Published in: BMC Musculoskeletal Disorders 1/2022

Open Access 01-12-2022 | Spinal Stenosis | Research

Biomechanical comparison of different interspinous process devices in the treatment of lumbar spinal stenosis: a finite element analysis

Authors: Zhengpeng Liu, Shuyi Zhang, Jia Li, Hai Tang

Published in: BMC Musculoskeletal Disorders | Issue 1/2022

Login to get access

Abstract

Background

Lumbar spinal stenosis (LSS) is a common disease among elderly individuals, and surgery is an effective treatment. The development of minimally invasive surgical techniques, such as the lumbar interspinous process device (IPD), has provided patients with more surgical options.

Objective

To investigate the biomechanical properties of different IPDs, including BacFuse, X-Stop and Coflex, in the treatment of LSS.

Methods

Based on the computed tomography images of a patient with LSS, four finite element (FE) models of L3-S5 were created in this study. The FE models included a surgical model of the intact lumbar spine and surgical models of the lumbar IPDs BacFuse, X-Stop, and Coflex. After validating the models, they were simulated for four physiological motions: flexion, extension, lateral bending and axial rotation, and range of motion (ROM). Stress distribution of discs and facet joints in each segment, stress distribution of the spinous process in the operated section, and stress distribution of the internal fixation were compared and analysed.

Results

Compared to the model of the intact lumbar spine, the other three models showed a decrease in ROM and disc and facet joint stresses in the surgical segment during movement and an increase in ROM and disc and facet joint stresses in the adjacent segments. These effects were greater for the proximal adjacent segment with BacFuse and more pronounced for the distal adjacent segment with Coflex, while X-Stop had the greatest stress effect on the spinous process in the surgical segment.

