Published in: La radiologia medica 1/2016

01-01-2016 | MUSCULOSKELETAL RADIOLOGY

Usefulness of IDEAL T2 imaging for homogeneous fat suppression and reducing susceptibility artefacts in brachial plexus MRI at 3.0 T

Authors: Alberto Tagliafico, Bianca Bignotti, Giulio Tagliafico, Carlo Martinoli

Published in: La radiologia medica | Issue 1/2016

Login to get access

Abstract

Purpose

To quantitatively and qualitatively compare fat-suppressed MR imaging quality using iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) with that using frequency-selective fat-suppressed (FSFS) T2 images of the brachial plexus at 3.0 T.

Materials and methods

Prospective MR image analysis was performed in 40 volunteers and 40 patients at a single centre. Oblique-sagittal and coronal IDEAL fat-suppressed T2 images and FSFS T2 images were compared. Visual assessment was performed by two independent musculoskeletal radiologists with respect to: (1) susceptibility artefacts around the neck, (2) homogeneity of fat suppression, (3) image sharpness and (4) tissue resolution contrast of pathologies. The signal-to-noise ratios (SNR) for each image sequence were assessed.

Results

Compared to FSFS sequences, IDEAL fat-suppressed T2 images significantly reduced artefacts around the brachial plexus and significantly improved homogeneous fat suppression (p < 0.05). IDEAL significantly improved sharpness and lesion-to-tissue contrast (p < 0.05). The mean SNRs were significantly improved on T2-weighted IDEAL images (p < 0.05).

