Abstract

Purpose

We report our initial experiences with use of a new technique we developed for implantation of Sonoma Crx intramedullary rod in patients with displaced clavicle fractures.

Methods

A total of 35 patients (mean age 41.82 ± 13.65, male:female ratio = 21:14) having Robinson Types 1b, 2b and 3b displaced midshaft fractures with >2 cm clavicle shortening were included into the study. A single small incision (~1 cm) was made over the anteromedial aspect of the involved clavicle and an appropriate sized intramedullary nail was inserted in reverse (mirror) configuration of that has been suggested by the manufacturer. Functional assessment was made using Constant shoulder and disability of the arm shoulder and hand scoring.

Results

Mean time of operation was 51.20 ± 10.56 min and mean time of fluoroscopy was 2.33 ± 1.12 min. One patient had implant failure 2 months after the operation and was revised to a new implant. Superficial or deep wound infection, hematoma, neurovascular complication, substance irritation or implant failure did not occur. Follow-up ranged from 12 to 45 months (mean 28.5 ± 9.95 months). At the latest follow-up, mean Constant shoulder score was 93.14 ± 4.06 (ranging from 84.00 to 100.00) and mean disability of the arm shoulder and hand score was 3.68 ± 1.73 (ranging from 0.0 to 6.80).

Conclusion

The technique we described herein provided successful procedural outcomes, eliminated the need for deep dissection of the fracture site and reduced the operation time. Further study on larger populations is warranted to confirm these findings.
Literature
1.
go back to reference Nowak J, Mallmin H, Larsson S. The aetiology and epidemiology of clavicular fractures. A prospective study during a two-year period in Uppsala, Sweden. Injury. 2000;31:350–8.CrossRef Nowak J, Mallmin H, Larsson S. The aetiology and epidemiology of clavicular fractures. A prospective study during a two-year period in Uppsala, Sweden. Injury. 2000;31:350–8.CrossRef
2.
go back to reference Van Tassel D, Owens BD, Pointer L. Moriatis Wolf J. Incidence of clavicle fractures in sports: analysis of the NEISS Database. Int J Sports Med. 2014;35:83–6.PubMed Van Tassel D, Owens BD, Pointer L. Moriatis Wolf J. Incidence of clavicle fractures in sports: analysis of the NEISS Database. Int J Sports Med. 2014;35:83–6.PubMed
3.
go back to reference van Laarhoven JJ, Ferree S, Houwert RM, Hietbrink F, Verleisdonk EM, Leenen LP. Demographics of the injury pattern in severely injured patients with an associated clavicle fracture: a retrospective observational cohort study. World J Emerg Surg. 2013;8:36.CrossRefPubMedPubMedCentral van Laarhoven JJ, Ferree S, Houwert RM, Hietbrink F, Verleisdonk EM, Leenen LP. Demographics of the injury pattern in severely injured patients with an associated clavicle fracture: a retrospective observational cohort study. World J Emerg Surg. 2013;8:36.CrossRefPubMedPubMedCentral
4.
go back to reference Postacchini F, Gumina S, De Santis P, Albo F. Epidemiology of clavicle fractures. J Shoulder Elbow Surg. 2002;11:452–6.CrossRefPubMed Postacchini F, Gumina S, De Santis P, Albo F. Epidemiology of clavicle fractures. J Shoulder Elbow Surg. 2002;11:452–6.CrossRefPubMed
5.
go back to reference Khan LA, Bradnock TJ, Scott C, Robinson CM. Fractures of the clavicle. J Bone Joint Surg Am. 2009;91:447–60.CrossRefPubMed Khan LA, Bradnock TJ, Scott C, Robinson CM. Fractures of the clavicle. J Bone Joint Surg Am. 2009;91:447–60.CrossRefPubMed
6.
