Published in: International Orthopaedics 6/2021

01-06-2021 | Original Paper

Early experience using patient-specific instrumentation in opening wedge high tibial osteotomy

Authors: Vlad Predescu, Alina-Maria Grosu, Iulian Gherman, Catalin Prescura, Valentin Hiohi, Bogdan Deleanu

Published in: International Orthopaedics | Issue 6/2021

Login to get access

Abstract

Purpose

Patient-specific instruments (PSI) have been designed to improve the accuracy of performing opening-wedge high-tibial osteotomies (OW-HTO). This study aims to evaluate the lower limb alignment, by comparing pre-operative desired correction to post-operative achieved correction, the difference in surgical time and number of radiological exposures in OW-HTO using patient-specific instruments (PSI) versus conventional osteotomies and the specific and non-specific complications that occurred.

Methods

We performed a single-centre, retrospective, observational study, including 25 consecutive patients undergoing OW-HTO using PSI, from January 2019 to October 2020.

Results

Pre-operatively, the mean hip-knee angle (HKA) was 167°, the mean tibial slope was 7.9° and the mean medial proximal tibial angle (MPTA) was 82.5°. Post-operatively, the mean HKA was 182.2° (180.1–184.7°), the tibial slope was 6.5° (4.2–12.9°) and the MPTA was 92.8° (90.6–93°). In both coronal and sagittal plane, all knees were within 2° from the planned value. The mean tourniquet time, by which the surgical time was measured, was 40 minutes and the average number of intra-operative fluoroscopic images was 10 (range: 7–14), significantly less than when using conventional techniques.

Conclusions

The use of PSI in OW-HTO allows accurate achievement of the desired correction, while shortening the OR time. The number of radiological exposures is also decreased, in comparison with the conventional osteotomies.
Literature
10.
go back to reference So SY, Lee SS, Jung EY et al (2020) Difference in joint line convergence angle between the supine and standing positions is the most important predictive factor of coronal correction error after medial opening wedge high tibial osteotomy. Knee Surg Sport Traumatol Arthrosc 28:1516–1525. https://doi.org/10.1007/s00167-019-05555-7CrossRef So SY, Lee SS, Jung EY et al (2020) Difference in joint line convergence angle between the supine and standing positions is the most important predictive factor of coronal correction error after medial opening wedge high tibial osteotomy. Knee Surg Sport Traumatol Arthrosc 28:1516–1525. https://​doi.​org/​10.​1007/​s00167-019-05555-7CrossRef
14.
go back to reference Brosset T, Pasquier G, Migaud H, Gougeon F (2011) Opening wedge high tibial osteotomy performed without filling the defect but with locking plate fixation (TomoFixTM) and early weight-bearing: prospective evaluation of bone union, precision and maintenance of correction in 51 cases. Orthop Traumatol Surg Res 97:705–711. https://doi.org/10.1016/j.otsr.2011.06.011CrossRefPubMed Brosset T, Pasquier G, Migaud H, Gougeon F (2011) Opening wedge high tibial osteotomy performed without filling the defect but with locking plate fixation (TomoFixTM) and early weight-bearing: prospective evaluation of bone union, precision and maintenance of correction in 51 cases. Orthop Traumatol Surg Res 97:705–711. https://​doi.​org/​10.​1016/​j.​otsr.​2011.​06.​011CrossRefPubMed
25.
go back to reference Pape D, Lorbach O, Schmitz C, Busch LC, Nicolien Van Giffen RS, Kohn DM (2009) Effect of a biplanar osteotomy on primary stability following high tibial osteotomy: a biomechanical cadaver study. Knee Surg Sport Traumatol Arthrosc:204–211(2010) Pape D, Lorbach O, Schmitz C, Busch LC, Nicolien Van Giffen RS, Kohn DM (2009) Effect of a biplanar osteotomy on primary stability following high tibial osteotomy: a biomechanical cadaver study. Knee Surg Sport Traumatol Arthrosc:204–211(2010)
28.
go back to reference Jud L, Fürnstahl P, Vlachopoulos L et al (2020) Malpositioning of patient-specific instruments within the possible degrees of freedom in high-tibial osteotomy has no considerable influence on mechanical leg axis correction. Knee Surg Sport Traumatol Arthrosc 28:1356–1364. https://doi.org/10.1007/s00167-019-05432-3CrossRef Jud L, Fürnstahl P, Vlachopoulos L et al (2020) Malpositioning of patient-specific instruments within the possible degrees of freedom in high-tibial osteotomy has no considerable influence on mechanical leg axis correction. Knee Surg Sport Traumatol Arthrosc 28:1356–1364. https://​doi.​org/​10.​1007/​s00167-019-05432-3CrossRef
29.
go back to reference Akamatsu Y, Kumagai K, Kobayashi H, Tsuji M, Saito T (2018) Effect of increased coronal inclination of the tibial plateau after opening-wedge high tibial osteotomy. Arthrosc J Arthrosc Relat Surg 34:2158–2169.e2CrossRef Akamatsu Y, Kumagai K, Kobayashi H, Tsuji M, Saito T (2018) Effect of increased coronal inclination of the tibial plateau after opening-wedge high tibial osteotomy. Arthrosc J Arthrosc Relat Surg 34:2158–2169.e2CrossRef
30.
go back to reference Babis GC, An K-N, Chao EYS, Larson DR, Rand JA, Sim FH (2008) Upper tibia osteotomy: long term results — realignment analysis using OASIS computer software. J Orthop Sci 13 Babis GC, An K-N, Chao EYS, Larson DR, Rand JA, Sim FH (2008) Upper tibia osteotomy: long term results — realignment analysis using OASIS computer software. J Orthop Sci 13
31.
go back to reference Babis GC, An K-N, Chao EYS, Rand JA, Sim FH (2002) Double level osteotomy of the knee: a method to retain joint-line obliquity. Clinical results. J Bone Jt Surg Am 84:1380–1388CrossRef Babis GC, An K-N, Chao EYS, Rand JA, Sim FH (2002) Double level osteotomy of the knee: a method to retain joint-line obliquity. Clinical results. J Bone Jt Surg Am 84:1380–1388CrossRef
35.
go back to reference Brown CW, Orme TJ, Richardson HD (1986) The rate of pseudarthrosis (Surgical nonunion) in patients who are smokers and patients who are nonsmokers: a comparison study. Spine (Phila Pa 1976) 11:942–943CrossRef Brown CW, Orme TJ, Richardson HD (1986) The rate of pseudarthrosis (Surgical nonunion) in patients who are smokers and patients who are nonsmokers: a comparison study. Spine (Phila Pa 1976) 11:942–943CrossRef
Metadata
Title
Early experience using patient-specific instrumentation in opening wedge high tibial osteotomy
Authors
Vlad Predescu
Alina-Maria Grosu
Iulian Gherman
Catalin Prescura
Valentin Hiohi
Bogdan Deleanu
Publication date
01-06-2021
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 6/2021
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-021-04964-z

Other articles of this Issue 6/2021

International Orthopaedics 6/2021 Go to the issue