Optimal combination periprosthetic vasculature visualization and metal artifact reduction by spectral computed tomography using virtual monoenergetic images in total hip arthroplasty

Authors: Jie Zhao, Qiang Cheng, Chuan Liu, Qiqi Wang, Yuchan Lv, Ziyi Tang, Yuxi Luo, Haitao Yang

Published in: Insights into Imaging | Issue 1/2023

Login to get access

Abstract

Objectives

To investigate the optimal parameters of spectral CT for preferably visualizing the periprosthetic vasculature and metal artifact reduction (MAR) in total hip arthroplasty (THA).

Methods

A total of 34 THA of 30 patients were retrospectively included. Image reconstructions included conventional image (CI), CI combined with MAR (CIMAR), and virtual monoenergetic images (VMI) combined with MAR (VMIMAR) at 50–120 keV. The attenuation and standard deviation of the vessel and artifact, and the width of artifact were measured. Qualitative scoring was evaluated including the vascular contour, the extent of artifact, and overall diagnostic evaluation.

Results

The attenuation, noise of the vessel and artifact, and the width of artifact decreased as the energy level increased (p < 0.001). The downtrend was relatively flat at 80–120 keV, and the vascular attenuation dropped to 200 HU at 90 keV. The qualitative rating of vascular contour was significantly higher at CIMAR (3.47) and VMIMAR 60–80 keV (2.82–3.65) compared with CI (2.03) (p ≤ 0.029), and the highest score occurred at 70 and 80 keV (3.65 and 3.56). The score of the extent of artifact was higher at VMIMAR 80 keV than CIMAR (3.53 VS 3.12, p = 0.003). The score of the overall diagnostic evaluation was higher at VMIMAR 70 and 80 keV (3.32 and 3.53, respectively) than CIMAR (3.12) (p ≤ 0.035).

Conclusion

Eighty kiloelectron volts on VMIMAR, providing satisfactorily reduced metal artifacts and improved vascular visualization, can be an optimal recommended parameter of spectrum CT for the assessment of periprosthetic vasculature in THA patients.

Critical relevance statement

The metal artifact is gradually reducing with increasing energy level; however, the vascular visualization is worsening. The vascular visualization is terrible above 100 keV, while the vessel is disturbed by artifacts below 70 keV. The best performance is found at 80 keV.

Key points

• VMIMAR can provide both reduced metal artifacts and improved vascular visualization.
• Eighty kiloelectron volts on VMIMAR performs best in vascular visualization of total hip arthroplasty patients.
• Energy spectrum CT is recommended for routine use in patients with total hip arthroplasty.

