01-06-2011 | Technical Note

New Coaxial Transseptal Needle for Creation of Atrial Septal Defects in Adult Sheep

Authors: Barry T. Uchida, Dusan Pavcnik, Masashi Shimohira, Young Ho Choi, Miran Jeromel, Frederick S. Keller, Josef Rösch

Published in: CardioVascular and Interventional Radiology | Issue 3/2011

Login to get access

Abstract

Objectives

To introduce a new transseptal (TS) needle assembled in our laboratory—the coaxial TS (CTS) needle—and describe our experience with it in creating experimental atrial septal defects (ASD) in adult sheep.

Background

With commercially available TS needles, we were not able to consistently perform TS puncture at the fossa ovalis in adult sheep.

Material and Methods

Ten adult sheep with a mean weight of 63.5 kg were used. The CTS needle consists of four components: a 9F Teflon catheter, a 14-gauge blunt curved-tip metal cannula, a 4F tapered catheter, and a 20-gauge open needle. A transjugular 5F pigtail catheter was used to display the septal anatomy by angiocardiography and was left in place to mark the level of the fossa ovalis. The septum was then probed by a transfemoral 5F curved-tip end-hole catheter. The CTS needle was aligned with the tip of the transjugular catheter, and the TS puncture was performed under fluoroscopic guidance. After documenting a left atrial position, a balloon angioplasty catheter was used for creation of the ASD.

Results

A small patent foramen ovale was discovered by septal probing in one sheep. All sheep underwent successful TS punctures without complications. The ASD size ranged from 13 to 15 mm. In eight sheep, the ASD was in fossa ovalis. In the first two sheep where the needle was not well aligned with the marking catheter, the ASD was in the septum secundum. No damage to the atrial or other heart structures was found at necropsy.