Conclusion

BacFuse, Coflex and X-Stop could all be implemented to effectively reduce extension and disc and facet joint stresses, but they also increase the ROM and disc and facet joint stresses in adjacent segments, which may cause degeneration.
Literature
1.
go back to reference Tram J, Srinivas S, Wali AR, Lewis CS, Pham MH. Decompression Surgery versus Interspinous Devices for Lumbar Spinal Stenosis: A Systematic Review of the Literature. Asian Spine J. 2020;14(4):526–42.CrossRef Tram J, Srinivas S, Wali AR, Lewis CS, Pham MH. Decompression Surgery versus Interspinous Devices for Lumbar Spinal Stenosis: A Systematic Review of the Literature. Asian Spine J. 2020;14(4):526–42.CrossRef
2.
go back to reference Kong DS, Kim ES, Eoh W. One-year outcome evaluation after interspinous implantation for degenerative spinal stenosis with segmental instability. J Korean Med Sci. 2007;22(2):330–5.CrossRef Kong DS, Kim ES, Eoh W. One-year outcome evaluation after interspinous implantation for degenerative spinal stenosis with segmental instability. J Korean Med Sci. 2007;22(2):330–5.CrossRef
3.
go back to reference Moojen WA, Arts MP, Jacobs WC, van Zwet EW, van den Akker-van Marle ME, Koes BW, et al. Interspinous process device versus standard conventional surgical decompression for lumbar spinal stenosis: randomized controlled trial. BMJ. 2013;347:f6415.CrossRef Moojen WA, Arts MP, Jacobs WC, van Zwet EW, van den Akker-van Marle ME, Koes BW, et al. Interspinous process device versus standard conventional surgical decompression for lumbar spinal stenosis: randomized controlled trial. BMJ. 2013;347:f6415.CrossRef
4.
go back to reference Lønne G, Johnsen LG, Aas E, Lydersen S, Andresen H, Rønning R, et al. Comparing cost-effectiveness of X-Stop with minimally invasive decompression in lumbar spinal stenosis: a randomized controlled trial. Spine (Phila Pa 1976). 2015;40(8):514–20.CrossRef Lønne G, Johnsen LG, Aas E, Lydersen S, Andresen H, Rønning R, et al. Comparing cost-effectiveness of X-Stop with minimally invasive decompression in lumbar spinal stenosis: a randomized controlled trial. Spine (Phila Pa 1976). 2015;40(8):514–20.CrossRef
5.
go back to reference Mo Z, Li D, Zhang R, Chang M, Yang B, Tang S. Comparative effectiveness and safety of posterior lumbar interbody fusion, Coflex, Wallis, and X-stop for lumbar degenerative diseases: A systematic review and network meta-analysis. Clin Neurol Neurosurg. 2018;172:74–81.CrossRef Mo Z, Li D, Zhang R, Chang M, Yang B, Tang S. Comparative effectiveness and safety of posterior lumbar interbody fusion, Coflex, Wallis, and X-stop for lumbar degenerative diseases: A systematic review and network meta-analysis. Clin Neurol Neurosurg. 2018;172:74–81.CrossRef
6.
go back to reference Zhao XW, Ma JX, Ma XL, Li F, He WW, Jiang X, et al. Interspinous process devices (IPD) alone versus decompression surgery for lumbar spinal stenosis (LSS): A systematic review and meta-analysis of randomized controlled trials. Int J Surg. 2017;39:57–64.CrossRef Zhao XW, Ma JX, Ma XL, Li F, He WW, Jiang X, et al. Interspinous process devices (IPD) alone versus decompression surgery for lumbar spinal stenosis (LSS): A systematic review and meta-analysis of randomized controlled trials. Int J Surg. 2017;39:57–64.CrossRef
7.
go back to reference Mikhail CM, Dowdell JE, Hecht AC. Adult lumbar spondylolisthesis: Interspinous devices. Semin Spine Surg. 2020;32(3):100806.CrossRef Mikhail CM, Dowdell JE, Hecht AC. Adult lumbar spondylolisthesis: Interspinous devices. Semin Spine Surg. 2020;32(3):100806.CrossRef
8.
go back to reference Yan ZG, Wang YL, Shi JF, Yang LL, Liu Z. Clinical observation of transforaminal endoscopy combined with interspinous fusion in the treatment of lumbar spinal stenosis with instability in the elderly. Zhongguo Gu Shang. 2021;34(3):249–54.PubMed Yan ZG, Wang YL, Shi JF, Yang LL, Liu Z. Clinical observation of transforaminal endoscopy combined with interspinous fusion in the treatment of lumbar spinal stenosis with instability in the elderly. Zhongguo Gu Shang. 2021;34(3):249–54.PubMed
