Syringe and Needle Size, Syringe Type, Vacuum Generation, and Needle Control in Aspiration Procedures

Authors: Luke J. Haseler, Randy R. Sibbitt, Wilmer L. Sibbitt Jr., Adrian A. Michael, Charles M. Gasparovic, Arthur D. Bankhurst

Published in: CardioVascular and Interventional Radiology | Issue 3/2011

Login to get access

Abstract

Purpose

Syringes are used for diagnostic fluid aspiration and fine-needle aspiration biopsy in interventional procedures. We determined the benefits, disadvantages, and patient safety implications of syringe and needle size on vacuum generation, hand force requirements, biopsy/fluid yield, and needle control during aspiration procedures.

Materials and Methods

Different sizes (1, 3, 5, 10, and 20 ml) of the conventional syringe and aspirating mechanical safety syringe, the reciprocating procedure device, were studied. Twenty operators performed aspiration procedures with the following outcomes measured: (1) vacuum (torr), (2) time to vacuum (s), (3) hand force to generate vacuum (torr-cm2), (4) operator difficulty during aspiration, (5) biopsy yield (mg), and (6) operator control of the needle tip position (mm).

Results

Vacuum increased tissue biopsy yield at all needle diameters (P < 0.002). Twenty-milliliter syringes achieved a vacuum of −517 torr but required far more strength to aspirate, and resulted in significant loss of needle control (P < 0.002). The 10-ml syringe generated only 15% less vacuum (−435 torr) than the 20-ml device and required much less hand strength. The mechanical syringe generated identical vacuum at all syringe sizes with less hand force (P < 0.002) and provided significantly enhanced needle control (P < 0.002).

