Published in: Radiological Physics and Technology 1/2013

01-01-2013

Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models

Authors: Hideo Onishi, Nobutoku Motomura, Koichi Fujino, Takahiro Natsume, Yasuhiro Haramoto

Published in: Radiological Physics and Technology | Issue 1/2013

Login to get access

Abstract

Several resolution recovery (RR) methods have been developed. This study was aimed to validate the following performance of the advanced RR methods: Evolution™, Astonish™, Flash3D™, and 3D-OSEM. We compared the advanced RR method with filtered back projection (FBP) and standard order-subset expectation maximization (OSEM) using resolution (RES), cylinder/sphere (CYS), and myocardial (MYD) digital phantoms. The RES phantom was placed in three spheres. Sixteen spheres (hot and cold) were then placed in a concentric configuration (diameter: 96–9.6 mm) inside the CYS phantom. The MYD phantom was created by computer simulation with the use of an electron γ-shower 4 (EGS4) and it included two left ventricular defects in the myocardium. The performance was evaluated at source-to-detector distances (R-distance) of 166, 200, and 250 mm with reconstruction parameters (product of subset and iteration: SI) with use of the resolution recovery factor, count recovery, normalized mean square error (NMSE), and %CV. According to increased SI updates, the value of the FWHM decreased, and the effect was more obvious as the R-distance increased. The spatial resolution of the advanced RR method was 20 % better than that of FBP and OSEM. The resolution recovery ratio was 80 %, and the count recovery was maintained only in objects with a diameter of >30 mm in the advanced RR method. The NMSE and %CV was 50 and 30 % improved over FBP and OSEM, respectively. The advanced RR method caused overestimation due to Gibbs’s phenomenon in the marginal region when the diameter of the sphere was 16–28.8 mm.
Literature
1.
go back to reference Edholm RP, Lewitt M, Lindholm B. Novel properties of the Fourier decomposition of the sonogram. In: International Workshop on Physics and Engineering of Computerized Multidimensional Imaging and Processing, vol 671. 1986. p. 8–18. Edholm RP, Lewitt M, Lindholm B. Novel properties of the Fourier decomposition of the sonogram. In: International Workshop on Physics and Engineering of Computerized Multidimensional Imaging and Processing, vol 671. 1986. p. 8–18.
2.
go back to reference Kohli V, King AM, Glick JS, Pan TS. Comparison of frequency-distance relationship and Gaussian-diffusion-based methods of compensation for distance-dependent spatial resolution in SPECT imaging. Phys Med Biol. 1998;43:1025–37.PubMedCrossRef Kohli V, King AM, Glick JS, Pan TS. Comparison of frequency-distance relationship and Gaussian-diffusion-based methods of compensation for distance-dependent spatial resolution in SPECT imaging. Phys Med Biol. 1998;43:1025–37.PubMedCrossRef
3.
go back to reference Glick JS, Penney CB, King AM, Byrne CL. Noniterative compensation for the distance-dependent detector response and photon attenuation in SPECT imaging. IEEE Trans Med Imaging. 1994;13:363–74.PubMedCrossRef Glick JS, Penney CB, King AM, Byrne CL. Noniterative compensation for the distance-dependent detector response and photon attenuation in SPECT imaging. IEEE Trans Med Imaging. 1994;13:363–74.PubMedCrossRef
4.
go back to reference Pretorius HP, King AM, Pan TS, de Vries DJ, Click SJ, Byrne CL. Reduction the influence of the partial volume effect on SPECT activity quantitation with 3D modeling of spatial resolution in iterative reconstruction. Phys Med Biol. 1998;43:407–20.PubMedCrossRef Pretorius HP, King AM, Pan TS, de Vries DJ, Click SJ, Byrne CL. Reduction the influence of the partial volume effect on SPECT activity quantitation with 3D modeling of spatial resolution in iterative reconstruction. Phys Med Biol. 1998;43:407–20.PubMedCrossRef
5.
