Skip to main content
Top
Published in: Strahlentherapie und Onkologie 10/2013

01-10-2013 | Original article

Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy

Authors: C. Gaisberger, MSc, P. Steininger, B. Mitterlechner, S. Huber, H. Weichenberger, F. Sedlmayer, H. Deutschmann

Published in: Strahlentherapie und Onkologie | Issue 10/2013

Login to get access

Abstract

Purpose

Clinical evaluation of an optical three-dimensional surface scanning (3D-SS) system for patient positioning and monitoring during radiotherapy (RT) for breast cancer.

Materials and methods

A ceiling-mounted scanner was developed to acquire multiple 3D body surface images and tested in 14 conservatively operated breast cancer patients. A reference skin surface was derived from the planning computed tomography (CT) scan as basis for rigid registration with the surface scans. In addition to electronic portal images (EPIs), optical scans were acquired at three defined time points before and during daily RT. Patient setup was guided by laser alignments and corrected according to EPI findings. The accuracy of the 3D-SS system was validated by comparison of the optical scans to EPIs generated in parallel. Interfraction shifts were investigated by comparison of the first 3D-SS image with the reference body outline. Intrafractional motions were analysed by comparing the three daily surface scans with the first EPI.

Results

Comparison of EPIs and 3D-SS images revealed good accordance (− 0.05 ± 0.94 mm). Analysis of daily patient positions revealed average deviations of 0.4 ± 2.4 mm laterally, 0.3 ± 1.9 mm longitudinally and 0.2 ± 3.3 mm vertically. After 2 weeks, a systematic interfraction shift in patient positioning was noted, particularly in the vertical direction (4.9 ± 0.56 mm), which was attributed to patients progressively relaxing. 3D-SS images showed intrafractional shifts of 1.2 ± 0.7 mm over a time course of 2 min.

