Published in: BMC Anesthesiology 1/2021

Open Access 01-12-2021 | Malignant Hyperthermia | Research

Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients

Authors: Sebastian Heiderich, Christian Thoben, Nils Dennhardt, Terence Krauß, Robert Sümpelmann, Stefan Zimmermann, Michael Reitz, Henrik Rüffert

Published in: BMC Anesthesiology | Issue 1/2021

Login to get access

Abstract

Background

Patients at risk of malignant hyperthermia need trigger-free anesthesia. Therefore, anesthesia machines prepared for safe use in predisposed patients should be free of volatile anesthetics. The washout time depends on the composition of rubber and plastic in the anesthesia machine. Therefore, new anesthesia machines should be evaluated regarding the safe preparation for trigger-free anesthesia. This study investigates wash out procedures of volatile anesthetics for two new anesthetic workstations: Dräger Atlan A350 and General Electric Healthcare (GE) Carestation 650 and compare it with preparation using activated charcoal filters (ACF).

Methods

A Dräger Atlan and a Carestation 650 were contaminated with 4% sevoflurane for 90 min. The machines were decontaminated with method (M1): using ACF, method 2 (M2): a wash out method that included exchange of internal parts, breathing circuits and soda lime canister followed by ventilating a test lung using a preliminary protocol provided by Dräger or method 3 (M3): a universal wash out instruction of GE, method 4 (M4): M3 plus exchange of breathing system and bellows. Decontamination was followed by a simulated trigger-free ventilation. All experiments were repeated with 8% desflurane contaminated machines. Volatile anesthetics were detected with a closed gas loop high-resolution ion mobility spectrometer with gas chromatographic pre-separation attached to the bacterial filter of the breathing circuits. Primary outcome was time until < 5 ppm of volatile anesthetics and total preparation time.

Results

Time to < 5 ppm for the Atlan was 17 min (desflurane) and 50 min (sevoflurane), wash out continued for a total of 60 min according to protocol resulting in a total preparation time of 96-122 min. The Carestation needed 66 min (desflurane) and 24 min (sevoflurane) which could be abbreviated to 24 min (desflurane) if breathing system and bellows were changed. Total preparation time was 30-73 min. When using active charcoal filters time to < 5 ppm was 0 min for both machines, and total preparation time < 5 min.

