Skip to main content
Top
Published in: European Journal of Clinical Microbiology & Infectious Diseases 10/2017

Open Access 01-10-2017 | Original Article

A controlled comparison of the BacT/ALERT® 3D and VIRTUO™ microbial detection systems

Authors: H. Totty, M. Ullery, J. Spontak, J. Viray, M. Adamik, B. Katzin, W. M. Dunne Jr, P. Deol

Published in: European Journal of Clinical Microbiology & Infectious Diseases | Issue 10/2017

Login to get access

Abstract

The performance of the next-generation BacT/ALERT® VIRTUO™ Microbial Detection System (VIRTUO™, bioMérieux Inc., Hazelwood, MO) was compared to the BacT/ALERT® 3D Microbial Detection System (3D, bioMérieux Inc., Durham, NC) using BacT/ALERT® FA Plus (FA Plus), BacT/ALERT® PF Plus (PF Plus), BacT/ALERT® FN Plus (FN Plus), BacT/ALERT® Standard Aerobic (SA), and BacT/ALERT® Standard Anaerobic (SN) blood culture bottles (bioMérieux Inc., Durham, NC). A seeded limit of detection (LoD) study was performed for each bottle type in both systems. The LoD studies demonstrated that both systems were capable of detecting organisms at nearly identical levels [<10 colony-forming units (CFU) per bottle], with no significant difference. Following LoD determination, a seeded study was performed to compare the time to detection (TTD) between the systems using a panel of clinically relevant microorganisms inoculated at or near the LoD with 0, 4, or 10 mL of healthy human blood. VIRTUO™ exhibited a faster TTD by an average of 3.5 h, as well as demonstrated a significantly improved detection rate of 99.9% compared to 98.8% with 3D (p-value <0.05).
Appendix
Available only for authorised users
Literature
1.
go back to reference Magadia RR, Weinstein MP (2001) Laboratory diagnosis of bacteremia and fungemia. Infect Dis Clin North Am 15:1009–1024CrossRefPubMed Magadia RR, Weinstein MP (2001) Laboratory diagnosis of bacteremia and fungemia. Infect Dis Clin North Am 15:1009–1024CrossRefPubMed
2.
go back to reference Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34:1589–1596CrossRefPubMed Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34:1589–1596CrossRefPubMed
3.
go back to reference BacT/ALERT® VIRTUO™ Microbial Detection System User Manual (2014-06) 414998-EN BacT/ALERT® VIRTUO™ Microbial Detection System User Manual (2014-06) 414998-EN
4.
go back to reference BacT/ALERT® SA [package insert] (2010) bioMérieux, Inc., Durham, NC BacT/ALERT® SA [package insert] (2010) bioMérieux, Inc., Durham, NC
5.
go back to reference BacT/ALERT® SN [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® SN [package insert] (2011) bioMérieux, Inc., Durham, NC
6.
go back to reference BacT/ALERT® FA Plus [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® FA Plus [package insert] (2011) bioMérieux, Inc., Durham, NC
7.
go back to reference BacT/ALERT® PF Plus [package insert] (2016) bioMérieux, Inc., Durham, NC BacT/ALERT® PF Plus [package insert] (2016) bioMérieux, Inc., Durham, NC
8.
go back to reference BacT/ALERT® FN Plus [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® FN Plus [package insert] (2011) bioMérieux, Inc., Durham, NC
9.
go back to reference Clinical and Laboratory Standards Institute (CLSI) (2007) Principles and procedures for blood cultures; approved guideline. CLSI document M47-A. CLSI, Wayne Clinical and Laboratory Standards Institute (CLSI) (2007) Principles and procedures for blood cultures; approved guideline. CLSI document M47-A. CLSI, Wayne
10.
go back to reference Altun O, Almuhayawi M, Lüthje P, Taha R, Ullberg M, Özenci V (2016) Controlled evaluation of the new BacT/Alert Virtuo blood culture system for detection and time to detection of bacteria and yeasts. J Clin Microbiol 54:1148–1151CrossRefPubMedPubMedCentral Altun O, Almuhayawi M, Lüthje P, Taha R, Ullberg M, Özenci V (2016) Controlled evaluation of the new BacT/Alert Virtuo blood culture system for detection and time to detection of bacteria and yeasts. J Clin Microbiol 54:1148–1151CrossRefPubMedPubMedCentral
