Wider translations and rotations in posterior-stabilised mobile-bearing total knee arthroplasty compared to fixed-bearing both implanted with mechanical alignment: a dynamic RSA study

Authors: Raffaele Zinno, Domenico Alesi, Stefano Di Paolo, Nicola Pizza, Stefano Zaffagnini, Giulio Maria Marcheggiani Muccioli, Laura Bragonzoni

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2023

Login to get access

Abstract

Purpose

The purpose of this study was to investigate the in vivo kinematics of the same femoral design mechanically aligned posterior-stabilised (PS) total knee arthroplasty (TKA) with either fixed-bearing (FB) or mobile-bearing (MB) inlay, implanted by the same surgeon, using model-based dynamic radiostereometric analysis (RSA). The hypothesis of the present study was that the MB design would show wider axial rotation than the FB design, without affecting the clinical outcomes.

Materials and methods

A cohort of 21 non-randomised patients (21 DePuy Attune PS-FB) was evaluated by dynamic RSA analysis at a minimum 9-month follow-up, while performing differently demanding daily living activities such as sit to stand (STS) and deep knee lunge (DKL). Kinematic data were compared with those of a cohort of 22 patients implanted with the same prosthetic design but with MB inlay. Anterior–posterior (AP) translations, varus–valgus (VV) and internal–external (IE) rotations of the femoral component with respect to the tibial baseplate were investigated. Translation of medial and lateral compartment was analysed using the low point method according to Freeman et al. Questionnaires to calculate objective and subjective clinical scores were administered preoperatively and during follow-up visit by the same investigator.

Results

The FB TKA design showed lower AP translation during STS (6.8 ± 3.3 mm in FB vs 9.9 ± 3.7 mm in MB, p = 0.006*), lower VV rotation (1.9 ± 0.8° in FB vs 5.3 ± 3.3° in MB, p = 0.005) and lower IE rotation (2.8 ± 1.1° in FB vs 9.5 ± 4.3° in MB, p = 0.001) during DKL than the mobile-bearing TKA design. Posterior-stabilised FB group showed significant lower translation of the low point of the medial compartment than the MB group (p = 0.008). The percentage of patients performing medial pivot in the FB group was higher compared to MB group in the examined motor tasks. No significant differences in post-operative range of motion (117° ± 16° for FB group and 124° ± 13° for MB group) and in clinical outcomes emerged between the two cohort.

Conclusions

The FB and MB designs differed in AP translations, VV rotations and IE rotations of the femoral component with respect to the tibial component in STS and DKL. Furthermore, FB cohort reported a significant higher percentage of medial pivot with respect to MB cohort. Despite this, no differences in clinical outcomes were detected between groups. Both designs showed stable kinematics and represent a viable option in primary TKA.