Conclusion

IDEAL technique improved image quality by reducing artefacts around the brachial plexus while maintaining a high SNR and provided superior homogeneous fat suppression than FSFS sequences.
Literature
1.
go back to reference Tagliafico A, Succio G, Emanuele Neumaier C et al (2011) MR imaging of the brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary experience. Skeletal Radiol 40:717–724PubMedCrossRef Tagliafico A, Succio G, Emanuele Neumaier C et al (2011) MR imaging of the brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary experience. Skeletal Radiol 40:717–724PubMedCrossRef
2.
go back to reference Tagliafico A, Succio G, Serafini G, Martinoli C (2012) Diagnostic accuracy of MRI in adults with suspect brachial plexus lesions: a multicentre retrospective study with surgical findings and clinical follow-up as reference standard. Eur J Radiol 81:2666–2672PubMedCrossRef Tagliafico A, Succio G, Serafini G, Martinoli C (2012) Diagnostic accuracy of MRI in adults with suspect brachial plexus lesions: a multicentre retrospective study with surgical findings and clinical follow-up as reference standard. Eur J Radiol 81:2666–2672PubMedCrossRef
3.
go back to reference Aagaard BD, Maravilla KR, Kliot M (1998) MR neurography MR imaging of peripheral nerves. Magn Reson Imaging Clin N Am 6:179–194PubMed Aagaard BD, Maravilla KR, Kliot M (1998) MR neurography MR imaging of peripheral nerves. Magn Reson Imaging Clin N Am 6:179–194PubMed
4.
go back to reference Kim S, Choi JY, Huh YM et al (2007) Role of magnetic resonance imaging in entrapment and compressive neuropathy-what, where, and how to see the peripheral nerves on the musculoskeletal magnetic resonance image: part 2. Upper extremity. Eur Radiol 17:509–522PubMedCrossRef Kim S, Choi JY, Huh YM et al (2007) Role of magnetic resonance imaging in entrapment and compressive neuropathy-what, where, and how to see the peripheral nerves on the musculoskeletal magnetic resonance image: part 2. Upper extremity. Eur Radiol 17:509–522PubMedCrossRef
5.
go back to reference Todd M, Shah GV, Mukherji SK (2004) MR imaging of brachial plexus. Top Magn Reson Imaging 15:113–125PubMedCrossRef Todd M, Shah GV, Mukherji SK (2004) MR imaging of brachial plexus. Top Magn Reson Imaging 15:113–125PubMedCrossRef
7.
go back to reference Viallon M, Vargas MI, Jlassi H, Lövblad KO, Delavelle J (2008) High-resolution and functional magnetic resonance imaging of the brachial plexus using an isotropic 3D T2 STIR (short term inversion recovery) SPACE sequence and diffusion tensor imaging. Eur Radiol 18:1018–1023PubMedCrossRef Viallon M, Vargas MI, Jlassi H, Lövblad KO, Delavelle J (2008) High-resolution and functional magnetic resonance imaging of the brachial plexus using an isotropic 3D T2 STIR (short term inversion recovery) SPACE sequence and diffusion tensor imaging. Eur Radiol 18:1018–1023PubMedCrossRef
8.
go back to reference Zhou L, Yousem DM, Chaudhry V (2004) Role of magnetic resonance neurography in brachial plexus lesions. Muscle Nerve 30:305–309PubMedCrossRef Zhou L, Yousem DM, Chaudhry V (2004) Role of magnetic resonance neurography in brachial plexus lesions. Muscle Nerve 30:305–309PubMedCrossRef
9.
go back to reference Ross MR, Schomer DF, Chappell P, Enzmann DR (1994) MR imaging of head and neck tumors: comparison of T1-weighted contrast-enhanced fat-suppressed images with conventional T2-weighted and fast spin-echo T2-weighted images. AJR Am J Radiol 162:173–178 Ross MR, Schomer DF, Chappell P, Enzmann DR (1994) MR imaging of head and neck tumors: comparison of T1-weighted contrast-enhanced fat-suppressed images with conventional T2-weighted and fast spin-echo T2-weighted images. AJR Am J Radiol 162:173–178
10.
go back to reference Finkenzeller T, Zorger N, Kühnel T et al (2013) Novel application of T1-weighted BLADE sequences with fat suppression compared to TSE in contrast-enhanced T1-weighted imaging of the neck: cutting-edge images? J Magn Reson Imaging 37:660–668PubMedCrossRef Finkenzeller T, Zorger N, Kühnel T et al (2013) Novel application of T1-weighted BLADE sequences with fat suppression compared to TSE in contrast-enhanced T1-weighted imaging of the neck: cutting-edge images? J Magn Reson Imaging 37:660–668PubMedCrossRef
11.
