Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening

Authors: Peem Lorvidhaya, Ivan Mendoza, Sharmila Sehli, Michael K. Atalay, Michael H. Kim

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2013

Login to get access

Abstract

Introduction

Lead insulation defects with externalization of the conductors exist in Riata defibrillator leads. Cinefluoroscopy is currently the gold standard to detect such defects. Prospective evaluation of alternative screening options such as chest radiography (CXR), which has been recommended by the FDA, is not well described.

Methods and results

Patients with Riata leads underwent cinefluoroscopy, CXR, and device interrogation. Leads were classified as abnormal (clear cable separation), borderline, or normal by independent evaluation of cinefluoroscopy and CXR. CXR evaluation was done in two ways as follows: (1) routine CXR read by daily staff radiologists for lead screening and (2) CXR evaluation by a radiologist educated about the lead defect. One hundred two patients were evaluated at our institution. Cinefluoroscopy showed externalized conductors in 33 patients (32 %). Twenty-five of 33 patients (76 %) who had abnormal cinefluoroscopic findings had abnormal CXR findings on blinded review by the educated radiologist. All 25 patients with abnormal CXR had abnormal findings on cinefluoroscopy. Daily staff radiologists without direct education other than prompts for lead screening detected CXR abnormalities in only 8 out of 102 (8 %) cases.

