01-04-2021 | Temperature Management | Original Work

Comparison of a Continuous Noninvasive Temperature to Monitor Core Temperature Measures During Targeted Temperature Management

Authors: Melissa Wagner, Krista Lim-Hing, Mary Ann Bautista, Brigid Blaber, Taghi Ryder, Joseph Haymore, Neeraj Badjatia

Published in: Neurocritical Care | Issue 2/2021

Login to get access

Abstract

Background

Temperature modulating devices (TMD) currently utilize core temperature measurements during targeted temperature management (TTM) that are currently limited to esophageal (Et), bladder (Bt), or rectal (Rt) temperatures. We assessed the ability of a continuous noninvasive temperature monitor to accurately approximate core temperature during TTM.

Methods

All patients undergoing TTM using a gel pad surface TMD and an existing core temperature monitoring device were eligible for this study. Core and continuous noninvasive temperature monitoring values were simultaneously recorded for up to 72 h of TTM. The two sets of temperature data were downloaded from a clinical data acquisition storage system at 1-min intervals. The Bland–Altman method assessed agreement between the core and continuous noninvasive temperature monitor values, by measuring the mean difference (± 2 SD) between these values.

Results

There were 20 subjects that underwent study between January 2018 and March 2018 (55% women, age: 57 ± 14 years old, BMI: 28.9 + 9.8 kg/m2, 100% mechanically ventilated). The comparison patient temperature source was predominantly esophageal (n = 10) followed by bladder (n = 5) or rectal (n = 5). There were a total of 999 h of paired patient temperature data from esophageal (50%), bladder (25%), and rectal (25%) temperatures. Bland–Altman analysis demonstrated good agreement with the superficial temperature monitor and core temperature measures in all patients overall, with a difference mean of 0.06 ± 0.39 C (P = 0.99) and no proportional bias noted (β =0.002, P = 0.917).

Conclusions

Continuous noninvasive temperature monitoring is a suitable alternative method for assessing core temperature during TTM. Future studies should focus on developing connectivity with a continuous noninvasive temperature monitor to approximate core temperature during TTM.
Literature
9.
go back to reference Suleman MI, Doufas AG, Akca O, Ducharme M, Sessler DI. Insufficiency in a new temporal-artery thermometer for adult and pediatric patients. Anesth Analg. 2002;95:67–71.CrossRef Suleman MI, Doufas AG, Akca O, Ducharme M, Sessler DI. Insufficiency in a new temporal-artery thermometer for adult and pediatric patients. Anesth Analg. 2002;95:67–71.CrossRef
10.
go back to reference Moran JL, Peter JV, Solomon PJ, et al. Tympanic temperature measurements: are they reliable in the critically ill? A clinical study of measures of agreement. Crit Care Med. 2007;35:155–64.CrossRef Moran JL, Peter JV, Solomon PJ, et al. Tympanic temperature measurements: are they reliable in the critically ill? A clinical study of measures of agreement. Crit Care Med. 2007;35:155–64.CrossRef
11.
go back to reference Kimberger O, Cohen D, Illievich U, Lenhardt R. Temporal artery versus bladder thermometry during perioperative and intensive care unit monitoring. Anesth Analg. 2007;105:1042–7.CrossRef Kimberger O, Cohen D, Illievich U, Lenhardt R. Temporal artery versus bladder thermometry during perioperative and intensive care unit monitoring. Anesth Analg. 2007;105:1042–7.CrossRef
12.
go back to reference Tayefeh F, Plattner O, Sessler DI, Ikeda T, Marder D. Circadian changes in the sweating-to-vasoconstriction interthreshold range. Pflugers Arch. 1998;435:402–6.CrossRef Tayefeh F, Plattner O, Sessler DI, Ikeda T, Marder D. Circadian changes in the sweating-to-vasoconstriction interthreshold range. Pflugers Arch. 1998;435:402–6.CrossRef
13.
go back to reference Sessler DI, Lee KA, McGuire J. Isoflurane anesthesia and circadian temperature cycles in humans. Anesthesiology. 1991;75:985–9.CrossRef Sessler DI, Lee KA, McGuire J. Isoflurane anesthesia and circadian temperature cycles in humans. Anesthesiology. 1991;75:985–9.CrossRef
14.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef
15.
go back to reference Platt M, Vicario S. 141/Heat illness. Rosen’s emergency medicine, vol. 2. 8th ed. Philadelphia: Saunders; 2009. p. 1904. Platt M, Vicario S. 141/Heat illness. Rosen’s emergency medicine, vol. 2. 8th ed. Philadelphia: Saunders; 2009. p. 1904.
Metadata
Title
Comparison of a Continuous Noninvasive Temperature to Monitor Core Temperature Measures During Targeted Temperature Management
Authors
Melissa Wagner
Krista Lim-Hing
Mary Ann Bautista
Brigid Blaber
Taghi Ryder
Joseph Haymore
Neeraj Badjatia
Publication date
01-04-2021
Publisher
Springer US
Published in
Neurocritical Care / Issue 2/2021
Print ISSN: 1541-6933
Electronic ISSN: 1556-0961
DOI
https://doi.org/10.1007/s12028-020-01036-9

Other articles of this Issue 2/2021

Neurocritical Care 2/2021 Go to the issue