01-06-2018 | Original Article

Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures

Authors: Nadine Marheineke, Uta Scherer, Martin Rücker, Constantin von See, Björn Rahlf, Nils-Claudius Gellrich, Marcus Stoetzer

Published in: Clinical Oral Investigations | Issue 5/2018

Login to get access

Abstract

Objectives

Dental implant failure and insufficient osseointegration are proven results of mechanical and thermal damage during the surgery process. We herein performed a comparative study of a less invasive single-step drilling preparation protocol and a conventional multiple drilling sequence. Accuracy of drilling holes was precisely analyzed and the influence of different levels of expertise of the handlers and additional use of drill template guidance was evaluated.

Material and methods

Six experimental groups, deployed in an osseous study model, were representing template-guided and freehanded drilling actions in a stepwise drilling procedure in comparison to a single-drill protocol. Each experimental condition was studied by the drilling actions of respectively three persons without surgical knowledge as well as three highly experienced oral surgeons. Drilling actions were performed and diameters were recorded with a precision measuring instrument.

Results

Less experienced operators were able to significantly increase the drilling accuracy using a guiding template, especially when multi-step preparations are performed. Improved accuracy without template guidance was observed when experienced operators were executing single-step versus multi-step technique.

Conclusion

Single-step drilling protocols have shown to produce more accurate results than multi-step procedures. The outcome of any protocol can be further improved by use of guiding templates. Operator experience can be a contributing factor.