Conclusions

To optimize patient safety and control of the needle, and to maximize fluid and tissue yield during aspiration procedures, a two-handed technique and the smallest syringe size adequate for the procedure should be used. If precise needle control or one-handed operation is required, a mechanical safety syringe should be considered.
Literature
1.
go back to reference Tani E, Seregard S, Rupp G et al (2006) Fine-needle aspiration cytology and immunocytochemistry of orbital masses. Diagn Cytopathol 34:1–5PubMedCrossRef Tani E, Seregard S, Rupp G et al (2006) Fine-needle aspiration cytology and immunocytochemistry of orbital masses. Diagn Cytopathol 34:1–5PubMedCrossRef
2.
go back to reference Quinn SF, Nelson HA, Demlow TA (1994) Thyroid biopsies: fine-needle aspiration biopsy versus spring-activated core biopsy needle in 102 patients. J Vasc Interv Radiol 5:619–623PubMedCrossRef Quinn SF, Nelson HA, Demlow TA (1994) Thyroid biopsies: fine-needle aspiration biopsy versus spring-activated core biopsy needle in 102 patients. J Vasc Interv Radiol 5:619–623PubMedCrossRef
3.
go back to reference Althoff CE, Hermann KG, Wiechen K et al (2006) Formalin-fixed blood clots—additional histological findings on computed tomography-guided fine-needle aspiration biopsies in comparison with core biopsies. J Comput Assist Tomogr 30:386–390PubMedCrossRef Althoff CE, Hermann KG, Wiechen K et al (2006) Formalin-fixed blood clots—additional histological findings on computed tomography-guided fine-needle aspiration biopsies in comparison with core biopsies. J Comput Assist Tomogr 30:386–390PubMedCrossRef
4.
go back to reference Jain D (2002) Diagnosis of hepatocellular carcinoma: fine needle aspiration cytology or needle core biopsy. J Clin Gastroenterol 35(5 suppl 2):S101–S108PubMedCrossRef Jain D (2002) Diagnosis of hepatocellular carcinoma: fine needle aspiration cytology or needle core biopsy. J Clin Gastroenterol 35(5 suppl 2):S101–S108PubMedCrossRef
5.
go back to reference Coldewey J, Stewart IS (2002) Comparison of fine needle aspiration cytology and needle core biopsy in the diagnosis of radiologically detected abdominal lesions. J Clin Pathol 55:93–97PubMedCrossRef Coldewey J, Stewart IS (2002) Comparison of fine needle aspiration cytology and needle core biopsy in the diagnosis of radiologically detected abdominal lesions. J Clin Pathol 55:93–97PubMedCrossRef
6.
go back to reference Aviram G, Schwartz DS, Meirsdorf S et al (2005) Transthoracic needle biopsy of lung masses: a survey of techniques. Clin Radiol 60:370–374PubMedCrossRef Aviram G, Schwartz DS, Meirsdorf S et al (2005) Transthoracic needle biopsy of lung masses: a survey of techniques. Clin Radiol 60:370–374PubMedCrossRef
7.
go back to reference Oyama T, Koibuchi Y, McKee G (2004) Core needle biopsy (CNB) as a diagnostic method for breast lesions: comparison with fine needle aspiration cytology (FNA). Breast Cancer 11:339–342PubMedCrossRef Oyama T, Koibuchi Y, McKee G (2004) Core needle biopsy (CNB) as a diagnostic method for breast lesions: comparison with fine needle aspiration cytology (FNA). Breast Cancer 11:339–342PubMedCrossRef
8.
go back to reference Masood S (2005) Core needle biopsy versus fine needle aspiration biopsy: are there similar sampling and diagnostic issues? Clin Lab Med 25:679–688PubMedCrossRef Masood S (2005) Core needle biopsy versus fine needle aspiration biopsy: are there similar sampling and diagnostic issues? Clin Lab Med 25:679–688PubMedCrossRef
9.
go back to reference Cochrane L, Ainscough M, Alfirevic Z (2003) The influence of needle and syringe size on chorionic villus sampling of term placentae: a randomised trial. Prenat Diagn 23:1049–1051PubMedCrossRef Cochrane L, Ainscough M, Alfirevic Z (2003) The influence of needle and syringe size on chorionic villus sampling of term placentae: a randomised trial. Prenat Diagn 23:1049–1051PubMedCrossRef
10.
go back to reference Cannon CR, Richardson LD, Replogle W, Halloran R (1996) Quantitative evaluation of fine-needle aspiration. Otolaryngol Head Neck Surg 114:407–412PubMedCrossRef Cannon CR, Richardson LD, Replogle W, Halloran R (1996) Quantitative evaluation of fine-needle aspiration. Otolaryngol Head Neck Surg 114:407–412PubMedCrossRef
11.
go back to reference Yankelevitz DF, Hayt D, Henschke CI (1995) Transthoracic needle biopsy. What size syringe? Clin Imaging 19:208–209PubMedCrossRef Yankelevitz DF, Hayt D, Henschke CI (1995) Transthoracic needle biopsy. What size syringe? Clin Imaging 19:208–209PubMedCrossRef
12.
