Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy

Authors: F. Stieler, PhD, F. Wenz, D. Scherrer, M. Bernhardt, F. Lohr

Published in: Strahlentherapie und Onkologie | Issue 12/2012

Login to get access

Abstract

Background

Laser scanning-based patient surface positioning and surveillance may complement image-guided radiotherapy (IGRT) as a nonradiation-based approach. We investigated the performance of an optical system compared to standard kilovoltage cone-beam computed tomography (CBCT) and its potential to reduce the number of daily CBCTs.

Patients and methods

We analyzed the patient positioning of 153 treatment fractions in 21 patients applied to three different treatment regions. Patients were first scanned with CBCT, shifted to the optimal isocenter position, and an optical scan was performed to verify the matching in relation to CBCT.

Results

For the head-and-neck region, the lateral/longitudinal/vertical/rotational/roll and pitch shift was 0.9 ± 1.8 mm/−2.7± 3.8 mm/−0.8± 3.6 mm/0.0± 1.1°/−0.5± 2.1°/0.2± 1.6°. For the thorax, the lateral/longitudinal/vertical/roll and pitch shift was −1.2± 3.6 mm/0.8± 5.1 mm/0.8± 4.3 mm/0.6± 1.4°/0.1± 0.9°/0.3± 1.0°. For the pelvis, the respective values were −2.5± 4.1 mm/4.6± 7.3 mm/−5.1± 7.4 mm/0.3± 1.1°/-0.5± 1.0°/0.3± 2.1°. In total, the recorded disagreement was −1.0± 3.6 mm/1.0± 6.3 mm/−1.8± 5.9 mm/0.3± 1.2°/−0.3± 1.5°/0.2± 1.7°.

Conclusion

This analysis showed good agreement between the optical scanner approach and CBCT. The optical system holds potential to ensure precise patient positioning and reduced CBCT frequency in tumor locations with fixed relation to surface structures.
Literature
1.
go back to reference Boda-Heggemann J et al (2011) kV cone-beam CT-based IGRT: a clinical review. Strahlenther Onkol 187(5):284–291PubMedCrossRef Boda-Heggemann J et al (2011) kV cone-beam CT-based IGRT: a clinical review. Strahlenther Onkol 187(5):284–291PubMedCrossRef
2.
go back to reference Walter C et al (2007) Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kV portal images vs. cone-beam CT. Radiother Oncol 85(3):418–423PubMedCrossRef Walter C et al (2007) Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kV portal images vs. cone-beam CT. Radiother Oncol 85(3):418–423PubMedCrossRef
3.
go back to reference Amer A et al (2007) Imaging doses from the Elekta Synergy X-ray cone beam CT system. Br J Radiol 80(954):476–482PubMedCrossRef Amer A et al (2007) Imaging doses from the Elekta Synergy X-ray cone beam CT system. Br J Radiol 80(954):476–482PubMedCrossRef
4.
go back to reference Pallotta S et al (2012) A phantom evaluation of Sentinel(), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef Pallotta S et al (2012) A phantom evaluation of Sentinel(), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef
5.
go back to reference Gopan, O, Wu Q (2012) Evaluation of the Accuracy of a 3D Surface Imaging System for Patient Setup in Head and Neck Cancer Radiotherapy. Int J Radiat Oncol Biol Phys Gopan, O, Wu Q (2012) Evaluation of the Accuracy of a 3D Surface Imaging System for Patient Setup in Head and Neck Cancer Radiotherapy. Int J Radiat Oncol Biol Phys
6.
go back to reference Kauweloa KI et al (2012) GateCT surface tracking system for respiratory signal reconstruction in 4DCT imaging. Med Phys 39(1):492–502PubMedCrossRef Kauweloa KI et al (2012) GateCT surface tracking system for respiratory signal reconstruction in 4DCT imaging. Med Phys 39(1):492–502PubMedCrossRef
7.
go back to reference Placht S et al (2012) Fast time-of-flight camera based surface registration for radiotherapy patient positioning. Med Phys 39(1):4–17PubMedCrossRef Placht S et al (2012) Fast time-of-flight camera based surface registration for radiotherapy patient positioning. Med Phys 39(1):4–17PubMedCrossRef
8.
go back to reference Besl PJ, McKay ND (1992) A method for registration of 3-D shapes. IEEE Trans. Pattern Anal Mach Intell 14:239–255CrossRef Besl PJ, McKay ND (1992) A method for registration of 3-D shapes. IEEE Trans. Pattern Anal Mach Intell 14:239–255CrossRef
9.
go back to reference Moser T et al (2012) Clinical Evaluation of a Laser Surface Scanning System in 120 Patients for Improving Daily Setup Accuracy in Fractionated Radiation Therapy. Int J Radiat Oncol Biol Phys Moser T et al (2012) Clinical Evaluation of a Laser Surface Scanning System in 120 Patients for Improving Daily Setup Accuracy in Fractionated Radiation Therapy. Int J Radiat Oncol Biol Phys
10.
go back to reference Hansen EK et al (2006) Repeat CT imaging and replanning during the course of IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys 64(2):355–362PubMedCrossRef Hansen EK et al (2006) Repeat CT imaging and replanning during the course of IMRT for head-and-neck cancer. Int J Radiat Oncol Biol Phys 64(2):355–362PubMedCrossRef
11.
go back to reference Ho KF et al (2012) Monitoring dosimetric impact of weight loss with kilovoltage (kV) cone beam CT (CBCT) during parotid-sparing IMRT and concurrent chemotherapy. Int J Radiat Oncol Biol Phys 82(3):e375–e382PubMedCrossRef Ho KF et al (2012) Monitoring dosimetric impact of weight loss with kilovoltage (kV) cone beam CT (CBCT) during parotid-sparing IMRT and concurrent chemotherapy. Int J Radiat Oncol Biol Phys 82(3):e375–e382PubMedCrossRef
12.
go back to reference Giraud P et al (2003) Respiration-gated radiotherapy: current techniques and potential benefits. Cancer Radiother 7(Suppl 1):15s–25sPubMed Giraud P et al (2003) Respiration-gated radiotherapy: current techniques and potential benefits. Cancer Radiother 7(Suppl 1):15s–25sPubMed
13.
go back to reference Lorchel F et al (2006) Dosimetric consequences of breath-hold respiration in conformal radiotherapy of esophageal cancer. Phys Med 22(4):119–126PubMedCrossRef Lorchel F et al (2006) Dosimetric consequences of breath-hold respiration in conformal radiotherapy of esophageal cancer. Phys Med 22(4):119–126PubMedCrossRef
14.
go back to reference Guckenberger M et al (2012) Motion Compensation in Radiotherapy. Biomedical Engineering 40(3):187–197 Guckenberger M et al (2012) Motion Compensation in Radiotherapy. Biomedical Engineering 40(3):187–197
Metadata
Title
Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy
Authors
F. Stieler, PhD
F. Wenz
D. Scherrer
M. Bernhardt
F. Lohr
Publication date
01-12-2012
Publisher
Springer-Verlag
Published in
Strahlentherapie und Onkologie / Issue 12/2012
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-012-0244-7

Other articles of this Issue 12/2012

Strahlentherapie und Onkologie 12/2012 Go to the issue

Mitteilungen der Fachgesellschaften

Mitteilungen