Skip to main content
Top
Published in: Pediatric Cardiology 7/2013

01-10-2013 | Original Article

Transcatheter Closure of Abnormal Vessels and Arteriovenous Fistulas With the Amplatzer Vascular Plug 4 in Patients With Congenital Heart Disease

Authors: Gesa Wiegand, Ludger Sieverding, Wolfgang Bocksch, Michael Hofbeck

Published in: Pediatric Cardiology | Issue 7/2013

Login to get access

Abstract

Although vascular plugs allow the interventional closure of medium-sized to large abnormal vessels, their application is limited by the need for long sheaths or large guiding catheters. The authors report their experience with the new Amplatzer vascular plug 4 (AVP 4), a self-expanding spindle-shaped occluder made of Nitinol wire mash, which can be placed through 4-Fr catheters with an internal diameter of 0.038 in. or larger. From October 2009 until June 2012, 14 AVP 4 devices were deployed in 12 patients (ages, 0.3–48.8 years). Nine patients had venovenous or arteriovenous collaterals in functional univentricular hearts. One patient had pulmonary atresia with a ventricular septal defect and major aortopulmonary collateral arteries, and one patient had a pulmonary arteriovenous fistula. One child had a large coronary artery fistula to the right atrium. The authors used AVP 4 devices with diameters of 4–8 mm. In all the patients, the AVP 4 was implanted successfully. No occluder dislocations and no complications related to the procedure occurred. Complete vessel occlusion was achieved in seven cases. In seven additional cases, a residual shunt was present at the end of the procedure while the patients were still fully heparinized. In 2 of 14 vessels, the decision was made to place additional devices to abolish residual shunting. According to the authors’ experience, the AVP 4 allows safe and effective occlusion of medium-size and large abnormal vessels. It is also well suited for tortuous high-flow vessels such as coronary or pulmonary arteriovenous fistulas. In case of a suboptimal position, it is possible to reposition the occluder with ease. Further studies are needed to determine whether initial residual shunting in heparinized patients disappears during follow-up care. The AVP 4 represents a valuable new device for the interventional treatment of complex congenital vessel malformations.
Literature
1.
go back to reference Adelmann R, Windfuhr A, Bennink G, Emmel M, Sreeram N (2011) Extended applications of the Amplatzer vascular plug IV in infants. Cardiol Young 21:178–181CrossRefPubMed Adelmann R, Windfuhr A, Bennink G, Emmel M, Sreeram N (2011) Extended applications of the Amplatzer vascular plug IV in infants. Cardiol Young 21:178–181CrossRefPubMed
2.
go back to reference Baspinar O, Mete A, Davutoglu V (2012) Transcatheter closure of a large left ventricular pseudoaneurysm using an Amplatzer vascular plug 4 and stenting of the inferior caval vein in a child. Cardiol Young 22:106–109CrossRefPubMed Baspinar O, Mete A, Davutoglu V (2012) Transcatheter closure of a large left ventricular pseudoaneurysm using an Amplatzer vascular plug 4 and stenting of the inferior caval vein in a child. Cardiol Young 22:106–109CrossRefPubMed
3.
go back to reference Brzezinska-Rajszys G (2007) Systemic arteriovenous fistulas. In: Sievert H, Qureshi A, Hijazi ZM (eds) Percutaneous interventions for congenital heart disease. Informa Healthcare, London, pp 411–418CrossRef Brzezinska-Rajszys G (2007) Systemic arteriovenous fistulas. In: Sievert H, Qureshi A, Hijazi ZM (eds) Percutaneous interventions for congenital heart disease. Informa Healthcare, London, pp 411–418CrossRef
4.