Graphical Abstract

Literature
1.
go back to reference Melvin JS, Karthikeyan T, Cope R, Fehring TK (2014) Early failures in total hip arthroplasty – a changing paradigm. J Arthroplasty 29:1285–1288PubMed Melvin JS, Karthikeyan T, Cope R, Fehring TK (2014) Early failures in total hip arthroplasty – a changing paradigm. J Arthroplasty 29:1285–1288PubMed
3.
go back to reference Tsantes AG, Papadopoulos DV, Lytras T et al (2019) Association of malnutrition with periprosthetic joint and surgical site infections after total joint arthroplasty: a systematic review and meta-analysis. J Hosp Infect 103:69–77PubMed Tsantes AG, Papadopoulos DV, Lytras T et al (2019) Association of malnutrition with periprosthetic joint and surgical site infections after total joint arthroplasty: a systematic review and meta-analysis. J Hosp Infect 103:69–77PubMed
4.
go back to reference Street MWJ, Howard LC, Neufeld ME, Masri BA (2022) Vascular injuries during hip and knee replacement. Orthop Clin North Am 53:1–12PubMed Street MWJ, Howard LC, Neufeld ME, Masri BA (2022) Vascular injuries during hip and knee replacement. Orthop Clin North Am 53:1–12PubMed
5.
go back to reference Regensteiner JG, Hiatt WR (2002) Current medical therapies for patients with peripheral arterial disease: a critical review. Am J Med 112:49–57PubMed Regensteiner JG, Hiatt WR (2002) Current medical therapies for patients with peripheral arterial disease: a critical review. Am J Med 112:49–57PubMed
6.
go back to reference Lussana F, Squizzato A, Permunian ET, Cattaneo M (2014) A systematic review on the effect of aspirin in the prevention of post-operative arterial thrombosis in patients undergoing total hip and total knee arthroplasty. Thromb Res 134:599–603PubMed Lussana F, Squizzato A, Permunian ET, Cattaneo M (2014) A systematic review on the effect of aspirin in the prevention of post-operative arterial thrombosis in patients undergoing total hip and total knee arthroplasty. Thromb Res 134:599–603PubMed
7.
go back to reference Leung AG, Cabanela ME (1998) Aortic thrombosis after acetabular revision of a total hip arthroplasty. J Arthroplasty 13:961–965PubMed Leung AG, Cabanela ME (1998) Aortic thrombosis after acetabular revision of a total hip arthroplasty. J Arthroplasty 13:961–965PubMed
8.
go back to reference Trousdale RT, Donnelly RS, Hallett JW (1999) Thrombosis of an aortobifemoral bypass graft after total hip arthroplasty. J Arthroplasty 14:386–390PubMed Trousdale RT, Donnelly RS, Hallett JW (1999) Thrombosis of an aortobifemoral bypass graft after total hip arthroplasty. J Arthroplasty 14:386–390PubMed
9.
go back to reference Chou TY, Su TW, Jou HJ et al (2015) Increased risk of peripheral arterial disease after hip replacement: an 11-year retrospective population-based cohort study. Medicine (Baltimore) 94:e870PubMedPubMedCentral Chou TY, Su TW, Jou HJ et al (2015) Increased risk of peripheral arterial disease after hip replacement: an 11-year retrospective population-based cohort study. Medicine (Baltimore) 94:e870PubMedPubMedCentral
10.
go back to reference Mori I, Machida Y, Osanai M, Iinuma K (2013) Photon starvation artifacts of X-ray CT: their true cause and a solution. Radiol Phys Technol 6:130–141PubMed Mori I, Machida Y, Osanai M, Iinuma K (2013) Photon starvation artifacts of X-ray CT: their true cause and a solution. Radiol Phys Technol 6:130–141PubMed
11.
go back to reference Brooks RA, Di Chiro G (1976) Beam hardening in x-ray reconstructive tomography. Phys Med Biol 21:390–398PubMed Brooks RA, Di Chiro G (1976) Beam hardening in x-ray reconstructive tomography. Phys Med Biol 21:390–398PubMed
12.
go back to reference Wellenberg RHH, Hakvoort ET, Slump CH, Boomsma MF, Maas M, Streekstra GJ (2018) Metal artifact reduction techniques in musculoskeletal CT-imaging. Eur J Radiol 107:60–69PubMed Wellenberg RHH, Hakvoort ET, Slump CH, Boomsma MF, Maas M, Streekstra GJ (2018) Metal artifact reduction techniques in musculoskeletal CT-imaging. Eur J Radiol 107:60–69PubMed
13.