go back to reference Gerdes CM, Kijowski R, Reeder SB (2007) IDEAL imaging of the musculoskeletal system: robust water fat separation for homogeneous fat suppression, marrow evaluation, and cartilage imaging. AJR Am J Roentgenol 189:284–291CrossRef Gerdes CM, Kijowski R, Reeder SB (2007) IDEAL imaging of the musculoskeletal system: robust water fat separation for homogeneous fat suppression, marrow evaluation, and cartilage imaging. AJR Am J Roentgenol 189:284–291CrossRef
12.
go back to reference Chen CA, Lu W, John CT et al (2009) Multiecho IDEAL gradient-echo water-fat separation for rapid assessment of cartilage volume at 1.5T: initial experience. Radiology 252:561–567PubMedPubMedCentralCrossRef Chen CA, Lu W, John CT et al (2009) Multiecho IDEAL gradient-echo water-fat separation for rapid assessment of cartilage volume at 1.5T: initial experience. Radiology 252:561–567PubMedPubMedCentralCrossRef
13.
go back to reference Kijowski R, Woods MA, Lee KS et al (2009) Improved fat suppression using multipeak reconstruction for IDEAL chemical shift fat-water separation: application with fast spin echo imaging. J Magn Reson Imaging 29:436–442PubMedPubMedCentralCrossRef Kijowski R, Woods MA, Lee KS et al (2009) Improved fat suppression using multipeak reconstruction for IDEAL chemical shift fat-water separation: application with fast spin echo imaging. J Magn Reson Imaging 29:436–442PubMedPubMedCentralCrossRef
14.
go back to reference Grayev A, Shimakawa A, Cousins J, Turski P, Brittain J, Reeder S (2009) Improved time-of-flight magnetic resonance angiography with IDEAL water-fat separation. J Magn Reson Imaging 29:1367–1374PubMedCrossRef Grayev A, Shimakawa A, Cousins J, Turski P, Brittain J, Reeder S (2009) Improved time-of-flight magnetic resonance angiography with IDEAL water-fat separation. J Magn Reson Imaging 29:1367–1374PubMedCrossRef
15.
go back to reference Aoki T, Yamashita Y, Oki H et al (2012) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) of the wrist and finger at 3T: comparison with chemical shift selective fat suppression images. J Magn Reson Imaging 37:733–738PubMedCrossRef Aoki T, Yamashita Y, Oki H et al (2012) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) of the wrist and finger at 3T: comparison with chemical shift selective fat suppression images. J Magn Reson Imaging 37:733–738PubMedCrossRef
16.
go back to reference Reeder SB, Pineda AR, Wen Z et al (2005) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med 54:636–644PubMedCrossRef Reeder SB, Pineda AR, Wen Z et al (2005) Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): application with fast spin-echo imaging. Magn Reson Med 54:636–644PubMedCrossRef
17.
go back to reference Tagliafico A, Bignotti B, Tagliafico G et al (2014) Breast density assessment using a 3T MRI system: comparison among different sequences. PLoS One 9:e99027PubMedPubMedCentralCrossRef Tagliafico A, Bignotti B, Tagliafico G et al (2014) Breast density assessment using a 3T MRI system: comparison among different sequences. PLoS One 9:e99027PubMedPubMedCentralCrossRef
18.
go back to reference Barger AV, DeLone DR, Bernstein MA, Welker KM (2006) Fat signal suppression in head and neck imaging using fast spin-eho-IDEAL technique. AJNR Am J Neuroradiol 27:1292–1294PubMed Barger AV, DeLone DR, Bernstein MA, Welker KM (2006) Fat signal suppression in head and neck imaging using fast spin-eho-IDEAL technique. AJNR Am J Neuroradiol 27:1292–1294PubMed
19.
go back to reference Chang HC, Juan CJ, Chiu HC et al (2013) Parotid fat contents in healthy subjects evaluated with iterative decomposition with echo asymmetry and least squares fat–water separation. Radiology 267:918–923PubMedCrossRef Chang HC, Juan CJ, Chiu HC et al (2013) Parotid fat contents in healthy subjects evaluated with iterative decomposition with echo asymmetry and least squares fat–water separation. Radiology 267:918–923PubMedCrossRef
20.
go back to reference Tomura N, Saginoya T, Kokubun M, Tsuchiya M, Yamamoto T (2014) T2-weighted IDEAL fast spin echo imaging of the brachial plexus: comparison with STIR. Acta Radiol (pii:0284185114552292, Epub ahead of print) Tomura N, Saginoya T, Kokubun M, Tsuchiya M, Yamamoto T (2014) T2-weighted IDEAL fast spin echo imaging of the brachial plexus: comparison with STIR. Acta Radiol (pii:0284185114552292, Epub ahead of print)
21.
go back to reference Reeder SB, McKenzie CA, Pineda AR et al (2007) Water-fat separation with IDEAL gradient-echo imaging. J Magn Reson Imaging 25:644–652PubMedCrossRef Reeder SB, McKenzie CA, Pineda AR et al (2007) Water-fat separation with IDEAL gradient-echo imaging. J Magn Reson Imaging 25:644–652PubMedCrossRef