Level of evidence

Prospective cohort study, II.
Literature
1.
go back to reference Alesi D, MarcheggianiMuccioli GM, Roberti di Sarsina T, Bontempi M, Pizza N, Zinno R, Di Paolo S, Zaffagnini S, Bragonzoni L (2020) In vivo femorotibial kinematics of medial-stabilized total knee arthroplasty correlates to post-operative clinical outcomes. Knee Surg Sports Traumatol Arthrosc 29(2):491–497CrossRefPubMed Alesi D, MarcheggianiMuccioli GM, Roberti di Sarsina T, Bontempi M, Pizza N, Zinno R, Di Paolo S, Zaffagnini S, Bragonzoni L (2020) In vivo femorotibial kinematics of medial-stabilized total knee arthroplasty correlates to post-operative clinical outcomes. Knee Surg Sports Traumatol Arthrosc 29(2):491–497CrossRefPubMed
2.
go back to reference Arauz P, Klemt C, Limmahakhun S, An S, Kwon Y-M (2019) Stair climbing and high knee flexion activities in bi-cruciate retaining total knee arthroplasty: in vivo kinematics and articular contact analysis. J Arthroplasty 34(3):570–576CrossRefPubMed Arauz P, Klemt C, Limmahakhun S, An S, Kwon Y-M (2019) Stair climbing and high knee flexion activities in bi-cruciate retaining total knee arthroplasty: in vivo kinematics and articular contact analysis. J Arthroplasty 34(3):570–576CrossRefPubMed
3.
go back to reference Banks SA, Hodge WA (2004) 2003 Hap Paul award paper of the international society for technology in arthroplasty. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplasty 19(7):809–816CrossRefPubMed Banks SA, Hodge WA (2004) 2003 Hap Paul award paper of the international society for technology in arthroplasty. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplasty 19(7):809–816CrossRefPubMed
4.
go back to reference Batra S, Sugumar PAA, Kumar V, Malhotra R (2021) Which one restores in vivo knee kinematics effectively-medial or lateral pivot? J Clin Orthop Trauma 13:70–73CrossRefPubMed Batra S, Sugumar PAA, Kumar V, Malhotra R (2021) Which one restores in vivo knee kinematics effectively-medial or lateral pivot? J Clin Orthop Trauma 13:70–73CrossRefPubMed
5.
go back to reference Bontempi M, Cardinale U, Bragonzoni L, Muccioli GMM, Alesi D, di Matteo B, Marcacci M, Zaffagnini S (2020) A computer simulation protocol to assess the accuracy of a radio stereometric analysis (RSA) image processor according to the ISO-5725. ArXiv200603913 Phys Bontempi M, Cardinale U, Bragonzoni L, Muccioli GMM, Alesi D, di Matteo B, Marcacci M, Zaffagnini S (2020) A computer simulation protocol to assess the accuracy of a radio stereometric analysis (RSA) image processor according to the ISO-5725. ArXiv200603913 Phys
6.
go back to reference Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop 468(1):57–63CrossRefPubMed Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop 468(1):57–63CrossRefPubMed
7.
go back to reference Callaghan JJ, Squire MW, Goetz DD, Sullivan PM, Johnston RC (2000) Cemented rotating-platform total knee replacement. A nine to twelve-year follow-up study. J Bone Joint Surg Am 82(5):705–711CrossRefPubMed Callaghan JJ, Squire MW, Goetz DD, Sullivan PM, Johnston RC (2000) Cemented rotating-platform total knee replacement. A nine to twelve-year follow-up study. J Bone Joint Surg Am 82(5):705–711CrossRefPubMed
8.
go back to reference Cardinale U, Bragonzoni L, Bontempi M, Alesi D, Roberti di Sarsina T, Lo Presti M, Zaffagnini S, MarcheggianiMuccioli GM, Iacono F (2019) Knee kinematics after cruciate retaining highly congruent mobile bearing total knee arthroplasty: an in vivo dynamic RSA study. Knee 27(2):341–347CrossRefPubMed Cardinale U, Bragonzoni L, Bontempi M, Alesi D, Roberti di Sarsina T, Lo Presti M, Zaffagnini S, MarcheggianiMuccioli GM, Iacono F (2019) Knee kinematics after cruciate retaining highly congruent mobile bearing total knee arthroplasty: an in vivo dynamic RSA study. Knee 27(2):341–347CrossRefPubMed
