Published in: International Journal of Implant Dentistry 1/2017

Open Access 01-12-2017 | Research

ISQ calculation evaluation of in vitro laser scanning vibrometry-captured resonance frequency

Authors: Stijn Debruyne, Nicolas Grognard, Gino Verleye, Korneel Van Massenhove, Dimitrios Mavreas, Bart Vande Vannet

Published in: International Journal of Implant Dentistry | Issue 1/2017

Login to get access

Abstract

Background

Implant stability testing at various stages of implant therapy by means of resonance frequency analysis is extensively used. The overall measurement outcome is a function of the resulting stiffness of three entities: surrounding bone, bone-implant complex, and implant-Smartpeg complex. The influence of the latter on the overall measurement results is presently unknown. It can be investigated in vitro by use of imbedded implants with mounted Smartpegs. This enables to keep the influence of the two other entities constant and controlled.
The purpose of this study is to verify if a laboratory laser Doppler vibrometry technology-based procedure results in comparable ISQ results after calculation of captured resonance frequency spectra by aid of the Osstell algorithm with direct Osstell IDX device measurements.

Methods

A laboratory procedure was engineered to record frequency spectra of resin-imbedded test implants with mounted Smartpegs, after electromagnetic excitation with the Osstell IDX device and laser Doppler vibrometry response detection. Fast Fourier transformation data processing of resonance frequency data resulted in determination of a maximum resonance frequency values allowing calculation of implant stability quotient (ISQ) values using the Osstell algorithm.

Results

Laboratory-based ISQ values were compared to Osstell IDx device-generated ISQ values for Straumann tissue level, Ankylos, and 3i Certain implant systems. For both systems, a correlation coefficient r = 0.99 was found. Furthermore, a clinically rejectable mean difference of 0.09 ISQ units was noted between both datasets.