go back to reference Riehl JT, Athans BJ, Munro MW, Langford JR, Kupiszewski SJ, Haidukewych GJ, et al. Minimally displaced clavicle fracture after high-energy injury: are they likely to displace? Can J Surg. 2014;57:169–74.CrossRefPubMedPubMedCentral Riehl JT, Athans BJ, Munro MW, Langford JR, Kupiszewski SJ, Haidukewych GJ, et al. Minimally displaced clavicle fracture after high-energy injury: are they likely to displace? Can J Surg. 2014;57:169–74.CrossRefPubMedPubMedCentral
7.
go back to reference Liu GD, Tong SL, Ou S, Zhou LS, Fei J, Nan GX, Gu JW. Operative versus non-operative treatment for clavicle fracture: a meta-analysis. Int Orthop. 2013;37:1495–500.CrossRefPubMedPubMedCentral Liu GD, Tong SL, Ou S, Zhou LS, Fei J, Nan GX, Gu JW. Operative versus non-operative treatment for clavicle fracture: a meta-analysis. Int Orthop. 2013;37:1495–500.CrossRefPubMedPubMedCentral
8.
9.
go back to reference Lee YS, Lau MJ, Tseng YC, Chen WC, Kao HY, Wei JD. Comparison of the efficacy of hook plate versus tension band wire in the treatment of unstable fractures of the distal clavicle. Int Orthop. 2009;33:1401–5.CrossRefPubMed Lee YS, Lau MJ, Tseng YC, Chen WC, Kao HY, Wei JD. Comparison of the efficacy of hook plate versus tension band wire in the treatment of unstable fractures of the distal clavicle. Int Orthop. 2009;33:1401–5.CrossRefPubMed
10.
go back to reference Robinson CM. Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br. 1998;80:476–84.CrossRefPubMed Robinson CM. Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br. 1998;80:476–84.CrossRefPubMed
11.
go back to reference Assobhi JE. Reconstruction plate versus minimal invasive retrograde titanium elastic nail fixation for displaced midclavicular fractures. J Orthop Traumatol. 2011;12:185–92.CrossRefPubMedPubMedCentral Assobhi JE. Reconstruction plate versus minimal invasive retrograde titanium elastic nail fixation for displaced midclavicular fractures. J Orthop Traumatol. 2011;12:185–92.CrossRefPubMedPubMedCentral
12.
go back to reference King PR, Ikram A. Lamberts RP The treatment of clavicular shaft fractures with an innovative locked intramedullary device. J Shoulder Elbow Surg. 2014;17:e1–6. King PR, Ikram A. Lamberts RP The treatment of clavicular shaft fractures with an innovative locked intramedullary device. J Shoulder Elbow Surg. 2014;17:e1–6.
13.
go back to reference Richardson M, Asadollahi S, Richardson L. Management of acute displaced midshaft clavicular fractures using Herbert cannulated screw: technique and results in 114 patients. Int J Shoulder Surg. 2013;7:52–8.CrossRefPubMedPubMedCentral Richardson M, Asadollahi S, Richardson L. Management of acute displaced midshaft clavicular fractures using Herbert cannulated screw: technique and results in 114 patients. Int J Shoulder Surg. 2013;7:52–8.CrossRefPubMedPubMedCentral
14.
go back to reference Wu CL, Chang HC, Lu KH. Risk factors for nonunion in 337 displaced midshaft clavicular fractures treated with Knowles pin fixation. Arch Orthop Trauma Surg. 2013;133:15–22.CrossRefPubMed Wu CL, Chang HC, Lu KH. Risk factors for nonunion in 337 displaced midshaft clavicular fractures treated with Knowles pin fixation. Arch Orthop Trauma Surg. 2013;133:15–22.CrossRefPubMed
15.
go back to reference Strauss EJ, Egol KA, France MA, Koval KJ, Zuckerman JD. Complications of intramedullary Hagie pin fixation for acute midshaft clavicle fractures. J Shoulder Elbow Surg. 2007;16:280–4.CrossRefPubMed Strauss EJ, Egol KA, France MA, Koval KJ, Zuckerman JD. Complications of intramedullary Hagie pin fixation for acute midshaft clavicle fractures. J Shoulder Elbow Surg. 2007;16:280–4.CrossRefPubMed
17.
