Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective

Authors: Christian Fabry, Jean Langlois, Moussa Hamadouche, Rainer Bader

Published in: International Orthopaedics | Issue 5/2016

Login to get access

Abstract

Introduction

Recurrent dislocation of total hip arthroplasty is a frequent indication for revision surgery. Hip joint stability depends on implant design, cup position and crucially on femoral head diameter. Due to an effective ultra-large diameter femoral head, dual-mobility cups are considered an attractive solution to prevent dislocation in unstable conditions. Although patients obviously benefit for many years in terms of mobility and pain, an increase of intra-prosthetic dislocation reports using dual-mobility cups has been recently observed. However, the failure mechanism of this implant-specific complication, which is characterized by the loss of the positive-locking between the femoral head and the mobile liner, is not yet completely understood.

Methods

A comprehensive search was performed with the PubMed database and a search engine to overview this topic and to identify potential causes for this implant-specific failure from a clinical and biomechanical perspective.

Results

Peri-operative findings indicate extensive fibrosis at the large articulation as well as cup loosening as potential causes. In addition, current research has shown that the failure mechanism is affected by the surface topography of the femoral neck and in particular by the design of the mobile liner.

Discussion

In clinical practice it is necessary to differentiate a classic dislocation between the mobile liner and the metallic shell from an intra-prosthetic dislocation between the femoral head and the liner.