go back to reference Buheitel G, Ludwig J, Hofbeck M (2001) Transcatheter occlusion of a large coronary arterial fistula with new detachable platinum microcoils. Cardiol Young 11:571–573CrossRefPubMed Buheitel G, Ludwig J, Hofbeck M (2001) Transcatheter occlusion of a large coronary arterial fistula with new detachable platinum microcoils. Cardiol Young 11:571–573CrossRefPubMed
5.
go back to reference Buttarelli L, Capocasale E, Marcato C, Mazzoni MP, Iaria M, Rossi C (2011) Embolization of pancreatic allograft arteriovenous fistula with the Amplatzer vascular plug 4: case report and literature analysis. Transplant Proc 43:4044–4047CrossRefPubMed Buttarelli L, Capocasale E, Marcato C, Mazzoni MP, Iaria M, Rossi C (2011) Embolization of pancreatic allograft arteriovenous fistula with the Amplatzer vascular plug 4: case report and literature analysis. Transplant Proc 43:4044–4047CrossRefPubMed
6.
go back to reference Ferro C, Rossi UG, Bovio G, Petrocelli F, Seitun S (2010) The Amplatzer vascular plug 4: preliminary experience. Cardiovasc Interv Radiol 33:844–848CrossRef Ferro C, Rossi UG, Bovio G, Petrocelli F, Seitun S (2010) The Amplatzer vascular plug 4: preliminary experience. Cardiovasc Interv Radiol 33:844–848CrossRef
7.
go back to reference Fischer G, Apostolopoulou SC, Rammos S, Kiaffas M, Kramer HH (2007) Transcatheter closure of coronary arterial fistulas using the new Amplatzer vascular plug. Cardiol Young 17:283–287CrossRefPubMed Fischer G, Apostolopoulou SC, Rammos S, Kiaffas M, Kramer HH (2007) Transcatheter closure of coronary arterial fistulas using the new Amplatzer vascular plug. Cardiol Young 17:283–287CrossRefPubMed
8.
go back to reference Galal MO, Bulbul Z, Kakadekar A, Fatani AE, de Moor M, el-Oufi S, Solymar L, al-Fadley F, Fawzy ME (2001) Comparison between the safety profile and clinical results of the Cook detachable and Gianturco coils for transcatheter closure of patent ductus arteriosus in 272 patients. J Interv Cardiol 14:169–177CrossRefPubMed Galal MO, Bulbul Z, Kakadekar A, Fatani AE, de Moor M, el-Oufi S, Solymar L, al-Fadley F, Fawzy ME (2001) Comparison between the safety profile and clinical results of the Cook detachable and Gianturco coils for transcatheter closure of patent ductus arteriosus in 272 patients. J Interv Cardiol 14:169–177CrossRefPubMed
9.
go back to reference Girisch M, Sieverding L, Rauch R, Kaulitz R, Gass M, Ziemer G, Hofbeck M (2005) Recanalisation of bilateral superior vena cava after total cavopulmonary connection: interventional occlusion with the Amplatzer VSD occluder. Z Kardiol 94:469–473CrossRefPubMed Girisch M, Sieverding L, Rauch R, Kaulitz R, Gass M, Ziemer G, Hofbeck M (2005) Recanalisation of bilateral superior vena cava after total cavopulmonary connection: interventional occlusion with the Amplatzer VSD occluder. Z Kardiol 94:469–473CrossRefPubMed
10.
go back to reference Khan MD, Qureshi SA, Rosenthal E, Sharland GK (2003) Neonatal transcatheter occlusion of a large coronary artery fistula with the Amplatzer duct occluder. Catheter Cardiovasc Interv 60:282–286CrossRefPubMed Khan MD, Qureshi SA, Rosenthal E, Sharland GK (2003) Neonatal transcatheter occlusion of a large coronary artery fistula with the Amplatzer duct occluder. Catheter Cardiovasc Interv 60:282–286CrossRefPubMed
11.
go back to reference Libicher M, Reichert V, Schwabe H, Matoussevitch V, Gawenda M (2011) Occlusion of arteriovenous fistulas of in situ saphenous vein bypass grafts using the Amplatzer vascular plug 4: initial experience. Cardiovasc Interv Radiol 34:502–507CrossRef Libicher M, Reichert V, Schwabe H, Matoussevitch V, Gawenda M (2011) Occlusion of arteriovenous fistulas of in situ saphenous vein bypass grafts using the Amplatzer vascular plug 4: initial experience. Cardiovasc Interv Radiol 34:502–507CrossRef