9.
go back to reference Chen M, Tang H, Shan J, Chen H, Jia P, Bao L, et al. A new interspinous process distraction device BacFuse in the treatment of lumbar spinal stenosis with 5 years follow-up study. Medicine (Baltimore). 2020;99(26):e20925.CrossRef Chen M, Tang H, Shan J, Chen H, Jia P, Bao L, et al. A new interspinous process distraction device BacFuse in the treatment of lumbar spinal stenosis with 5 years follow-up study. Medicine (Baltimore). 2020;99(26):e20925.CrossRef
10.
go back to reference Choi J, Shin DA, Kim S. Biomechanical Effects of the Geometry of Ball-and-Socket Artificial Disc on Lumbar Spine: A Finite Element Study. Spine (Phila Pa 1976). 2017;42(6):E332–9.CrossRef Choi J, Shin DA, Kim S. Biomechanical Effects of the Geometry of Ball-and-Socket Artificial Disc on Lumbar Spine: A Finite Element Study. Spine (Phila Pa 1976). 2017;42(6):E332–9.CrossRef
11.
go back to reference Guo TM, Lu J, Xing YL, Liu GX, Zhu HY, Yang L, et al. A 3-Dimensional Finite Element Analysis of Adjacent Segment Disk Degeneration Induced by Transforaminal Lumbar Interbody Fusion After Pedicle Screw Fixation. World Neurosurg. 2018;S1878-8750(18):32754–2. Guo TM, Lu J, Xing YL, Liu GX, Zhu HY, Yang L, et al. A 3-Dimensional Finite Element Analysis of Adjacent Segment Disk Degeneration Induced by Transforaminal Lumbar Interbody Fusion After Pedicle Screw Fixation. World Neurosurg. 2018;S1878-8750(18):32754–2.
12.
go back to reference Zhang Z, Fogel GR, Liao Z, Sun Y, Liu W. Biomechanical Analysis of Lateral Lumbar Interbody Fusion Constructs with Various Fixation Options: Based on a Validated Finite Element Model. World Neurosurg. 2018;114:e1120–9.CrossRef Zhang Z, Fogel GR, Liao Z, Sun Y, Liu W. Biomechanical Analysis of Lateral Lumbar Interbody Fusion Constructs with Various Fixation Options: Based on a Validated Finite Element Model. World Neurosurg. 2018;114:e1120–9.CrossRef
13.
go back to reference Zhao X, Du L, Xie Y, Zhao J. Effect of Lumbar Lordosis on the Adjacent Segment in Transforaminal Lumbar Interbody Fusion: A Finite Element Analysis. World Neurosurg. 2018;114:e114–20.CrossRef Zhao X, Du L, Xie Y, Zhao J. Effect of Lumbar Lordosis on the Adjacent Segment in Transforaminal Lumbar Interbody Fusion: A Finite Element Analysis. World Neurosurg. 2018;114:e114–20.CrossRef
14.
go back to reference Najarian S, Dargahi J, Heidari B. Biomechanical effect of posterior elements and ligamentous tissues of lumbar spine on load sharing. Biomed Mater Eng. 2005;15(3):145–58.PubMed Najarian S, Dargahi J, Heidari B. Biomechanical effect of posterior elements and ligamentous tissues of lumbar spine on load sharing. Biomed Mater Eng. 2005;15(3):145–58.PubMed
15.
go back to reference Fan W, Guo LX. A comparison of the influence of three different lumbar interbody fusion approaches on stress in the pedicle screw fixation system: Finite element static and vibration analyses. Int J Numer Method Biomed Eng. 2019;35(3):e3162.CrossRef Fan W, Guo LX. A comparison of the influence of three different lumbar interbody fusion approaches on stress in the pedicle screw fixation system: Finite element static and vibration analyses. Int J Numer Method Biomed Eng. 2019;35(3):e3162.CrossRef
16.
go back to reference Lo CC, Tsai KJ, Chen SH, Zhong ZC, Hung C. Biomechanical effect after Coflex and Coflex rivet implantation for segmental instability at surgical and adjacent segments: a finite element analysis. Comput Methods Biomech Biomed Engin. 2011;14(11):969–78.CrossRef Lo CC, Tsai KJ, Chen SH, Zhong ZC, Hung C. Biomechanical effect after Coflex and Coflex rivet implantation for segmental instability at surgical and adjacent segments: a finite element analysis. Comput Methods Biomech Biomed Engin. 2011;14(11):969–78.CrossRef
17.
go back to reference Panjabi MM. Hybrid multidirectional test method to evaluate spinal adjacent-level effects. Clin Biomech (Bristol, Avon). 2007;22(3):257–65.CrossRef Panjabi MM. Hybrid multidirectional test method to evaluate spinal adjacent-level effects. Clin Biomech (Bristol, Avon). 2007;22(3):257–65.CrossRef