Conclusion

Due to the increasing popularity of dual-mobility cups in total hip arthroplasty, the understanding of which implant-specific features or tissue response may increase the risk of intra-prosthetic dislocation is of major importance for reduced revision rates by using optimized surgical techniques and implant designs.
Literature
3.
go back to reference Abu-Amer Y, Darwech I, Clohisy JC (2007) Aseptic loosening of total joint replacements: mechanisms underlying osteolysis and potential therapies. Arthritis Res Ther 9 Suppl 1:S6 Abu-Amer Y, Darwech I, Clohisy JC (2007) Aseptic loosening of total joint replacements: mechanisms underlying osteolysis and potential therapies. Arthritis Res Ther 9 Suppl 1:S6
4.
go back to reference Prudhon JL, Steffann F, Ferreira A, Verdier R, Aslanian T, Caton J (2014) Cementless dual-mobility cup in total hip arthroplasty revision. Int Orthop 38(12):2463–8CrossRefPubMed Prudhon JL, Steffann F, Ferreira A, Verdier R, Aslanian T, Caton J (2014) Cementless dual-mobility cup in total hip arthroplasty revision. Int Orthop 38(12):2463–8CrossRefPubMed
5.
go back to reference Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W (2004) The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand 75(4):415–21CrossRefPubMed Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W (2004) The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand 75(4):415–21CrossRefPubMed
6.
go back to reference Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R (2007) Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys 29(4):465–71CrossRefPubMed Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R (2007) Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys 29(4):465–71CrossRefPubMed
7.
go back to reference Blom AW, Rogers M, Taylor AH, Pattison G, Whitehouse S, Bannister GC (2008) Dislocation following total hip replacement: the Avon orthopaedic centre experience. Ann R Coll Surg Engl 90(8):658–62CrossRefPubMedPubMedCentral Blom AW, Rogers M, Taylor AH, Pattison G, Whitehouse S, Bannister GC (2008) Dislocation following total hip replacement: the Avon orthopaedic centre experience. Ann R Coll Surg Engl 90(8):658–62CrossRefPubMedPubMedCentral
8.
go back to reference Mahomed NN, Barrett JA, Katz JN, Phillips CB, Losina E, Lew RA et al (2003) Rates and outcomes of primary and revision total hip replacement in the United States medicare population. J Bone Joint Surg Am 85-A(1):27–32PubMed Mahomed NN, Barrett JA, Katz JN, Phillips CB, Losina E, Lew RA et al (2003) Rates and outcomes of primary and revision total hip replacement in the United States medicare population. J Bone Joint Surg Am 85-A(1):27–32PubMed
9.
go back to reference Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty 20(1):11–9CrossRefPubMed Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH (2005) Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes. J Arthroplasty 20(1):11–9CrossRefPubMed
10.
go back to reference D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82(3):315–21PubMed D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82(3):315–21PubMed
11.
go back to reference Bader R, Steinhauser E, Scholz R, Simnacher M, Mittelmeier W (2004) Experimental analysis of neutral, asymmetric and constraint liners for total hip replacement: Investigation of range of motion and stability against joint instability. Z Orthop Grenzgeb 142(5):577–85CrossRef Bader R, Steinhauser E, Scholz R, Simnacher M, Mittelmeier W (2004) Experimental analysis of neutral, asymmetric and constraint liners for total hip replacement: Investigation of range of motion and stability against joint instability. Z Orthop Grenzgeb 142(5):577–85CrossRef
12.
go back to reference Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J (2009) Use of a dual mobility socket to manage total Hip arthroplasty instability. Clin Orthop Relat Res 467(2):465–72CrossRefPubMedPubMedCentral Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J (2009) Use of a dual mobility socket to manage total Hip arthroplasty instability. Clin Orthop Relat Res 467(2):465–72CrossRefPubMedPubMedCentral
13.
14.
go back to reference Thomsen M, Breusch SJ, Schneider U, Kubein-Meesenburg D, Nägerl H (2001) Developments in hip hemi-arthroplasty and theory of the link-chain dimeric hip prosthesis. Unfallchirurg 104(11):1061–7CrossRefPubMed Thomsen M, Breusch SJ, Schneider U, Kubein-Meesenburg D, Nägerl H (2001) Developments in hip hemi-arthroplasty and theory of the link-chain dimeric hip prosthesis. Unfallchirurg 104(11):1061–7CrossRefPubMed
15.
go back to reference Fabry C, Woernle C, Bader R (2014) Self-centering dual-mobility total hip systems: prediction of relative movements and realignment of different intermediate components. Proc Inst Mech Eng H 228(5):477–485CrossRefPubMed Fabry C, Woernle C, Bader R (2014) Self-centering dual-mobility total hip systems: prediction of relative movements and realignment of different intermediate components. Proc Inst Mech Eng H 228(5):477–485CrossRefPubMed
16.
go back to reference Chen QS, Lazennec JY, Guyen O, Kinbrum A, Berry DJ, An KN (2005) Technical note: validation of a motion analysis system for measuring the relative motion of the intermediate component of a tripolar total hip arthroplasty prosthesis. Med Eng Phys 27(6):505–12CrossRefPubMed Chen QS, Lazennec JY, Guyen O, Kinbrum A, Berry DJ, An KN (2005) Technical note: validation of a motion analysis system for measuring the relative motion of the intermediate component of a tripolar total hip arthroplasty prosthesis. Med Eng Phys 27(6):505–12CrossRefPubMed
17.
go back to reference Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F (2009) The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop 33(4):927–32CrossRefPubMedPubMedCentral Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F (2009) The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop 33(4):927–32CrossRefPubMedPubMedCentral
18.
go back to reference Hamadouche M, Biau DJ, Huten D, Musset T, Gaucher F (2010) The Use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 468(12):3248–54CrossRefPubMedPubMedCentral Hamadouche M, Biau DJ, Huten D, Musset T, Gaucher F (2010) The Use of a cemented dual mobility socket to treat recurrent dislocation. Clin Orthop Relat Res 468(12):3248–54CrossRefPubMedPubMedCentral
19.
20.
21.
