Effect of resolution recovery using graph plots on regional cerebral blood flow in healthy volunteers

Authors: Nobuhiro Yada, Hideo Onishi, Masahiro Miyai, Kentarou Ozasa, Takashi Katsube, Keiichi Onoda, Masuo Haramoto, Yasushi Yamamoto, Shuhei Yamaguchi, Hajime Kitagaki

Published in: Annals of Nuclear Medicine | Issue 7/2017

Login to get access

Abstract

Purpose

We evaluated the effect of resolution recovery (RR) using graph plots on regional cerebral blood flow (rCBF) in brain perfusion single-photon emission computed tomography (SPECT) images derived from healthy volunteers and patients diagnosed with probable Alzheimer’s disease.

Method

We acquired brain perfusion SPECT images with scatter correction (SC), computed tomography-based attenuation correction (CTAC), and RR from a three-dimensional brain phantom and from healthy volunteers. We then compared contrast-to-noise ratio, count density ratios, increase maps, and rCBF using statistical parametric mapping 8.

Results

Regional brain counts were significantly increased from 20–24% with SC, CTAC, and RR compared with SC and CTAC. Mean CBF in healthy volunteers was 42.5 ± 5.4 mL/100 g/min. Average rCBF determined using SC, CTAC and RR increased 7.5, 2.0, and 3.7% at the thalamus, posterior cingulate, and whole brain, respectively, compared with SC and CTAC.

