Open Access 01-12-2015 | Research article

Novel sirolimus-eluting stent Prolim® with a biodegradable polymer in the all-comers population: one year clinical results with quantitative coronary angiography and optical coherence tomography analysis

Authors: Jacek Bil, Robert J. Gil, Adam Kern, Tomasz Pawłowski, Piotr Seweryniak, Zbigniew Śliwiński

Published in: BMC Cardiovascular Disorders | Issue 1/2015

Login to get access

Abstract

Background

The aim of this study was to assess the safety and the efficacy of the novel sirolimus-eluting Prolim® stent with a biodegradable polymer in the all-comers population.

Methods

We prospectively enrolled all patients with stable coronary artery disease or acute coronary syndrome treated with Prolim® stent between January and December 2013 in two interventional cardiology centers in Poland. Angiographic control was planned at 12 months, in which 15 % of patients (randomly chosen) underwent optical coherence tomography imaging. The primary end-point was the cumulative rate of cardiac death, myocardial infarction, and target lesion revascularization at 12 months.

Results

There were 204 patients enrolled, in whom 238 Prolim® stents were deployed (1.17 stent per patient). The mean age was 68 ± 10 years and 32.8 % were females. The examined stent was implanted in 5.9 % in STEMI patients, in 21.6 % - in NSTE-ACS and in 72.5 % - in patients with stable coronary artery disease. The Prolim® stent was most frequently implanted in right coronary artery (38.2 %) followed by left anterior descending artery (34.0 %). The cumulative major adverse cardiovascular events rate at 12 months was 6.9 %, and the clinically-driven target lesion revascularization rate – 5.4 %. At 12 months in quantitative coronary angiography the late lumen loss was 0.21 ± 0.18 mm, and in optical coherence tomography the mean neointima burden was 24.6 ± 8.6 %.

Conclusions

Sirolimus-eluting Prolim® stent with a biodegradable polymer is a feasible device with a very good safety profile and long-term clinical effectiveness.