go back to reference Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R (2014) A comparative and retrospective study of three hundred and twenty primary charnley type hip replacements with a minimum follow up of ten years to assess whether a dual mobility cup has a decreased dislocation risk. Int Orthop 38(6):1125–9CrossRefPubMedPubMedCentral Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R (2014) A comparative and retrospective study of three hundred and twenty primary charnley type hip replacements with a minimum follow up of ten years to assess whether a dual mobility cup has a decreased dislocation risk. Int Orthop 38(6):1125–9CrossRefPubMedPubMedCentral
22.
go back to reference Leclercq S, Benoit JY, de Rosa JP, Tallier E, Leteurtre C, Girardin PH (2013) Evora® chromium-cobalt dual mobility socket: results at a minimum 10 years’ follow-up. Orthop Traumatol Surg Res 99(8):923–8CrossRefPubMed Leclercq S, Benoit JY, de Rosa JP, Tallier E, Leteurtre C, Girardin PH (2013) Evora® chromium-cobalt dual mobility socket: results at a minimum 10 years’ follow-up. Orthop Traumatol Surg Res 99(8):923–8CrossRefPubMed
23.
go back to reference Odland AN, Sierra RJ (2014) Intraprosthetic dislocation of a contemporary dual-mobility design used during conversion THA. Orthopedics 37(12):e1124–8CrossRefPubMed Odland AN, Sierra RJ (2014) Intraprosthetic dislocation of a contemporary dual-mobility design used during conversion THA. Orthopedics 37(12):e1124–8CrossRefPubMed
24.
go back to reference Langlois J, El Hage S, Hamadouche M (2014) Intraprosthetic dislocation: a potentially serious complication of dual mobility acetabular cups. Skelet Radiol 43(7):1013–6 Langlois J, El Hage S, Hamadouche M (2014) Intraprosthetic dislocation: a potentially serious complication of dual mobility acetabular cups. Skelet Radiol 43(7):1013–6
25.
go back to reference Philippot R, Boyer B, Farizon F (2013) Intraprosthetic dislocation: a specific complication of the dual-mobility system. Clin Orthop Relat Res 471(3):965–70CrossRefPubMedPubMedCentral Philippot R, Boyer B, Farizon F (2013) Intraprosthetic dislocation: a specific complication of the dual-mobility system. Clin Orthop Relat Res 471(3):965–70CrossRefPubMedPubMedCentral
26.
go back to reference Combes A, Migaud H, Girard J, Duhamel A, Fessy MH (2013) Low rate of dislocation of dual-mobility cups in primary total hip arthroplasty. Clin Orthop Relat Res 471(12):3891–900CrossRefPubMedPubMedCentral Combes A, Migaud H, Girard J, Duhamel A, Fessy MH (2013) Low rate of dislocation of dual-mobility cups in primary total hip arthroplasty. Clin Orthop Relat Res 471(12):3891–900CrossRefPubMedPubMedCentral
27.
go back to reference Lecuire F, Benareau I, Rubini J, Basso M (2004) Intra-prosthetic dislocation of the bousquet dual mobility socket. Rev Chir Orthop Reparatrice Appar Mot 90(3):249–55CrossRefPubMed Lecuire F, Benareau I, Rubini J, Basso M (2004) Intra-prosthetic dislocation of the bousquet dual mobility socket. Rev Chir Orthop Reparatrice Appar Mot 90(3):249–55CrossRefPubMed
28.
go back to reference Hamadouche M, Arnould H, Bouxin B (2012) Is a cementless dual mobility socket in primary THA a reasonable option? Clin Orthop Relat Res 470(11):3048–53CrossRefPubMedPubMedCentral Hamadouche M, Arnould H, Bouxin B (2012) Is a cementless dual mobility socket in primary THA a reasonable option? Clin Orthop Relat Res 470(11):3048–53CrossRefPubMedPubMedCentral
29.
go back to reference Massin P, Orain V, Philippot R, Farizon F, Fessy MH (2012) Fixation failures of dual mobility cups: a mid-term study of 2601 hip replacements. Clin Orthop Relat Res 470(7):1932–40CrossRefPubMedPubMedCentral Massin P, Orain V, Philippot R, Farizon F, Fessy MH (2012) Fixation failures of dual mobility cups: a mid-term study of 2601 hip replacements. Clin Orthop Relat Res 470(7):1932–40CrossRefPubMedPubMedCentral
30.
go back to reference Lautridou C, Lebel B, Burdin G, Vielpeau C (2008) Survival of the cementless bousquet dual mobility cup: minimum 15-year follow-up of 437 total hip arthroplasties. Rev Chir Orthop Reparatrice Appar Mot 94(8):731–9CrossRefPubMed Lautridou C, Lebel B, Burdin G, Vielpeau C (2008) Survival of the cementless bousquet dual mobility cup: minimum 15-year follow-up of 437 total hip arthroplasties. Rev Chir Orthop Reparatrice Appar Mot 94(8):731–9CrossRefPubMed
31.
go back to reference Mertl P, Combes A, Leiber-Wackenheim F, Fessy MH, Girard J, Migaud H (2012) Recurrence of dislocation following total hip arthroplasty revision using dual mobility cups was rare in 180 hips followed over 7 years. HSS J 8(3):251–6CrossRefPubMedPubMedCentral Mertl P, Combes A, Leiber-Wackenheim F, Fessy MH, Girard J, Migaud H (2012) Recurrence of dislocation following total hip arthroplasty revision using dual mobility cups was rare in 180 hips followed over 7 years. HSS J 8(3):251–6CrossRefPubMedPubMedCentral
32.
go back to reference Lee HH, Lo YC, Lin LC, Wu SS (2008) Disassembly and dislocation of a bipolar hip prosthesis. J Formos Med Assoc 107(1):84–8CrossRefPubMed Lee HH, Lo YC, Lin LC, Wu SS (2008) Disassembly and dislocation of a bipolar hip prosthesis. J Formos Med Assoc 107(1):84–8CrossRefPubMed
33.
go back to reference Banka TR, Ast MP, Parks ML (2014) Early intraprosthetic dislocation in a revision dual-mobility hip prosthesis. Orthopedics 37(4):e395–7CrossRefPubMed Banka TR, Ast MP, Parks ML (2014) Early intraprosthetic dislocation in a revision dual-mobility hip prosthesis. Orthopedics 37(4):e395–7CrossRefPubMed
35.
go back to reference Langlais FL, Ropars M, Gaucher F, Musset T, Chaix O (2008) Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res 466(2):389–95CrossRefPubMedPubMedCentral Langlais FL, Ropars M, Gaucher F, Musset T, Chaix O (2008) Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res 466(2):389–95CrossRefPubMedPubMedCentral
36.
go back to reference Fabry C, Kaehler M, Herrmann S, Woernle C, Bader R (2014) Dynamic behavior of tripolar hip endoprostheses under physiological conditions and their effect on stability. Med Eng Phys 36(1):65–71CrossRefPubMed Fabry C, Kaehler M, Herrmann S, Woernle C, Bader R (2014) Dynamic behavior of tripolar hip endoprostheses under physiological conditions and their effect on stability. Med Eng Phys 36(1):65–71CrossRefPubMed
Metadata
Title
Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective
Authors
Christian Fabry
Jean Langlois
Moussa Hamadouche
Rainer Bader
Publication date
01-05-2016
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 5/2016
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-3000-7

Other articles of this Issue 5/2016

International Orthopaedics 5/2016 Go to the issue