Published in: Abdominal Radiology 5/2017

01-05-2017

Accurate simultaneous quantification of liver steatosis and iron overload in diffuse liver diseases with MRI

Authors: Manuela França, Ángel Alberich-Bayarri, Luis Martí-Bonmatí, Pedro Oliveira, Francisca Emanuel Costa, Graça Porto, José Ramón Vizcaíno, Javier Sanchez Gonzalez, Eduardo Ribeiro, João Oliveira, Helena Pessegueiro Miranda

Published in: Abdominal Radiology | Issue 5/2017

Login to get access

Abstract

Purpose

To evaluate the diagnostic performances of 3 Tesla multi-echo chemical shift-encoded gradient echo magnetic resonance (MECSE-MR) imaging to simultaneously quantify liver steatosis and iron overload in a wide spectrum of diffuse liver diseases having biopsy as reference standard.

Methods

MECSE-MR-acquired images were used to calculate fat fraction and iron content in a single breath-hold in 109 adult patients. Proton density fat fraction (PDFF) was prospectively estimated using complex-based data reconstruction with multipeak fat modeling. Water R2* was used to estimate iron content. Biopsy was obtained in all cases, grading liver steatosis, siderosis, inflammation, and fibrosis. Differences in PDFF and R2* values across histopathological grades were analyzed, and ROC curves analyses evaluated the MR diagnostic performance.

Results

Calculated fat fraction measurements showed significant differences (p < 0.001) among steatosis grades, being unaffected by the presence of inflammation or fibrosis (p ≥ 0.05). A strong correlation was found between fat fraction and steatosis grade (R S = 0.718, p < 0.001). Iron deposits did not affect fat fraction quantitation (p ≥ 0.05), except in cases with severe iron overload (grade 4). A strong positive correlation was also observed between R2* measurements and iron grades (R S = 0.704, p < 0.001). Calculated R2* values were not different across grades of steatosis, inflammation, and fibrosis (p ≥ 0.05).