Clinical relevance

Single-step preparations are less invasive and are promoting osseointegration. Even highly experienced surgeons are achieving higher levels of accuracy by combining this technique with template guidance. Hereby template guidance enables a reduction of hands-on time and side effects during surgery and lead to a more predictable clinical diameter.
Literature
2.
go back to reference Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11:35–37PubMed Brisman DL (1996) The effect of speed, pressure, and time on bone temperature during the drilling of implant sites. Int J Oral Maxillofac Implants 11:35–37PubMed
3.
go back to reference Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed
4.
go back to reference Davidson SR, James DF (2003) Drilling in bone: modeling heat generation and temperature distribution. J Biomech Eng 125:305–314CrossRefPubMed Davidson SR, James DF (2003) Drilling in bone: modeling heat generation and temperature distribution. J Biomech Eng 125:305–314CrossRefPubMed
6.
go back to reference Abouzgia MB, Symington JM (1996) Effect of drill speed on bone temperature. Int J Oral Maxillofac Surg 25:394–399CrossRefPubMed Abouzgia MB, Symington JM (1996) Effect of drill speed on bone temperature. Int J Oral Maxillofac Surg 25:394–399CrossRefPubMed
8.
go back to reference dos Santos PL, Queiroz TP, Margonar R, de Souza Carvalho AC, Betoni W Jr, Rezende RR, dos Santos PH, Garcia IR Jr (2014) Evaluation of bone heating, drill deformation, and drill roughness after implant osteotomy: guided surgery and classic drilling procedure. Int J Oral Maxillofac Implants 29(1):51–58. https://doi.org/10.11607/jomi.2919 CrossRefPubMed dos Santos PL, Queiroz TP, Margonar R, de Souza Carvalho AC, Betoni W Jr, Rezende RR, dos Santos PH, Garcia IR Jr (2014) Evaluation of bone heating, drill deformation, and drill roughness after implant osteotomy: guided surgery and classic drilling procedure. Int J Oral Maxillofac Implants 29(1):51–58. https://​doi.​org/​10.​11607/​jomi.​2919 CrossRefPubMed
10.
go back to reference Fuchsberger A (1988) Damaging temperature during the machining of bone. Unfallchirurgie 14:173–183PubMed Fuchsberger A (1988) Damaging temperature during the machining of bone. Unfallchirurgie 14:173–183PubMed
12.
go back to reference Iyer S, Weiss C, Mehta A (1997) Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies. Part II: relationship between drill speed and healing. Int J Prosthodont 10:536–540PubMed Iyer S, Weiss C, Mehta A (1997) Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies. Part II: relationship between drill speed and healing. Int J Prosthodont 10:536–540PubMed
14.
go back to reference Gehrke SA, Bettach R, Taschieri S, Boukhris G, Corbella S, Del Fabbro M (2015) Temperature changes in cortical bone after implant site preparation using a single bur versus multiple drilling steps: an in vitro investigation. Clin Implant Dent Relat Res 17:700–707. https://doi.org/10.1111/cid.12172 CrossRefPubMed Gehrke SA, Bettach R, Taschieri S, Boukhris G, Corbella S, Del Fabbro M (2015) Temperature changes in cortical bone after implant site preparation using a single bur versus multiple drilling steps: an in vitro investigation. Clin Implant Dent Relat Res 17:700–707. https://​doi.​org/​10.​1111/​cid.​12172 CrossRefPubMed
17.
go back to reference Valente F, Schiroli G, Sbrenna A (2009) Accuracy of computer-aided oral implant surgery: a clinical and radiographic study. Int J Oral Maxillofac Implants 24:234–242PubMed Valente F, Schiroli G, Sbrenna A (2009) Accuracy of computer-aided oral implant surgery: a clinical and radiographic study. Int J Oral Maxillofac Implants 24:234–242PubMed
18.
go back to reference Bulloch SE, Olsen RG, Bulloch B (2012) Comparison of heat generation between internally guided (cannulated) single drill and traditional sequential drilling with and without a drill guide for dental implants. Int J Oral Maxillofac Implants 27:1456–1460PubMed Bulloch SE, Olsen RG, Bulloch B (2012) Comparison of heat generation between internally guided (cannulated) single drill and traditional sequential drilling with and without a drill guide for dental implants. Int J Oral Maxillofac Implants 27:1456–1460PubMed
19.
go back to reference Bardyn T, Gedet P, Hallermann W, Buchler P (2009) Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials. Int J Oral Maxillofac Implants 24:1006–1014PubMed Bardyn T, Gedet P, Hallermann W, Buchler P (2009) Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials. Int J Oral Maxillofac Implants 24:1006–1014PubMed
27.
go back to reference Friberg B, Ekestubbe A, Sennerby L (2002) Clinical outcome of Branemark system implants of various diameters: a retrospective study. Int J Oral Maxillofac Implants 17:671–677PubMed Friberg B, Ekestubbe A, Sennerby L (2002) Clinical outcome of Branemark system implants of various diameters: a retrospective study. Int J Oral Maxillofac Implants 17:671–677PubMed
38.
go back to reference Anitua E, Alkhraisat MH, Pinas L, Orive G (2014) Efficacy of biologically guided implant site preparation to obtain adequate primary implant stability. Ann Anat 199:9–15CrossRefPubMed Anitua E, Alkhraisat MH, Pinas L, Orive G (2014) Efficacy of biologically guided implant site preparation to obtain adequate primary implant stability. Ann Anat 199:9–15CrossRefPubMed
40.
go back to reference Penarrocha M, Garcia B, Marti E, Balaguer J (2006) Pain and inflammation after periapical surgery in 60 patients. J Oral Maxillofac Surg 64:429–433CrossRefPubMed Penarrocha M, Garcia B, Marti E, Balaguer J (2006) Pain and inflammation after periapical surgery in 60 patients. J Oral Maxillofac Surg 64:429–433CrossRefPubMed
49.
go back to reference Allsobrook OF, Leichter J, Holborrow D, Swain M (2011) Descriptive study of the longevity of dental implant surgery drills. Clin Implant Dent Relat Res 13:244–254CrossRefPubMed Allsobrook OF, Leichter J, Holborrow D, Swain M (2011) Descriptive study of the longevity of dental implant surgery drills. Clin Implant Dent Relat Res 13:244–254CrossRefPubMed
52.
go back to reference Fortin T, Bosson JL, Coudert JL, Isidori M (2003) Reliability of preoperative planning of an image-guided system for oral implant placement based on 3-dimensional images: an in vivo study. Int J Oral Maxillofac Implants 18(6):886–893PubMed Fortin T, Bosson JL, Coudert JL, Isidori M (2003) Reliability of preoperative planning of an image-guided system for oral implant placement based on 3-dimensional images: an in vivo study. Int J Oral Maxillofac Implants 18(6):886–893PubMed
53.
go back to reference Nickenig HJ, Eitner S, Rothamel D, Wichmann M, Zoller JE (2012) Possibilities and limitations of implant placement by virtual planning data and surgical guide templates. Int J Comput Dent 15:9–21PubMed Nickenig HJ, Eitner S, Rothamel D, Wichmann M, Zoller JE (2012) Possibilities and limitations of implant placement by virtual planning data and surgical guide templates. Int J Comput Dent 15:9–21PubMed
54.
go back to reference Sarment DP, Sukovic P, Clinthorne N (2003) Accuracy of implant placement with a stereolithographic surgical guide. Int J Oral Maxillofac Implants 18:571–577PubMed Sarment DP, Sukovic P, Clinthorne N (2003) Accuracy of implant placement with a stereolithographic surgical guide. Int J Oral Maxillofac Implants 18:571–577PubMed
56.
go back to reference van Steenberghe D, Glauser R, Blomback U, Andersson M, Schutyser F, Pettersson A, Wendelhag I (2005) A computed tomographic scan-derived customized surgical template and fixed prosthesis for flapless surgery and immediate loading of implants in fully edentulous maxillae: a prospective multicenter study. Clin Implant Dent Relat Res 7(Suppl 1):S111–S120CrossRefPubMed van Steenberghe D, Glauser R, Blomback U, Andersson M, Schutyser F, Pettersson A, Wendelhag I (2005) A computed tomographic scan-derived customized surgical template and fixed prosthesis for flapless surgery and immediate loading of implants in fully edentulous maxillae: a prospective multicenter study. Clin Implant Dent Relat Res 7(Suppl 1):S111–S120CrossRefPubMed
Metadata
Title
Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures
Authors
Nadine Marheineke
Uta Scherer
Martin Rücker
Constantin von See
Björn Rahlf
Nils-Claudius Gellrich
Marcus Stoetzer
Publication date
01-06-2018
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 5/2018
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-017-2312-y

Other articles of this Issue 5/2018

Clinical Oral Investigations 5/2018 Go to the issue