go back to reference Narsaria N, Singh AK, Arun GR, Seth RR. Surgical fixation of displaced midshaft clavicle fractures: elastic intramedullary nailing versus precontoured plating. J Orthop Traumatol. 2014;25:165–71.CrossRef Narsaria N, Singh AK, Arun GR, Seth RR. Surgical fixation of displaced midshaft clavicle fractures: elastic intramedullary nailing versus precontoured plating. J Orthop Traumatol. 2014;25:165–71.CrossRef
18.
go back to reference Jubel A, Andermahr J, Schiffer G, Tsironis K, Rehm KE. Elastic stable intramedullary nailing of midclavicular fractures with a titanium nail. Clin Orthop Relat Res. 2003;408:279–85.CrossRef Jubel A, Andermahr J, Schiffer G, Tsironis K, Rehm KE. Elastic stable intramedullary nailing of midclavicular fractures with a titanium nail. Clin Orthop Relat Res. 2003;408:279–85.CrossRef
19.
go back to reference Lu CC, Liu PC, Huang SH, Hsieh CH, Tien YC, Chien SH. Complications and technical pitfalls of titanium elastic nail fixation for midclavicular fractures. Orthopedics. 2014;37:e377–83.CrossRefPubMed Lu CC, Liu PC, Huang SH, Hsieh CH, Tien YC, Chien SH. Complications and technical pitfalls of titanium elastic nail fixation for midclavicular fractures. Orthopedics. 2014;37:e377–83.CrossRefPubMed
20.
go back to reference King R, Ikram A. Intramedullary locked fixation of clavicle shaft fractures-review of early results. Bone Joint J Orthopaedic Proc Suppl. 2013;95:71–71. King R, Ikram A. Intramedullary locked fixation of clavicle shaft fractures-review of early results. Bone Joint J Orthopaedic Proc Suppl. 2013;95:71–71.
21.
go back to reference Wilson DJ, Weaver DL, Balog TP. Arrington ED Early postoperative failure of a new intramedullary fixation device for midshaft clavicle fractures. Orthopedics. 2013;36:1450–3.CrossRef Wilson DJ, Weaver DL, Balog TP. Arrington ED Early postoperative failure of a new intramedullary fixation device for midshaft clavicle fractures. Orthopedics. 2013;36:1450–3.CrossRef
22.
go back to reference Palmer DK, Husain A, Phipatanakul WP, Wongworawat MD. Failure of a new intramedullary device in fixation of clavicle fractures: a report of two cases and review of the literature. J Shoulder Elbow Surg. 2011;20:1–4.CrossRef Palmer DK, Husain A, Phipatanakul WP, Wongworawat MD. Failure of a new intramedullary device in fixation of clavicle fractures: a report of two cases and review of the literature. J Shoulder Elbow Surg. 2011;20:1–4.CrossRef
23.
go back to reference Payne DE, Wray WH, Ruch DS, Zura RD, Moorman CT. Outcome of intramedullary fixation of clavicular fractures. Am J Orthop (Belle Mead NJ]. 2011;40:E99–104.PubMed Payne DE, Wray WH, Ruch DS, Zura RD, Moorman CT. Outcome of intramedullary fixation of clavicular fractures. Am J Orthop (Belle Mead NJ]. 2011;40:E99–104.PubMed
24.
go back to reference King PR, Scheepers S, Ikram A. Anatomy of the clavicle and its medullary canal: a computed tomography study. Eur J Orthop Surg Traumatol. 2014;24:37–42.CrossRefPubMed King PR, Scheepers S, Ikram A. Anatomy of the clavicle and its medullary canal: a computed tomography study. Eur J Orthop Surg Traumatol. 2014;24:37–42.CrossRefPubMed
Metadata
Title
Minimally invasive implantation of a novel flexible intramedullary nail in patients with displaced midshaft clavicle fractures
Authors
M. Calbiyik
S. Zehir
D. Ipek
Publication date
01-12-2016
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Trauma and Emergency Surgery / Issue 6/2016
Print ISSN: 1863-9933
Electronic ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-015-0567-z

Other articles of this Issue 6/2016

European Journal of Trauma and Emergency Surgery 6/2016 Go to the issue

Letter to the Editor

Letter to the editor