Intra- and inter-examination repeatability of magnetic resonance spectroscopy, magnitude-based MRI, and complex-based MRI for estimation of hepatic proton density fat fraction in overweight and obese children and adults

Authors: Avishkar Tyagi, Omid Yeganeh, Yakir Levin, Jonathan C. Hooker, Gavin C. Hamilton, Tanya Wolfson, Anthony Gamst, Amir K. Zand, Elhamy Heba, Rohit Loomba, Jeffrey Schwimmer, Michael S. Middleton, Claude B. Sirlin

Published in: Abdominal Radiology | Issue 8/2015

Login to get access

Abstract

Purpose

Determine intra- and inter-examination repeatability of magnitude-based magnetic resonance imaging (MRI-M), complex-based magnetic resonance imaging (MRI-C), and magnetic resonance spectroscopy (MRS) at 3T for estimating hepatic proton density fat fraction (PDFF), and using MRS as a reference, confirm MRI-M and MRI-C accuracy.

Methods

Twenty-nine overweight and obese pediatric (n = 20) and adult (n = 9) subjects (23 male, 6 female) underwent three same-day 3T MR examinations. In each examination MRI-M, MRI-C, and single-voxel MRS were acquired three times. For each MRI acquisition, hepatic PDFF was estimated at the MRS voxel location. Intra- and inter-examination repeatability were assessed by computing standard deviations (SDs) and intra-class correlation coefficients (ICCs). Aggregate SD was computed for each method as the square root of the average of first repeat variances. MRI-M and MRI-C PDFF estimation accuracy was assessed using linear regression with MRS as a reference.

Results

For MRI-M, MRI-C, and MRS acquisitions, respectively, mean intra-examination SDs were 0.25%, 0.42%, and 0.49%; mean intra-examination ICCs were 0.999, 0.997, and 0.995; mean inter-examination SDs were 0.42%, 0.45%, and 0.46%; and inter-examination ICCs were 0.995, 0.992, and 0.990. Aggregate SD for each method was <0.9%. Using MRS as a reference, regression slope, intercept, average bias, and R 2, respectively, for MRI-M were 0.99%, 1.73%, 1.61%, and 0.986, and for MRI-C were 0.96%, 0.43%, 0.40%, and 0.991.

