Published in: BMC Cardiovascular Disorders 1/2021

Open Access 01-12-2021 | Computed Tomography | Research

Subclinical cardiac perforation by cardiac implantable electronic device leads detected by cardiac computed tomography

Authors: Yeong-Min Lim, Jae-Sun Uhm, Min Kim, In-Soo Kim, Moo-Nyun Jin, Hee Tae Yu, Tae-Hoon Kim, Hye-Jeong Lee, Young-Jin Kim, Boyoung Joung, Hui-Nam Pak, Moon-Hyoung Lee

Published in: BMC Cardiovascular Disorders | Issue 1/2021

Login to get access

Abstract

Background

The relationship between the characteristics of cardiac implantable electronic device (CIED) leads and subclinical cardiac perforations remains unclear. This study aimed to evaluate the incidence of subclinical cardiac perforation among various CIED leads using cardiac computed tomography (CT).

Methods

A total of 271 consecutive patients with 463 CIED leads, who underwent cardiac CT after CIED implantation, were included in this retrospective observational study. Cardiac CT images were reviewed by one radiologist and two cardiologists. Subclinical perforation was defined as traversal of the lead tip past the outer myocardial layer without symptoms and signs related to cardiac perforation. We compared the subclinical cardiac perforation rates of the available lead types.

Results

A total of 219, 49, and 3 patients had pacemakers, implantable cardioverter-defibrillators, and cardiac resynchronization therapy, respectively. The total subclinical cardiac perforation rate was 5.6%. Subclinical cardiac perforation by screw-in ventricular leads was significantly more frequent than that caused by tined ventricular leads (13.3% vs 3.3%, respectively, p = 0.002). There were no significant differences in the incidence of cardiac perforation between atrial and ventricular leads, screw-in and tined atrial leads, pacing and defibrillator ventricular leads, nor between magnetic resonance (MR)-conditional and MR-unsafe screw-in ventricular leads. Screw-in ventricular leads were significantly associated with subclinical cardiac perforation [odds ratio, 4.554; 95% confidence interval, 1.587–13.065, p = 0.005]. There was no case subclinical cardiac perforation by septal ventricular leads.