10.
go back to reference Delport HP, Banks SA, De Schepper J, Bellemans J (2006) A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br 88(8):1016–1021CrossRefPubMed Delport HP, Banks SA, De Schepper J, Bellemans J (2006) A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br 88(8):1016–1021CrossRefPubMed
11.
go back to reference Dennis DA, Komistek RD (2006) Mobile-bearing total knee arthroplasty: design factors in minimizing wear. Clin Orthop 452:70–77CrossRefPubMed Dennis DA, Komistek RD (2006) Mobile-bearing total knee arthroplasty: design factors in minimizing wear. Clin Orthop 452:70–77CrossRefPubMed
12.
go back to reference Dennis DA, Komistek RD, Colwell CE, Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop 356:47–57CrossRef Dennis DA, Komistek RD, Colwell CE, Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop 356:47–57CrossRef
13.
go back to reference Dennis DA, Komistek RD, Mahfouz MR, Outten JT, Sharma A (2005) Mobile-bearing total knee arthroplasty: do the polyethylene bearings rotate? Clin Orthop 440:88–95CrossRefPubMed Dennis DA, Komistek RD, Mahfouz MR, Outten JT, Sharma A (2005) Mobile-bearing total knee arthroplasty: do the polyethylene bearings rotate? Clin Orthop 440:88–95CrossRefPubMed
14.
go back to reference Dennis DA, Komistek RD, Mahfouz MR, Walker SA, Tucker A (2004) A multicenter analysis of axial femorotibial rotation after total knee arthroplasty. Clin Orthop 428:180–189CrossRef Dennis DA, Komistek RD, Mahfouz MR, Walker SA, Tucker A (2004) A multicenter analysis of axial femorotibial rotation after total knee arthroplasty. Clin Orthop 428:180–189CrossRef
15.
go back to reference Freeman MR, Pinskerova V (2005) The movement of the normal tibio-femoral joint. J Biomech 38(2):197–208CrossRefPubMed Freeman MR, Pinskerova V (2005) The movement of the normal tibio-femoral joint. J Biomech 38(2):197–208CrossRefPubMed
16.
go back to reference Ghirardelli S, Asay JL, Leonardi EA, Amoroso T, Andriacchi TP, Indelli PF (2021) Kinematic comparison between medially congruent and posterior-stabilized third-generation TKA designs. J Funct Morphol Kinesiol 6(1):27CrossRefPubMedPubMedCentral Ghirardelli S, Asay JL, Leonardi EA, Amoroso T, Andriacchi TP, Indelli PF (2021) Kinematic comparison between medially congruent and posterior-stabilized third-generation TKA designs. J Funct Morphol Kinesiol 6(1):27CrossRefPubMedPubMedCentral
17.
go back to reference Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144CrossRefPubMed Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105(2):136–144CrossRefPubMed
18.
go back to reference Hamilton DF, Burnett R, Patton JT, Howie CR, Moran M, Simpson AHRW, Gaston P (2015) Implant design influences patient outcome after total knee arthroplasty: a prospective double-blind randomised controlled trial. Bone Jt J 97(1):64–70CrossRef Hamilton DF, Burnett R, Patton JT, Howie CR, Moran M, Simpson AHRW, Gaston P (2015) Implant design influences patient outcome after total knee arthroplasty: a prospective double-blind randomised controlled trial. Bone Jt J 97(1):64–70CrossRef
19.
go back to reference Hantouly AT, Ahmed AF, Alzobi O, Toubasi A, Salameh M, Elmhiregh A, Hameed S, Ahmed GO, Alvand A, Al Dosari MAA (2022) Mobile-bearing versus fixed-bearing total knee arthroplasty: a meta-analysis of randomized controlled trials. Eur J Orthop Surg Traumatol 32(3):481–495CrossRefPubMed Hantouly AT, Ahmed AF, Alzobi O, Toubasi A, Salameh M, Elmhiregh A, Hameed S, Ahmed GO, Alvand A, Al Dosari MAA (2022) Mobile-bearing versus fixed-bearing total knee arthroplasty: a meta-analysis of randomized controlled trials. Eur J Orthop Surg Traumatol 32(3):481–495CrossRefPubMed