Conclusion

Both wash out protocols resulted in a significant reduction of trace gas concentrations. However, due to the complexity of the protocols and prolonged total preparation time, feasibility in clinical practice remains questionable. Especially when time is limited preparation of the anesthetic machines using ACF remain superior.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kim TW, Nemergut ME. Preparation of modern anesthesia workstations for malignant hyperthermia-susceptible patients: a review of past and present practice. Anesthesiology. 2011;114(1):205–12.CrossRef Kim TW, Nemergut ME. Preparation of modern anesthesia workstations for malignant hyperthermia-susceptible patients: a review of past and present practice. Anesthesiology. 2011;114(1):205–12.CrossRef
2.
go back to reference Rüffert H, Bastian B, Bendixen D, Girard T, Heiderich S, Hellblom A, et al. Consensus guidelines on perioperative management of malignant hyperthermia suspected or susceptible patients from the European malignant hyperthermia group. Br J Anaesth. 2021;126(1):120–30.CrossRef Rüffert H, Bastian B, Bendixen D, Girard T, Heiderich S, Hellblom A, et al. Consensus guidelines on perioperative management of malignant hyperthermia suspected or susceptible patients from the European malignant hyperthermia group. Br J Anaesth. 2021;126(1):120–30.CrossRef
3.
go back to reference Neira VM, Al Madhoun W, Ghaffari K, Barrowman N, Berrigan P, Splinter W. Efficacy of malignant hyperthermia Association of the United States-Recommended Methods of preparation for malignant hyperthermia-susceptible patients using Drager Zeus anesthesia workstations and associated costs. Anesth Analg. 2019;129(1):74–83.CrossRef Neira VM, Al Madhoun W, Ghaffari K, Barrowman N, Berrigan P, Splinter W. Efficacy of malignant hyperthermia Association of the United States-Recommended Methods of preparation for malignant hyperthermia-susceptible patients using Drager Zeus anesthesia workstations and associated costs. Anesth Analg. 2019;129(1):74–83.CrossRef
4.
go back to reference Muller-Wirtz LM, Meiser A, Kermad A, Godsch C, Sessler DI, Volk T, et al. Response by the authors, reflection of volatile anesthetic by the vapor-clean filter. Acta Anaesthesiol Scand. 2020;64(9):1383–4.CrossRef Muller-Wirtz LM, Meiser A, Kermad A, Godsch C, Sessler DI, Volk T, et al. Response by the authors, reflection of volatile anesthetic by the vapor-clean filter. Acta Anaesthesiol Scand. 2020;64(9):1383–4.CrossRef
5.
go back to reference Cottron N, Larcher C, Sommet A, Fesseau R, Alacoque X, Minville V, et al. The sevoflurane washout profile of seven recent anesthesia workstations for malignant hyperthermia-susceptible adults and infants: a bench test study. Anesth Analg. 2014;119(1):67–75.CrossRef Cottron N, Larcher C, Sommet A, Fesseau R, Alacoque X, Minville V, et al. The sevoflurane washout profile of seven recent anesthesia workstations for malignant hyperthermia-susceptible adults and infants: a bench test study. Anesth Analg. 2014;119(1):67–75.CrossRef
6.
go back to reference Kim T, Tham R. GE Healthcare non-clinical internal lab test report number DOC0994043; 2011. Kim T, Tham R. GE Healthcare non-clinical internal lab test report number DOC0994043; 2011.
7.
go back to reference Thoben C, Dennhardt N, Krauß T, Sümpelmann R, Zimmermann S, Rüffert H, et al. Preparation of anaesthesia workstation for trigger-free anaesthesia: an observational laboratory study. Eur J Anaesthesiol. 2019;36(11):851–6.CrossRef Thoben C, Dennhardt N, Krauß T, Sümpelmann R, Zimmermann S, Rüffert H, et al. Preparation of anaesthesia workstation for trigger-free anaesthesia: an observational laboratory study. Eur J Anaesthesiol. 2019;36(11):851–6.CrossRef
8.
go back to reference Kirk AT, Bohnhorst A, Raddatz C-R, Allers M, Zimmermann S. Ultra-high-resolution ion mobility spectrometry—current instrumentation, limitations, and future developments. Anal Bioanal Chem. 2019;411(24):6229–46.CrossRef Kirk AT, Bohnhorst A, Raddatz C-R, Allers M, Zimmermann S. Ultra-high-resolution ion mobility spectrometry—current instrumentation, limitations, and future developments. Anal Bioanal Chem. 2019;411(24):6229–46.CrossRef
9.
go back to reference Speckbacher V, Zeilinger S, Zimmermann S, Mayhew CA, Wiesenhofer H, Ruzsanyi V. Monitoring the volatile language of fungi using gas chromatography-ion mobility spectrometry. Anal Bioanal Chem. 2021;413(11):3055–67.CrossRef Speckbacher V, Zeilinger S, Zimmermann S, Mayhew CA, Wiesenhofer H, Ruzsanyi V. Monitoring the volatile language of fungi using gas chromatography-ion mobility spectrometry. Anal Bioanal Chem. 2021;413(11):3055–67.CrossRef
10.
go back to reference Langejuergen J, Wagner C, Beutel S, Hopmeier T, Scheper T, Zimmermann S. Non-invasive monitoring of bacterial growth and auto-induced protein production in a bioreactor with a closed-loop GC-IMS. Int J Ion Mobil Spectrom. 2015;18(1–2):9–15.CrossRef Langejuergen J, Wagner C, Beutel S, Hopmeier T, Scheper T, Zimmermann S. Non-invasive monitoring of bacterial growth and auto-induced protein production in a bioreactor with a closed-loop GC-IMS. Int J Ion Mobil Spectrom. 2015;18(1–2):9–15.CrossRef
11.
go back to reference Heiderich S, Thoben C, Dennhardt N, Koppert W, Krauß T, Sümpelmann R, et al. Low anaesthetic waste gas concentrations in postanaesthesia care unit: a prospective observational study. Eur J Anaesthesiol. 2018;35(7):534–8.CrossRef Heiderich S, Thoben C, Dennhardt N, Koppert W, Krauß T, Sümpelmann R, et al. Low anaesthetic waste gas concentrations in postanaesthesia care unit: a prospective observational study. Eur J Anaesthesiol. 2018;35(7):534–8.CrossRef
Metadata
Title
Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients
Authors
Sebastian Heiderich
Christian Thoben
Nils Dennhardt
Terence Krauß
Robert Sümpelmann
Stefan Zimmermann
Michael Reitz
Henrik Rüffert
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2021
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-021-01533-0