Conclusion

The CTS needle is a suitable needle for TS puncture and ASD creation in adult sheep. Proper alignment of the CTS needle with a catheter marking the fossa ovalis is essential for successful puncture.
Literature
1.
go back to reference Ross J Jr (1959) Transeptal left heart catheterization: a new method of left atrial puncture. Ann Surg 149:395–401PubMedCrossRef Ross J Jr (1959) Transeptal left heart catheterization: a new method of left atrial puncture. Ann Surg 149:395–401PubMedCrossRef
2.
go back to reference Ross J Jr (2008) Transseptal left heart catheterization: a 50-year odyssey. J Am Coll Cardiol 51:2107–2115PubMedCrossRef Ross J Jr (2008) Transseptal left heart catheterization: a 50-year odyssey. J Am Coll Cardiol 51:2107–2115PubMedCrossRef
3.
go back to reference Roelke M, Smith AJC, Palacios IF (1994) The technique and safety of transseptal left heart catheterization: the Massachusetts General Hospital experience with 1,279 procedures. Cathet Cardiovasc Diagnosis 32:332–339CrossRef Roelke M, Smith AJC, Palacios IF (1994) The technique and safety of transseptal left heart catheterization: the Massachusetts General Hospital experience with 1,279 procedures. Cathet Cardiovasc Diagnosis 32:332–339CrossRef
4.
go back to reference Babaliaros VC, Green JT, Lerakis S et al (2008) Emerging applications for transseptal left heart catheterization: old technique for new procedures. J Am Coll Cardiol 51:2116–2122PubMedCrossRef Babaliaros VC, Green JT, Lerakis S et al (2008) Emerging applications for transseptal left heart catheterization: old technique for new procedures. J Am Coll Cardiol 51:2116–2122PubMedCrossRef
5.
go back to reference Lock JE, Rome JJ, Davis R et al (1989) Transcatheter closure of atrial septal defects. Experimental studies. Circulation 79:1091–1099PubMed Lock JE, Rome JJ, Davis R et al (1989) Transcatheter closure of atrial septal defects. Experimental studies. Circulation 79:1091–1099PubMed
6.
go back to reference Jux C, Bertram H, Wohlsein P et al (2002) Experimental ASD closure using autologous cell-seeded interventional closure devices. Cardiovasc Res 53:181–191PubMedCrossRef Jux C, Bertram H, Wohlsein P et al (2002) Experimental ASD closure using autologous cell-seeded interventional closure devices. Cardiovasc Res 53:181–191PubMedCrossRef
7.
go back to reference Jux C, Bertram H, Wohlsein P et al (2006) Interventional atrial septal defect closure using a totally bioresorbable occluder matrix. J Am Coll Cardiol 48:161–169PubMedCrossRef Jux C, Bertram H, Wohlsein P et al (2006) Interventional atrial septal defect closure using a totally bioresorbable occluder matrix. J Am Coll Cardiol 48:161–169PubMedCrossRef
8.
go back to reference Thomsen AB, Schneider M, Baandrup U et al (1998) Animal experimental implantation of an atrial septal defect occluder system. Heart 80:606–611 Thomsen AB, Schneider M, Baandrup U et al (1998) Animal experimental implantation of an atrial septal defect occluder system. Heart 80:606–611
9.
go back to reference Kuhn MA, Latson LA, Cheatham JP et al (1996) Biological response to Bard Clamshell septal occluders in the canine heart. Circulation 93:1459–1463PubMed Kuhn MA, Latson LA, Cheatham JP et al (1996) Biological response to Bard Clamshell septal occluders in the canine heart. Circulation 93:1459–1463PubMed
10.
go back to reference Naqvi TZ, Zarbatany D, Molloy MD et al (2006) Intracardiac echocardiography for percutaneous mitral valve repair in a swine model. J Am Soc Echocardiogr 19:147–153PubMedCrossRef Naqvi TZ, Zarbatany D, Molloy MD et al (2006) Intracardiac echocardiography for percutaneous mitral valve repair in a swine model. J Am Soc Echocardiogr 19:147–153PubMedCrossRef
11.
go back to reference Naqvi TZ, Buchbinder M, Zarbatany D et al (2007) Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model. Catheter Cardiovasc Interv 69:525–531PubMedCrossRef Naqvi TZ, Buchbinder M, Zarbatany D et al (2007) Beating-heart percutaneous mitral valve repair using a transcatheter endovascular suturing device in an animal model. Catheter Cardiovasc Interv 69:525–531PubMedCrossRef
12.
go back to reference Petersen B, Uchida BT, Timmermans H et al (2001) Intravascular US-guided direct intrahepatic portacaval shunt with a PTFE-covered stent-graft: feasibility study in swine and initial clinical results. J Vasc Interv Radiol 12:475–486PubMedCrossRef Petersen B, Uchida BT, Timmermans H et al (2001) Intravascular US-guided direct intrahepatic portacaval shunt with a PTFE-covered stent-graft: feasibility study in swine and initial clinical results. J Vasc Interv Radiol 12:475–486PubMedCrossRef
13.
go back to reference Ni RF, Kranokpiraksa P, Pavcnik D et al (2009) Testing percutaneous arterial closure devices: an animal model. Cardiovasc Interv Radiol 32:313–316CrossRef Ni RF, Kranokpiraksa P, Pavcnik D et al (2009) Testing percutaneous arterial closure devices: an animal model. Cardiovasc Interv Radiol 32:313–316CrossRef
14.
go back to reference Narayanaswamy M, Wright KC, Kandarpa K (2000) Animal models for atherosclerosis, restenosis and endovascular graft research. J Vasc Interv Radiol 11:5–17PubMedCrossRef Narayanaswamy M, Wright KC, Kandarpa K (2000) Animal models for atherosclerosis, restenosis and endovascular graft research. J Vasc Interv Radiol 11:5–17PubMedCrossRef
Metadata
Title
New Coaxial Transseptal Needle for Creation of Atrial Septal Defects in Adult Sheep
Authors
Barry T. Uchida
Dusan Pavcnik
Masashi Shimohira
Young Ho Choi
Miran Jeromel
Frederick S. Keller
Josef Rösch
Publication date
01-06-2011
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 3/2011
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-010-9948-1

Other articles of this Issue 3/2011

CardioVascular and Interventional Radiology 3/2011 Go to the issue