Serial myocardial perfusion imaging. JACC Cardiovasc Imaging. 2014;7:79-96.CrossRefPubMed
23.
go back to reference Berman DS, Abidov A, Kang X, Hayes SW, Friedman JD, Sciammarella MG, et al. Prognostic validation of a 17-segment score derived from a 20-segment score for myocardial perfusion SPECT interpretation. J Nucl Cardiol. 2004;11:414-23.CrossRefPubMed Berman DS, Abidov A, Kang X, Hayes SW, Friedman JD, Sciammarella MG, et al. Prognostic validation of a 17-segment score derived from a 20-segment score for myocardial perfusion SPECT interpretation. J Nucl Cardiol. 2004;11:414-23.CrossRefPubMed
24.
go back to reference Shaw LJ, Berman DS, Maron DJ, Mancini GB, Hayes SW, Hartigan PM, et al. Optimal medical therapy with or without percutaneous coronary intervention to reduce ischemic burden: Results from the Clinical Outcomes Utilizing Revascularization and Aggressive Drug Evaluation (COURAGE) trial nuclear substudy. Circulation. 2008;117:1283-91.CrossRefPubMed Shaw LJ, Berman DS, Maron DJ, Mancini GB, Hayes SW, Hartigan PM, et al. Optimal medical therapy with or without percutaneous coronary intervention to reduce ischemic burden: Results from the Clinical Outcomes Utilizing Revascularization and Aggressive Drug Evaluation (COURAGE) trial nuclear substudy. Circulation. 2008;117:1283-91.CrossRefPubMed
25.
go back to reference Clerc OF, Fuchs TA, Possner M, Vontobel J, Mikulicic F, Stehli J, et al. Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: Impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction. Eur Heart J Cardiovasc Imaging. 2016. doi:10.1093/ehjci/jew031. Clerc OF, Fuchs TA, Possner M, Vontobel J, Mikulicic F, Stehli J, et al. Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: Impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction. Eur Heart J Cardiovasc Imaging. 2016. doi:10.​1093/​ehjci/​jew031.
26.
go back to reference van Dijk JD, Jager PL, Ottervanger JP, Slump CH, de Boer J, Oostdijk AH, van Dalen JA. Minimizing patient-specific tracer dose in myocardial perfusion imaging using CZT SPECT. J Nucl Med Technol. 2015;43:36-40.CrossRefPubMed van Dijk JD, Jager PL, Ottervanger JP, Slump CH, de Boer J, Oostdijk AH, van Dalen JA. Minimizing patient-specific tracer dose in myocardial perfusion imaging using CZT SPECT. J Nucl Med Technol. 2015;43:36-40.CrossRefPubMed
27.
go back to reference Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, et al. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med. 2013;54:1-7.CrossRef Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, et al. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med. 2013;54:1-7.CrossRef
28.
go back to reference Engbers EM, Timmer JR, Ottervanger JP, Mouden M, Oostdijk AH, Knollema S, Jager PL. Sequential SPECT/CT imaging for detection of coronary artery disease in a large cohort: evaluation of the need for additional imaging and radiation exposure. J Nucl Cardiol. 2015. doi:10.1007/s12350-015-0243-z. Engbers EM, Timmer JR, Ottervanger JP, Mouden M, Oostdijk AH, Knollema S, Jager PL. Sequential SPECT/CT imaging for detection of coronary artery disease in a large cohort: evaluation of the need for additional imaging and radiation exposure. J Nucl Cardiol. 2015. doi:10.​1007/​s12350-015-0243-z.
29.
go back to reference Pijls NHJ, Sels J-W. Functional measurement of coronary stenosis. J Am Coll Cardiol. 2012;59:1045-57.CrossRefPubMed Pijls NHJ, Sels J-W. Functional measurement of coronary stenosis. J Am Coll Cardiol. 2012;59:1045-57.CrossRefPubMed
30.
go back to reference Liga R, Vontobel J, Rovai D, Marinelli M, Caselli C, Pietila M, et al. Multicentre multi-device hybrid imaging study of coronary artery disease: Results from the evaluation of integrated cardiac imaging for the detection and characterization of ischaemic heart disease (EVINCI) hybrid imaging population. Eur Heart J Cardiovasc Imaging. 2016. doi:10.1093/ehjci/jew038. Liga R, Vontobel J, Rovai D, Marinelli M, Caselli C, Pietila M, et al. Multicentre multi-device hybrid imaging study of coronary artery disease: Results from the evaluation of integrated cardiac imaging for the detection and characterization of ischaemic heart disease (EVINCI) hybrid imaging population. Eur Heart J Cardiovasc Imaging. 2016. doi:10.​1093/​ehjci/​jew038.
31.
go back to reference Dey D, Slomka PJ, Berman DS. Achieving very-low-dose radiation exposure in cardiac computed tomography, single-photon emission computed tomography, and positron emission tomography. Circ Imaging. 2014;7:723-34.CrossRef Dey D, Slomka PJ, Berman DS. Achieving very-low-dose radiation exposure in cardiac computed tomography, single-photon emission computed tomography, and positron emission tomography. Circ Imaging. 2014;7:723-34.CrossRef
32.
go back to reference van Dijk JD, Jager PL, Ottervanger JP, Slump CH, Knollema S, van Dalen JA. Patient-specific tracer activity in MPI SPECT: A hands-on approach. J Nucl Cardiol. 2016;23:145-8.CrossRefPubMed van Dijk JD, Jager PL, Ottervanger JP, Slump CH, Knollema S, van Dalen JA. Patient-specific tracer activity in MPI SPECT: A hands-on approach. J Nucl Cardiol. 2016;23:145-8.CrossRefPubMed
Metadata
Title
Value of automatic patient motion detection and correction in myocardial perfusion imaging using a CZT-based SPECT camera
Authors
Joris D. van Dijk, MSc
Jorn A. van Dalen, PhD
Mohamed Mouden, MD, PhD
Jan Paul Ottervanger, MD, PhD
Siert Knollema, MD, PhD, MSc
Cornelis H. Slump, PhD
Pieter L. Jager, MD, PhD
Publication date
01-04-2018
Publisher
Springer US
Published in
Journal of Nuclear Cardiology / Issue 2/2018
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-016-0571-7

Other articles of this Issue 2/2018

Journal of Nuclear Cardiology 2/2018 Go to the issue