Trial registration number

ClinicalTrials.gov NCT02545985.
Literature
1.
go back to reference Stettler C, Wandel S, Allemann S, Kastrati A, Morice MC, Schomig A, et al. Outcomes associated with drug-eluting and bare-metal stents: a collaborative network meta-analysis. Lancet. 2007;9591:937–48.CrossRef Stettler C, Wandel S, Allemann S, Kastrati A, Morice MC, Schomig A, et al. Outcomes associated with drug-eluting and bare-metal stents: a collaborative network meta-analysis. Lancet. 2007;9591:937–48.CrossRef
2.
go back to reference Buszman P, Trznadel S, Milewski K, Rzezniczak J, Przewlocki T, Kosmider M, et al. Novel paclitaxel-eluting, biodegradable polymer coated stent in the treatment of de novo coronary lesions: a prospective multicenter registry. Catheter Cardiovasc Interv. 2008;1:51–7.CrossRef Buszman P, Trznadel S, Milewski K, Rzezniczak J, Przewlocki T, Kosmider M, et al. Novel paclitaxel-eluting, biodegradable polymer coated stent in the treatment of de novo coronary lesions: a prospective multicenter registry. Catheter Cardiovasc Interv. 2008;1:51–7.CrossRef
3.
go back to reference Kandzari DE, Leon MB, Meredith I, Fajadet J, Wijns W, Mauri L. Final 5-year outcomes from the Endeavor zotarolimus-eluting stent clinical trial program: comparison of safety and efficacy with first-generation drug-eluting and bare-metal stents. J Am Coll Cardiol Intv. 2013;5:504–12.CrossRef Kandzari DE, Leon MB, Meredith I, Fajadet J, Wijns W, Mauri L. Final 5-year outcomes from the Endeavor zotarolimus-eluting stent clinical trial program: comparison of safety and efficacy with first-generation drug-eluting and bare-metal stents. J Am Coll Cardiol Intv. 2013;5:504–12.CrossRef
4.
go back to reference Shahar L, Džavík V. Biodegradable Stent Platforms: Are We Heading in the Right Direction? Can J Cardiol. 2015;8:957–9. Shahar L, Džavík V. Biodegradable Stent Platforms: Are We Heading in the Right Direction? Can J Cardiol. 2015;8:957–9.
5.
go back to reference Milewski K, Gorycki B, Buszman PP, Jelonek M, Beaudry D, Lapointe JM, et al. Vascular response and mechanical integrity of the new biodegradable polymer coated sirolimus-eluting PROLIM stent implanted in porcine coronary arteries. Kardiol Pol. 2012;7:703–11. Milewski K, Gorycki B, Buszman PP, Jelonek M, Beaudry D, Lapointe JM, et al. Vascular response and mechanical integrity of the new biodegradable polymer coated sirolimus-eluting PROLIM stent implanted in porcine coronary arteries. Kardiol Pol. 2012;7:703–11.
6.
go back to reference Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, et al. Third universal definition of myocardial infarction. Circulation. 2012;16:2020–35.CrossRef Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, et al. Third universal definition of myocardial infarction. Circulation. 2012;16:2020–35.CrossRef
7.
go back to reference Vranckx P, Kint PP, Morel MA, Van Es GA, Serruys PW, Cutlip DE. Identifying stent thrombosis, a critical appraisal of the academic research consortium (ARC) consensus definitions: a lighthouse and as a toe in the water. EuroIntervention. 2008;4(Suppl C):C39-44. Vranckx P, Kint PP, Morel MA, Van Es GA, Serruys PW, Cutlip DE. Identifying stent thrombosis, a critical appraisal of the academic research consortium (ARC) consensus definitions: a lighthouse and as a toe in the water. EuroIntervention. 2008;4(Suppl C):C39-44.
8.
go back to reference Bil J, Gil RJ, Vassilev D, Rzezak J, Kulawik T, Pawlowski T. Dedicated bifurcation paclitaxel-eluting stent BiOSS Expert(R) in the treatment of distal left main stem stenosis. J Interv Cardiol. 2014;3:242–51.CrossRef Bil J, Gil RJ, Vassilev D, Rzezak J, Kulawik T, Pawlowski T. Dedicated bifurcation paclitaxel-eluting stent BiOSS Expert(R) in the treatment of distal left main stem stenosis. J Interv Cardiol. 2014;3:242–51.CrossRef
9.