20.
go back to reference Heiberg KE, Bruun-Olsen V, Mengshoel AM (2010) Pain and recovery of physical functioning nine months after total knee arthroplasty. J Rehabil Med 42(7):614–619CrossRefPubMed Heiberg KE, Bruun-Olsen V, Mengshoel AM (2010) Pain and recovery of physical functioning nine months after total knee arthroplasty. J Rehabil Med 42(7):614–619CrossRefPubMed
21.
go back to reference Jones VC, Barton DC, Fitzpatrick DP, Auger DD, Stone MH, Fisher J (1999) An experimental model of tibial counterface polyethylene wear in mobile bearing knees: the influence of design and kinematics. Biomed Mater Eng 9(3):189–196PubMed Jones VC, Barton DC, Fitzpatrick DP, Auger DD, Stone MH, Fisher J (1999) An experimental model of tibial counterface polyethylene wear in mobile bearing knees: the influence of design and kinematics. Biomed Mater Eng 9(3):189–196PubMed
22.
go back to reference Kage T, Inui H, Tomita T, Yamazaki T, Taketomi S, Yamagami R, Kono K, Kawaguchi K, Takagi K, Sameshima S, Tanaka S (2022) The association between in vivo knee kinematics and patient-reported outcomes during squatting in bicruciate-stabilized total knee arthroplasty. J Knee Surg 35(12):1342–1348CrossRefPubMed Kage T, Inui H, Tomita T, Yamazaki T, Taketomi S, Yamagami R, Kono K, Kawaguchi K, Takagi K, Sameshima S, Tanaka S (2022) The association between in vivo knee kinematics and patient-reported outcomes during squatting in bicruciate-stabilized total knee arthroplasty. J Knee Surg 35(12):1342–1348CrossRefPubMed
23.
go back to reference Komistek RD, Dennis DA, Mahfouz MR, Walker S, Outten J (2004) In vivo polyethylene bearing mobility is maintained in posterior stabilized total knee arthroplasty. Clin Orthop 428:207–213CrossRef Komistek RD, Dennis DA, Mahfouz MR, Walker S, Outten J (2004) In vivo polyethylene bearing mobility is maintained in posterior stabilized total knee arthroplasty. Clin Orthop 428:207–213CrossRef
24.
go back to reference Li G, Suggs J, Hanson G, Durbhakula S, Johnson T, Freiberg A (2006) Three-dimensional tibiofemoral articular contact kinematics of a cruciate-retaining total knee arthroplasty. J Bone Joint Surg Am 88(2):395–402CrossRefPubMed Li G, Suggs J, Hanson G, Durbhakula S, Johnson T, Freiberg A (2006) Three-dimensional tibiofemoral articular contact kinematics of a cruciate-retaining total knee arthroplasty. J Bone Joint Surg Am 88(2):395–402CrossRefPubMed
25.
go back to reference Lu Y-C, Huang C-H, Chang T-K, Ho F-Y, Cheng C-K, Huang C-H (2010) Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Br 92(4):500–507CrossRefPubMed Lu Y-C, Huang C-H, Chang T-K, Ho F-Y, Cheng C-K, Huang C-H (2010) Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Br 92(4):500–507CrossRefPubMed
26.
go back to reference MarcheggianiMuccioli GM, Pizza N, Di Paolo S, Zinno R, Alesi D, Roberti Di Sarsina T, Bontempi M, Zaffagnini S, Bragonzoni L (2020) Multi-radius posterior-stabilized mobile-bearing total knee arthroplasty partially produces in-vivo medial pivot during activity of daily living and high demanding motor task. Knee Surg Sports Traumatol Arthrosc 28(12):3773–3779CrossRef MarcheggianiMuccioli GM, Pizza N, Di Paolo S, Zinno R, Alesi D, Roberti Di Sarsina T, Bontempi M, Zaffagnini S, Bragonzoni L (2020) Multi-radius posterior-stabilized mobile-bearing total knee arthroplasty partially produces in-vivo medial pivot during activity of daily living and high demanding motor task. Knee Surg Sports Traumatol Arthrosc 28(12):3773–3779CrossRef
27.
