Abstract

Purpose

Accurate and comprehensive data on cervical endplates is essential for developing and improving cervical devices. However, current literature on vertebral disc geometry is scarce or not suitable. The aim of this study was to obtain quantitative parameters of cervical endplates and provide morphometric references for designing cervical devices.

Methods

In this study, 19 human cervical spine cadaveric specimens were considered. Employing a reverse engineering system, the surface information of each endplate was recorded in digital cloud and then 3D reconstructed. A measurement protocol that included three sagittal and three frontal surface curves was developed. The information of surface curves and endplate concavity were obtained and analyzed. The parametric equations of endplate surfaces were deduced based on coordinates of landmarks, and the reliability was verified.

Results

The cervical endplate surface had a trend that to be transversely elongated gradually. The concavity depths of inferior endplates (1.88 to 2.13 mm) were significantly larger than those of superior endplates (0.62 to 0.84 mm). The most-concave points in inferior endplates were concentrated in the central portion, while always located in post-median region in superior endplates.

Conclusion

These results will give appropriate guidelines to design cervical prostheses without sacrificing valuable bone stock. The parametric equations applied for generating surface profile of cervical endplates may provide great convenience for subsequent studies.
Appendix
Available only for authorised users
Literature
9.
go back to reference Steffen T, Tsantrizos A, Aebi M (2000) Effect of implant design and endplate preparation on the compressive strength of interbody fusion constructs. Spine 25(9):1077–1084CrossRefPubMed Steffen T, Tsantrizos A, Aebi M (2000) Effect of implant design and endplate preparation on the compressive strength of interbody fusion constructs. Spine 25(9):1077–1084CrossRefPubMed
11.
go back to reference Panjabi MM, Duranceau J, Goel V, Oxland T, Takata K (1991) Cervical human vertebrae. Quantitative three-dimensional anatomy of the middle and lower regions. Spine 16(8):861–869CrossRefPubMed Panjabi MM, Duranceau J, Goel V, Oxland T, Takata K (1991) Cervical human vertebrae. Quantitative three-dimensional anatomy of the middle and lower regions. Spine 16(8):861–869CrossRefPubMed
14.
go back to reference Panjabi MM, Chen NC, Shin EK, Wang JL (2001) The cortical shell architecture of human cervical vertebral bodies. Spine 26(22):2478–2484CrossRefPubMed Panjabi MM, Chen NC, Shin EK, Wang JL (2001) The cortical shell architecture of human cervical vertebral bodies. Spine 26(22):2478–2484CrossRefPubMed
16.
go back to reference Silva MJ, Wang C, Keaveny TM, Hayes WC (1994) Direct and computed tomography thickness measurements of the human, lumbar vertebral shell and endplate. Bone 15(4):409–414CrossRefPubMed Silva MJ, Wang C, Keaveny TM, Hayes WC (1994) Direct and computed tomography thickness measurements of the human, lumbar vertebral shell and endplate. Bone 15(4):409–414CrossRefPubMed
18.
go back to reference Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86(2):420–428CrossRef Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86(2):420–428CrossRef
19.
go back to reference Bland JM, Altman DG (1997) Cronbach’s alpha. BMJ (Clinical research ed) 314(7080):572CrossRef Bland JM, Altman DG (1997) Cronbach’s alpha. BMJ (Clinical research ed) 314(7080):572CrossRef
26.
go back to reference Dmitriev AE, Cunningham BW, Hu N, Sell G, Vigna F, McAfee PC (2005) Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty: an in vitro human cadaveric model. Spine 30(10):1165–1172CrossRefPubMed Dmitriev AE, Cunningham BW, Hu N, Sell G, Vigna F, McAfee PC (2005) Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty: an in vitro human cadaveric model. Spine 30(10):1165–1172CrossRefPubMed
Metadata
Title
Morphometry evaluations of cervical osseous endplates based on three dimensional reconstructions
Authors
Hang Feng
Haoxi Li
Zhaoyu Ba
Zhaoxiong Chen
Xinhua Li
Desheng Wu
Publication date
01-06-2019
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 6/2019
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-018-4053-1

Other articles of this Issue 6/2019

International Orthopaedics 6/2019 Go to the issue