Conclusion

Cinefluoroscopy appears to be more sensitive than CXR for the detection of Riata cable extrusion. Interpretation of CXR by a radiologist with education in lead defects correlates highly with cinefluoroscopy with very high specificity. Depending on available resources for screening, CXR may be a reasonable alternative to cinefluoroscopy. Multidisciplinary collaboration across specialties (radiology and electrophysiology) can lead to improved diagnostic capability and thus the potential for enhanced quality of care.
Literature
1.
go back to reference Duray, G. Z., Israel, C. W., Schmitt, J., & Hohnloser, S. H. (2008). Implantable cardioverter–defibrillator lead disintegration at the level of the tricuspid valve. Heart Rhythm, 5(8), 1224–1225.PubMedCrossRef Duray, G. Z., Israel, C. W., Schmitt, J., & Hohnloser, S. H. (2008). Implantable cardioverter–defibrillator lead disintegration at the level of the tricuspid valve. Heart Rhythm, 5(8), 1224–1225.PubMedCrossRef
2.
go back to reference Jalal, Z., Derval, N., Ploux, S., & Bordachar, P. (2010). Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead. Heart Rhythm, 7(8), 1166–1167.PubMedCrossRef Jalal, Z., Derval, N., Ploux, S., & Bordachar, P. (2010). Unusual failure of a multilumen, small-diameter implantable cardioverter-defibrillator lead. Heart Rhythm, 7(8), 1166–1167.PubMedCrossRef
3.
go back to reference Richards, M. W., Warren, C. E., & Anderson, M. H. (2010). Late failure of a single-coil transvenous implantable cardioverter–defibrillator lead associated with conductor separation. Europace, 12(8), 1191–1192.PubMedCrossRef Richards, M. W., Warren, C. E., & Anderson, M. H. (2010). Late failure of a single-coil transvenous implantable cardioverter–defibrillator lead associated with conductor separation. Europace, 12(8), 1191–1192.PubMedCrossRef
4.
go back to reference Valk, S. D., Theuns, D. A., & Jordaens, L. (2013). Long-term performance of the St Jude Riata 1580–1582 ICD lead family. Neth Heart J, 21(3), 127–134.PubMedCrossRef Valk, S. D., Theuns, D. A., & Jordaens, L. (2013). Long-term performance of the St Jude Riata 1580–1582 ICD lead family. Neth Heart J, 21(3), 127–134.PubMedCrossRef
5.
go back to reference Chan, C. W., & Chiang, C. S. (2012). An ICD lead with failure of outer insulation goes undetected by regular measurements. Pacing and Clinical Electrophysiology, 35(9), e261–262.PubMedCrossRef Chan, C. W., & Chiang, C. S. (2012). An ICD lead with failure of outer insulation goes undetected by regular measurements. Pacing and Clinical Electrophysiology, 35(9), e261–262.PubMedCrossRef
6.
go back to reference Erkapic, D., Duray, G. Z., Bauernfeind, T., De Rosa, S., & Hohnloser, S. H. (2011). Insulation defects of thin high-voltage ICD leads: an underestimated problem? Journal of Cardiovascular Electrophysiology, 22(9), 1018–1022.PubMedCrossRef Erkapic, D., Duray, G. Z., Bauernfeind, T., De Rosa, S., & Hohnloser, S. H. (2011). Insulation defects of thin high-voltage ICD leads: an underestimated problem? Journal of Cardiovascular Electrophysiology, 22(9), 1018–1022.PubMedCrossRef
7.
go back to reference Krebsbach, A., Alhumaid, F., Henrikson, C. A., Calkins, H., Berger, R. D., & Cheng, A. (2011). Premature failure of a Riata defibrillator lead without impedance change or inappropriate sensing: a case report and review of the literature. Journal of Cardiovascular Electrophysiology, 22(9), 1070–1072.PubMedCrossRef Krebsbach, A., Alhumaid, F., Henrikson, C. A., Calkins, H., Berger, R. D., & Cheng, A. (2011). Premature failure of a Riata defibrillator lead without impedance change or inappropriate sensing: a case report and review of the literature. Journal of Cardiovascular Electrophysiology, 22(9), 1070–1072.PubMedCrossRef
8.
go back to reference Porterfield, J. G., Porterfield, L. M., Kuck, K. H., Corbisiero, R., Greenberg, S. M., Hindricks, G., et al. (2010). Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. Journal of Cardiovascular Electrophysiology, 21(5), 551–556.PubMedCrossRef Porterfield, J. G., Porterfield, L. M., Kuck, K. H., Corbisiero, R., Greenberg, S. M., Hindricks, G., et al. (2010). Clinical performance of the St. Jude Medical Riata defibrillation lead in a large patient population. Journal of Cardiovascular Electrophysiology, 21(5), 551–556.PubMedCrossRef
9.
go back to reference Schmutz, M., Delacrétaz, E., Schwick, N., Roten, L., Fuhrer, J., Boesch, C., et al. (2012). Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads. International Journal of Cardiology. Schmutz, M., Delacrétaz, E., Schwick, N., Roten, L., Fuhrer, J., Boesch, C., et al. (2012). Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata® and Riata® ST implantable cardioverter-defibrillator leads. International Journal of Cardiology.
10.
go back to reference Shen, S., Bhave, P., Giedrimas, E., Patel, T., Arora, R., Chicos, A. B., et al. (2012). Prevalence and predictors of cable extrusion and loss of electrical integrity with the Riata defibrillator lead. Journal of Cardiovascular Electrophysiology, 23(11), 1207–1212.PubMedCrossRef Shen, S., Bhave, P., Giedrimas, E., Patel, T., Arora, R., Chicos, A. B., et al. (2012). Prevalence and predictors of cable extrusion and loss of electrical integrity with the Riata defibrillator lead. Journal of Cardiovascular Electrophysiology, 23(11), 1207–1212.PubMedCrossRef
11.
go back to reference Liu, J., Rattan, R., Adelstein, E., Barrington, W., Bazaz, R., Brode, S., et al. (2012). Fluoroscopic screening of asymptomatic patients implanted with the recalled Riata lead family. Circulation. Arrhythmia and Electrophysiology, 5(4), 809–814.PubMedCrossRef Liu, J., Rattan, R., Adelstein, E., Barrington, W., Bazaz, R., Brode, S., et al. (2012). Fluoroscopic screening of asymptomatic patients implanted with the recalled Riata lead family. Circulation. Arrhythmia and Electrophysiology, 5(4), 809–814.PubMedCrossRef
12.
go back to reference Parvathaneni, S. V., Ellis, C. R., & Rottman, J. N. (2012). High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm, 9(8), 1218–1224.PubMedCrossRef Parvathaneni, S. V., Ellis, C. R., & Rottman, J. N. (2012). High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm, 9(8), 1218–1224.PubMedCrossRef
13.
go back to reference Larsen, J. M., Riahi, S., Nielsen, J. C., Videbaek, R., Haarbo, J., Due, K. M., et al. (2013). Nationwide fluoroscopic screening of recalled Riata defibrillator leads in Denmark. Heart Rhythm. Larsen, J. M., Riahi, S., Nielsen, J. C., Videbaek, R., Haarbo, J., Due, K. M., et al. (2013). Nationwide fluoroscopic screening of recalled Riata defibrillator leads in Denmark. Heart Rhythm.
14.
go back to reference Cronin, E. M., Baranowski, B. J., & Martin, D. O. (2012). Failure of fluoroscopy to detect “inside-out” insulation failure and externalized conductors in a Riata ICD lead. Heart Rhythm. Cronin, E. M., Baranowski, B. J., & Martin, D. O. (2012). Failure of fluoroscopy to detect “inside-out” insulation failure and externalized conductors in a Riata ICD lead. Heart Rhythm.
15.
go back to reference Steinberg, C., Sarrazin, J. F., Philippon, F., Bouchard, M. A., O’Hara, G., Molin, F., et al. (2013). Detection of high incidence of Riata lead breaches by systematic posteroanterior and lateral chest X-ray in a large cohort. Europace, 15(3), 402–408.PubMedCrossRef Steinberg, C., Sarrazin, J. F., Philippon, F., Bouchard, M. A., O’Hara, G., Molin, F., et al. (2013). Detection of high incidence of Riata lead breaches by systematic posteroanterior and lateral chest X-ray in a large cohort. Europace, 15(3), 402–408.PubMedCrossRef
16.
go back to reference Pachon, M., Arias, M. A., Nadal, R., & Jimenez-Lopez, J. (2013). Standard x-ray is inadequate for the identification of conductor cable externalization in Riata leads. Journal of Cardiovascular Electrophysiology, 24(4), 471.PubMedCrossRef Pachon, M., Arias, M. A., Nadal, R., & Jimenez-Lopez, J. (2013). Standard x-ray is inadequate for the identification of conductor cable externalization in Riata leads. Journal of Cardiovascular Electrophysiology, 24(4), 471.PubMedCrossRef
Metadata
Title
Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening
Authors
Peem Lorvidhaya
Ivan Mendoza
Sharmila Sehli
Michael K. Atalay
Michael H. Kim
Publication date
01-11-2013
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2013
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-013-9822-6

Other articles of this Issue 2/2013

Journal of Interventional Cardiac Electrophysiology 2/2013 Go to the issue