Published in: Radiation Oncology 1/2009

Open Access 01-12-2009 | Research

Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer

Authors: Marco Krengli, Simone Gaiano, Eleonora Mones, Andrea Ballarè, Debora Beldì, Cesare Bolchini, Gianfranco Loi

Published in: Radiation Oncology | Issue 1/2009

Login to get access

Abstract

Background

The reproducibility of patient setup for radiotherapy is based on various methods including external markers, X-rays with planar or computerized image acquisition, and, more recently, surface matching imaging. We analyzed the setup reproducibility of 16 patients affected by prostate cancer who underwent conformal radiotherapy with curative intent by using a surface image registration system.

Methods

We analyzed the setup reproducibility of 16 patients affected by prostate cancer candidates for conformal radiotherapy by using a surface image registration system. At the initial setup, EPID images were compared with DRRs and a reference 3D surface image was obtained by the AlignRT system (Vision RT, London, UK). Surface images were acquired prior to every subsequent setup procedure. EPID acquisition was repeated when errors > 5 mm were reported.

Results

The mean random and systematic errors were 1.2 ± 2.3 mm and 0.3 ± 3.0 mm along the X axis, 0.0 ± 1.4 mm and 0.5 ± 2.0 mm along the Y axis, and 2.0 ± 1.8 mm and -0.7 ± 2.4 mm along the Z axis respectively. The positioning error detected by AlignRT along the 3 axes X, Y, and Z exceeded the value of 5 mm in 14.1%, 2.0%, and 5.1% measurements and the value of 3 mm in 36.9%, 13.6% and 27.8% measurements, respectively. Correlation factors calculated by linear regression between the errors measured by AlignRT and EPID ranged from 0.77 to 0.92 with a mean of 0.85 and SD of 0.13. The setup measurements by surface imaging are highly reproducible and correlate with the setup errors detected by EPID.