go back to reference Bastard JP, Cuevas J, Cohen S et al (1994) Percutaneous adipose tissue biopsy by mini-liposuction for metabolic studies. JPEN J Parenter Enteral Nutr 18:466–468PubMedCrossRef Bastard JP, Cuevas J, Cohen S et al (1994) Percutaneous adipose tissue biopsy by mini-liposuction for metabolic studies. JPEN J Parenter Enteral Nutr 18:466–468PubMedCrossRef
13.
go back to reference Sandrucci F, Vismara L, Molinari S et al (1998) Percutaneous needle biopsy guided with computerized tomography of the chest. Personal experience with 1, 605 cases. Radiol Med 96:375–383PubMed Sandrucci F, Vismara L, Molinari S et al (1998) Percutaneous needle biopsy guided with computerized tomography of the chest. Personal experience with 1, 605 cases. Radiol Med 96:375–383PubMed
14.
go back to reference Hopper KD, Abendroth CS, Sturtz KW et al (1992) Fine-needle aspiration biopsy for cytopathologic analysis: utility of syringe handles, automated guns, and the nonsuction method. Radiology 185:819–824PubMed Hopper KD, Abendroth CS, Sturtz KW et al (1992) Fine-needle aspiration biopsy for cytopathologic analysis: utility of syringe handles, automated guns, and the nonsuction method. Radiology 185:819–824PubMed
15.
16.
go back to reference Bates T, Davidson T, Mansel RE (2002) Litigation for pneumothorax as a complication of fine-needle aspiration of the breast. Br J Surg 89:134–137PubMed Bates T, Davidson T, Mansel RE (2002) Litigation for pneumothorax as a complication of fine-needle aspiration of the breast. Br J Surg 89:134–137PubMed
17.
go back to reference Gordon CE, Feller-Kopman D, Balk EM, Smetana GW (2010) Pneumothorax following thoracentesis: a systematic review and meta-analysis. Arch Intern Med 170:332–339PubMedCrossRef Gordon CE, Feller-Kopman D, Balk EM, Smetana GW (2010) Pneumothorax following thoracentesis: a systematic review and meta-analysis. Arch Intern Med 170:332–339PubMedCrossRef
18.
go back to reference Drinkovic I, Brkljacic B (1996) Cases of lethal complications following ultrasound guided percutaneous fine-needle biopsy of the liver. Cardiovasc Intervent Radiol 19:360–363PubMedCrossRef Drinkovic I, Brkljacic B (1996) Cases of lethal complications following ultrasound guided percutaneous fine-needle biopsy of the liver. Cardiovasc Intervent Radiol 19:360–363PubMedCrossRef
19.
go back to reference Rocke DA (1984) Percutaneous lung biopsy. Management of tracheobronchial haemorrhage. Anaesthesia 39:888–890PubMedCrossRef Rocke DA (1984) Percutaneous lung biopsy. Management of tracheobronchial haemorrhage. Anaesthesia 39:888–890PubMedCrossRef
20.
go back to reference Windsor RE, Storm S, Sugar R (2003) Prevention and management of complications resulting from common spinal injections. Pain Physician 6:473–483PubMed Windsor RE, Storm S, Sugar R (2003) Prevention and management of complications resulting from common spinal injections. Pain Physician 6:473–483PubMed
21.
go back to reference Shevland JE (1991) Right ventricular perforation: a rare complication of percutaneous lung biopsy. J Thorac Imaging 6:85–86PubMedCrossRef Shevland JE (1991) Right ventricular perforation: a rare complication of percutaneous lung biopsy. J Thorac Imaging 6:85–86PubMedCrossRef
25.
go back to reference Heget JR, Bagian JP, Lee CZ et al (2002) Eisenberg Patient Safety Awards. System innovation: Veterans Health Administration National Center for Patient Safety. Jt Comm J Qual Improv 28:660–665PubMed Heget JR, Bagian JP, Lee CZ et al (2002) Eisenberg Patient Safety Awards. System innovation: Veterans Health Administration National Center for Patient Safety. Jt Comm J Qual Improv 28:660–665PubMed
26.
go back to reference Thomas AN, Galvin I (2008) Patient safety incidents associated with equipment in critical care: a review of reports to the UK National Patient Safety Agency. Anaesthesia 63:1193–1197PubMedCrossRef Thomas AN, Galvin I (2008) Patient safety incidents associated with equipment in critical care: a review of reports to the UK National Patient Safety Agency. Anaesthesia 63:1193–1197PubMedCrossRef
27.
go back to reference Duncan JR (2008) Strategies for improving safety and quality in interventional radiology. J Vasc Interv Radiol 19:3–7PubMedCrossRef Duncan JR (2008) Strategies for improving safety and quality in interventional radiology. J Vasc Interv Radiol 19:3–7PubMedCrossRef
28.