Conclusion

A MECSE-MR sequence simultaneously quantifies liver steatosis and siderosis, regardless coexisting liver inflammation or fibrosis, with high accuracy in a wide spectrum of diffuse liver disorders. This sequence can be acquired within a single breath-hold and can be implemented in the routine MR evaluation of the liver.
Literature
4.
go back to reference Ratziu V, Charlotte F, Heurtier A, et al. (2005) Sampling variability of liver biopsy in nonalcoholic fatty liver disease. Gastroenterology 128:1898–1906CrossRefPubMed Ratziu V, Charlotte F, Heurtier A, et al. (2005) Sampling variability of liver biopsy in nonalcoholic fatty liver disease. Gastroenterology 128:1898–1906CrossRefPubMed
10.
15.
go back to reference Hines CDG, Frydrychowicz A, Hamilton G, et al. (2011) T1 independent, T2* corrected chemical shift based fat-water separation with multi-peak fat spectral modeling is an accurate and precise measure of hepatic steatosis. J Magn Reson Imaging 33:873–881. doi:10.1002/jmri.22514 CrossRefPubMedPubMedCentral Hines CDG, Frydrychowicz A, Hamilton G, et al. (2011) T1 independent, T2* corrected chemical shift based fat-water separation with multi-peak fat spectral modeling is an accurate and precise measure of hepatic steatosis. J Magn Reson Imaging 33:873–881. doi:10.​1002/​jmri.​22514 CrossRefPubMedPubMedCentral
16.
24.
25.
go back to reference Ishak K, Baptista A, Bianchi L, et al. (1995) Histological grading and staging of chronic hepatitis. J Hepatol 22:696–699CrossRefPubMed Ishak K, Baptista A, Bianchi L, et al. (1995) Histological grading and staging of chronic hepatitis. J Hepatol 22:696–699CrossRefPubMed
27.
go back to reference Hijona E, Sánchez-González J, Alústiza JM, et al. (2012) Accurate fat fraction quantification by multiecho gradient-recalled-echo magnetic resonance at 1.5 T in rats with nonalcoholic fatty liver disease. Eur J Radiol 81:1122–1127. doi:10.1016/j.ejrad.2011.02.065 CrossRefPubMed Hijona E, Sánchez-González J, Alústiza JM, et al. (2012) Accurate fat fraction quantification by multiecho gradient-recalled-echo magnetic resonance at 1.5 T in rats with nonalcoholic fatty liver disease. Eur J Radiol 81:1122–1127. doi:10.​1016/​j.​ejrad.​2011.​02.​065 CrossRefPubMed
28.
go back to reference Kang B-K, Yu ES, Lee SS, et al. (2012) Hepatic fat quantification: a prospective comparison of magnetic resonance spectroscopy and analysis methods for chemical-shift gradient echo magnetic resonance imaging with histologic assessment as the reference standard. Invest Radiol 47:368–375. doi:10.1097/RLI.0b013e31824baff3 CrossRefPubMed Kang B-K, Yu ES, Lee SS, et al. (2012) Hepatic fat quantification: a prospective comparison of magnetic resonance spectroscopy and analysis methods for chemical-shift gradient echo magnetic resonance imaging with histologic assessment as the reference standard. Invest Radiol 47:368–375. doi:10.​1097/​RLI.​0b013e31824baff3​ CrossRefPubMed
29.
go back to reference Tang A, Desai A, Hamilton G, et al. (2015) Accuracy of MR imaging-estimated proton density fat fraction for classification of dichotomized histologic steatosis grades in nonalcoholic fatty liver disease. Radiology 274:416–425. doi:10.1148/radiol.14140754 CrossRefPubMed Tang A, Desai A, Hamilton G, et al. (2015) Accuracy of MR imaging-estimated proton density fat fraction for classification of dichotomized histologic steatosis grades in nonalcoholic fatty liver disease. Radiology 274:416–425. doi:10.​1148/​radiol.​14140754 CrossRefPubMed
32.
go back to reference Chandarana H, Lim RP, Jensen JH, et al. (2009) Hepatic iron deposition in patients with liver disease: preliminary experience with breath-hold multiecho T2*-weighted sequence. Am J Roentgenol 193:1261–1267. doi:10.2214/AJR.08.1996 CrossRef Chandarana H, Lim RP, Jensen JH, et al. (2009) Hepatic iron deposition in patients with liver disease: preliminary experience with breath-hold multiecho T2*-weighted sequence. Am J Roentgenol 193:1261–1267. doi:10.​2214/​AJR.​08.​1996 CrossRef
36.
37.
go back to reference Heba ER, Desai A, Zand KA, et al. (2016) Accuracy and the effect of possible subject-based confounders of magnitude-based MRI for estimating hepatic proton density fat fraction in adults, using MR spectroscopy as reference. J Magn Reson Imaging 43:398–406. doi:10.1002/jmri.25006 CrossRefPubMed Heba ER, Desai A, Zand KA, et al. (2016) Accuracy and the effect of possible subject-based confounders of magnitude-based MRI for estimating hepatic proton density fat fraction in adults, using MR spectroscopy as reference. J Magn Reson Imaging 43:398–406. doi:10.​1002/​jmri.​25006 CrossRefPubMed
40.
go back to reference Meloni A, Positano V, Keilberg P, et al. (2012) Feasibility, reproducibility, and reliability for the T*2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68:543–551. doi:10.1002/mrm.23236 CrossRefPubMed Meloni A, Positano V, Keilberg P, et al. (2012) Feasibility, reproducibility, and reliability for the T*2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68:543–551. doi:10.​1002/​mrm.​23236 CrossRefPubMed
41.
go back to reference Tyagi A, Yeganeh O, Levin Y, et al. (2015) Intra- and inter-examination repeatability of magnetic resonance spectroscopy, magnitude-based MRI, and complex-based MRI for estimation of hepatic proton density fat fraction in overweight and obese children and adults. Abdom Imaging 40:3070–3077. doi:10.1007/s00261-015-0542-5 CrossRefPubMedPubMedCentral Tyagi A, Yeganeh O, Levin Y, et al. (2015) Intra- and inter-examination repeatability of magnetic resonance spectroscopy, magnitude-based MRI, and complex-based MRI for estimation of hepatic proton density fat fraction in overweight and obese children and adults. Abdom Imaging 40:3070–3077. doi:10.​1007/​s00261-015-0542-5 CrossRefPubMedPubMedCentral
42.
go back to reference Hernando D, Sharma SD, Aliyari Ghasabeh M et al (2016) Multisite, multivendor validation of the accuracy and reproducibility of proton-density fat-fraction quantification at 1.5 T and 3 T using a fat-water phantom. Magn Reson Med. doi: 10.1002/mrm.26228.CrossRefPubMedPubMedCentral Hernando D, Sharma SD, Aliyari Ghasabeh M et al (2016) Multisite, multivendor validation of the accuracy and reproducibility of proton-density fat-fraction quantification at 1.5 T and 3 T using a fat-water phantom. Magn Reson Med. doi: 10.​1002/​mrm.​26228.CrossRefPubMedPubMedCentral
Metadata
Title
Accurate simultaneous quantification of liver steatosis and iron overload in diffuse liver diseases with MRI
Authors
Manuela França
Ángel Alberich-Bayarri
Luis Martí-Bonmatí
Pedro Oliveira
Francisca Emanuel Costa
Graça Porto
José Ramón Vizcaíno
Javier Sanchez Gonzalez
Eduardo Ribeiro
João Oliveira
Helena Pessegueiro Miranda
Publication date
01-05-2017
Publisher
Springer US
Published in
Abdominal Radiology / Issue 5/2017
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-017-1048-0

Other articles of this Issue 5/2017

Abdominal Radiology 5/2017 Go to the issue

Classics in Abdominal Imaging

The T2-dark spot sign

Classics in Abdominal Imaging

The esophageal bird’s beak sign