The Lateral Plane Delivers Higher Dose than the Frontal Plane in Biplane Cardiac Catheterization Systems
Authors:
Osamah Aldoss, Sonali Patel, Kyle Harris, Abhay Divekar
Published in:
Pediatric Cardiology
|
Issue 5/2015
Login to get access
Abstract
The objective of the study is to compare radiation dose between the frontal and lateral planes in a biplane cardiac catheterization laboratory. Tube angulation progressively increases patient and operator radiation dose in single-plane cardiac catheterization laboratories. This retrospective study captured biplane radiation dose in a pediatric cardiac catheterization laboratory between April 2010 and January 2014. Raw and time-indexed fluoroscopic, cineangiographic and total (fluoroscopic + cineangiographic) air kerma (AK, mGy) and kerma area product (PKA, µGym2/Kg) for each plane were compared. Data for 716 patients were analyzed: 408 (56.98 %) were male, the median age was 4.86 years, and the median weight was 17.35 kg. Although median beam-on time (minutes) was 4.2 times greater in the frontal plane, there was no difference in raw median total PKA between the two planes. However, when indexed to beam-on time, the lateral plane had a higher median-indexed fluoroscopic (0.75 vs. 1.70), cineangiographic (16.03 vs. 24.92), and total (1.43 vs. 5.15) PKA (p < 0.0001). The median time-indexed total PKA in the lateral plane is 3.6 times the frontal plane. This is the first report showing that the lateral plane delivers a higher dose than the frontal plane per unit time. Operators should consciously reduce the lateral plane beam-on time and incorporate this practice in radiation reduction protocols.