go back to reference Liang BA, Riley W, Hamman W, Rutherford W (2007) The patient safety and quality improvement act of 2005: provisions and potential. Am J Med Qual 22:8–12PubMedCrossRef Liang BA, Riley W, Hamman W, Rutherford W (2007) The patient safety and quality improvement act of 2005: provisions and potential. Am J Med Qual 22:8–12PubMedCrossRef
29.
go back to reference Sibbitt RR, Sibbitt WL Jr, Nunez SE et al (2006) Control and performance characteristics of eight different suction biopsy devices. J Vasc Interv Radiol 17:1657–1669PubMedCrossRef Sibbitt RR, Sibbitt WL Jr, Nunez SE et al (2006) Control and performance characteristics of eight different suction biopsy devices. J Vasc Interv Radiol 17:1657–1669PubMedCrossRef
30.
go back to reference Sibbitt WL Jr, Sibbitt RR, Michael AA et al (2006) Physician control of needle and syringe during aspiration–injection procedures with the new reciprocating syringe. J Rheumatol 33:771–778PubMed Sibbitt WL Jr, Sibbitt RR, Michael AA et al (2006) Physician control of needle and syringe during aspiration–injection procedures with the new reciprocating syringe. J Rheumatol 33:771–778PubMed
31.
go back to reference Hopper KD, Grenko RT, Fisher AI et al (1996) Capillary versus aspiration biopsy: effect of needle size and length on the cytopathological specimen quality. Cardiovasc Intervent Radiol 19:341–344PubMedCrossRef Hopper KD, Grenko RT, Fisher AI et al (1996) Capillary versus aspiration biopsy: effect of needle size and length on the cytopathological specimen quality. Cardiovasc Intervent Radiol 19:341–344PubMedCrossRef
32.
go back to reference Savage CA, Hopper KD, Abendroth CS et al (1995) Fine-needle aspiration biopsy versus fine-needle capillary (nonaspiration) biopsy: in vivo comparison. Radiology 195:815–819PubMed Savage CA, Hopper KD, Abendroth CS et al (1995) Fine-needle aspiration biopsy versus fine-needle capillary (nonaspiration) biopsy: in vivo comparison. Radiology 195:815–819PubMed
33.
go back to reference Haddadi-Nezhad S, Larijani B, Tavangar SM, Nouraei SM (2003) Comparison of fine-needle-nonaspiration with fine-needle-aspiration technique in the cytologic studies of thyroid nodules. Endocr Pathol 14:369–373PubMedCrossRef Haddadi-Nezhad S, Larijani B, Tavangar SM, Nouraei SM (2003) Comparison of fine-needle-nonaspiration with fine-needle-aspiration technique in the cytologic studies of thyroid nodules. Endocr Pathol 14:369–373PubMedCrossRef
34.
go back to reference Ciatto S, Catania S, Bravetti P et al (1991) Fine-needle cytology of the breast: a controlled study of aspiration versus nonaspiration. Diagn Cytopathol 7:125–127PubMedCrossRef Ciatto S, Catania S, Bravetti P et al (1991) Fine-needle cytology of the breast: a controlled study of aspiration versus nonaspiration. Diagn Cytopathol 7:125–127PubMedCrossRef
35.
go back to reference Moorjani GR, Michael AA, Peisjovich A et al (2008) Patient pain and tissue trauma during syringe procedures: a randomized controlled trial. J Rheumatology 35:1124–1129 Moorjani GR, Michael AA, Peisjovich A et al (2008) Patient pain and tissue trauma during syringe procedures: a randomized controlled trial. J Rheumatology 35:1124–1129
36.
go back to reference Sibbitt WL Jr, Peisajovich A, Michael AA et al (2009) Does sonographic needle guidance affect the clinical outcome of intraarticular injections? J Rheumatol 36:1892–1902PubMedCrossRef Sibbitt WL Jr, Peisajovich A, Michael AA et al (2009) Does sonographic needle guidance affect the clinical outcome of intraarticular injections? J Rheumatol 36:1892–1902PubMedCrossRef
37.
go back to reference Calda P, Brestak M (2009) Amniovacucentesis vs standard syringe technique for amniocentesis: experience with 1219 cases. Am J Obstet Gynecol 201:593.e1–593.e3CrossRef Calda P, Brestak M (2009) Amniovacucentesis vs standard syringe technique for amniocentesis: experience with 1219 cases. Am J Obstet Gynecol 201:593.e1–593.e3CrossRef
38.
go back to reference Kahveci R, Rehimli M, Esmer A et al (2009) A useful technique to obtain adequate negative pressure for liposuction. J Plast Reconstr Aesthet Surg 62:e604–e605PubMedCrossRef Kahveci R, Rehimli M, Esmer A et al (2009) A useful technique to obtain adequate negative pressure for liposuction. J Plast Reconstr Aesthet Surg 62:e604–e605PubMedCrossRef
39.
go back to reference Battagliarin G, Lanna M, Coviello D et al (2009) A randomized study to assess two different techniques of aspiration while performing transabdominal chorionic villus sampling. Ultrasound Obstet Gynecol 33:169–172PubMedCrossRef Battagliarin G, Lanna M, Coviello D et al (2009) A randomized study to assess two different techniques of aspiration while performing transabdominal chorionic villus sampling. Ultrasound Obstet Gynecol 33:169–172PubMedCrossRef