Conclusion

Optical surface scanning is a simple, fast and reproducible method for breast cancer patient alignment. Particularly for more sophisticated irradiation techniques, it helps to improve accuracy in patient positioning during radiotherapy without the exposure to additional ionizing radiation.
Literature
1.
go back to reference Bartoncini S, Fiandra C, Ruo Redda MG et al (2012) Target registration errors with surface imaging system in conformal radiotherapy for prostate cancer: study on 19 patients. Radiol Med 117(8):1419–1428PubMedCrossRef Bartoncini S, Fiandra C, Ruo Redda MG et al (2012) Target registration errors with surface imaging system in conformal radiotherapy for prostate cancer: study on 19 patients. Radiol Med 117(8):1419–1428PubMedCrossRef
2.
go back to reference Bert C, Metheany KG, Doppke K et al (2005) A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2:2753–2762CrossRef Bert C, Metheany KG, Doppke K et al (2005) A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2:2753–2762CrossRef
3.
go back to reference Bert C, Metheany KG, Doppke KP et al (2006) Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 64:1265–1274PubMedCrossRef Bert C, Metheany KG, Doppke KP et al (2006) Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 64:1265–1274PubMedCrossRef
4.
go back to reference Besl P, McKay N (1992) A method for registration of 3-D shapes. IEEE trans. Pat Anal Mach Intel 14(2):239–256CrossRef Besl P, McKay N (1992) A method for registration of 3-D shapes. IEEE trans. Pat Anal Mach Intel 14(2):239–256CrossRef
5.
go back to reference Brahme A, Nyman P, Skatt B (2008) 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 35:1670–1681PubMedCrossRef Brahme A, Nyman P, Skatt B (2008) 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 35:1670–1681PubMedCrossRef
6.
go back to reference Calow R, Gademann G, Krell G et al (2002) Photogrammetric measurement of patients in radiotherapy. ISPRS J Photogrammet Remote Sensing 56:347–359CrossRef Calow R, Gademann G, Krell G et al (2002) Photogrammetric measurement of patients in radiotherapy. ISPRS J Photogrammet Remote Sensing 56:347–359CrossRef
7.
go back to reference Czigner K, Agoston P, Forgács G et al (2012) Patient positioning variations to reduce dose to normal tissues during 3D conformal radiotherapy for high-risk prostate cancer. Strahlenther Onkol 188(9):816–822PubMedCrossRef Czigner K, Agoston P, Forgács G et al (2012) Patient positioning variations to reduce dose to normal tissues during 3D conformal radiotherapy for high-risk prostate cancer. Strahlenther Onkol 188(9):816–822PubMedCrossRef
8.
go back to reference Boer HC de, Sörnsen de Koste JR van, Senan S et al (2001) Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys 49(3):857–868PubMedCrossRef Boer HC de, Sörnsen de Koste JR van, Senan S et al (2001) Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys 49(3):857–868PubMedCrossRef
9.
go back to reference Deutschmann H, Steininger P, Nairz O et al (2008) “Augmented reality” in conventional simulation by projection of 3-D structures into 2-D images: a comparison with virtual methods. Strahlenther Onkol 184(2):93–99PubMedCrossRef Deutschmann H, Steininger P, Nairz O et al (2008) “Augmented reality” in conventional simulation by projection of 3-D structures into 2-D images: a comparison with virtual methods. Strahlenther Onkol 184(2):93–99PubMedCrossRef
10.
go back to reference Dickie-Euler C, Parent A, Griffin A et al (2008) Measuring interfraction and intrafraction motion with Cone Beam ComputedTomography (CBCT) and an Optical Localization System (OLS) for lower extremity soft tissue sarcoma patients treated with Preoperative Intensity Modulated Radiation Therapy (IMRT). Int J Radiat Oncol Biol Phys 72:S105–S105CrossRef Dickie-Euler C, Parent A, Griffin A et al (2008) Measuring interfraction and intrafraction motion with Cone Beam ComputedTomography (CBCT) and an Optical Localization System (OLS) for lower extremity soft tissue sarcoma patients treated with Preoperative Intensity Modulated Radiation Therapy (IMRT). Int J Radiat Oncol Biol Phys 72:S105–S105CrossRef
11.
go back to reference Gierga DP, Riboldi M, Turcotte JC et al (2008) Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 70(4):1239–1246PubMedCrossRef Gierga DP, Riboldi M, Turcotte JC et al (2008) Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 70(4):1239–1246PubMedCrossRef
12.
go back to reference Giovanna S, Carocci M, Rodella R (1999) Three-dimensional vision based on a combination of gray-code and phase-shift light projection: analysis and compensation of the systematic errors. Appl Optics 38(31):6565–6573CrossRef Giovanna S, Carocci M, Rodella R (1999) Three-dimensional vision based on a combination of gray-code and phase-shift light projection: analysis and compensation of the systematic errors. Appl Optics 38(31):6565–6573CrossRef
13.
go back to reference Gong GZ, Yin Y, Xing LG et al (2012) RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma. Strahlenther Onkol 188(3):262–268PubMedCrossRef Gong GZ, Yin Y, Xing LG et al (2012) RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma. Strahlenther Onkol 188(3):262–268PubMedCrossRef
14.
go back to reference Gulybán A, Kovács P, Sebestyén Z et al (2008) Multisegmented tangential breast fields: a rational way to treat breast cancer. Strahlenther Onkol 184(5):262–269PubMedCrossRef Gulybán A, Kovács P, Sebestyén Z et al (2008) Multisegmented tangential breast fields: a rational way to treat breast cancer. Strahlenther Onkol 184(5):262–269PubMedCrossRef
15.
go back to reference Hasan Y, Kim L, Martinez A et al (2008) Image guidance in external beam accelerated partial breast irradiation: comparison of surrogates for the lumpectomy cavity. Int J Radiat Oncol Biol Phys 70(2):619–625PubMedCrossRef Hasan Y, Kim L, Martinez A et al (2008) Image guidance in external beam accelerated partial breast irradiation: comparison of surrogates for the lumpectomy cavity. Int J Radiat Oncol Biol Phys 70(2):619–625PubMedCrossRef