go back to reference Leithner D, Mahmoudi S, Wichmann JL et al (2018) Evaluation of virtual monoenergetic imaging algorithms for dual-energy carotid and intracerebral CT angiography: effects on image quality, artefacts and diagnostic performance for the detection of stenosis. Eur J Radiol 99:111–117PubMed Leithner D, Mahmoudi S, Wichmann JL et al (2018) Evaluation of virtual monoenergetic imaging algorithms for dual-energy carotid and intracerebral CT angiography: effects on image quality, artefacts and diagnostic performance for the detection of stenosis. Eur J Radiol 99:111–117PubMed
14.
go back to reference Hamid S, Nasir MU, So A, Andrews G, Nicolaou S, Qamar SR (2021) Clinical applications of dual-energy CT. Korean J Radiol 22:970–982PubMedPubMedCentral Hamid S, Nasir MU, So A, Andrews G, Nicolaou S, Qamar SR (2021) Clinical applications of dual-energy CT. Korean J Radiol 22:970–982PubMedPubMedCentral
15.
go back to reference Albrecht MH, Trommer J, Wichmann JL et al (2016) Comprehensive comparison of virtual monoenergetic and linearly blended reconstruction techniques in third-generation dual-source dual-energy computed tomography angiography of the thorax and abdomen. Invest Radiol 51:582–590PubMed Albrecht MH, Trommer J, Wichmann JL et al (2016) Comprehensive comparison of virtual monoenergetic and linearly blended reconstruction techniques in third-generation dual-source dual-energy computed tomography angiography of the thorax and abdomen. Invest Radiol 51:582–590PubMed
16.
go back to reference Albrecht MH, Scholtz JE, Husers K et al (2016) Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 26:1863–1870PubMed Albrecht MH, Scholtz JE, Husers K et al (2016) Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 26:1863–1870PubMed
17.
go back to reference Neuhaus V, Grosse Hokamp N, Zopfs D et al (2019) Reducing artifacts from total hip replacements in dual layer detector CT: combination of virtual monoenergetic images and orthopedic metal artifact reduction. Eur J Radiol 111:14–20PubMed Neuhaus V, Grosse Hokamp N, Zopfs D et al (2019) Reducing artifacts from total hip replacements in dual layer detector CT: combination of virtual monoenergetic images and orthopedic metal artifact reduction. Eur J Radiol 111:14–20PubMed
18.
go back to reference Park J, Kim SH, Han JK (2019) Combined application of virtual monoenergetic high keV images and the orthopedic metal artifact reduction algorithm (O-MAR): effect on image quality. Abdom Radiol (NY) 44:756–765PubMed Park J, Kim SH, Han JK (2019) Combined application of virtual monoenergetic high keV images and the orthopedic metal artifact reduction algorithm (O-MAR): effect on image quality. Abdom Radiol (NY) 44:756–765PubMed
19.
go back to reference Bongers MN, Schabel C, Thomas C et al (2015) Comparison and combination of dual-energy- and iterative-based metal artefact reduction on hip prosthesis and dental implants. PLoS One 10:e0143584PubMedPubMedCentral Bongers MN, Schabel C, Thomas C et al (2015) Comparison and combination of dual-energy- and iterative-based metal artefact reduction on hip prosthesis and dental implants. PLoS One 10:e0143584PubMedPubMedCentral
20.
go back to reference Han SC, Chung YE, Lee YH, Park KK, Kim MJ, Kim KW (2014) Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol 203:788–795PubMed Han SC, Chung YE, Lee YH, Park KK, Kim MJ, Kim KW (2014) Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol 203:788–795PubMed
21.
go back to reference Sudarski S, Apfaltrer P, Nance JW Jr et al (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–581PubMed Sudarski S, Apfaltrer P, Nance JW Jr et al (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–581PubMed
23.
go back to reference Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15PubMed Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15PubMed
24.
go back to reference Delesalle MA, Pontana F, Duhamel A et al (2013) Spectral optimization of chest CT angiography with reduced iodine load: experience in 80 patients evaluated with dual-source, dual-energy CT. Radiology 267:256–266PubMed Delesalle MA, Pontana F, Duhamel A et al (2013) Spectral optimization of chest CT angiography with reduced iodine load: experience in 80 patients evaluated with dual-source, dual-energy CT. Radiology 267:256–266PubMed