go back to reference Yokoi T, Shinohara H, Onishi H. Performance evaluation of OSEM reconstruction algorithm incorporating three-dimensional distance-dependent resolution compensation for brain SPECT: a simulation study. Ann Nucl Med. 2002;16:11–8.PubMedCrossRef Yokoi T, Shinohara H, Onishi H. Performance evaluation of OSEM reconstruction algorithm incorporating three-dimensional distance-dependent resolution compensation for brain SPECT: a simulation study. Ann Nucl Med. 2002;16:11–8.PubMedCrossRef
6.
go back to reference Blinder S, Celler A, Wells RG, Thomson D, Harrop R. Experimental verification of 3D detector response compensation using the OSEM reconstruction method. In: 2001 IEEE nuclear science symposium conference record, 2001, San Diego, USA, p. 2174–2178. Blinder S, Celler A, Wells RG, Thomson D, Harrop R. Experimental verification of 3D detector response compensation using the OSEM reconstruction method. In: 2001 IEEE nuclear science symposium conference record, 2001, San Diego, USA, p. 2174–2178.
7.
go back to reference Frey EC, Gilland KL, Tsui BM. Application of task-based measures of image quality to optimization and evaluation of three dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging. 2002;21:1040–50.PubMedCrossRef Frey EC, Gilland KL, Tsui BM. Application of task-based measures of image quality to optimization and evaluation of three dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging. 2002;21:1040–50.PubMedCrossRef
8.
go back to reference Jinghan Y, Xiyun S, Zuo Z, Da Silva. Iterative SPECT Reconstruction using matched filtering for improved image quality. In: Nuclear science symposium conference IEEE. 2006. p. 2285–2287. Jinghan Y, Xiyun S, Zuo Z, Da Silva. Iterative SPECT Reconstruction using matched filtering for improved image quality. In: Nuclear science symposium conference IEEE. 2006. p. 2285–2287.
9.
go back to reference Vija H, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: phantom studies. In: 2003 IEEE nuclear science symposium, Medical imaging conference. 2003, Portland, USA, p. 2662–2666. Vija H, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: phantom studies. In: 2003 IEEE nuclear science symposium, Medical imaging conference. 2003, Portland, USA, p. 2662–2666.
10.
go back to reference Myers RW, Haseman MK, Durbin M. Halftime acquisition of MPI with maintained image quality. J Nucl Cardiol. 2006;13:S14–5.CrossRef Myers RW, Haseman MK, Durbin M. Halftime acquisition of MPI with maintained image quality. J Nucl Cardiol. 2006;13:S14–5.CrossRef
11.
go back to reference DePuey EG, Gadiraju R, Clark J, Thompson L, Ansetett F, Shwartz SC. Ordered subset expectation maximization and wide beam reconstruction “half-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” filtered backprojection. J Nucl Cardiol. 2008;15:547–63.PubMedCrossRef DePuey EG, Gadiraju R, Clark J, Thompson L, Ansetett F, Shwartz SC. Ordered subset expectation maximization and wide beam reconstruction “half-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” filtered backprojection. J Nucl Cardiol. 2008;15:547–63.PubMedCrossRef
12.
go back to reference Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med. 2009;50:544–62.CrossRef Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med. 2009;50:544–62.CrossRef
13.
go back to reference DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol. 2009;16:736–52.PubMedCrossRef DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol. 2009;16:736–52.PubMedCrossRef
14.
go back to reference Zeintl J, Vija AH, Yahil A, Homegger J, Kuwent T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921–8.PubMedCrossRef Zeintl J, Vija AH, Yahil A, Homegger J, Kuwent T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921–8.PubMedCrossRef
15.
go back to reference Tashiro K, Tomiguchi S, Shiraishi S, Yoshida M, Sakaguchi F, Yamashita Y. Clinical usefulness of a collimator distance dependent resolution recovery in myocardial perfusion SPECT: a clinical report from a single institute. Ann Nucl Med. 2011;25:133–7.PubMedCrossRef Tashiro K, Tomiguchi S, Shiraishi S, Yoshida M, Sakaguchi F, Yamashita Y. Clinical usefulness of a collimator distance dependent resolution recovery in myocardial perfusion SPECT: a clinical report from a single institute. Ann Nucl Med. 2011;25:133–7.PubMedCrossRef