Conclusions

Subclinical cardiac perforation by screw-in ventricular leads is not rare. Septal pacing may be helpful in avoiding cardiac perforation.
Appendix
Available only for authorised users
Literature
1.
go back to reference Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007;30(1):28–32.CrossRef Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007;30(1):28–32.CrossRef
2.
go back to reference Ellenbogen KA, Hellkamp AS, Wilkoff BL, Camunas JL, Love JC, Hadjis TA, Lee KL, Lamas GA. Complications arising after implantation of DDD pacemakers: the MOST experience. Am J Cardiol. 2003;92(6):740–1.CrossRef Ellenbogen KA, Hellkamp AS, Wilkoff BL, Camunas JL, Love JC, Hadjis TA, Lee KL, Lamas GA. Complications arising after implantation of DDD pacemakers: the MOST experience. Am J Cardiol. 2003;92(6):740–1.CrossRef
3.
go back to reference Wilkoff BL. Lead failures: dealing with even less perfect. Heart Rhythm. 2007;4(7):897–9.CrossRef Wilkoff BL. Lead failures: dealing with even less perfect. Heart Rhythm. 2007;4(7):897–9.CrossRef
4.
go back to reference Burri H, Sunthorn H, Dorsaz PA, Viera I, Shah D. Thresholds and complications with right ventricular septal pacing compared to apical pacing. Pacing Clin Electrophysiol. 2007;30(Suppl 1):S75-78.PubMed Burri H, Sunthorn H, Dorsaz PA, Viera I, Shah D. Thresholds and complications with right ventricular septal pacing compared to apical pacing. Pacing Clin Electrophysiol. 2007;30(Suppl 1):S75-78.PubMed
5.
go back to reference Cano O, Andres A, Alonso P, Osca J, Sancho-Tello MJ, Olague J, Martinez-Dolz L. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads. Europace. 2017;19(1):96–102.PubMed Cano O, Andres A, Alonso P, Osca J, Sancho-Tello MJ, Olague J, Martinez-Dolz L. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads. Europace. 2017;19(1):96–102.PubMed
6.
go back to reference Banaszewski M, Stepinska J. Right heart perforation by pacemaker leads. Arch Med Sci. 2012;8(1):11–3.CrossRef Banaszewski M, Stepinska J. Right heart perforation by pacemaker leads. Arch Med Sci. 2012;8(1):11–3.CrossRef
7.
go back to reference Khan MN, Joseph G, Khaykin Y, Ziada KM, Wilkoff BL. Delayed lead perforation: a disturbing trend. Pacing Clin Electrophysiol. 2005;28(3):251–3.CrossRef Khan MN, Joseph G, Khaykin Y, Ziada KM, Wilkoff BL. Delayed lead perforation: a disturbing trend. Pacing Clin Electrophysiol. 2005;28(3):251–3.CrossRef
8.
go back to reference Liang JJ, Killu AM, Osborn MJ, Jaffe AS. Pacemaker lead perforation causing hemopericardium eight years after implantation. Indian Heart J. 2013;65(3):331–3.CrossRef Liang JJ, Killu AM, Osborn MJ, Jaffe AS. Pacemaker lead perforation causing hemopericardium eight years after implantation. Indian Heart J. 2013;65(3):331–3.CrossRef
9.
go back to reference Nichols J, Berger N, Joseph P, Datta D. Subacute right ventricle perforation by pacemaker lead presenting with left hemothorax and shock. Case Rep Cardiol. 2015;2015:983930.PubMedPubMedCentral Nichols J, Berger N, Joseph P, Datta D. Subacute right ventricle perforation by pacemaker lead presenting with left hemothorax and shock. Case Rep Cardiol. 2015;2015:983930.PubMedPubMedCentral
10.
go back to reference Oikonomou A, Foutzitzi S, Prassopoulos P. Delayed cardiac perforation due to migrating pacemaker lead. CMAJ. 2010;182(9):E379.CrossRef Oikonomou A, Foutzitzi S, Prassopoulos P. Delayed cardiac perforation due to migrating pacemaker lead. CMAJ. 2010;182(9):E379.CrossRef
11.
go back to reference Refaat MM, Hashash JG, Shalaby AA. Late perforation by cardiac implantable electronic device leads: clinical presentation, diagnostic clues, and management. Clin Cardiol. 2010;33(8):466–75.CrossRef Refaat MM, Hashash JG, Shalaby AA. Late perforation by cardiac implantable electronic device leads: clinical presentation, diagnostic clues, and management. Clin Cardiol. 2010;33(8):466–75.CrossRef
12.
go back to reference Lin YS, Chen TH, Hung SP, Chen DY, Mao CT, Tsai ML, Chang ST, Wang CC, Wen MS, Chen MC. Impact of Pacemaker Lead Characteristics on Pacemaker Related Infection and Heart Perforation: A Nationwide Population-Based Cohort Study. PLoS ONE. 2015;10(6):e0128320.CrossRef Lin YS, Chen TH, Hung SP, Chen DY, Mao CT, Tsai ML, Chang ST, Wang CC, Wen MS, Chen MC. Impact of Pacemaker Lead Characteristics on Pacemaker Related Infection and Heart Perforation: A Nationwide Population-Based Cohort Study. PLoS ONE. 2015;10(6):e0128320.CrossRef
13.
go back to reference Acha MR, Keaney JJ, Lubitz SA, Milan DJ, Mansour M, Heist KE, Ptaszek LM, Singh JP, Blendea D, Mela T. Increased perforation risk with an MRI-conditional pacing lead: a single-center study. Pacing Clin Electrophysiol. 2015;38(3):334–42.CrossRef Acha MR, Keaney JJ, Lubitz SA, Milan DJ, Mansour M, Heist KE, Ptaszek LM, Singh JP, Blendea D, Mela T. Increased perforation risk with an MRI-conditional pacing lead: a single-center study. Pacing Clin Electrophysiol. 2015;38(3):334–42.CrossRef
14.
go back to reference Kwon CH, Choi JH, Kim J, Jo U, Lee JH, Lee WS, Kim YR, Lee SY, Whang KW, Yang J, et al. Complications of cardiac perforation and lead dislodgement with an MRI-conditional pacing lead: a Korean multi-center experience. J Korean Med Sci. 2016;31(9):1397–402.CrossRef Kwon CH, Choi JH, Kim J, Jo U, Lee JH, Lee WS, Kim YR, Lee SY, Whang KW, Yang J, et al. Complications of cardiac perforation and lead dislodgement with an MRI-conditional pacing lead: a Korean multi-center experience. J Korean Med Sci. 2016;31(9):1397–402.CrossRef