25.
go back to reference Laukamp KR, Lennartz S, Neuhaus VF et al (2018) CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol 28:4524–4533PubMed Laukamp KR, Lennartz S, Neuhaus VF et al (2018) CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar. Eur Radiol 28:4524–4533PubMed
26.
go back to reference Neuhaus V, Grosse Hokamp N, Abdullayev N et al (2017) Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images. Eur J Radiol 93:143–148PubMed Neuhaus V, Grosse Hokamp N, Abdullayev N et al (2017) Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images. Eur J Radiol 93:143–148PubMed
27.
go back to reference Yin XP, Gao BL, Li CY et al (2018) Optimal monochromatic imaging of spectral computed tomography potentially improves the quality of hepatic vascular imaging. Korean J Radiol 19:578–584PubMedPubMedCentral Yin XP, Gao BL, Li CY et al (2018) Optimal monochromatic imaging of spectral computed tomography potentially improves the quality of hepatic vascular imaging. Korean J Radiol 19:578–584PubMedPubMedCentral
28.
go back to reference Kosmas C, Hojjati M, Young PC, Abedi A, Gholamrezanezhad A, Rajiah P (2019) Dual-layer spectral computerized tomography for metal artifact reduction: small versus large orthopedic devices. Skeletal Radiol 48:1981–1990PubMed Kosmas C, Hojjati M, Young PC, Abedi A, Gholamrezanezhad A, Rajiah P (2019) Dual-layer spectral computerized tomography for metal artifact reduction: small versus large orthopedic devices. Skeletal Radiol 48:1981–1990PubMed
29.
go back to reference Grosse Hokamp N, Neuhaus V, Abdullayev N et al (2018) Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms. Skeletal Radiol 47:195–201PubMed Grosse Hokamp N, Neuhaus V, Abdullayev N et al (2018) Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms. Skeletal Radiol 47:195–201PubMed
30.
go back to reference Hilgers G, Nuver T, Minken A (2014) The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants. J Appl Clin Med Phys 15:4597PubMed Hilgers G, Nuver T, Minken A (2014) The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants. J Appl Clin Med Phys 15:4597PubMed
31.
go back to reference Lewis M, Reid K, Toms AP (2013) Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements. Skeletal Radiol 42:275–282PubMed Lewis M, Reid K, Toms AP (2013) Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements. Skeletal Radiol 42:275–282PubMed
32.
go back to reference Dangelmaier J, Schwaiger BJ, Gersing AS et al (2018) Dual layer computed tomography: reduction of metal artefacts from posterior spinal fusion using virtual monoenergetic imaging. Eur J Radiol 105:195–203PubMed Dangelmaier J, Schwaiger BJ, Gersing AS et al (2018) Dual layer computed tomography: reduction of metal artefacts from posterior spinal fusion using virtual monoenergetic imaging. Eur J Radiol 105:195–203PubMed
33.
go back to reference Lee YH, Park KK, Song HT, Kim S, Suh JS (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340PubMed Lee YH, Park KK, Song HT, Kim S, Suh JS (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340PubMed
34.
go back to reference Vlahos I, Godoy MC, Naidich DP (2010) Dual-energy computed tomography imaging of the aorta. J Thorac Imaging 25:289–300PubMed Vlahos I, Godoy MC, Naidich DP (2010) Dual-energy computed tomography imaging of the aorta. J Thorac Imaging 25:289–300PubMed
35.
go back to reference Bae K, Jeon KN, Cho SB et al (2018) Improved opacification of a suboptimally enhanced pulmonary artery in chest CT: experience using a dual-layer detector spectral CT. AJR Am J Roentgenol 210:734–741PubMed Bae K, Jeon KN, Cho SB et al (2018) Improved opacification of a suboptimally enhanced pulmonary artery in chest CT: experience using a dual-layer detector spectral CT. AJR Am J Roentgenol 210:734–741PubMed
36.