Conclusion

Resolution recovery caused variations in normal rCBF because counts increased in cerebral regions.
Literature
1.
go back to reference Kuhl DE, Barrio JR, Huang SC, Selin C, Ackermann RF, Lear JL, et al. Quantifying local cerebral blood flow by N-isopropyl-p-[123I] iodoamphetamine (IMP) tomography. J Nucl Med. 1982;23:196–203.PubMed Kuhl DE, Barrio JR, Huang SC, Selin C, Ackermann RF, Lear JL, et al. Quantifying local cerebral blood flow by N-isopropyl-p-[123I] iodoamphetamine (IMP) tomography. J Nucl Med. 1982;23:196–203.PubMed
2.
go back to reference Mosconi L, Tsui WH, Herholz K, Pupi A, Drzezga A, Lucignani G, et al. Multicenter standardized 18F-FDG PET diagnosis of mild cognitive impairment, Alzheimer’s disease, and other dementia. J Nucl Med. 2008;49:390–8.CrossRefPubMedPubMedCentral Mosconi L, Tsui WH, Herholz K, Pupi A, Drzezga A, Lucignani G, et al. Multicenter standardized 18F-FDG PET diagnosis of mild cognitive impairment, Alzheimer’s disease, and other dementia. J Nucl Med. 2008;49:390–8.CrossRefPubMedPubMedCentral
3.
go back to reference Staffen W, Schonauer U, Zauner H, Spindler I, Mair A, Iglseder B, et al. Brain perfusion SPECT in patients with mild cognitive impairment and Alzheimer’s disease: comparison of a semiquantitative and a visual evaluation. J Neural Transm. 2006;113:195–203.CrossRefPubMed Staffen W, Schonauer U, Zauner H, Spindler I, Mair A, Iglseder B, et al. Brain perfusion SPECT in patients with mild cognitive impairment and Alzheimer’s disease: comparison of a semiquantitative and a visual evaluation. J Neural Transm. 2006;113:195–203.CrossRefPubMed
4.
go back to reference Chida K, Ogasawara K, Kuroda H, Aso K, Kobayashi M, Fujiwara S, et al. Central benzodiazepine receptor binding potential and CBF images on SPECT correlate with oxygen extraction fraction images on PET in the cerebral cortex with unilateral major cerebral artery occlusive disease. J Nucl Med. 2011;52:511–8.CrossRefPubMed Chida K, Ogasawara K, Kuroda H, Aso K, Kobayashi M, Fujiwara S, et al. Central benzodiazepine receptor binding potential and CBF images on SPECT correlate with oxygen extraction fraction images on PET in the cerebral cortex with unilateral major cerebral artery occlusive disease. J Nucl Med. 2011;52:511–8.CrossRefPubMed
5.
go back to reference Iida H, Ito H, Nakazawa M, Hatazawa J, Nishimura H, Onishi Y, et al. Quantitative mapping of regional cerebral blood flow using iodine-123-IMP and SPECT. J Nucl Med. 1994;35(12):2019–30.PubMed Iida H, Ito H, Nakazawa M, Hatazawa J, Nishimura H, Onishi Y, et al. Quantitative mapping of regional cerebral blood flow using iodine-123-IMP and SPECT. J Nucl Med. 1994;35(12):2019–30.PubMed
6.
go back to reference Iida H, Akutsu T, Endo K, Fukuda H, Inoue T, Ito H, et al. A multicenter validation of regional cerebral blood flow quantitation using [123I] iodoamphetamine and single photon emission computed tomography. J Cereb Blood Flow Metab. 1996;16(5):781–93.CrossRefPubMed Iida H, Akutsu T, Endo K, Fukuda H, Inoue T, Ito H, et al. A multicenter validation of regional cerebral blood flow quantitation using [123I] iodoamphetamine and single photon emission computed tomography. J Cereb Blood Flow Metab. 1996;16(5):781–93.CrossRefPubMed
7.
go back to reference Hatazawa J, Iida H, Shimosegawa E, Sato T, Murakami M, Miura Y. Regional cerebral blood flow measurement with iodine-123-IMP autoradiography: normal values, reproducibility and sensitivity to hypoperfusion. J Nucl Med. 1997;38(7):1102–8.PubMed Hatazawa J, Iida H, Shimosegawa E, Sato T, Murakami M, Miura Y. Regional cerebral blood flow measurement with iodine-123-IMP autoradiography: normal values, reproducibility and sensitivity to hypoperfusion. J Nucl Med. 1997;38(7):1102–8.PubMed
8.
go back to reference Okamoto K, Ushijima Y, Okuyama C, Nakamura T, Nishimura T. Measurement of cerebral blood flow using graph plot analysis and I-123 iodoamphetamine. Clin Nucl Med. 2002;27:191–6.CrossRefPubMed Okamoto K, Ushijima Y, Okuyama C, Nakamura T, Nishimura T. Measurement of cerebral blood flow using graph plot analysis and I-123 iodoamphetamine. Clin Nucl Med. 2002;27:191–6.CrossRefPubMed
9.
go back to reference Ishii K, Umemura T, Miyamoto N, Yoshikawa T, Yamaguchi T, Ashihara T, et al. Regional cerebral blood flow in healthy volunteers measured by the graph plot method with iodoamphetamine SPECT. Ann Nucl Med. 2011;25:255–60.CrossRefPubMed Ishii K, Umemura T, Miyamoto N, Yoshikawa T, Yamaguchi T, Ashihara T, et al. Regional cerebral blood flow in healthy volunteers measured by the graph plot method with iodoamphetamine SPECT. Ann Nucl Med. 2011;25:255–60.CrossRefPubMed
10.