Conclusion

Surface image registration system appears to be a simple, fast, non-invasive, and reproducible method to analyze the set-up alignment in 3DCRT of prostate cancer patients.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kupelian PA, Lee C, Langen KM, Zeidan OA, Mañon RR, Willoughby TR, Meeks SL: Evaluation of image-guidance strategies in the treatment of localized prostate cancer. Int J Radiat Oncol Biol Phys 2008, 70: 1151-1157.CrossRefPubMed Kupelian PA, Lee C, Langen KM, Zeidan OA, Mañon RR, Willoughby TR, Meeks SL: Evaluation of image-guidance strategies in the treatment of localized prostate cancer. Int J Radiat Oncol Biol Phys 2008, 70: 1151-1157.CrossRefPubMed
2.
go back to reference Vargas C, Wagner M, Indelicato D, Fryer A, Horne D, Chellini A, McKenzie C, Lawlor P, Mahajan C, Li Z, Lin L, Keole S: Image Guidance Based on Prostate Position for Prostate Cancer Proton Therapy. Int J Radiat Oncol Biol Phys 2008, 71: 1322-1328.CrossRefPubMed Vargas C, Wagner M, Indelicato D, Fryer A, Horne D, Chellini A, McKenzie C, Lawlor P, Mahajan C, Li Z, Lin L, Keole S: Image Guidance Based on Prostate Position for Prostate Cancer Proton Therapy. Int J Radiat Oncol Biol Phys 2008, 71: 1322-1328.CrossRefPubMed
3.
go back to reference Ploeger LS, Frenay M, Betgen A, de Bois JA, Gilhuijs KG, van Herk M: Application of video imaging for improvement of patient set-up. Radiother Oncol 2003, 68: 277-284. 10.1016/S0167-8140(03)00230-5CrossRefPubMed Ploeger LS, Frenay M, Betgen A, de Bois JA, Gilhuijs KG, van Herk M: Application of video imaging for improvement of patient set-up. Radiother Oncol 2003, 68: 277-284. 10.1016/S0167-8140(03)00230-5CrossRefPubMed
4.
go back to reference Trichter F, Ennis RD: Prostate localization using transabdominal ultrasound imaging. Int J Radiat Oncol Biol Phys 2003, 56: 1225-1233. 10.1016/S0360-3016(03)00269-4CrossRefPubMed Trichter F, Ennis RD: Prostate localization using transabdominal ultrasound imaging. Int J Radiat Oncol Biol Phys 2003, 56: 1225-1233. 10.1016/S0360-3016(03)00269-4CrossRefPubMed
5.
go back to reference Brahme A, Nyman P, Skatt B: 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 2008, 35: 1670-1681. 10.1118/1.2889720CrossRefPubMed Brahme A, Nyman P, Skatt B: 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 2008, 35: 1670-1681. 10.1118/1.2889720CrossRefPubMed
6.
go back to reference Wong JR, Gao Z, Uematsu M, Merrick S, Machernis NP, Chen T, Cheng CW: Interfractional Prostate Shifts: Review of 1870 Computed Tomography (CT) Scans Obtained During Image-Guided Radiotherapy Using CT-on-Rails for the Treatment of Prostate Cancer. Int J Radiat Oncol Biol Phys 2008, 72: 1396-401.CrossRefPubMed Wong JR, Gao Z, Uematsu M, Merrick S, Machernis NP, Chen T, Cheng CW: Interfractional Prostate Shifts: Review of 1870 Computed Tomography (CT) Scans Obtained During Image-Guided Radiotherapy Using CT-on-Rails for the Treatment of Prostate Cancer. Int J Radiat Oncol Biol Phys 2008, 72: 1396-401.CrossRefPubMed
7.
go back to reference Bert C, Metheany KG, Doppke KP, Taghian AG, Powell SN, Chen GTY: Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 2006, 64: 1265-1274.CrossRefPubMed Bert C, Metheany KG, Doppke KP, Taghian AG, Powell SN, Chen GTY: Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 2006, 64: 1265-1274.CrossRefPubMed
8.
go back to reference Spadea MF, Baroni G, Riboldi M, Tagaste B, Garibaldi C, Orecchia R, Pedotti A: Patient set-up verification by infrared optical localization and body surface sensing in breast radiation therapy. Radiother Oncol 2006, 79: 170-178. 10.1016/j.radonc.2006.02.011CrossRefPubMed Spadea MF, Baroni G, Riboldi M, Tagaste B, Garibaldi C, Orecchia R, Pedotti A: Patient set-up verification by infrared optical localization and body surface sensing in breast radiation therapy. Radiother Oncol 2006, 79: 170-178. 10.1016/j.radonc.2006.02.011CrossRefPubMed
9.
go back to reference Schöffel PJ, Harms W, Sroka-Perez G, Schlegel W, Karger CP: Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax. Phys Med Biol 2007, 52: 3949-3963. 10.1088/0031-9155/52/13/019CrossRefPubMed Schöffel PJ, Harms W, Sroka-Perez G, Schlegel W, Karger CP: Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax. Phys Med Biol 2007, 52: 3949-3963. 10.1088/0031-9155/52/13/019CrossRefPubMed
10.
go back to reference Gierga DP, Riboldi M, Turcotte JC, Sharp GC, Jiang SB, Taghian AG, Chen GTY: Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 2008, 70: 1239-1246.CrossRefPubMed Gierga DP, Riboldi M, Turcotte JC, Sharp GC, Jiang SB, Taghian AG, Chen GTY: Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 2008, 70: 1239-1246.CrossRefPubMed
11.
go back to reference Bert C, Metheany KG, Doppke K, Chen GTY: A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2005, 32: 2753-2762. 10.1118/1.1984263CrossRefPubMed Bert C, Metheany KG, Doppke K, Chen GTY: A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2005, 32: 2753-2762. 10.1118/1.1984263CrossRefPubMed
12.
go back to reference Pagella S, Loi G, Krengli M: Study of set-up and internal margins (ICRU 62) for the delineation of the PTV in 3D conformal radiotherapy for prostate cancer: preliminary results. Tumori 2003,6(suppl 1):S31. Pagella S, Loi G, Krengli M: Study of set-up and internal margins (ICRU 62) for the delineation of the PTV in 3D conformal radiotherapy for prostate cancer: preliminary results. Tumori 2003,6(suppl 1):S31.
13.
go back to reference van Herk M, Remeijer P, Rasch C, Lebesque GV: The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 2000, 47: 1121-1135.CrossRefPubMed van Herk M, Remeijer P, Rasch C, Lebesque GV: The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 2000, 47: 1121-1135.CrossRefPubMed
14.
go back to reference Johnson LS, Milliken BD, Hadley SW, Pelizzari CA, Haraf DJ, Chen GTY: Initial clinical experience with a video-based patient positioning system. Int J Radiat Oncol Biol Phys 1999, 45: 205-213.CrossRefPubMed Johnson LS, Milliken BD, Hadley SW, Pelizzari CA, Haraf DJ, Chen GTY: Initial clinical experience with a video-based patient positioning system. Int J Radiat Oncol Biol Phys 1999, 45: 205-213.CrossRefPubMed
15.
go back to reference Hurkmans CW, Remeijer P, Lebesque JV, Mijnheer BJ: Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol 2001, 58: 105-120. 10.1016/S0167-8140(00)00260-7CrossRefPubMed Hurkmans CW, Remeijer P, Lebesque JV, Mijnheer BJ: Set-up verification using portal imaging: review of current clinical practice. Radiother Oncol 2001, 58: 105-120. 10.1016/S0167-8140(00)00260-7CrossRefPubMed
Metadata
Title
Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer
Authors
Marco Krengli
Simone Gaiano
Eleonora Mones
Andrea Ballarè
Debora Beldì
Cesare Bolchini
Gianfranco Loi
Publication date
01-12-2009
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2009
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-4-9

Other articles of this Issue 1/2009

Radiation Oncology 1/2009 Go to the issue