go back to reference Ozawa Y, Hara M, Shibamoto Y (2011) The frequency of insufficient contrast enhancement of the pulmonary artery in routine contrast-enhanced chest CT and its improvement with an increased injection rate: a prospective study. J Thorac Imaging 26:42–47PubMed Ozawa Y, Hara M, Shibamoto Y (2011) The frequency of insufficient contrast enhancement of the pulmonary artery in routine contrast-enhanced chest CT and its improvement with an increased injection rate: a prospective study. J Thorac Imaging 26:42–47PubMed
37.
go back to reference Mangold S, De Cecco CN, Schoepf UJ et al (2016) A noise-optimized virtual monochromatic reconstruction algorithm improves stent visualization and diagnostic accuracy for detection of in-stent re-stenosis in lower extremity run-off CT angiography. Eur Radiol 26:4380–4389PubMed Mangold S, De Cecco CN, Schoepf UJ et al (2016) A noise-optimized virtual monochromatic reconstruction algorithm improves stent visualization and diagnostic accuracy for detection of in-stent re-stenosis in lower extremity run-off CT angiography. Eur Radiol 26:4380–4389PubMed
38.
go back to reference Zhang D, Xie Y, Wang Y et al (2020) Initial clinical experience of virtual monoenergetic imaging improves stent visualization in lower extremity run-off CT angiography by dual-layer spectral detector CT. Acad Radiol 27:825–832PubMed Zhang D, Xie Y, Wang Y et al (2020) Initial clinical experience of virtual monoenergetic imaging improves stent visualization in lower extremity run-off CT angiography by dual-layer spectral detector CT. Acad Radiol 27:825–832PubMed
39.
go back to reference Selles M, Stuivenberg VH, Wellenberg RHH et al (2021) Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study. Insights Imaging 12:171PubMedPubMedCentral Selles M, Stuivenberg VH, Wellenberg RHH et al (2021) Quantitative analysis of metal artifact reduction in total hip arthroplasty using virtual monochromatic imaging and orthopedic metal artifact reduction, a phantom study. Insights Imaging 12:171PubMedPubMedCentral
40.
go back to reference Wellenberg RH, Boomsma MF, van Osch JA et al (2017) Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction. Skeletal Radiol 46:623–632PubMedPubMedCentral Wellenberg RH, Boomsma MF, van Osch JA et al (2017) Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction. Skeletal Radiol 46:623–632PubMedPubMedCentral
41.
go back to reference Krishna S, Sadoughi N, McInnes MDF, Chatelain R, MacDonald DB, Schieda N (2018) Attenuation and degree of enhancement with conventional 120-kVp polychromatic CT and 70-keV monochromatic rapid kilovoltage-switching dual-energy CT in cystic and solid renal masses. AJR Am J Roentgenol 211:789–796PubMed Krishna S, Sadoughi N, McInnes MDF, Chatelain R, MacDonald DB, Schieda N (2018) Attenuation and degree of enhancement with conventional 120-kVp polychromatic CT and 70-keV monochromatic rapid kilovoltage-switching dual-energy CT in cystic and solid renal masses. AJR Am J Roentgenol 211:789–796PubMed
42.
go back to reference Wellenberg RH, Boomsma MF, van Osch JA et al (2017) Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur J Radiol 88:61–70PubMed Wellenberg RH, Boomsma MF, van Osch JA et al (2017) Quantifying metal artefact reduction using virtual monochromatic dual-layer detector spectral CT imaging in unilateral and bilateral total hip prostheses. Eur J Radiol 88:61–70PubMed
43.
go back to reference Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379PubMedPubMedCentral Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379PubMedPubMedCentral
44.
go back to reference Pessis E, Campagna R, Sverzut JM et al (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583PubMed Pessis E, Campagna R, Sverzut JM et al (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583PubMed
Metadata
Title
Optimal combination periprosthetic vasculature visualization and metal artifact reduction by spectral computed tomography using virtual monoenergetic images in total hip arthroplasty
Authors
Jie Zhao
Qiang Cheng
Chuan Liu
Qiqi Wang
Yuchan Lv
Ziyi Tang
Yuxi Luo
Haitao Yang
Publication date
01-12-2023
Publisher
Springer Vienna
Published in
Insights into Imaging / Issue 1/2023
Electronic ISSN: 1869-4101
DOI
https://doi.org/10.1186/s13244-023-01533-3

Other articles of this Issue 1/2023

Insights into Imaging 1/2023 Go to the issue