Effectiveness of VF induction with DC fibber versus conventional induction methods in patients on chronic amiodarone therapy

Authors: Jason C. Rubenstein, Manish S. Gupta, Michael H. Kim

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2013

Login to get access

Abstract

Background

Amiodarone therapy, especially chronic, can result in difficult ventricular fibrillation (VF) inductions during implantable cardioverter defibrillator (ICD) testing. The efficacy of various VF induction methods on patients treated with amiodarone has not been well described. This prospective analysis evaluated the impact of direct current (DC) fibber, burst fibber, and synchronized T-wave shock VF induction methods.

Methods

Data were collected from one study and two enrolling centers totaling 14 ICD patients (92.9 % male, age 64.3 years (range 23.6–81.2)). A minimum of two successful VF inductions from each patient were required to be included in the study analysis. Each VF induction method was attempted in each patient. Non-sustained VF resulting in an aborted shock was considered an induction failure. All 14 patients were on chronic amiodarone therapy.

Results

From a total of 42 attempted inductions, 25 (59.5 %) were successful at inducing VF. The success rates of effective induction of VF during the first attempt using DC, burst, and Shock-on-T are 100, 50, and 28.6 %, respectively. DC fibber had significantly higher VF induction rates than Shock-on-T (p = 0.004) or burst fibber (p = 0.02). There was no difference between Shock-on-T and burst fibber (p = 0.45).