16.
go back to reference Jin J, Ajlouni M, Ryu S et al (2007) A technique of quantitatively monitoring both respiratory and nonrespiratory motion in patients using external body markers. Med Phys 34:2875–2881PubMedCrossRef Jin J, Ajlouni M, Ryu S et al (2007) A technique of quantitatively monitoring both respiratory and nonrespiratory motion in patients using external body markers. Med Phys 34:2875–2881PubMedCrossRef
17.
go back to reference Kang HJ, Grelewicz Z, Wiersma RD (2012) Development of an automated region of interest selection method for 3D surface monitoring of head motion. Med Phys 39(6):3270–3282PubMedCrossRef Kang HJ, Grelewicz Z, Wiersma RD (2012) Development of an automated region of interest selection method for 3D surface monitoring of head motion. Med Phys 39(6):3270–3282PubMedCrossRef
18.
go back to reference Krengli M, Gaiano S, Mones E et al (2009) Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer. Radiat Oncol 4:9PubMedCrossRef Krengli M, Gaiano S, Mones E et al (2009) Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer. Radiat Oncol 4:9PubMedCrossRef
19.
go back to reference Liebich J, Licher J, Scherf C et al (2011) Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules. Strahlenther Onkol 187(5):316–321PubMedCrossRef Liebich J, Licher J, Scherf C et al (2011) Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules. Strahlenther Onkol 187(5):316–321PubMedCrossRef
20.
go back to reference Moore C, Lilley F, Sauret V et al (2003) Opto-electronic sensing of body surface topology changes during radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys 56(1):248–258PubMedCrossRef Moore C, Lilley F, Sauret V et al (2003) Opto-electronic sensing of body surface topology changes during radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys 56(1):248–258PubMedCrossRef
21.
go back to reference Mulchrone K (2003) Application of Delaunay triangulation to the nearest neighbour method of strain analysis. J Struct Geol 25:689–702CrossRef Mulchrone K (2003) Application of Delaunay triangulation to the nearest neighbour method of strain analysis. J Struct Geol 25:689–702CrossRef
22.
go back to reference Nairz O, Merz F, Deutschmann H et al (2008) A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlenther Onkol 184:663–667PubMedCrossRef Nairz O, Merz F, Deutschmann H et al (2008) A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlenther Onkol 184:663–667PubMedCrossRef
23.
go back to reference Pallotta S, Marrazzo L, Ceroti M et al (2012) A phantom evaluation of Sentinel(™), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef Pallotta S, Marrazzo L, Ceroti M et al (2012) A phantom evaluation of Sentinel(™), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef
24.
go back to reference Pili G, Grimaldi L, Fidanza C et al (2011) Geometric and dosimetric approach to determine probability of late cardiac mortality in left tangential breast irradiation: comparison between wedged beams and field-in-field technique. Int J Radiat Oncol Biol Phys 81(3):894–900PubMedCrossRef Pili G, Grimaldi L, Fidanza C et al (2011) Geometric and dosimetric approach to determine probability of late cardiac mortality in left tangential breast irradiation: comparison between wedged beams and field-in-field technique. Int J Radiat Oncol Biol Phys 81(3):894–900PubMedCrossRef
25.
go back to reference Remouchamps VM, Letts N, Yan D et al (2003) Three-dimensional evaluation of intra- and interfraction immobilization of lung and chest wall using active breathing control: a reproducibility study with breast cancer patients. Int J Radiat Oncol Biol Phys 57(4):968–978PubMedCrossRef Remouchamps VM, Letts N, Yan D et al (2003) Three-dimensional evaluation of intra- and interfraction immobilization of lung and chest wall using active breathing control: a reproducibility study with breast cancer patients. Int J Radiat Oncol Biol Phys 57(4):968–978PubMedCrossRef
26.
go back to reference Stieler F, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188(12):1080–1084PubMedCrossRef Stieler F, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188(12):1080–1084PubMedCrossRef
27.
go back to reference Subramaniam S, Thirumalaiswamy S, Srinivas C et al (2012) Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol 188(6):484–490PubMedCrossRef Subramaniam S, Thirumalaiswamy S, Srinivas C et al (2012) Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol 188(6):484–490PubMedCrossRef
28.
go back to reference Tsai R (1987) A versatile camera calibration techniaue for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses. IEEE J Robotics Automation 3(4):323–344CrossRef Tsai R (1987) A versatile camera calibration techniaue for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses. IEEE J Robotics Automation 3(4):323–344CrossRef
29.
go back to reference Vaerenbergh K et al (2005) Automatic generation of a plan optimization volume for tangential field breast cancer radiation therapy. Strahlenther Onkol 181(2):82–88PubMedCrossRef Vaerenbergh K et al (2005) Automatic generation of a plan optimization volume for tangential field breast cancer radiation therapy. Strahlenther Onkol 181(2):82–88PubMedCrossRef
Metadata
Title
Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy
Authors
C. Gaisberger, MSc
P. Steininger
B. Mitterlechner
S. Huber
H. Weichenberger
F. Sedlmayer
H. Deutschmann
Publication date
01-10-2013
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 10/2013
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-013-0358-6

Other articles of this Issue 10/2013

Strahlentherapie und Onkologie 10/2013 Go to the issue

Mitteilungen der Fachgesellschaften

Mitteilungen der Fachgesellschaften