Quantitative CBF measurement using an integrated SPECT/CT system: validation of three-dimensional ordered-subset expectation maximization and CT-based attenuation correction by comparing with O-15 water PET

Authors: Eku Shimosegawa, Kouichi Fujino, Hiroki Kato, Jun Hatazawa

Published in: Annals of Nuclear Medicine | Issue 9/2013

Login to get access

Abstract

Rationale and objectives

As spreading integrated SPECT/CT scanners, reconstruction method based on three-dimensional ordered-subset expectation maximization (3D-OSEM) with attenuation correction (AC) using an X-ray CT image (CTAC) is easily available in brain imaging. For quantitative cerebral blood flow (CBF) measurements, however, the accuracy of this method has not been fully evaluated clinically. To validate this new algorithm, we sequentially studied quantitative PET-CBF and SPECT-CBF measurements in the same healthy volunteers and compared CBF values.

Methods

Ten healthy subjects underwent quantitative PET-CBF and SPECT-CBF measurements on the same day. PET-CBF data were obtained by the intravenous injection of O-15 water and the quantitative IMP-ARG method followed. Three types of SPECT images were reconstructed: (1) filtered back projection (FBP) with Chang’s AC (FBP + Chang’s AC), (2) 3D-OSEM with Chang’s AC (3D-OSEM + Chang’s AC), and (3) 3D-OSEM with CTAC (3D-OSEM + CTAC). The mean CBF difference, the linearity, and the correlation between the PET-CBF and SPECT-CBF values were compared among the SPECT reconstruction algorithms.

Results

The mean SPECT-CBF values of all the algorithms were significantly lower in the pons (P = 0.000–0.007) and higher in the frontal lobe (P = 0.002–0.022). All the SPECT-CBF values were significantly correlated with the PET-CBF values (r = 0.749–0.829, P < 0.001). The SPECT-CBF values obtained using the 3D-OSEM + CTAC method showed the best regression with the PET-CBF values.