18.
go back to reference Chen HC, Wu JL, Huang SC, Zhong ZC, Chiu SL, Lai YS, et al. Biomechanical evaluation of a novel pedicle screw-based interspinous spacer: A finite element analysis. Med Eng Phys. 2017;46:27–32.CrossRef Chen HC, Wu JL, Huang SC, Zhong ZC, Chiu SL, Lai YS, et al. Biomechanical evaluation of a novel pedicle screw-based interspinous spacer: A finite element analysis. Med Eng Phys. 2017;46:27–32.CrossRef
19.
go back to reference Byun DH, Shin DA, Kim JM, Kim SH, Kim HI. Finite element analysis of the biomechanical effect of coflex™ on the lumbar spine. Korean J Spine. 2012;9(3):131–6.CrossRef Byun DH, Shin DA, Kim JM, Kim SH, Kim HI. Finite element analysis of the biomechanical effect of coflex™ on the lumbar spine. Korean J Spine. 2012;9(3):131–6.CrossRef
20.
go back to reference Lo HJ, Chen CS, Chen HM, Yang SW. Application of an interspinous process device after minimally invasive lumbar decompression could lead to stress redistribution at the pars interarticularis: a finite element analysis. BMC Musculoskelet Disord. 2019;20(1):213.CrossRef Lo HJ, Chen CS, Chen HM, Yang SW. Application of an interspinous process device after minimally invasive lumbar decompression could lead to stress redistribution at the pars interarticularis: a finite element analysis. BMC Musculoskelet Disord. 2019;20(1):213.CrossRef
21.
go back to reference Lo HJ, Chen HM, Kuo YJ, Yang SW. Effect of different designs of interspinous process devices on the instrumented and adjacent levels after double-level lumbar decompression surgery: A finite element analysis. PLoS ONE. 2020;15(12):e0244571.CrossRef Lo HJ, Chen HM, Kuo YJ, Yang SW. Effect of different designs of interspinous process devices on the instrumented and adjacent levels after double-level lumbar decompression surgery: A finite element analysis. PLoS ONE. 2020;15(12):e0244571.CrossRef
22.
go back to reference Whitesides TE. The effect of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976). 2003;28(16):1906–7. (author reply 1907-8).CrossRef Whitesides TE. The effect of an interspinous implant on intervertebral disc pressures. Spine (Phila Pa 1976). 2003;28(16):1906–7. (author reply 1907-8).CrossRef
23.
go back to reference Tsai KJ, Murakami H, Lowery GL, Hutton WC. A biomechanical evaluation of an interspinous device (Coflex) used to stabilize the lumbar spine. J Surg Orthop Adv. 2006;15(3):167–72.PubMed Tsai KJ, Murakami H, Lowery GL, Hutton WC. A biomechanical evaluation of an interspinous device (Coflex) used to stabilize the lumbar spine. J Surg Orthop Adv. 2006;15(3):167–72.PubMed
24.
go back to reference Chow DH, Luk KD, Evans JH, Leong JC. Effects of short anterior lumbar interbody fusion on biomechanics of neighboring unfused segments. Spine (Phila Pa 1976). 1996;21(5):549–55.CrossRef Chow DH, Luk KD, Evans JH, Leong JC. Effects of short anterior lumbar interbody fusion on biomechanics of neighboring unfused segments. Spine (Phila Pa 1976). 1996;21(5):549–55.CrossRef
25.
go back to reference Sim HB, Murovic JA, Cho BY, Lim TJ, Park J. Biomechanical comparison of single-level posterior versus transforaminal lumbar interbody fusions with bilateral pedicle screw fixation: segmental stability and the effects on adjacent motion segments. J Neurosurg Spine. 2010;12(6):700–8.CrossRef Sim HB, Murovic JA, Cho BY, Lim TJ, Park J. Biomechanical comparison of single-level posterior versus transforaminal lumbar interbody fusions with bilateral pedicle screw fixation: segmental stability and the effects on adjacent motion segments. J Neurosurg Spine. 2010;12(6):700–8.CrossRef
26.
go back to reference Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA. The effect of an interspinous process implant on facet loading during extension. Spine (Phila Pa 1976). 2005;30(8):903–7.CrossRef Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA. The effect of an interspinous process implant on facet loading during extension. Spine (Phila Pa 1976). 2005;30(8):903–7.CrossRef