Conclusion

MRI-M, MRI-C, and MRS showed high intra- and inter-examination hepatic PDFF estimation repeatability in overweight and obese subjects. Longitudinal hepatic PDFF change >1.8% (twice the maximum aggregate SD) may represent real change rather than measurement imprecision. Further research is needed to assess whether examinations performed on different days or with different MR technologists affect repeatability of MRS voxel placement and MRS-based PDFF measurements.
Appendix
Available only for authorised users
Literature
5.
go back to reference Thomsen C, Becker U, Winkler K, et al. (1994) Quantification of liver fat using magnetic resonance spectroscopy. Magn Reson Imaging 12:487–495CrossRefPubMed Thomsen C, Becker U, Winkler K, et al. (1994) Quantification of liver fat using magnetic resonance spectroscopy. Magn Reson Imaging 12:487–495CrossRefPubMed
6.
go back to reference Johnson NA, Walton DW, Sachinwalla T, et al. (2008) Noninvasive assessment of hepatic lipid composition: advancing understanding and management of fatty liver disorders. Hepatology 47:1513–1523. doi:10.1002/hep.22220 CrossRefPubMed Johnson NA, Walton DW, Sachinwalla T, et al. (2008) Noninvasive assessment of hepatic lipid composition: advancing understanding and management of fatty liver disorders. Hepatology 47:1513–1523. doi:10.​1002/​hep.​22220 CrossRefPubMed
12.
go back to reference Artz NS, Haufe WM, Hooker CA, et al. (2015) Reproducibility of MR-based liver fat quantification across field strength: same-day comparison between 1.5T and 3T in obese subjects. J Magn Reson Imaging 42:811–817. doi:10.1002/jmri.24842 CrossRefPubMed Artz NS, Haufe WM, Hooker CA, et al. (2015) Reproducibility of MR-based liver fat quantification across field strength: same-day comparison between 1.5T and 3T in obese subjects. J Magn Reson Imaging 42:811–817. doi:10.​1002/​jmri.​24842 CrossRefPubMed
13.
go back to reference Guaraldi G, Besutti G, Stentarelli C, et al. (2012) Magnetic resonance for quantitative assessment of liver steatosis: a new potential tool to monitor antiretroviral-drug-related toxicities. Antivir Ther 17:965–971. doi:10.3851/IMP2228 CrossRefPubMed Guaraldi G, Besutti G, Stentarelli C, et al. (2012) Magnetic resonance for quantitative assessment of liver steatosis: a new potential tool to monitor antiretroviral-drug-related toxicities. Antivir Ther 17:965–971. doi:10.​3851/​IMP2228 CrossRefPubMed
14.
go back to reference Johnson BL, Schroeder ME, Wolfson T, et al. (2014) Effect of flip angle on the accuracy and repeatability of hepatic proton density fat fraction estimation by complex data-based, T1-independent, T2*-corrected, spectrum-modeled MRI. J Magn Reson Imaging 39:440–447. doi:10.1002/jmri.24153 PubMedCentralCrossRefPubMed Johnson BL, Schroeder ME, Wolfson T, et al. (2014) Effect of flip angle on the accuracy and repeatability of hepatic proton density fat fraction estimation by complex data-based, T1-independent, T2*-corrected, spectrum-modeled MRI. J Magn Reson Imaging 39:440–447. doi:10.​1002/​jmri.​24153 PubMedCentralCrossRefPubMed
15.
go back to reference Kang GH, Cruite I, Shiehmorteza M, et al. (2011) Reproducibility of MRI-determined proton density fat fraction across two different MR scanner platforms. J Magn Reson Imaging 34:928–934. doi:10.1002/jmri.22701 CrossRefPubMed Kang GH, Cruite I, Shiehmorteza M, et al. (2011) Reproducibility of MRI-determined proton density fat fraction across two different MR scanner platforms. J Magn Reson Imaging 34:928–934. doi:10.​1002/​jmri.​22701 CrossRefPubMed
23.
go back to reference Hines CD, Frydrychowicz A, Hamilton G, et al. (2011) T(1) independent, T(2) (*) corrected chemical shift based fat-water separation with multi-peak fat spectral modeling is an accurate and precise measure of hepatic steatosis. J Magn Reson Imaging 33:873–881. doi:10.1002/jmri.22514 PubMedCentralCrossRefPubMed Hines CD, Frydrychowicz A, Hamilton G, et al. (2011) T(1) independent, T(2) (*) corrected chemical shift based fat-water separation with multi-peak fat spectral modeling is an accurate and precise measure of hepatic steatosis. J Magn Reson Imaging 33:873–881. doi:10.​1002/​jmri.​22514 PubMedCentralCrossRefPubMed
26.
go back to reference Noureddin M, Lam J, Peterson MR, et al. (2013) Utility of magnetic resonance imaging versus histology for quantifying changes in liver fat in nonalcoholic fatty liver disease trials. Hepatology 58:1930–1940. doi:10.1002/hep.26455 CrossRefPubMed Noureddin M, Lam J, Peterson MR, et al. (2013) Utility of magnetic resonance imaging versus histology for quantifying changes in liver fat in nonalcoholic fatty liver disease trials. Hepatology 58:1930–1940. doi:10.​1002/​hep.​26455 CrossRefPubMed
29.
go back to reference Naressi A, Couturier C, Devos JM, et al. (2001) Java-based graphical user interface for MRUI, a software package for quantitation of in vivo/medical magnetic resonance spectroscopy signals. Comput Biol Med 31:269–286CrossRefPubMed Naressi A, Couturier C, Devos JM, et al. (2001) Java-based graphical user interface for MRUI, a software package for quantitation of in vivo/medical magnetic resonance spectroscopy signals. Comput Biol Med 31:269–286CrossRefPubMed
30.
go back to reference Vanhamme L, van den Boogaart A, Van Huffel S (1997) Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson 129:35–43CrossRefPubMed Vanhamme L, van den Boogaart A, Van Huffel S (1997) Improved method for accurate and efficient quantification of MRS data with use of prior knowledge. J Magn Reson 129:35–43CrossRefPubMed
Metadata
Title
Intra- and inter-examination repeatability of magnetic resonance spectroscopy, magnitude-based MRI, and complex-based MRI for estimation of hepatic proton density fat fraction in overweight and obese children and adults
Authors
Avishkar Tyagi
Omid Yeganeh
Yakir Levin
Jonathan C. Hooker
Gavin C. Hamilton
Tanya Wolfson
Anthony Gamst
Amir K. Zand
Elhamy Heba
Rohit Loomba
Jeffrey Schwimmer
Michael S. Middleton
Claude B. Sirlin
Publication date
01-10-2015
Publisher
Springer US
Published in
Abdominal Radiology / Issue 8/2015
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-015-0542-5

Other articles of this Issue 8/2015

Abdominal Radiology 8/2015 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.