Conclusion

The present clinical study validated accuracy of CBF image reconstructed by the 3-D OSEM method with CTAC and the integrated SPECT/CT system.
Literature
1.
go back to reference Ljungberg M, Strand SE. Scatter and attenuation correction in SPECT using density maps and Monte Carlo simulated scatter functions. J Nucl Med. 1990;31:1560–7.PubMed Ljungberg M, Strand SE. Scatter and attenuation correction in SPECT using density maps and Monte Carlo simulated scatter functions. J Nucl Med. 1990;31:1560–7.PubMed
2.
go back to reference Tsui BM, Zhao X, Frey EC, MaCartney WH. Quantitative single-photon emission computed tomography: basics and clinical considerations. Semin Nucl Med. 1994;24:38–65.CrossRefPubMed Tsui BM, Zhao X, Frey EC, MaCartney WH. Quantitative single-photon emission computed tomography: basics and clinical considerations. Semin Nucl Med. 1994;24:38–65.CrossRefPubMed
3.
go back to reference Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, et al. Scatter and attenuation correction in technetium-99 m brain SPECT. J Nucl Med. 1997;38:157–62.PubMed Hashimoto J, Kubo A, Ogawa K, Amano T, Fukuuchi Y, Motomura N, et al. Scatter and attenuation correction in technetium-99 m brain SPECT. J Nucl Med. 1997;38:157–62.PubMed
4.
go back to reference Liang Z, Ye J, Harrington DP. An analytical approach to quantitative reconstruction of non-uniform attenuated brain SPECT. Phys Med Biol. 1994;39:2023–41.CrossRefPubMed Liang Z, Ye J, Harrington DP. An analytical approach to quantitative reconstruction of non-uniform attenuated brain SPECT. Phys Med Biol. 1994;39:2023–41.CrossRefPubMed
5.
go back to reference Chang LT. A method for attenuation correction in radionuclide computed tomography. IEEE Trans Nuci Sci. 1978;25:638–43.CrossRef Chang LT. A method for attenuation correction in radionuclide computed tomography. IEEE Trans Nuci Sci. 1978;25:638–43.CrossRef
6.
go back to reference Leong LK, Kruger RL, O'Connor MK. A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OSEM techniques. J Nucl Med Technol. 2001;29:79–83. Leong LK, Kruger RL, O'Connor MK. A comparison of the uniformity requirements for SPECT image reconstruction using FBP and OSEM techniques. J Nucl Med Technol. 2001;29:79–83.
7.
go back to reference Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med. 2000;27:1380–6.CrossRefPubMed Kauppinen T, Koskinen MO, Alenius S, Vanninen E, Kuikka JT. Improvement of brain perfusion SPET using iterative reconstruction with scatter and non-uniform attenuation correction. Eur J Nucl Med. 2000;27:1380–6.CrossRefPubMed
8.
go back to reference Brambilla M, Cannillo B, Dominietto M, Leval L, Secco C, Inglese E. Characterization of ordered-subsets expectation maximization with 3D post-reconstruction Gauss filtering and comparison with filtered backprojection in 99mTc SPECT. Ann Nucl Med. 2005;19:75–82.CrossRefPubMed Brambilla M, Cannillo B, Dominietto M, Leval L, Secco C, Inglese E. Characterization of ordered-subsets expectation maximization with 3D post-reconstruction Gauss filtering and comparison with filtered backprojection in 99mTc SPECT. Ann Nucl Med. 2005;19:75–82.CrossRefPubMed
9.
go back to reference Olsson A, Ärlig Ǻ, Carlsson GA, Gustafsson A. Evaluation of reconstruction techniques in regional cerebral blood flow SPECT using trade-off plots: a Monte Carlo study. Nucl Med Commun. 2007;28:719–25.CrossRefPubMed Olsson A, Ärlig Ǻ, Carlsson GA, Gustafsson A. Evaluation of reconstruction techniques in regional cerebral blood flow SPECT using trade-off plots: a Monte Carlo study. Nucl Med Commun. 2007;28:719–25.CrossRefPubMed
10.
go back to reference Miller TR, Walhis JR. Clinically important characteristics of maximum likelihood reconstruction. J NucI Med. 1992;33:1678–84. Miller TR, Walhis JR. Clinically important characteristics of maximum likelihood reconstruction. J NucI Med. 1992;33:1678–84.
11.
go back to reference Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921–8.CrossRefPubMed Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921–8.CrossRefPubMed
12.
go back to reference Vija AH, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: phantom studies. IEEE Nucl Sci Symp Conf Rec. 2003;4:2662–6. Vija AH, Hawman EG, Engdahl JC. Analysis of a SPECT OSEM reconstruction method with 3D beam modeling and optional attenuation correction: phantom studies. IEEE Nucl Sci Symp Conf Rec. 2003;4:2662–6.
13.
go back to reference Römer W, Reichel N, Vija HA, Nickel I, Hornegger J, Bauts W, et al. Isotropic reconstruction of SPECT data using OSEM3D: correlation with CT. Acad Radiol. 2006;13:496–502.CrossRefPubMed Römer W, Reichel N, Vija HA, Nickel I, Hornegger J, Bauts W, et al. Isotropic reconstruction of SPECT data using OSEM3D: correlation with CT. Acad Radiol. 2006;13:496–502.CrossRefPubMed
14.
go back to reference Iida H, Akutsu T, Endo K, Fukuda H, Inoue T, Ito H, et al. A multicenter validation of regional cerebral blood flow quantitation using [123I]Iodoamphetamine and single photon emission computed tomography. J Cereb Blood Flow Metab. 1996;16:781–93.CrossRefPubMed Iida H, Akutsu T, Endo K, Fukuda H, Inoue T, Ito H, et al. A multicenter validation of regional cerebral blood flow quantitation using [123I]Iodoamphetamine and single photon emission computed tomography. J Cereb Blood Flow Metab. 1996;16:781–93.CrossRefPubMed