go back to reference Pawlowski T, Prati F, Kulawik T, Ficarra E, Bil J, Gil R. Optical coherence tomography criteria for defining functional severity of intermediate lesions: a comparative study with FFR. Int J Cardiovasc Imaging. 2013;8:1685–91.CrossRef Pawlowski T, Prati F, Kulawik T, Ficarra E, Bil J, Gil R. Optical coherence tomography criteria for defining functional severity of intermediate lesions: a comparative study with FFR. Int J Cardiovasc Imaging. 2013;8:1685–91.CrossRef
10.
go back to reference Ali ZA, Roleder T, Narula J, Mohanty BD, Baber U, Kovacic JC, et al. Increased thin-cap neoatheroma and periprocedural myocardial infarction in drug-eluting stent restenosis: multimodality intravascular imaging of drug-eluting and bare-metal stents. Circ Cardiovasc Interv. 2013;5:507–17.CrossRef Ali ZA, Roleder T, Narula J, Mohanty BD, Baber U, Kovacic JC, et al. Increased thin-cap neoatheroma and periprocedural myocardial infarction in drug-eluting stent restenosis: multimodality intravascular imaging of drug-eluting and bare-metal stents. Circ Cardiovasc Interv. 2013;5:507–17.CrossRef
11.
go back to reference Nagoshi R, Shinke T, Otake H, Shite J, Matsumoto D, Kawamori H, et al. Qualitative and quantitative assessment of stent restenosis by optical coherence tomography: comparison between drug-eluting and bare-metal stents. Circ J. 2013;3:652–60.CrossRef Nagoshi R, Shinke T, Otake H, Shite J, Matsumoto D, Kawamori H, et al. Qualitative and quantitative assessment of stent restenosis by optical coherence tomography: comparison between drug-eluting and bare-metal stents. Circ J. 2013;3:652–60.CrossRef
12.
go back to reference Windecker S, Serruys PW, Wandel S, Buszman P, Trznadel S, Linke A, et al. Biolimus-eluting stent with biodegradable polymer versus sirolimus-eluting stent with durable polymer for coronary revascularisation (LEADERS): a randomised non-inferiority trial. Lancet. 2008;9644:1163–73.CrossRef Windecker S, Serruys PW, Wandel S, Buszman P, Trznadel S, Linke A, et al. Biolimus-eluting stent with biodegradable polymer versus sirolimus-eluting stent with durable polymer for coronary revascularisation (LEADERS): a randomised non-inferiority trial. Lancet. 2008;9644:1163–73.CrossRef
13.
go back to reference Zhang Q, Qiu JP, Kirtane AJ, Zhu TQ, Zhang RY, Yang ZK, et al. Comparison of biodegradable polymer versus durable polymer sirolimus-eluting stenting in patients with acute st-elevation myocardial infarction undergoing primary percutaneous coronary intervention: results of the RESOLVE study. J Interv Cardiol. 2014;2:131–41.CrossRef Zhang Q, Qiu JP, Kirtane AJ, Zhu TQ, Zhang RY, Yang ZK, et al. Comparison of biodegradable polymer versus durable polymer sirolimus-eluting stenting in patients with acute st-elevation myocardial infarction undergoing primary percutaneous coronary intervention: results of the RESOLVE study. J Interv Cardiol. 2014;2:131–41.CrossRef
14.
go back to reference Dani S, Kukreja N, Parikh P, Joshi H, Prajapati J, Jain S, et al. Biodegradable-polymer-based, sirolimus-eluting Supralimus stent: 6-month angiographic and 30-month clinical follow-up results from the series I prospective study. EuroIntervention. 2008;1:59–63.CrossRef Dani S, Kukreja N, Parikh P, Joshi H, Prajapati J, Jain S, et al. Biodegradable-polymer-based, sirolimus-eluting Supralimus stent: 6-month angiographic and 30-month clinical follow-up results from the series I prospective study. EuroIntervention. 2008;1:59–63.CrossRef
15.
go back to reference Han Y, Jing Q, Chen X, Wang S, Ma Y, Liu H, et al. Long-term clinical, angiographic, and intravascular ultrasound outcomes of biodegradable polymer-coated sirolimus-eluting stents. Catheter Cardiovasc Interv. 2008;2:177–83.CrossRef Han Y, Jing Q, Chen X, Wang S, Ma Y, Liu H, et al. Long-term clinical, angiographic, and intravascular ultrasound outcomes of biodegradable polymer-coated sirolimus-eluting stents. Catheter Cardiovasc Interv. 2008;2:177–83.CrossRef
16.