go back to reference Mugnai R, Digennaro V, Ensini A, Leardini A, Catani F (2014) Can TKA design affect the clinical outcome? Comparison between two guided-motion systems. Knee Surg Sports Traumatol Arthrosc 22(3):581–589CrossRefPubMed Mugnai R, Digennaro V, Ensini A, Leardini A, Catani F (2014) Can TKA design affect the clinical outcome? Comparison between two guided-motion systems. Knee Surg Sports Traumatol Arthrosc 22(3):581–589CrossRefPubMed
28.
go back to reference Nakamura E, Banks SA, Tanaka A, Sei A, Mizuta H (2009) Three-dimensional tibiofemoral kinematics during deep flexion kneeling in a mobile-bearing total knee arthroplasty. J Arthroplasty 24(7):1120–1124CrossRefPubMed Nakamura E, Banks SA, Tanaka A, Sei A, Mizuta H (2009) Three-dimensional tibiofemoral kinematics during deep flexion kneeling in a mobile-bearing total knee arthroplasty. J Arthroplasty 24(7):1120–1124CrossRefPubMed
29.
go back to reference Pataky TC, Vanrenterghem J, Robinson MA (2015) Zero- vs one-dimensional, parametric vs non-parametric, and confidence interval vs hypothesis testing procedures in one-dimensional biomechanical trajectory analysis. J Biomech 48(7):1277–1285CrossRefPubMed Pataky TC, Vanrenterghem J, Robinson MA (2015) Zero- vs one-dimensional, parametric vs non-parametric, and confidence interval vs hypothesis testing procedures in one-dimensional biomechanical trajectory analysis. J Biomech 48(7):1277–1285CrossRefPubMed
30.
go back to reference Pizza N, Di Paolo S, Zinno R, MarcheggianiMuccioli GM, Agostinone P, Alesi D, Bontempi M, Zaffagnini S, Bragonzoni L (2021) Over-constrained kinematic of the medial compartment leads to lower clinical outcomes after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 30(2):661–667CrossRefPubMedPubMedCentral Pizza N, Di Paolo S, Zinno R, MarcheggianiMuccioli GM, Agostinone P, Alesi D, Bontempi M, Zaffagnini S, Bragonzoni L (2021) Over-constrained kinematic of the medial compartment leads to lower clinical outcomes after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 30(2):661–667CrossRefPubMedPubMedCentral
31.
go back to reference Poirier N, Graf P, Dubrana F (2015) Mobile-bearing versus fixed-bearing total knee implants. Results of a series of 100 randomised cases after 9 years follow-up. Orthop Traumatol Surg Res 101(4):S187-192CrossRefPubMed Poirier N, Graf P, Dubrana F (2015) Mobile-bearing versus fixed-bearing total knee implants. Results of a series of 100 randomised cases after 9 years follow-up. Orthop Traumatol Surg Res 101(4):S187-192CrossRefPubMed
32.
go back to reference Ranawat CS, Komistek RD, Rodriguez JA, Dennis DA, Anderle M (2004) In vivo kinematics for fixed and mobile-bearing posterior stabilized knee prostheses. Clin Orthop 2:184–190CrossRef Ranawat CS, Komistek RD, Rodriguez JA, Dennis DA, Anderle M (2004) In vivo kinematics for fixed and mobile-bearing posterior stabilized knee prostheses. Clin Orthop 2:184–190CrossRef
33.
go back to reference Rivière C, Iranpour F, Auvinet E, Howell S, Vendittoli P-A, Cobb J, Parratte S (2017) Alignment options for total knee arthroplasty: a systematic review. Orthop Traumatol Surg Res 103(7):1047–1056CrossRefPubMed Rivière C, Iranpour F, Auvinet E, Howell S, Vendittoli P-A, Cobb J, Parratte S (2017) Alignment options for total knee arthroplasty: a systematic review. Orthop Traumatol Surg Res 103(7):1047–1056CrossRefPubMed
34.
go back to reference Smith LA, Nachtrab J, LaCour M, Cates H, Freeman MG, Komistek RD (2021) In vivo knee kinematics: how important are the roles of femoral geometry and the cruciate ligaments? J Arthroplasty 36(4):1445–1454CrossRefPubMed Smith LA, Nachtrab J, LaCour M, Cates H, Freeman MG, Komistek RD (2021) In vivo knee kinematics: how important are the roles of femoral geometry and the cruciate ligaments? J Arthroplasty 36(4):1445–1454CrossRefPubMed