15.
go back to reference Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S, et al. Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard. J Nucl Med. 2006;47:83–90.PubMed Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S, et al. Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard. J Nucl Med. 2006;47:83–90.PubMed
16.
go back to reference Ishikawa A, Kitamura K, Mizuta T, Tanaka K, Amano M. Implementation of on-the-fly scatter correction using dual energy window method in continuous 3D whole body PET scanning. Conf Rec IEEE NSS & MIC. 2005;5:2497–500. Ishikawa A, Kitamura K, Mizuta T, Tanaka K, Amano M. Implementation of on-the-fly scatter correction using dual energy window method in continuous 3D whole body PET scanning. Conf Rec IEEE NSS & MIC. 2005;5:2497–500.
17.
go back to reference Ferreira NC, Trebossen R, Lartizien C, Brulon V, Merceron P, Bendriem B. A hybrid scatter correction for 3D PET based on an estimation of the distribution of unscattered coincidences: implementation on the ECAT EXACT HR+. Phys Med Biol. 2002;47:1555–71.CrossRefPubMed Ferreira NC, Trebossen R, Lartizien C, Brulon V, Merceron P, Bendriem B. A hybrid scatter correction for 3D PET based on an estimation of the distribution of unscattered coincidences: implementation on the ECAT EXACT HR+. Phys Med Biol. 2002;47:1555–71.CrossRefPubMed
18.
go back to reference Defrise M, Kinahan PE, Townsend DW, Michel C, Sibomana M, Newport DF. Exact and approximate rebinning algorithms for 3-D PET data. IEEE Trans Med Imaging. 1997;16:145–58.CrossRefPubMed Defrise M, Kinahan PE, Townsend DW, Michel C, Sibomana M, Newport DF. Exact and approximate rebinning algorithms for 3-D PET data. IEEE Trans Med Imaging. 1997;16:145–58.CrossRefPubMed
19.
go back to reference Bailey DL, Meikle SR. A convolution-subtraction scatter correction method for 3D PET. Phys Med Biol. 1994;39:411–24.CrossRefPubMed Bailey DL, Meikle SR. A convolution-subtraction scatter correction method for 3D PET. Phys Med Biol. 1994;39:411–24.CrossRefPubMed
20.
go back to reference Ibaraki M, Miura S, Shimosegawa E, Sugawara S, Mizuta T, Ishikawa A, et al. Quantification of cerebral blood flow and oxygen metabolism with 3-dimensional PET and 15O: validation by comparison with 2-dimensional PET. J Nucl Med. 2008;49:50–9.CrossRefPubMed Ibaraki M, Miura S, Shimosegawa E, Sugawara S, Mizuta T, Ishikawa A, et al. Quantification of cerebral blood flow and oxygen metabolism with 3-dimensional PET and 15O: validation by comparison with 2-dimensional PET. J Nucl Med. 2008;49:50–9.CrossRefPubMed
21.
go back to reference Iida H, Kanno I, Miura S, Murakami M, Takahashi K, Uemura K. Error analysis of a quantitative cerebral blood flow measurement using H 2 15 O autoradiography and positron emission tomography, with respect to the dispersion of the input function. J Cereb Blood Flow Metab. 1986;6:536–45.CrossRefPubMed Iida H, Kanno I, Miura S, Murakami M, Takahashi K, Uemura K. Error analysis of a quantitative cerebral blood flow measurement using H 2 15 O autoradiography and positron emission tomography, with respect to the dispersion of the input function. J Cereb Blood Flow Metab. 1986;6:536–45.CrossRefPubMed
22.
go back to reference Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed
23.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
24.
go back to reference Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shouji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shouji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed
25.
go back to reference Shiga T, Kubo N, Takano A, Kobayashi J, Takeda Y, Nakamura F, et al. The effect of scatter correction on 123I-IMP brain perfusion SPET with the triple energy window method in normal subjects using SPM analysis. Eur J Nucl Med Mol Imaging. 2002;29:342–5.CrossRefPubMed Shiga T, Kubo N, Takano A, Kobayashi J, Takeda Y, Nakamura F, et al. The effect of scatter correction on 123I-IMP brain perfusion SPET with the triple energy window method in normal subjects using SPM analysis. Eur J Nucl Med Mol Imaging. 2002;29:342–5.CrossRefPubMed
26.
go back to reference Axelsson B, Msaki P, Israelsson A. Subtraction of Compton-scattered photon in single-photon emission computed tomography. J Nucl Med. 1984;25:490–4.PubMed Axelsson B, Msaki P, Israelsson A. Subtraction of Compton-scattered photon in single-photon emission computed tomography. J Nucl Med. 1984;25:490–4.PubMed
27.
go back to reference Ito H, Inoue K, Goto R, Kinomura S, Okada K, Sato K, et al. Detabase of normal human cerebral blood flow measured by SPECT: I. comparison between I-123-IMP, Tc-99m-HMPAO, and Tc-99m-ECD as referred with O-15 labeled water PET and voxel-based morphometry. Ann Nucl Med. 2006;20:131–8.CrossRefPubMed Ito H, Inoue K, Goto R, Kinomura S, Okada K, Sato K, et al. Detabase of normal human cerebral blood flow measured by SPECT: I. comparison between I-123-IMP, Tc-99m-HMPAO, and Tc-99m-ECD as referred with O-15 labeled water PET and voxel-based morphometry. Ann Nucl Med. 2006;20:131–8.CrossRefPubMed