go back to reference Waltenberger J, Brachmann J, van der Heyden J, Richardt G, Frobert O, Seige M, et al.: Real-world experience with a novel biodegradable polymer sirolimus-eluting stent: twelve-month results of the BIOFLOW-III registry. In PubMed there is: EuroIntervention. 2015 Mar 17;10(11). pii: 20140318-06. doi: 10.4244/EIJY15M03_08. Waltenberger J, Brachmann J, van der Heyden J, Richardt G, Frobert O, Seige M, et al.: Real-world experience with a novel biodegradable polymer sirolimus-eluting stent: twelve-month results of the BIOFLOW-III registry. In PubMed there is: EuroIntervention. 2015 Mar 17;10(11). pii: 20140318-06. doi: 10.​4244/​EIJY15M03_​08.
17.
go back to reference Windecker S, Haude M, Neumann FJ, Stangl K, Witzenbichler B, Slagboom T, et al. Comparison of a novel biodegradable polymer sirolimus-eluting stent with a durable polymer everolimus-eluting stent: results of the randomized BIOFLOW-II trial. Circ Cardiovasc Interv. 2015;2, e001441.CrossRef Windecker S, Haude M, Neumann FJ, Stangl K, Witzenbichler B, Slagboom T, et al. Comparison of a novel biodegradable polymer sirolimus-eluting stent with a durable polymer everolimus-eluting stent: results of the randomized BIOFLOW-II trial. Circ Cardiovasc Interv. 2015;2, e001441.CrossRef
18.
go back to reference Byrne RA, Kastrati A, Kufner S, Massberg S, Birkmeier KA, Laugwitz KL, et al. Randomized, non-inferiority trial of three limus agent-eluting stents with different polymer coatings: the Intracoronary Stenting and Angiographic Results: Test Efficacy of 3 Limus-Eluting Stents (ISAR-TEST-4) Trial. Eur Heart J. 2009;20:2441–9.CrossRef Byrne RA, Kastrati A, Kufner S, Massberg S, Birkmeier KA, Laugwitz KL, et al. Randomized, non-inferiority trial of three limus agent-eluting stents with different polymer coatings: the Intracoronary Stenting and Angiographic Results: Test Efficacy of 3 Limus-Eluting Stents (ISAR-TEST-4) Trial. Eur Heart J. 2009;20:2441–9.CrossRef
19.
go back to reference Mehilli J, Byrne RA, Wieczorek A, Iijima R, Schulz S, Bruskina O, et al. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis. Eur Heart J. 2008;16:1975–82.CrossRef Mehilli J, Byrne RA, Wieczorek A, Iijima R, Schulz S, Bruskina O, et al. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis. Eur Heart J. 2008;16:1975–82.CrossRef
20.
go back to reference Pache J, Kastrati A, Mehilli J, Schuhlen H, Dotzer F, Hausleiter J, et al. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2) trial. J Am Coll Cardiol. 2003;8:1283–8.CrossRef Pache J, Kastrati A, Mehilli J, Schuhlen H, Dotzer F, Hausleiter J, et al. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2) trial. J Am Coll Cardiol. 2003;8:1283–8.CrossRef
21.
go back to reference Rittersma SZ, de Winter RJ, Koch KT, Bax M, Schotborgh CE, Mulder KJ, et al. Impact of strut thickness on late luminal loss after coronary artery stent placement. Am J Cardiol. 2004;4:477–80.CrossRef Rittersma SZ, de Winter RJ, Koch KT, Bax M, Schotborgh CE, Mulder KJ, et al. Impact of strut thickness on late luminal loss after coronary artery stent placement. Am J Cardiol. 2004;4:477–80.CrossRef
22.
go back to reference Colmenarez H, Fernandez C, Escaned J. Impact of technological developments in drug-eluting stents on patient-focused outcomes: a pooled direct and indirect comparison of randomised trials comparing first- and second-generation drug-eluting stents. EuroIntervention. 2014;8:942–52.CrossRef Colmenarez H, Fernandez C, Escaned J. Impact of technological developments in drug-eluting stents on patient-focused outcomes: a pooled direct and indirect comparison of randomised trials comparing first- and second-generation drug-eluting stents. EuroIntervention. 2014;8:942–52.CrossRef
23.
go back to reference Naidu SS, Krucoff MW, Rutledge DR, Mao VW, Zhao W, Zheng Q, et al. Contemporary incidence and predictors of stent thrombosis and other major adverse cardiac events in the year after XIENCE V implantation: results from the 8,061-patient XIENCE V United States study. J Am Coll Cardiol Intv. 2012;6:626–35.CrossRef Naidu SS, Krucoff MW, Rutledge DR, Mao VW, Zhao W, Zheng Q, et al. Contemporary incidence and predictors of stent thrombosis and other major adverse cardiac events in the year after XIENCE V implantation: results from the 8,061-patient XIENCE V United States study. J Am Coll Cardiol Intv. 2012;6:626–35.CrossRef