12.
go back to reference MacDonald ST, Carminati M, Butera G (2011) Initial experience with the Amplatzer vascular plug IV in congenital heart disease: coronary artery fistula and aortopulmonary collateral artery embolization. J Invasive Cardiol 23:120–124PubMed MacDonald ST, Carminati M, Butera G (2011) Initial experience with the Amplatzer vascular plug IV in congenital heart disease: coronary artery fistula and aortopulmonary collateral artery embolization. J Invasive Cardiol 23:120–124PubMed
13.
go back to reference Mordasini P, Szucs-Farkas Z, Do DD, Gralla J, Kettenbach J, Hoppe H (2010) Use of a latest-generation vascular plug for peripheral vascular embolization with use of a diagnostic catheter: preliminary clinical experience. J Vasc Interv Radiol 21:1185–1190CrossRefPubMed Mordasini P, Szucs-Farkas Z, Do DD, Gralla J, Kettenbach J, Hoppe H (2010) Use of a latest-generation vascular plug for peripheral vascular embolization with use of a diagnostic catheter: preliminary clinical experience. J Vasc Interv Radiol 21:1185–1190CrossRefPubMed
14.
go back to reference Pech M, Mohnike K, Wieners G, Seidensticker R, Seidensticker M, Zapasnik A, Ricke J, Dudeck O (2011) Advantages and disadvantages of the Amplatzer vascular plug IV in visceral embolization: report of 50 placements. Cardiovasc Interv Radiol 34:1069–1073CrossRef Pech M, Mohnike K, Wieners G, Seidensticker R, Seidensticker M, Zapasnik A, Ricke J, Dudeck O (2011) Advantages and disadvantages of the Amplatzer vascular plug IV in visceral embolization: report of 50 placements. Cardiovasc Interv Radiol 34:1069–1073CrossRef
15.
go back to reference Peynircioglu B, Yurttutan N, Gulek B, Cil B, Yilmaz M (2011) Vertebral artery occlusion with Amplatzer vascular plug 4 to prevent subsequent endoleak in stent-graft treatment of subclavian artery gunshot injury. Acta Radiol 52:850–853CrossRefPubMed Peynircioglu B, Yurttutan N, Gulek B, Cil B, Yilmaz M (2011) Vertebral artery occlusion with Amplatzer vascular plug 4 to prevent subsequent endoleak in stent-graft treatment of subclavian artery gunshot injury. Acta Radiol 52:850–853CrossRefPubMed
17.
go back to reference Qureshi SA (2007) Transcatheter closure of coronary artery fistulas. In: Sievert H, Qureshi SA, Hijazi ZM (eds) Percutaneous interventions for congenital heart disease. Informa Healthcare, London, pp 423–430 Qureshi SA (2007) Transcatheter closure of coronary artery fistulas. In: Sievert H, Qureshi SA, Hijazi ZM (eds) Percutaneous interventions for congenital heart disease. Informa Healthcare, London, pp 423–430
18.
go back to reference Wiegand G, Sieverding L, Kaulitz R, Hofbeck M (2009) Transarterial and transvenous approach for transcatheter closure of a large coronary artery fistula with the Amplatzer vascular plug. Pediatr Cardiol 30:172–175CrossRefPubMed Wiegand G, Sieverding L, Kaulitz R, Hofbeck M (2009) Transarterial and transvenous approach for transcatheter closure of a large coronary artery fistula with the Amplatzer vascular plug. Pediatr Cardiol 30:172–175CrossRefPubMed
Metadata
Title
Transcatheter Closure of Abnormal Vessels and Arteriovenous Fistulas With the Amplatzer Vascular Plug 4 in Patients With Congenital Heart Disease
Authors
Gesa Wiegand
Ludger Sieverding
Wolfgang Bocksch
Michael Hofbeck
Publication date
01-10-2013
Publisher
Springer US
Published in
Pediatric Cardiology / Issue 7/2013
Print ISSN: 0172-0643
Electronic ISSN: 1432-1971
DOI
https://doi.org/10.1007/s00246-013-0701-9

Other articles of this Issue 7/2013

Pediatric Cardiology 7/2013 Go to the issue