Iterative reconstruction does not substantially delay CT imaging in an emergency setting

Authors: Martin J. Willemink, Arnold M. R. Schilham, Tim Leiner, Willem P. Th. M. Mali, Pim A. de Jong, Ricardo P. J. Budde

Published in: Insights into Imaging | Issue 3/2013

Login to get access

Abstract

Objectives

To evaluate the effects of iterative reconstruction (IR) on reconstruction time and speed in two commonly encountered acquisition protocols in an emergency setting: pulmonary CT angiography (CTA) and total body trauma CT.

Methods

Twenty-five patients underwent a pulmonary CTA for evaluation of pulmonary embolisms and 15 patients underwent a total body CT after a traumatic event on a 256-slice CT. Images were reconstructed with filtered back-projection (FBP) and two IR levels. Reconstruction time and speed were quantified using custom written software.

Results

Mean reconstruction time delays for pulmonary CTAs were 10 ± 10 s and 12 ± 12 s for IR levels 2 and 4, respectively, and 44 ± 8 s and 45 ± 7 s for total body trauma CTs for IR levels 1 and 6, respectively. Mean reconstruction times and speeds for pulmonary CTAs were 26 ± 7 s, 36 ± 9 s and 38 ± 12 s, and 26.7 ± 5.6 slices/s, 18.7 ± 2.3 slices/s and 18.0 ± 2.8 slices/s for FBP, IR levels 2 and 4, respectively. For total body trauma CTs these values were 87 ± 15 s, 132 ± 17 s and 132 ± 18 s, and 20.1 ± 1.6 slices/s, 13.2 ± 0.8 slices/s and 13.2 ± 0.6 slices/s for FBP, IR levels 1 and 6, respectively.

Conclusions

IR does not result in clinically important CT image reconstruction delays in an emergency setting. No substantial differences in reconstruction time and speed were found between different IR levels.