27.
go back to reference Poetscher AW, Gentil AF, Ferretti M, Lenza M. Interspinous process devices for treatment of degenerative lumbar spine stenosis: A systematic review and meta-analysis. PLoS ONE. 2018;13(7):e0199623.CrossRef Poetscher AW, Gentil AF, Ferretti M, Lenza M. Interspinous process devices for treatment of degenerative lumbar spine stenosis: A systematic review and meta-analysis. PLoS ONE. 2018;13(7):e0199623.CrossRef
28.
go back to reference Shepherd DE, Leahy JC, Mathias KJ, Wilkinson SJ, Hukins DW. Spinous process strength. Spine (Phila Pa 1976). 2000;25(3):319–23.CrossRef Shepherd DE, Leahy JC, Mathias KJ, Wilkinson SJ, Hukins DW. Spinous process strength. Spine (Phila Pa 1976). 2000;25(3):319–23.CrossRef
29.
go back to reference Chen CS, Shih SL. Biomechanical analysis of a new lumbar interspinous device with optimized topology. Med Biol Eng Comput. 2018;56(8):1333–41.CrossRef Chen CS, Shih SL. Biomechanical analysis of a new lumbar interspinous device with optimized topology. Med Biol Eng Comput. 2018;56(8):1333–41.CrossRef
30.
go back to reference Guo LX, Yin JY. Finite element analysis and design of an interspinous device using topology optimization. Med Biol Eng Comput. 2019;57(1):89–98.CrossRef Guo LX, Yin JY. Finite element analysis and design of an interspinous device using topology optimization. Med Biol Eng Comput. 2019;57(1):89–98.CrossRef
31.
go back to reference Aggarwal N, Chow R. Real world adverse events of interspinous spacers using Manufacturer and User Facility Device Experience data. Anesth Pain Med (Seoul). 2021;16(2):177–83.CrossRef Aggarwal N, Chow R. Real world adverse events of interspinous spacers using Manufacturer and User Facility Device Experience data. Anesth Pain Med (Seoul). 2021;16(2):177–83.CrossRef
32.
go back to reference Gazzeri R, Galarza M, Neroni M, Fiore C, Faiola A, Puzzilli F, et al. Failure rates and complications of interspinous process decompression devices: a European multicenter study. Neurosurg Focus. 2015;39(4):E14.CrossRef Gazzeri R, Galarza M, Neroni M, Fiore C, Faiola A, Puzzilli F, et al. Failure rates and complications of interspinous process decompression devices: a European multicenter study. Neurosurg Focus. 2015;39(4):E14.CrossRef
33.
go back to reference Verhoof OJ, Bron JL, Wapstra FH, van Royen BJ. High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J. 2008;17(2):188–92.CrossRef Verhoof OJ, Bron JL, Wapstra FH, van Royen BJ. High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis. Eur Spine J. 2008;17(2):188–92.CrossRef
34.
go back to reference Bowers C, Amini A, Dailey AT, Schmidt MH. Dynamic interspinous process stabilization: review of complications associated with the X-Stop device. Neurosurg Focus. 2010;28(6):E8.CrossRef Bowers C, Amini A, Dailey AT, Schmidt MH. Dynamic interspinous process stabilization: review of complications associated with the X-Stop device. Neurosurg Focus. 2010;28(6):E8.CrossRef
35.
go back to reference Fan W, Guo LX. The Role of Posterior Screw Fixation in Single-Level Transforaminal Lumbar Interbody Fusion During Whole Body Vibration: A Finite Element Study. World Neurosurg. 2018;114:e1086–93.CrossRef Fan W, Guo LX. The Role of Posterior Screw Fixation in Single-Level Transforaminal Lumbar Interbody Fusion During Whole Body Vibration: A Finite Element Study. World Neurosurg. 2018;114:e1086–93.CrossRef
Metadata
Title
Biomechanical comparison of different interspinous process devices in the treatment of lumbar spinal stenosis: a finite element analysis
Authors
Zhengpeng Liu
Shuyi Zhang
Jia Li
Hai Tang
Publication date
01-12-2022
Publisher
BioMed Central
Keyword
Spinal Stenosis
Published in
BMC Musculoskeletal Disorders / Issue 1/2022
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-022-05543-y

Other articles of this Issue 1/2022

BMC Musculoskeletal Disorders 1/2022 Go to the issue