24.
go back to reference Zhu L, Lv YN, Wang LY. Stent thrombosis with biodegradable polymer drug-eluting stents versus durable polymer sirolimus-eluting stents: an update meta-analysis. Cardiology. 2015;2:96–105. Zhu L, Lv YN, Wang LY. Stent thrombosis with biodegradable polymer drug-eluting stents versus durable polymer sirolimus-eluting stents: an update meta-analysis. Cardiology. 2015;2:96–105.
25.
go back to reference Lv J, Wu Y, Zhang X, Jing T, Zhang L, Tong S, et al. Comparison of the safety and efficacy of biodegradable polymer drug-eluting stents versus durable polymer drug-eluting stents: a meta-analysis. Eur J Med Res. 2015;1:21.CrossRef Lv J, Wu Y, Zhang X, Jing T, Zhang L, Tong S, et al. Comparison of the safety and efficacy of biodegradable polymer drug-eluting stents versus durable polymer drug-eluting stents: a meta-analysis. Eur J Med Res. 2015;1:21.CrossRef
26.
go back to reference Commandeur S, van Beusekom HM, van der Giessen WJ. Polymers, drug release, and drug-eluting stents. J Interv Cardiol. 2006;6:500–6.CrossRef Commandeur S, van Beusekom HM, van der Giessen WJ. Polymers, drug release, and drug-eluting stents. J Interv Cardiol. 2006;6:500–6.CrossRef
27.
go back to reference Gutierrez-Chico JL, Juni P, Garcia-Garcia HM, Regar E, Nuesch E, Borgia F, et al. Long-term tissue coverage of a biodegradable polylactide polymer-coated biolimus-eluting stent: comparative sequential assessment with optical coherence tomography until complete resorption of the polymer. Am Heart J. 2011;5:922–31.CrossRef Gutierrez-Chico JL, Juni P, Garcia-Garcia HM, Regar E, Nuesch E, Borgia F, et al. Long-term tissue coverage of a biodegradable polylactide polymer-coated biolimus-eluting stent: comparative sequential assessment with optical coherence tomography until complete resorption of the polymer. Am Heart J. 2011;5:922–31.CrossRef
28.
go back to reference Gutierrez-Chico JL, van Geuns RJ, Regar E, van der Giessen WJ, Kelbaek H, Saunamaki K, et al. Tissue coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13-month follow-up: an optical coherence tomography substudy from the RESOLUTE All Comers trial. Eur Heart J. 2011;19:2454–63.CrossRef Gutierrez-Chico JL, van Geuns RJ, Regar E, van der Giessen WJ, Kelbaek H, Saunamaki K, et al. Tissue coverage of a hydrophilic polymer-coated zotarolimus-eluting stent vs. a fluoropolymer-coated everolimus-eluting stent at 13-month follow-up: an optical coherence tomography substudy from the RESOLUTE All Comers trial. Eur Heart J. 2011;19:2454–63.CrossRef
29.
go back to reference Barlis P, Regar E, Serruys PW, Dimopoulos K, van der Giessen WJ, van Geuns RJ, et al. An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-study. Eur Heart J. 2010;2:165–76.CrossRef Barlis P, Regar E, Serruys PW, Dimopoulos K, van der Giessen WJ, van Geuns RJ, et al. An optical coherence tomography study of a biodegradable vs. durable polymer-coated limus-eluting stent: a LEADERS trial sub-study. Eur Heart J. 2010;2:165–76.CrossRef
30.
go back to reference Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, et al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;18:2435–41.CrossRef Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, et al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;18:2435–41.CrossRef
31.
go back to reference Kim BK, Ha J, Mintz GS, Kim JS, Shin DH, Ko YG, et al. Randomised comparison of strut coverage between Nobori biolimus-eluting and sirolimus-eluting stents: an optical coherence tomography analysis. EuroIntervention. 2014;12:1389–97.CrossRef Kim BK, Ha J, Mintz GS, Kim JS, Shin DH, Ko YG, et al. Randomised comparison of strut coverage between Nobori biolimus-eluting and sirolimus-eluting stents: an optical coherence tomography analysis. EuroIntervention. 2014;12:1389–97.CrossRef
32.