Conclusion

The success rate of VF induction using DC fibber method is significantly higher than either Shock-on-T or burst fibber induction methods for patients on chronic amiodarone therapy. This may facilitate defibrillation threshold testing in such patients.
Literature
1.
go back to reference Bhandari, A. K., Isber, N., Estioko, M., et al. (1998). Efficacy of low-energy T wave shocks for induction of ventricular fibrillation in patients with implantable cardioverter defibrillators. Journal of Electrocardiology, 31, 31–37.PubMedCrossRef Bhandari, A. K., Isber, N., Estioko, M., et al. (1998). Efficacy of low-energy T wave shocks for induction of ventricular fibrillation in patients with implantable cardioverter defibrillators. Journal of Electrocardiology, 31, 31–37.PubMedCrossRef
2.
go back to reference Fujimura, O., Jones, D. L., & Klein, G. J. (1989). The defibrillation threshold: how many measurements are enough? American Heart Journal, 117, 977–979.PubMedCrossRef Fujimura, O., Jones, D. L., & Klein, G. J. (1989). The defibrillation threshold: how many measurements are enough? American Heart Journal, 117, 977–979.PubMedCrossRef
3.
go back to reference Singer, I., van der Laken, J., Edmonds, H. L., Jr., et al. (1991). Is defibrillation testing safe? Pacing and Clinical Electrophysiology, 14, 1899–1904.PubMedCrossRef Singer, I., van der Laken, J., Edmonds, H. L., Jr., et al. (1991). Is defibrillation testing safe? Pacing and Clinical Electrophysiology, 14, 1899–1904.PubMedCrossRef
4.
go back to reference Moss, A. J., Hall, W. J., Cannom, D. S., et al. (1996). Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. The New England Journal of Medicine, 335, 1933–1940.PubMedCrossRef Moss, A. J., Hall, W. J., Cannom, D. S., et al. (1996). Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. The New England Journal of Medicine, 335, 1933–1940.PubMedCrossRef
5.
go back to reference Moss, A. J., Zareba, W., Hall, W. J., et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. The New England Journal of Medicine, 346, 877–883.PubMedCrossRef Moss, A. J., Zareba, W., Hall, W. J., et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. The New England Journal of Medicine, 346, 877–883.PubMedCrossRef
6.
go back to reference Birnie, D., Tung, S., Simpson, C., et al. (2008). Complications associated with defibrillation threshold testing: the Canadian experience. Hear Rhythm, 5, 387–390.CrossRef Birnie, D., Tung, S., Simpson, C., et al. (2008). Complications associated with defibrillation threshold testing: the Canadian experience. Hear Rhythm, 5, 387–390.CrossRef
7.
go back to reference Mollerus, M., & Naslund, L. (2007). Myocardial stunning following defibrillation threshold testing. Journal of Interventional Cardiac Electrophysiology, 19, 213–216.PubMedCrossRef Mollerus, M., & Naslund, L. (2007). Myocardial stunning following defibrillation threshold testing. Journal of Interventional Cardiac Electrophysiology, 19, 213–216.PubMedCrossRef
8.
go back to reference Kleiman, R. B., Callans, D. J., Hook, B. G., & Marchlinski, F. E. (1994). Effectiveness of noninvasive programmed stimulation for initiating ventricular tachyarrhythmias in patients with third-generation implantable cardioverter defibrillators. Pacing and Clinical Electrophysiology, 17, 1462–1468.PubMedCrossRef Kleiman, R. B., Callans, D. J., Hook, B. G., & Marchlinski, F. E. (1994). Effectiveness of noninvasive programmed stimulation for initiating ventricular tachyarrhythmias in patients with third-generation implantable cardioverter defibrillators. Pacing and Clinical Electrophysiology, 17, 1462–1468.PubMedCrossRef
9.
go back to reference Weismuller, P., Richter, P., Binner, L., et al. (1992). Direct current application: easy induction of ventricular fibrillation for the determination of the defibrillation threshold in patients with implantable cardioverter defibrillators. Pacing and Clinical Electrophysiology, 15, 1137–1143.PubMedCrossRef Weismuller, P., Richter, P., Binner, L., et al. (1992). Direct current application: easy induction of ventricular fibrillation for the determination of the defibrillation threshold in patients with implantable cardioverter defibrillators. Pacing and Clinical Electrophysiology, 15, 1137–1143.PubMedCrossRef
10.
go back to reference Sharma, A. D., Fain, E., O’Neill, P. G., et al. (2004). Shock on T versus direct current voltage for induction of ventricular fibrillation: a randomized prospective comparison. Pacing and Clinical Electrophysiology, 27, 89–94.PubMedCrossRef Sharma, A. D., Fain, E., O’Neill, P. G., et al. (2004). Shock on T versus direct current voltage for induction of ventricular fibrillation: a randomized prospective comparison. Pacing and Clinical Electrophysiology, 27, 89–94.PubMedCrossRef
11.
go back to reference Euler, D. E., Whitman, T. A., Roberts, P. R., & Kallok, M. J. (1999). Low voltage direct current delivered through unipolar transvenous leads: an alternate method for the induction of ventricular fibrillation. Pacing and Clinical Electrophysiology, 22, 908–914.PubMedCrossRef Euler, D. E., Whitman, T. A., Roberts, P. R., & Kallok, M. J. (1999). Low voltage direct current delivered through unipolar transvenous leads: an alternate method for the induction of ventricular fibrillation. Pacing and Clinical Electrophysiology, 22, 908–914.PubMedCrossRef
12.
go back to reference Rivergo Ayerza, M., Geelen, P., De Vusser, P., et al. (2003). Results of the DC fibber multicenter registry: effectiveness of direct current pulses to induce ventricular fibrillation. Europace, 4, A48. Rivergo Ayerza, M., Geelen, P., De Vusser, P., et al. (2003). Results of the DC fibber multicenter registry: effectiveness of direct current pulses to induce ventricular fibrillation. Europace, 4, A48.
13.
go back to reference Mueller, C., Meine, M., Kutscha, A., & Trappe, H. (2003). A new feature for induction of ventricular fibrillation. DC fibber vs. shock on T. Results of the prospective, randomized fib-induction-trial. Europace, 4, B146.CrossRef Mueller, C., Meine, M., Kutscha, A., & Trappe, H. (2003). A new feature for induction of ventricular fibrillation. DC fibber vs. shock on T. Results of the prospective, randomized fib-induction-trial. Europace, 4, B146.CrossRef
14.
go back to reference Pelosi, F., Jr., Oral, H., Kim, M. H., et al. (2000). Effect of chronic amiodarone therapy on defibrillation energy requirements in humans. Journal of Cardiovascular Electrophysiology, 11, 736–740.PubMedCrossRef Pelosi, F., Jr., Oral, H., Kim, M. H., et al. (2000). Effect of chronic amiodarone therapy on defibrillation energy requirements in humans. Journal of Cardiovascular Electrophysiology, 11, 736–740.PubMedCrossRef
15.
go back to reference Summitt, J., Morady, F., & Kadish, A. (1992). A comparison of standard and high-dose regimens for the initiation of amiodarone therapy. American Heart Journal, 124, 366–373.PubMedCrossRef Summitt, J., Morady, F., & Kadish, A. (1992). A comparison of standard and high-dose regimens for the initiation of amiodarone therapy. American Heart Journal, 124, 366–373.PubMedCrossRef
Metadata
Title
Effectiveness of VF induction with DC fibber versus conventional induction methods in patients on chronic amiodarone therapy
Authors
Jason C. Rubenstein
Manish S. Gupta
Michael H. Kim
Publication date
01-11-2013
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2013
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-013-9819-1

Other articles of this Issue 2/2013

Journal of Interventional Cardiac Electrophysiology 2/2013 Go to the issue