22.
go back to reference Glover G (1991) Multipoint Dixon technique for water and fat proton and susceptibility imaging. J Magn Reson Imaging 1:521–530PubMedCrossRef Glover G (1991) Multipoint Dixon technique for water and fat proton and susceptibility imaging. J Magn Reson Imaging 1:521–530PubMedCrossRef
23.
go back to reference Pineda AR, Reeder SB, Wen Z, Pelc NJ (2005) Cramér-Rao bounds for three-point decomposition of water and fat. Magn Reson Med 54:625–635PubMedCrossRef Pineda AR, Reeder SB, Wen Z, Pelc NJ (2005) Cramér-Rao bounds for three-point decomposition of water and fat. Magn Reson Med 54:625–635PubMedCrossRef
24.
go back to reference Starcuková J, Starcuk Z Jr, Hubálková H, Linetskiy I (2008) Magnetic susceptibility and electrical conductivity of metallic dental materials and their impact on MR imaging artifacts. Dent Mater 24:715–723PubMedCrossRef Starcuková J, Starcuk Z Jr, Hubálková H, Linetskiy I (2008) Magnetic susceptibility and electrical conductivity of metallic dental materials and their impact on MR imaging artifacts. Dent Mater 24:715–723PubMedCrossRef
25.
go back to reference Truong TK, Clymer BD, Chakeres DW, Schmalbrock P (2002) Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head. Magn Reson Imaging 20:759–770PubMedCrossRef Truong TK, Clymer BD, Chakeres DW, Schmalbrock P (2002) Three-dimensional numerical simulations of susceptibility-induced magnetic field inhomogeneities in the human head. Magn Reson Imaging 20:759–770PubMedCrossRef
26.
go back to reference Tagliafico A, Calabrese M, Puntoni M, Pace D, Baio G, Neumaier CE et al (2011) Brachial plexus MR imaging: accuracy and reproducibility of DTI-derived measurements and fibre tractography at 3.0-T. Eur Radiol 21:1764–1771PubMedCrossRef Tagliafico A, Calabrese M, Puntoni M, Pace D, Baio G, Neumaier CE et al (2011) Brachial plexus MR imaging: accuracy and reproducibility of DTI-derived measurements and fibre tractography at 3.0-T. Eur Radiol 21:1764–1771PubMedCrossRef
28.
go back to reference Tagliafico A, Succio G, Neumaier CE et al (2012) Brachial plexus assessment with three-dimensional isotropic resolution fast spin echo MRI: comparison with conventional MRI at 3.0 T. Br J Radiol 85:e110–e116PubMedPubMedCentralCrossRef Tagliafico A, Succio G, Neumaier CE et al (2012) Brachial plexus assessment with three-dimensional isotropic resolution fast spin echo MRI: comparison with conventional MRI at 3.0 T. Br J Radiol 85:e110–e116PubMedPubMedCentralCrossRef
29.
go back to reference Huo D, Li Z, Aboussouan E, Karis JP, Pipe JG (2009) Turboprop IDEAL: a motion-resistant fat-water separation technique. Magn Reson Med 61:188–195PubMedCrossRef Huo D, Li Z, Aboussouan E, Karis JP, Pipe JG (2009) Turboprop IDEAL: a motion-resistant fat-water separation technique. Magn Reson Med 61:188–195PubMedCrossRef
30.
go back to reference Tagliafico AS, Tagliafico G (2014) Fascicular ratio: a new parameter to evaluate peripheral nerve pathology on magnetic resonance imaging: a feasibility study on a 3T MRI system. Medicine (Baltimore) 93:e68CrossRef Tagliafico AS, Tagliafico G (2014) Fascicular ratio: a new parameter to evaluate peripheral nerve pathology on magnetic resonance imaging: a feasibility study on a 3T MRI system. Medicine (Baltimore) 93:e68CrossRef
31.
go back to reference Tagliafico AS (2011) Ulnar neuropathy at the elbow: is MR imaging reliable? Radiology 261:659–660PubMedCrossRef Tagliafico AS (2011) Ulnar neuropathy at the elbow: is MR imaging reliable? Radiology 261:659–660PubMedCrossRef
32.
go back to reference Husarik DB, Saupe N, Pfirrmann CW, Jost B, Hodler J, Zanetti M (2009) Elbow nerves: MR findings in 60 asymptomatic subjects—normal anatomy, variants, and pitfalls. Radiology 252:148–156PubMedCrossRef Husarik DB, Saupe N, Pfirrmann CW, Jost B, Hodler J, Zanetti M (2009) Elbow nerves: MR findings in 60 asymptomatic subjects—normal anatomy, variants, and pitfalls. Radiology 252:148–156PubMedCrossRef
Metadata
Title
Usefulness of IDEAL T2 imaging for homogeneous fat suppression and reducing susceptibility artefacts in brachial plexus MRI at 3.0 T
Authors
Alberto Tagliafico
Bianca Bignotti
Giulio Tagliafico
Carlo Martinoli
Publication date
01-01-2016
Publisher
Springer Milan
Published in
La radiologia medica / Issue 1/2016
Print ISSN: 0033-8362
Electronic ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-015-0576-3

Other articles of this Issue 1/2016

La radiologia medica 1/2016 Go to the issue