go back to reference Nishizawa S, Shiozaki T, Ueno M, Toyoda H, Shimono T, Kamoto Y, et al. A new method to estimate rCBF using IMP and SPECT without any blood sampling. Ann Nucl Med. 2000;14:433–40.CrossRefPubMed Nishizawa S, Shiozaki T, Ueno M, Toyoda H, Shimono T, Kamoto Y, et al. A new method to estimate rCBF using IMP and SPECT without any blood sampling. Ann Nucl Med. 2000;14:433–40.CrossRefPubMed
11.
go back to reference Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shoji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shoji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed
12.
go back to reference Ghoorun S, Baete K, Nuyts J, Groenewald W, Dupont P. The influence of attenuation correction and reconstruction techniques on the detection of hypo-perfused lesions in brain SPECT images. Nucl Med Commun. 2006;27:765–72.CrossRefPubMed Ghoorun S, Baete K, Nuyts J, Groenewald W, Dupont P. The influence of attenuation correction and reconstruction techniques on the detection of hypo-perfused lesions in brain SPECT images. Nucl Med Commun. 2006;27:765–72.CrossRefPubMed
13.
go back to reference Shiga T, Kubo N, Takano A, Kobayashi J, Takeda Y, Nakamura F, et al. The effect of scatter correction on 123I-IMP brain perfusion SPET with the triple energy window method in normal subjects using SPM analysis. Eur J Nucl Mol Imaging. 2002;29:342–5.CrossRef Shiga T, Kubo N, Takano A, Kobayashi J, Takeda Y, Nakamura F, et al. The effect of scatter correction on 123I-IMP brain perfusion SPET with the triple energy window method in normal subjects using SPM analysis. Eur J Nucl Mol Imaging. 2002;29:342–5.CrossRef
14.
go back to reference Ishii K, Hanaoka K, Okada M, Kumano S, Komeya Y, Tsuchiya M, et al. Impact of CT attenuation correction by SPECT/CT in brain perfusion images. Ann Nucl Med. 2012;26:241–7.CrossRefPubMed Ishii K, Hanaoka K, Okada M, Kumano S, Komeya Y, Tsuchiya M, et al. Impact of CT attenuation correction by SPECT/CT in brain perfusion images. Ann Nucl Med. 2012;26:241–7.CrossRefPubMed
15.
go back to reference Yokoi T, Shinohara H, Onishi H. Performance evaluation of OSEM reconstruction algorithm incorporating three-dimensional distance-dependent resolution compensation for brain SPECT: a phantom study. Ann Nucl Med. 2002;16:11–8.CrossRefPubMed Yokoi T, Shinohara H, Onishi H. Performance evaluation of OSEM reconstruction algorithm incorporating three-dimensional distance-dependent resolution compensation for brain SPECT: a phantom study. Ann Nucl Med. 2002;16:11–8.CrossRefPubMed
16.
go back to reference Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed
17.
go back to reference Knoll P, Kotalova D, Kochle G, Kuzelka I, Minear G, Mizaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys. 2012;22:58–69.CrossRefPubMed Knoll P, Kotalova D, Kochle G, Kuzelka I, Minear G, Mizaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys. 2012;22:58–69.CrossRefPubMed
18.
go back to reference Shimosegawa E, Fujino K, Kato H, Hatazawa J. Quantitative CBF measurement using an integrated SPECT/CT system: validation of three-dimensional ordered-subset expectation maximization and CT-based attenuation correction by comparing with O-15 water PET. Ann Nucl Med. 2013;27:822–33.CrossRefPubMed Shimosegawa E, Fujino K, Kato H, Hatazawa J. Quantitative CBF measurement using an integrated SPECT/CT system: validation of three-dimensional ordered-subset expectation maximization and CT-based attenuation correction by comparing with O-15 water PET. Ann Nucl Med. 2013;27:822–33.CrossRefPubMed
19.
go back to reference Okada M, Hayashi M, Tsuji H, Akagi H, Okayama K, Narumi Y. Application of collimator broad correction three dimensional ordered subset expectation maximization for regional cerebral blood flow measurement. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68:573–83 (Japanese).CrossRefPubMed Okada M, Hayashi M, Tsuji H, Akagi H, Okayama K, Narumi Y. Application of collimator broad correction three dimensional ordered subset expectation maximization for regional cerebral blood flow measurement. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68:573–83 (Japanese).CrossRefPubMed
20.
go back to reference Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Evaluation of commercial resolution recovery techniques in four state-of-the-art single photon emission computed tomography systems using a digital phantom model. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68:686–96 (Japanese).CrossRefPubMed Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Evaluation of commercial resolution recovery techniques in four state-of-the-art single photon emission computed tomography systems using a digital phantom model. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68:686–96 (Japanese).CrossRefPubMed