Main Messages

  • IR delayed total pulmonary CTA reconstruction with 10–12 s and total-body trauma CT with 44–45 s
  • IR is not substantially delaying reconstruction in emergency CT imaging
  • Reconstruction time and speed are similar for different levels of IR
Literature
1.
go back to reference Sato J, Akahane M, Inano S et al (2011) Effect of radiation dose and adaptive statistical iterative reconstruction on image quality of pulmonary computed tomography. Jpn J Radiol 30:146–153PubMedCrossRef Sato J, Akahane M, Inano S et al (2011) Effect of radiation dose and adaptive statistical iterative reconstruction on image quality of pulmonary computed tomography. Jpn J Radiol 30:146–153PubMedCrossRef
2.
go back to reference Moscariello A, Takx RA, Schoepf UJ et al (2011) Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol 21:2130–2138PubMedCrossRef Moscariello A, Takx RA, Schoepf UJ et al (2011) Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol 21:2130–2138PubMedCrossRef
3.
go back to reference Singh S, Kalra MK, Gilman MD et al (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology 259:565–573PubMedCrossRef Singh S, Kalra MK, Gilman MD et al (2011) Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study. Radiology 259:565–573PubMedCrossRef
4.
go back to reference Pontana F, Duhamel A, Pagniez J et al (2011) Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 21:636–643PubMedCrossRef Pontana F, Duhamel A, Pagniez J et al (2011) Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 21:636–643PubMedCrossRef
5.
go back to reference Leipsic J, Labounty TM, Heilbron B et al (2010) Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study. AJR Am J Roentgenol 195:655–660PubMedCrossRef Leipsic J, Labounty TM, Heilbron B et al (2010) Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study. AJR Am J Roentgenol 195:655–660PubMedCrossRef
6.
7.
go back to reference Willemink MJ, Leiner T, de Jong PA et al (2013) Iterative reconstruction techniques for computed tomography part 2: initial results in dose reduction and image quality. Eur Radiol. doi:10.1007/s00330-012-2764-z Willemink MJ, Leiner T, de Jong PA et al (2013) Iterative reconstruction techniques for computed tomography part 2: initial results in dose reduction and image quality. Eur Radiol. doi:10.​1007/​s00330-012-2764-z
8.
go back to reference Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623PubMedCrossRef Katsura M, Matsuda I, Akahane M et al (2012) Model-based iterative reconstruction technique for radiation dose reduction in chest CT: comparison with the adaptive statistical iterative reconstruction technique. Eur Radiol 22:1613–1623PubMedCrossRef
9.
go back to reference Nelson RC, Feuerlein S, Boll DT (2011) New iterative reconstruction techniques for cardiovascular computed tomography: how do they work, and what are the advantages and disadvantages? J Cardiovasc Comput Tomogr 5:286–292PubMedCrossRef Nelson RC, Feuerlein S, Boll DT (2011) New iterative reconstruction techniques for cardiovascular computed tomography: how do they work, and what are the advantages and disadvantages? J Cardiovasc Comput Tomogr 5:286–292PubMedCrossRef
10.
go back to reference Pan X, Sidky EY, Vannier M (2009) Why do commercial CT scanners still employ traditional, filtered back-projection for image reconstruction? Inverse Probl 25:1230009PubMedPubMedCentralCrossRef Pan X, Sidky EY, Vannier M (2009) Why do commercial CT scanners still employ traditional, filtered back-projection for image reconstruction? Inverse Probl 25:1230009PubMedPubMedCentralCrossRef
11.
go back to reference Fleischmann D, Boas FE (2011) Computed tomography—old ideas and new technology. Eur Radiol 21:510–517PubMedCrossRef Fleischmann D, Boas FE (2011) Computed tomography—old ideas and new technology. Eur Radiol 21:510–517PubMedCrossRef
12.
go back to reference Yanagawa M, Honda O, Kikuyama A et al (2012) Pulmonary nodules: effect of adaptive statistical iterative reconstruction (ASIR) technique on performance of a computer-aided detection (CAD) system—comparison of performance between different-dose CT scans. Eur J Radiol 81:2877–2886PubMedCrossRef Yanagawa M, Honda O, Kikuyama A et al (2012) Pulmonary nodules: effect of adaptive statistical iterative reconstruction (ASIR) technique on performance of a computer-aided detection (CAD) system—comparison of performance between different-dose CT scans. Eur J Radiol 81:2877–2886PubMedCrossRef
13.
go back to reference Noel PB, Fingerle AA, Renger B, Munzel D, Rummeny EJ, Dobritz M (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197:1404–1409PubMedCrossRef Noel PB, Fingerle AA, Renger B, Munzel D, Rummeny EJ, Dobritz M (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197:1404–1409PubMedCrossRef
14.
go back to reference Kligerman S, Mehta D, Farnadesh M, Jeudy J, Olsen K, White C (2013) Use of a hybrid iterative reconstruction technique to reduce image noise and improve image quality in obese patients undergoing computed tomographic pulmonary angiography. J Thorac Imaging 28:49–59PubMedCrossRef Kligerman S, Mehta D, Farnadesh M, Jeudy J, Olsen K, White C (2013) Use of a hybrid iterative reconstruction technique to reduce image noise and improve image quality in obese patients undergoing computed tomographic pulmonary angiography. J Thorac Imaging 28:49–59PubMedCrossRef
15.
go back to reference Funama Y, Taguchi K, Utsunomiya D et al (2011) Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr 35:480–485PubMedPubMedCentralCrossRef Funama Y, Taguchi K, Utsunomiya D et al (2011) Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography. J Comput Assist Tomogr 35:480–485PubMedPubMedCentralCrossRef
16.
go back to reference Scheffel H, Stolzmann P, Schlett CL et al (2011) Coronary artery plaques: cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. Eur J Radiol 81:e363–e369PubMedCrossRef Scheffel H, Stolzmann P, Schlett CL et al (2011) Coronary artery plaques: cardiac CT with model-based and adaptive-statistical iterative reconstruction technique. Eur J Radiol 81:e363–e369PubMedCrossRef
17.
go back to reference Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm–initial clinical experience. Radiology 254:145–153PubMedCrossRef Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm–initial clinical experience. Radiology 254:145–153PubMedCrossRef
18.
go back to reference May MS, Wust W, Brand M et al (2011) Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography. Invest Radiol 46:465–470PubMedCrossRef May MS, Wust W, Brand M et al (2011) Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography. Invest Radiol 46:465–470PubMedCrossRef
19.
go back to reference Korn A, Fenchel M, Bender B et al (2011) Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol 33:218–224PubMedCrossRef Korn A, Fenchel M, Bender B et al (2011) Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol 33:218–224PubMedCrossRef
20.
go back to reference Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301PubMedCrossRef Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295–301PubMedCrossRef
21.
go back to reference Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47:482–489PubMedCrossRef Yamada Y, Jinzaki M, Tanami Y et al (2012) Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: a pilot study. Invest Radiol 47:482–489PubMedCrossRef
Metadata
Title
Iterative reconstruction does not substantially delay CT imaging in an emergency setting
Authors
Martin J. Willemink
Arnold M. R. Schilham
Tim Leiner
Willem P. Th. M. Mali
Pim A. de Jong
Ricardo P. J. Budde
Publication date
01-06-2013
Publisher
Springer Berlin Heidelberg
Published in
Insights into Imaging / Issue 3/2013
Electronic ISSN: 1869-4101
DOI
https://doi.org/10.1007/s13244-013-0226-z

Other articles of this Issue 3/2013

Insights into Imaging 3/2013 Go to the issue