15.
go back to reference Elmouchi DA, Rosema S, Vanoosterhout SM, Khan M, Davis AT, Gauri AJ, Finta B, Woelfel AK, Chalfoun NT. Cardiac perforation and lead dislodgement after implantation of a MR-conditional pacing lead: a single-center experience. Pacing Clin Electrophysiol. 2014;37(1):4–10.CrossRef Elmouchi DA, Rosema S, Vanoosterhout SM, Khan M, Davis AT, Gauri AJ, Finta B, Woelfel AK, Chalfoun NT. Cardiac perforation and lead dislodgement after implantation of a MR-conditional pacing lead: a single-center experience. Pacing Clin Electrophysiol. 2014;37(1):4–10.CrossRef
16.
go back to reference Pang BJ, Lui EH, Joshi SB, Tacey MA, Alison J, Seneviratne SK, Cameron JD, Mond HG. Pacing and implantable cardioverter defibrillator lead perforation as assessed by multiplanar reformatted ECG-gated cardiac computed tomography and clinical correlates. Pacing Clin Electrophysiol. 2014;37(5):537–45.CrossRef Pang BJ, Lui EH, Joshi SB, Tacey MA, Alison J, Seneviratne SK, Cameron JD, Mond HG. Pacing and implantable cardioverter defibrillator lead perforation as assessed by multiplanar reformatted ECG-gated cardiac computed tomography and clinical correlates. Pacing Clin Electrophysiol. 2014;37(5):537–45.CrossRef
17.
go back to reference Rosso R, Teh AW, Medi C, Hung TT, Balasubramaniam R, Mond HG. Right ventricular septal pacing: the success of stylet-driven active-fixation leads. Pacing Clin Electrophysiol. 2010;33(1):49–53.CrossRef Rosso R, Teh AW, Medi C, Hung TT, Balasubramaniam R, Mond HG. Right ventricular septal pacing: the success of stylet-driven active-fixation leads. Pacing Clin Electrophysiol. 2010;33(1):49–53.CrossRef
18.
go back to reference Moore P, Coucher J, Ngai S, Stanton T, Wahi S, Gould P, Booth C, Pratap J, Kaye G. Imaging and right ventricular pacing lead position: a comparison of CT, MRI, and echocardiography. Pacing Clin Electrophysiol. 2016;39(4):382–92.CrossRef Moore P, Coucher J, Ngai S, Stanton T, Wahi S, Gould P, Booth C, Pratap J, Kaye G. Imaging and right ventricular pacing lead position: a comparison of CT, MRI, and echocardiography. Pacing Clin Electrophysiol. 2016;39(4):382–92.CrossRef
19.
go back to reference Aliyev F, Çeliker C, Türkoğlu C, Karadağ B, Yıldız A. Catheter related perforation of right heart chambers with diagnostic electrophysiology catheters and temporary transvenous pacing leads: when to pull it out and when to keep it in a place? Turk Kardiyol Dern Ars. 2011;39(1):16–22.PubMed Aliyev F, Çeliker C, Türkoğlu C, Karadağ B, Yıldız A. Catheter related perforation of right heart chambers with diagnostic electrophysiology catheters and temporary transvenous pacing leads: when to pull it out and when to keep it in a place? Turk Kardiyol Dern Ars. 2011;39(1):16–22.PubMed
20.
go back to reference Rosso R, Medi C, Teh AW, Hung TT, Feldman A, Lee G, Mond HG. Right ventricular septal pacing: a comparative study of outflow tract and mid ventricular sites. Pacing Clin Electrophysiol. 2010;33(10):1169–73.CrossRef Rosso R, Medi C, Teh AW, Hung TT, Feldman A, Lee G, Mond HG. Right ventricular septal pacing: a comparative study of outflow tract and mid ventricular sites. Pacing Clin Electrophysiol. 2010;33(10):1169–73.CrossRef
21.
go back to reference Hillock RJ, Stevenson IH, Mond HG. The right ventricular outflow tract: a comparative study of septal, anterior wall, and free wall pacing. Pacing Clin Electrophysiol. 2007;30(8):942–7.CrossRef Hillock RJ, Stevenson IH, Mond HG. The right ventricular outflow tract: a comparative study of septal, anterior wall, and free wall pacing. Pacing Clin Electrophysiol. 2007;30(8):942–7.CrossRef
22.
go back to reference Witt CM, Lenz CJ, Shih HH, Ebrille E, Rosenbaum AN, van Zyl M, Aung H, Manocha KK, Deshmukh AJ, Hodge DO, et al. Right ventricular pacemaker lead position is associated with differences in long-term outcomes and complications. J Cardiovasc Electrophysiol. 2017;28(8):924–30.CrossRef Witt CM, Lenz CJ, Shih HH, Ebrille E, Rosenbaum AN, van Zyl M, Aung H, Manocha KK, Deshmukh AJ, Hodge DO, et al. Right ventricular pacemaker lead position is associated with differences in long-term outcomes and complications. J Cardiovasc Electrophysiol. 2017;28(8):924–30.CrossRef
23.
go back to reference Zou C, Song J, Li H, Huang X, Liu Y, Zhao C, Shi X, Yang X: Right ventricular outflow tract septal pacing is superior to right ventricular apical pacing. J Am Heart Assoc 2015, 4(4). Zou C, Song J, Li H, Huang X, Liu Y, Zhao C, Shi X, Yang X: Right ventricular outflow tract septal pacing is superior to right ventricular apical pacing. J Am Heart Assoc 2015, 4(4).
Metadata
Title
Subclinical cardiac perforation by cardiac implantable electronic device leads detected by cardiac computed tomography
Authors
Yeong-Min Lim
Jae-Sun Uhm
Min Kim
In-Soo Kim
Moo-Nyun Jin
Hee Tae Yu
Tae-Hoon Kim
Hye-Jeong Lee
Young-Jin Kim
Boyoung Joung
Hui-Nam Pak
Moon-Hyoung Lee
Publication date
01-12-2021

Other articles of this Issue 1/2021

BMC Cardiovascular Disorders 1/2021 Go to the issue