go back to reference Davlouros PA, Mavronasiou E, Xanthopoulou I, Karantalis V, Tsigkas G, Hahalis G, et al. An optical coherence tomography study of two new generation stents with biodegradable polymer carrier, eluting paclitaxel vs. biolimus-A9. Int J Cardiol. 2012;3:341–6.CrossRef Davlouros PA, Mavronasiou E, Xanthopoulou I, Karantalis V, Tsigkas G, Hahalis G, et al. An optical coherence tomography study of two new generation stents with biodegradable polymer carrier, eluting paclitaxel vs. biolimus-A9. Int J Cardiol. 2012;3:341–6.CrossRef
33.
go back to reference Adriaenssens T, Ughi GJ, Dubois C, De Cock D, Onsea K, Bennett J, et al. STACCATO (Assessment of Stent sTrut Apposition and Coverage in Coronary ArTeries with Optical coherence tomography in patients with STEMI, NSTEMI and stable/unstable angina undergoing everolimus vs. biolimus A9-eluting stent implantation): a randomised controlled trial. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2014;20130911–03. doi: 10.4244/EIJY14M11_11. Adriaenssens T, Ughi GJ, Dubois C, De Cock D, Onsea K, Bennett J, et al. STACCATO (Assessment of Stent sTrut Apposition and Coverage in Coronary ArTeries with Optical coherence tomography in patients with STEMI, NSTEMI and stable/unstable angina undergoing everolimus vs. biolimus A9-eluting stent implantation): a randomised controlled trial. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2014;20130911–03. doi: 10.​4244/​EIJY14M11_​11.
34.
go back to reference Kim JS, Lee JH, Shin DH, Kim BK, Ko YG, Choi D, et al. Long-term outcomes of neointimal hyperplasia without neoatherosclerosis after drug-eluting stent implantation. JACC Cardiovasc Imaging. 2014;8:788–95.CrossRef Kim JS, Lee JH, Shin DH, Kim BK, Ko YG, Choi D, et al. Long-term outcomes of neointimal hyperplasia without neoatherosclerosis after drug-eluting stent implantation. JACC Cardiovasc Imaging. 2014;8:788–95.CrossRef
35.
go back to reference Kim JS, Afari ME, Ha J, Tellez A, Milewski K, Conditt G, et al. Neointimal patterns obtained by optical coherence tomography correlate with specific histological components and neointimal proliferation in a swine model of restenosis. Eur Heart J Cardiovasc Imaging. 2014;3:292–8.CrossRef Kim JS, Afari ME, Ha J, Tellez A, Milewski K, Conditt G, et al. Neointimal patterns obtained by optical coherence tomography correlate with specific histological components and neointimal proliferation in a swine model of restenosis. Eur Heart J Cardiovasc Imaging. 2014;3:292–8.CrossRef
36.
go back to reference D’Ascenzo F, Barbero U, Cerrato E, Lipinski MJ, Omede P, Montefusco A, et al. Accuracy of intravascular ultrasound and optical coherence tomography in identifying functionally significant coronary stenosis according to vessel diameter: A meta-analysis of 2,581 patients and 2,807 lesions. Am Heart J. 2015;5:663–73.CrossRef D’Ascenzo F, Barbero U, Cerrato E, Lipinski MJ, Omede P, Montefusco A, et al. Accuracy of intravascular ultrasound and optical coherence tomography in identifying functionally significant coronary stenosis according to vessel diameter: A meta-analysis of 2,581 patients and 2,807 lesions. Am Heart J. 2015;5:663–73.CrossRef
37.
go back to reference Zhang L, Li Y, Jing QM, Wang XZ, Ma YY, Wang G, et al. Dual antiplatelet therapy over 6 months increases the risk of bleeding after biodegradable polymer-coated sirolimus eluting stents implantation: insights from the CREATE study. J Interv Cardiol. 2014;2:119–26.CrossRef Zhang L, Li Y, Jing QM, Wang XZ, Ma YY, Wang G, et al. Dual antiplatelet therapy over 6 months increases the risk of bleeding after biodegradable polymer-coated sirolimus eluting stents implantation: insights from the CREATE study. J Interv Cardiol. 2014;2:119–26.CrossRef
Metadata
Title
Novel sirolimus-eluting stent Prolim® with a biodegradable polymer in the all-comers population: one year clinical results with quantitative coronary angiography and optical coherence tomography analysis
Authors
Jacek Bil
Robert J. Gil
Adam Kern
Tomasz Pawłowski
Piotr Seweryniak
Zbigniew Śliwiński
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Cardiovascular Disorders / Issue 1/2015
Electronic ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-015-0139-5

Other articles of this Issue 1/2015

BMC Cardiovascular Disorders 1/2015 Go to the issue