21.
go back to reference Arosio M, Pasquali C, Crivellaro C, De Ponti E, Morzenti S, Guerra L, et al. Performance of a SPECT collimator-detector response reconstruction algorithm: phantom studies and validation in inflammation clinical studies. Q J Nucl Mol Imaging. 2011;55:671–9. Arosio M, Pasquali C, Crivellaro C, De Ponti E, Morzenti S, Guerra L, et al. Performance of a SPECT collimator-detector response reconstruction algorithm: phantom studies and validation in inflammation clinical studies. Q J Nucl Mol Imaging. 2011;55:671–9.
22.
go back to reference Yada N, Onishi H, Miyai M, Ozasa K, Haramoto M, Yamamoto Y, et al. Evaluation of resolution recovery for each collimator in brain perfusion image. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2016;72:81–9.CrossRef Yada N, Onishi H, Miyai M, Ozasa K, Haramoto M, Yamamoto Y, et al. Evaluation of resolution recovery for each collimator in brain perfusion image. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2016;72:81–9.CrossRef
23.
go back to reference Iida H, Hori Y, Ishida K, Imabayashi E, Matsuda H, Takahashi M, et al. Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour. Ann Nucl Med. 2013;27(1):25–36.CrossRefPubMed Iida H, Hori Y, Ishida K, Imabayashi E, Matsuda H, Takahashi M, et al. Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour. Ann Nucl Med. 2013;27(1):25–36.CrossRefPubMed
24.
go back to reference Onishi H, Matsutomo N, Kai Y, Kangai Y, Amijima H, Yamaguchi T. Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP. Ann Nucl Med. 2012;26:16–25.CrossRefPubMed Onishi H, Matsutomo N, Kai Y, Kangai Y, Amijima H, Yamaguchi T. Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP. Ann Nucl Med. 2012;26:16–25.CrossRefPubMed
25.
go back to reference Yada N, Onishi H. Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values in four computed tomography scanners. Asia Oceania J Nucl Med Biol. 2016;4:81–9. Yada N, Onishi H. Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values in four computed tomography scanners. Asia Oceania J Nucl Med Biol. 2016;4:81–9.
26.
go back to reference Yamamoto Y, Onoguchi M, Kawakami K, Haramoto M, Wake R, Horiguchi J, et al. Evaluation of the difference-correction effect of the gamma camera systems used by easy Z-score Imaging System (eZIS) analysis. Ann Nucl Med. 2014;28:263–75.CrossRefPubMedPubMedCentral Yamamoto Y, Onoguchi M, Kawakami K, Haramoto M, Wake R, Horiguchi J, et al. Evaluation of the difference-correction effect of the gamma camera systems used by easy Z-score Imaging System (eZIS) analysis. Ann Nucl Med. 2014;28:263–75.CrossRefPubMedPubMedCentral
27.
go back to reference Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated Talairach Atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.CrossRefPubMed Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated Talairach Atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.CrossRefPubMed
28.
go back to reference Lancaster JL, Rainey LH, Summerlin JL, Freitas CS, Fox PT, Evans AC, et al. Automated labeling of the human brain: a preliminary report on the development and evaluation of a forward-transform method. Hum Brain Mapp. 1997;5:238–42.CrossRefPubMedPubMedCentral Lancaster JL, Rainey LH, Summerlin JL, Freitas CS, Fox PT, Evans AC, et al. Automated labeling of the human brain: a preliminary report on the development and evaluation of a forward-transform method. Hum Brain Mapp. 1997;5:238–42.CrossRefPubMedPubMedCentral
29.
go back to reference Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol. 2009;16(5):714–25.CrossRefPubMed Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol. 2009;16(5):714–25.CrossRefPubMed
30.
go back to reference Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51(6):921–8.CrossRefPubMed Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51(6):921–8.CrossRefPubMed
31.
go back to reference Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models. Radiol Phys Technol. 2013;6:42–53.CrossRefPubMed Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models. Radiol Phys Technol. 2013;6:42–53.CrossRefPubMed
Metadata
Title
Effect of resolution recovery using graph plots on regional cerebral blood flow in healthy volunteers
Authors
Nobuhiro Yada
Hideo Onishi
Masahiro Miyai
Kentarou Ozasa
Takashi Katsube
Keiichi Onoda
Masuo Haramoto
Yasushi Yamamoto
Shuhei Yamaguchi
Hajime Kitagaki
Publication date
01-08-2017
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 7/2017
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-017-1186-z

Other articles of this Issue 7/2017

Annals of Nuclear Medicine 7/2017 Go to the issue