35.
go back to reference Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty: comparison of fixed and mobile-bearing inlay designs. Acta Orthop Scand 73(6):638–646PubMed Stukenborg-Colsman C, Ostermeier S, Hurschler C, Wirth CJ (2002) Tibiofemoral contact stress after total knee arthroplasty: comparison of fixed and mobile-bearing inlay designs. Acta Orthop Scand 73(6):638–646PubMed
36.
go back to reference Suzuki K, Hara N, Mikami S, Tomita T, Iwamoto K, Yamazaki T, Sugamoto K, Matsuno S (2014) In vivo kinematic analysis of posterior-stabilized total knee arthroplasty for the valgus knee operated by the gap-balancing technique. Knee 21(6):1124–1128CrossRefPubMed Suzuki K, Hara N, Mikami S, Tomita T, Iwamoto K, Yamazaki T, Sugamoto K, Matsuno S (2014) In vivo kinematic analysis of posterior-stabilized total knee arthroplasty for the valgus knee operated by the gap-balancing technique. Knee 21(6):1124–1128CrossRefPubMed
37.
go back to reference Wasielewski RC, Komistek RD, Zingde SM, Sheridan KC, Mahfouz MR (2008) Lack of axial rotation in mobile-bearing knee designs. Clin Orthop 466(11):2662–2668CrossRefPubMedPubMedCentral Wasielewski RC, Komistek RD, Zingde SM, Sheridan KC, Mahfouz MR (2008) Lack of axial rotation in mobile-bearing knee designs. Clin Orthop 466(11):2662–2668CrossRefPubMedPubMedCentral
38.
go back to reference Watanabe T, Ishizuki M, Muneta T, Banks SA (2012) Matched comparison of kinematics in knees with mild and severe varus deformity using fixed- and mobile-bearing total knee arthroplasty. Clin Biomech Bristol Avon 27(9):924–928CrossRefPubMed Watanabe T, Ishizuki M, Muneta T, Banks SA (2012) Matched comparison of kinematics in knees with mild and severe varus deformity using fixed- and mobile-bearing total knee arthroplasty. Clin Biomech Bristol Avon 27(9):924–928CrossRefPubMed
39.
go back to reference White PB, Ranawat AS, Ranawat CS (2015) Fixed bearings versus rotating platforms in total knee arthroplasty. J Knee Surg 28(5):358–362CrossRefPubMed White PB, Ranawat AS, Ranawat CS (2015) Fixed bearings versus rotating platforms in total knee arthroplasty. J Knee Surg 28(5):358–362CrossRefPubMed
40.
go back to reference Zihlmann MS, Stacoff A, Romero J, Quervain IK, Stüssi E (2005) Biomechanical background and clinical observations of rotational malalignment in TKA: literature review and consequences. Clin Biomech Bristol Avon 20(7):661–668CrossRefPubMed Zihlmann MS, Stacoff A, Romero J, Quervain IK, Stüssi E (2005) Biomechanical background and clinical observations of rotational malalignment in TKA: literature review and consequences. Clin Biomech Bristol Avon 20(7):661–668CrossRefPubMed
41.
go back to reference Zingde SM, Leszko F, Sharma A, Mahfouz MR, Komistek RD, Dennis DA (2014) In vivo determination of cam-post engagement in fixed and mobile-bearing TKA. Clin Orthop 472(1):254–262CrossRefPubMed Zingde SM, Leszko F, Sharma A, Mahfouz MR, Komistek RD, Dennis DA (2014) In vivo determination of cam-post engagement in fixed and mobile-bearing TKA. Clin Orthop 472(1):254–262CrossRefPubMed
Metadata
Title
Wider translations and rotations in posterior-stabilised mobile-bearing total knee arthroplasty compared to fixed-bearing both implanted with mechanical alignment: a dynamic RSA study
Authors
Raffaele Zinno
Domenico Alesi
Stefano Di Paolo
Nicola Pizza
Stefano Zaffagnini
Giulio Maria Marcheggiani Muccioli
Laura Bragonzoni
Publication date
24-08-2023
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2023
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-023-07541-6

Other articles of this Issue 11/2023

Knee Surgery, Sports Traumatology, Arthroscopy 11/2023 Go to the issue