40.
go back to reference Grillo A, Fusaroli P, Naik A et al (2006) Home-made vacuum syringe for endoscopic ultrasound-guided fine-needle aspiration. Endoscopy 38(suppl 2):E56PubMedCrossRef Grillo A, Fusaroli P, Naik A et al (2006) Home-made vacuum syringe for endoscopic ultrasound-guided fine-needle aspiration. Endoscopy 38(suppl 2):E56PubMedCrossRef
41.
go back to reference Freitas Júnior R, Moreira MA, Souza GA et al (2005) Fine-needle aspiration biopsy for breast lesions: a comparison between two devices for obtaining cytological samples. Sao Paulo Med J 123:271–276PubMedCrossRef Freitas Júnior R, Moreira MA, Souza GA et al (2005) Fine-needle aspiration biopsy for breast lesions: a comparison between two devices for obtaining cytological samples. Sao Paulo Med J 123:271–276PubMedCrossRef
42.
go back to reference Orbach D, Schaff E (2004) Which cannulae fit the Ipas manual vacuum aspiration syringe? Contraception 69:171–173PubMedCrossRef Orbach D, Schaff E (2004) Which cannulae fit the Ipas manual vacuum aspiration syringe? Contraception 69:171–173PubMedCrossRef
43.
go back to reference Jaques PF, Mauro MA, Keefe B (1992) US guidance for vascular access. Technical note. J Vasc Interv Radiol 3:427–430PubMedCrossRef Jaques PF, Mauro MA, Keefe B (1992) US guidance for vascular access. Technical note. J Vasc Interv Radiol 3:427–430PubMedCrossRef
44.
go back to reference Ezri T, Khazin V, Houri S et al (2002) Removal of dead space volume from arterial catheter: how much is enough? Pediatr Crit Care Med 3:141–143PubMedCrossRef Ezri T, Khazin V, Houri S et al (2002) Removal of dead space volume from arterial catheter: how much is enough? Pediatr Crit Care Med 3:141–143PubMedCrossRef
45.
go back to reference Gerstein NS, Martin HB, Toma G et al (2009) Introduction of new safety technologies into central venous access. J Clin Anesth 21:363–365PubMedCrossRef Gerstein NS, Martin HB, Toma G et al (2009) Introduction of new safety technologies into central venous access. J Clin Anesth 21:363–365PubMedCrossRef
46.
go back to reference Kreula J, Virkkunen P, Bondestam S (1990) Effect of suction on specimen size in fine-needle aspiration biopsy. Invest Radiol 25:1175–1181PubMedCrossRef Kreula J, Virkkunen P, Bondestam S (1990) Effect of suction on specimen size in fine-needle aspiration biopsy. Invest Radiol 25:1175–1181PubMedCrossRef
47.
go back to reference Sibbitt RR, Sibbitt WL Jr, Palmer DJ, Bankhurst AD (2008) Needle aspiration of peritonsillar abscess with the new safety technology: the reciprocating procedure device. Otolaryngol Head Neck Surg 139:307–309PubMedCrossRef Sibbitt RR, Sibbitt WL Jr, Palmer DJ, Bankhurst AD (2008) Needle aspiration of peritonsillar abscess with the new safety technology: the reciprocating procedure device. Otolaryngol Head Neck Surg 139:307–309PubMedCrossRef
48.
go back to reference Sibbitt RR, Palmer DJ, Sibbitt WL Jr (2008) Introduction of safety technologies into sclerotherapy of varicose veins. Vasc Endovascular Surg 42:446–455PubMedCrossRef Sibbitt RR, Palmer DJ, Sibbitt WL Jr (2008) Introduction of safety technologies into sclerotherapy of varicose veins. Vasc Endovascular Surg 42:446–455PubMedCrossRef
49.
go back to reference Sibbitt RR, Palmer DJ, Sibbitt WL Jr, Bankhurst AD (2009) Integration of new safety technologies into needle aspiration of breast cysts. Arch Gynecol Obstet 279:285–292PubMedCrossRef Sibbitt RR, Palmer DJ, Sibbitt WL Jr, Bankhurst AD (2009) Integration of new safety technologies into needle aspiration of breast cysts. Arch Gynecol Obstet 279:285–292PubMedCrossRef
50.
go back to reference Sibbitt RR, Palmer DJ, Sibbitt WL Jr (2009) Reciprocating procedure device for thyroid cyst aspiration and ablative sclerotherapy. J Laryngol Otol 123:343–345PubMedCrossRef Sibbitt RR, Palmer DJ, Sibbitt WL Jr (2009) Reciprocating procedure device for thyroid cyst aspiration and ablative sclerotherapy. J Laryngol Otol 123:343–345PubMedCrossRef
Metadata
Title
Syringe and Needle Size, Syringe Type, Vacuum Generation, and Needle Control in Aspiration Procedures
Authors
Luke J. Haseler
Randy R. Sibbitt
Wilmer L. Sibbitt Jr.
Adrian A. Michael
Charles M. Gasparovic
Arthur D. Bankhurst
Publication date
01-06-2011
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 3/2011
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-010-0011-z

Other articles of this Issue 3/2011

CardioVascular and Interventional Radiology 3/2011 Go to the issue