28.
go back to reference Nishizawa S, Tanada S, Yonekura Y, Fujita T, Mukai T, Saji H, et al. Regional dynamics of N-isopropyl-(123I)p-iodo-amphetamine in human brain. J Nucl Med. 1989;30:150–6.PubMed Nishizawa S, Tanada S, Yonekura Y, Fujita T, Mukai T, Saji H, et al. Regional dynamics of N-isopropyl-(123I)p-iodo-amphetamine in human brain. J Nucl Med. 1989;30:150–6.PubMed
29.
go back to reference Wilson DW, Tsui BMW. Noise properties of filtered-backprojection and ML−EM reconstructed emission tomographic images. IEEE Trans Nucl Sci. 1993;40:1198–1203 Wilson DW, Tsui BMW. Noise properties of filtered-backprojection and ML−EM reconstructed emission tomographic images. IEEE Trans Nucl Sci. 1993;40:1198–1203
30.
go back to reference Rapisarda E, Bettinardi V, Thielemans K, Gilardi MC. Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET. Phys Med Biol. 2010;55:4131–51. Rapisarda E, Bettinardi V, Thielemans K, Gilardi MC. Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET. Phys Med Biol. 2010;55:4131–51.
31.
go back to reference Gantet P, Payoux P, Celler A, Majorel C, Gourion D, Noll D, et al. Iterative three-dimensional expectation maximization restoration of single photon emission computed tomography images: application in striatal imaging. Med Phys. 2006;33:52–60. Gantet P, Payoux P, Celler A, Majorel C, Gourion D, Noll D, et al. Iterative three-dimensional expectation maximization restoration of single photon emission computed tomography images: application in striatal imaging. Med Phys. 2006;33:52–60.
32.
go back to reference Snyder DL, Miller MI, Thomas LJ, Politte DG. Noise and edge artefacts in maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging. 1987;6:228–38. Snyder DL, Miller MI, Thomas LJ, Politte DG. Noise and edge artefacts in maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging. 1987;6:228–38.
33.
go back to reference Turkington TG, Gilland DR, Jaszczk RJ, Geer KL, Coleman RE. A direct measurement of skull attenuation for quantitative SPECT. IEEE Trans Nucl Sci. 1993;40:1158–61. Turkington TG, Gilland DR, Jaszczk RJ, Geer KL, Coleman RE. A direct measurement of skull attenuation for quantitative SPECT. IEEE Trans Nucl Sci. 1993;40:1158–61.
34.
go back to reference Rajeevan N, Zubal IG, Ramsby SQ, Zoghbi SS, Seibyl J, Innis RB. Significance of nonuniform attenuation correction in quantitative brain SPECT imaging. J Nucl Med. 1998;39:1719–26. Rajeevan N, Zubal IG, Ramsby SQ, Zoghbi SS, Seibyl J, Innis RB. Significance of nonuniform attenuation correction in quantitative brain SPECT imaging. J Nucl Med. 1998;39:1719–26.
35.
go back to reference Van Laere K, Koole M, Versijpt J, Dierckx R. Non-uniform versus uniform attenuation correction in brain perfusion SPET of healthy volunteers. Eur J Nucl Med. 2001;28:90–8. Van Laere K, Koole M, Versijpt J, Dierckx R. Non-uniform versus uniform attenuation correction in brain perfusion SPET of healthy volunteers. Eur J Nucl Med. 2001;28:90–8.
36.
go back to reference Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med. 1998;39:190–7. Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med. 1998;39:190–7.
37.
go back to reference Licho R, Glick SJ, Xia W, Pan T-S, Penney BC, King MA. Attenuation compensation in 99mTc-SPECT brain imaging: a comparison of the use of attenuation maps derived from transmission versus emission data in normal scans. J Nucl Med. 1999;40:456–63. Licho R, Glick SJ, Xia W, Pan T-S, Penney BC, King MA. Attenuation compensation in 99mTc-SPECT brain imaging: a comparison of the use of attenuation maps derived from transmission versus emission data in normal scans. J Nucl Med. 1999;40:456–63.
38.
go back to reference Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9. Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.
39.
go back to reference Sheehy N, Tetrault TA, Zurakowski D, Vija AH, Fahey FH, Treves ST. Pediatric 99mTc-DMSA SPECT performed by using iterative reconstruction with isotropic resolution recovery: improved image quality and reduced radiopharmaceutical activity. Radiology. 2009;251:511–6. Sheehy N, Tetrault TA, Zurakowski D, Vija AH, Fahey FH, Treves ST. Pediatric 99mTc-DMSA SPECT performed by using iterative reconstruction with isotropic resolution recovery: improved image quality and reduced radiopharmaceutical activity. Radiology. 2009;251:511–6.
Metadata
Title
Quantitative CBF measurement using an integrated SPECT/CT system: validation of three-dimensional ordered-subset expectation maximization and CT-based attenuation correction by comparing with O-15 water PET
Authors
Eku Shimosegawa
Kouichi Fujino
Hiroki Kato
Jun Hatazawa
Publication date
01-11-2013
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 9/2013
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-013-0752-2

Other articles of this Issue 9/2013

Annals of Nuclear Medicine 9/2013 Go to the issue