11.
go back to reference Kirn TJ, Mirrett S, Reller LB, Weinstein MP (2014) Controlled clinical comparison of BacT/Alert FA Plus and FN Plus blood culture media with BacT/alert FA and FN blood culture media. J Clin Microbiol 52:839–843CrossRefPubMedPubMedCentral Kirn TJ, Mirrett S, Reller LB, Weinstein MP (2014) Controlled clinical comparison of BacT/Alert FA Plus and FN Plus blood culture media with BacT/alert FA and FN blood culture media. J Clin Microbiol 52:839–843CrossRefPubMedPubMedCentral
12.
go back to reference Terhes G, Piukovics K, Urbán E, Nagy E (2011) Four cases of bacteraemia caused by Fusobacterium nucleatum in febrile, neutropenic patients. J Med Microbiol 60:1046–1049CrossRefPubMed Terhes G, Piukovics K, Urbán E, Nagy E (2011) Four cases of bacteraemia caused by Fusobacterium nucleatum in febrile, neutropenic patients. J Med Microbiol 60:1046–1049CrossRefPubMed
13.
go back to reference Fiori B, D’Inzeo T, Di Florio V, De Maio F, De Angelis G, Giaquinto A, Campana L, Tanzarella E, Tumbarello M, Antonelli M, Sanguinetti M, Spanu T (2014) Performance of two resin-containing blood culture media in detection of bloodstream infections and in direct matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) broth assays for isolate identification: clinical comparison of the BacT/Alert Plus and BACTEC Plus systems. J Clin Microbiol 52:3558–3567 Fiori B, D’Inzeo T, Di Florio V, De Maio F, De Angelis G, Giaquinto A, Campana L, Tanzarella E, Tumbarello M, Antonelli M, Sanguinetti M, Spanu T (2014) Performance of two resin-containing blood culture media in detection of bloodstream infections and in direct matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) broth assays for isolate identification: clinical comparison of the BacT/Alert Plus and BACTEC Plus systems. J Clin Microbiol 52:3558–3567
14.
go back to reference Rosa C, Araujo R, Rodrigues AG, Pinto-de-Sousa MI, Pina-Vaz C (2011) Detection of Aspergillus species in BACTEC blood cultures. J Med Microbiol 60:1467–1471CrossRefPubMed Rosa C, Araujo R, Rodrigues AG, Pinto-de-Sousa MI, Pina-Vaz C (2011) Detection of Aspergillus species in BACTEC blood cultures. J Med Microbiol 60:1467–1471CrossRefPubMed
15.
go back to reference Lu PL, Hsueh PR, Hung CC, Teng LJ, Jang TN, Luh KT (2000) Infective endocarditis complicated with progressive heart failure due to beta-lactamase-producing Cardiobacterium hominis. J Clin Microbiol 38:2015–2017PubMedPubMedCentral Lu PL, Hsueh PR, Hung CC, Teng LJ, Jang TN, Luh KT (2000) Infective endocarditis complicated with progressive heart failure due to beta-lactamase-producing Cardiobacterium hominis. J Clin Microbiol 38:2015–2017PubMedPubMedCentral
16.
go back to reference Manso E, Strusi P, Massacci C, Ferrini L (1987) Endocarditis due to Cardiobacterium hominis. Boll Ist Sieroter Milan 66:489–490PubMed Manso E, Strusi P, Massacci C, Ferrini L (1987) Endocarditis due to Cardiobacterium hominis. Boll Ist Sieroter Milan 66:489–490PubMed
Metadata
Title
A controlled comparison of the BacT/ALERT® 3D and VIRTUO™ microbial detection systems
Authors
H. Totty
M. Ullery
J. Spontak
J. Viray
M. Adamik
B. Katzin
W. M. Dunne Jr
P. Deol
Publication date
01-10-2017
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Clinical Microbiology & Infectious Diseases / Issue 10/2017
Print ISSN: 0934-9723
Electronic ISSN: 1435-4373
DOI
https://doi.org/10.1007/s10096-017-2994-8

Other articles of this Issue 10/2017

European Journal of Clinical Microbiology & Infectious Diseases 10/2017 Go to the issue