16.
go back to reference Yamaki N, Natsume T, Takeda K, Maeda H, Hasebe S, Kinda A, et al. Simultaneous spatial resolution correction in SPECT reconstruction using OSEM algorithm. Jpn J Med Phys. 2004;24:61–71. (in Japanese). Yamaki N, Natsume T, Takeda K, Maeda H, Hasebe S, Kinda A, et al. Simultaneous spatial resolution correction in SPECT reconstruction using OSEM algorithm. Jpn J Med Phys. 2004;24:61–71. (in Japanese).
17.
go back to reference Penny BC, King MA, Schwinger RB, Baker SP, Stritzke P, Doherty PW. Constrained least-squares restoration of nuclear medicine images: selecting the coarseness function. Med Phys. 1987;14:849–58.CrossRef Penny BC, King MA, Schwinger RB, Baker SP, Stritzke P, Doherty PW. Constrained least-squares restoration of nuclear medicine images: selecting the coarseness function. Med Phys. 1987;14:849–58.CrossRef
18.
go back to reference Ohnishi H, Ota T, Takada M, Kida T, Noma K, Matsuo S, et al. Two optimal prefilter cutoff frequencies needed for SPECT images of myocardial perfusion in a one-day protocol. J Nucl Med Technol. 1997;25:256–60.PubMed Ohnishi H, Ota T, Takada M, Kida T, Noma K, Matsuo S, et al. Two optimal prefilter cutoff frequencies needed for SPECT images of myocardial perfusion in a one-day protocol. J Nucl Med Technol. 1997;25:256–60.PubMed
19.
go back to reference Hughes T, Scherbinin S, Celler A. A multi-center phantom study comparing image resolution from three state-of art SPECT-CT systems. J Nucl Cardiol. 2009;16:914–26.PubMedCrossRef Hughes T, Scherbinin S, Celler A. A multi-center phantom study comparing image resolution from three state-of art SPECT-CT systems. J Nucl Cardiol. 2009;16:914–26.PubMedCrossRef
20.
go back to reference Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol. 2009;16:714–25.PubMedCrossRef Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol. 2009;16:714–25.PubMedCrossRef
21.
go back to reference Stansfield EC, Sheehy N, Zurakowski D, Vija AH, Fahey FH, Treves ST. Rediatric 99mTc-MDP bone SPECT with ordered subset expectation maximization iterative reconstruction with isotropic 3D resolution recovery. Radiology. 2010;257:793–801.PubMedCrossRef Stansfield EC, Sheehy N, Zurakowski D, Vija AH, Fahey FH, Treves ST. Rediatric 99mTc-MDP bone SPECT with ordered subset expectation maximization iterative reconstruction with isotropic 3D resolution recovery. Radiology. 2010;257:793–801.PubMedCrossRef
22.
go back to reference Knull P, Katalova D, Kochle G, Kuzenlka I, Minear G, Mizrzaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys. 2012;22:58–69. Knull P, Katalova D, Kochle G, Kuzenlka I, Minear G, Mizrzaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys. 2012;22:58–69.
23.
go back to reference Arosio M, Pasquali C, Crivellaro C, De Ponti E, Morzent S, Guerra L, et al. Performance of a SPECT collimator-detector response reconstruction algorithm: phantom studies and validation in inflammation clinical studies. Q J Nucl Med Mol Imaging. 2011;55:1–9. Arosio M, Pasquali C, Crivellaro C, De Ponti E, Morzent S, Guerra L, et al. Performance of a SPECT collimator-detector response reconstruction algorithm: phantom studies and validation in inflammation clinical studies. Q J Nucl Med Mol Imaging. 2011;55:1–9.
24.
go back to reference Slomka PJ, Patton JA, Berman DS, Gemano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol. 2009;16:255–76.PubMedCrossRef Slomka PJ, Patton JA, Berman DS, Gemano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol. 2009;16:255–76.PubMedCrossRef
Metadata
Title
Quantitative performance of advanced resolution recovery strategies on SPECT images: evaluation with use of digital phantom models
Authors
Hideo Onishi
Nobutoku Motomura
Koichi Fujino
Takahiro Natsume
Yasuhiro Haramoto
Publication date
01-01-2013
Publisher
Springer Japan
Published in
Radiological Physics and Technology / Issue 1/2013
Print ISSN: 1865-0333
Electronic ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-012-0168-z

Other articles of this Issue 1/2013

Radiological Physics and Technology 1/2013 Go to the issue