Conclusions

The proposed laboratory method with the application of the Osstell algorithm for ISQ calculation is appropriate for future studies to in vitro research aspects of resonance frequency analysis implant stability measurements.
Literature
1.
go back to reference Adell R, Lekholm U, Brånemark PI. Surgical procedures. In: Brånemark PI, Zarb GA, Albrektsson T, editors. Tissue integrated prothese. Osseointegration in clinical dentistry. Chigaco: Quintessence; 1985. p. 211–32. Adell R, Lekholm U, Brånemark PI. Surgical procedures. In: Brånemark PI, Zarb GA, Albrektsson T, editors. Tissue integrated prothese. Osseointegration in clinical dentistry. Chigaco: Quintessence; 1985. p. 211–32.
2.
go back to reference Strid K. Radiographic procedures. In: Brånemark P-I, Zarb G, Albrektsson T, editors. Tissue integrated prothese. Osseointegration in clinical dentistry. Chigaco: Quintessence; 1985. p. 187–98. Strid K. Radiographic procedures. In: Brånemark P-I, Zarb G, Albrektsson T, editors. Tissue integrated prothese. Osseointegration in clinical dentistry. Chigaco: Quintessence; 1985. p. 187–98.
3.
go back to reference Johansson P, Strid K. Assessment of bone quality from cutting resistance during implant surgery. Int J Oral Maxillofac Implants. 1994;5:279–88. Johansson P, Strid K. Assessment of bone quality from cutting resistance during implant surgery. Int J Oral Maxillofac Implants. 1994;5:279–88.
4.
go back to reference Sullivan D, Sherwood R, Collins T, Krogh P. The reverse torque test: a clinical report. Int J Oral Maxillofac Implants. 1990;5:179–85. Sullivan D, Sherwood R, Collins T, Krogh P. The reverse torque test: a clinical report. Int J Oral Maxillofac Implants. 1990;5:179–85.
5.
go back to reference Olivé J, Aparicio C. The periotest as a measure of osseo-integrated implant stability. Int J Oral Maxillofac Implants. 1990;5:309–400. Olivé J, Aparicio C. The periotest as a measure of osseo-integrated implant stability. Int J Oral Maxillofac Implants. 1990;5:309–400.
6.
go back to reference Van Steenberghe D, Tricio J, Naert I, Nys M. Damping characteristics of bone-to-implant interfaces. A clinical study with the Periotest device. Clin Oral Implants Res. 1995;6:31–9.CrossRefPubMed Van Steenberghe D, Tricio J, Naert I, Nys M. Damping characteristics of bone-to-implant interfaces. A clinical study with the Periotest device. Clin Oral Implants Res. 1995;6:31–9.CrossRefPubMed
7.
go back to reference Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency analysis to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Impl Res. 1997;8:234.CrossRef Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency analysis to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Impl Res. 1997;8:234.CrossRef
8.
go back to reference Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1997;24:568–72. Meredith N, Alleyne D, Cawley P. Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res. 1997;24:568–72.
9.
go back to reference Aparicio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of the implant/bone interface. Clin Oral Impl Res. 2006;17(suppl 2):2–7.CrossRef Aparicio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of the implant/bone interface. Clin Oral Impl Res. 2006;17(suppl 2):2–7.CrossRef
10.
go back to reference Barewal R, Oates T, Meredith N, Cochran D. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants. 2003;18:641–51.PubMed Barewal R, Oates T, Meredith N, Cochran D. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants. 2003;18:641–51.PubMed
11.
go back to reference Bischof M, Nedir R, Szmuckler-Moncler S, Bernard JP, Samson S. Implant stability of delayed and immediately loaded implants of delayed and immediately loaded implants during healing. A clinical RFA study with SLA ITI implants. Clin Oral Impl Res. 2003;15:529–39.CrossRef Bischof M, Nedir R, Szmuckler-Moncler S, Bernard JP, Samson S. Implant stability of delayed and immediately loaded implants of delayed and immediately loaded implants during healing. A clinical RFA study with SLA ITI implants. Clin Oral Impl Res. 2003;15:529–39.CrossRef
12.
go back to reference Huwiler M, Pjetursson D, Bosshardt D, Salvi G, Lang NP. Resonance frequency analysis in relation to jawbone characteristics qnd during early healing of implant installation. Clin Oral Impl Res. 2006;18:275–80.CrossRef Huwiler M, Pjetursson D, Bosshardt D, Salvi G, Lang NP. Resonance frequency analysis in relation to jawbone characteristics qnd during early healing of implant installation. Clin Oral Impl Res. 2006;18:275–80.CrossRef
13.
go back to reference Han J, Lulic M, Lang NP. Factors influencing resonance frequency analyis assessed by Osstell Mentor during implant issue integration: II: implant surface modifications and implant diameter. Clin Oral Implants Res. 2010;6:605–11.CrossRef Han J, Lulic M, Lang NP. Factors influencing resonance frequency analyis assessed by Osstell Mentor during implant issue integration: II: implant surface modifications and implant diameter. Clin Oral Implants Res. 2010;6:605–11.CrossRef
14.
go back to reference Bornstein M, Hart C, Halbritter S, Morton D, Buser D. Early loading of nonsubmerged titanium implants with a chemically modified sand-blasted and acid-etched surface: 6-month results of a prospective case series study in the posterior mandible focusing on peri-implant crestal bone changes and implant stability quotien (ISQ) values. Clin Implant Dent Relat Res. 2009;11(4):338–47.CrossRefPubMed Bornstein M, Hart C, Halbritter S, Morton D, Buser D. Early loading of nonsubmerged titanium implants with a chemically modified sand-blasted and acid-etched surface: 6-month results of a prospective case series study in the posterior mandible focusing on peri-implant crestal bone changes and implant stability quotien (ISQ) values. Clin Implant Dent Relat Res. 2009;11(4):338–47.CrossRefPubMed
15.
go back to reference Guler A, Sumer M, Duran I, Sandikci E, Telcioglu N. Resonance frequency analysis of 208 Straumann dental implants during the healing period. J Oral Impl. 2013;39(2):161–7.CrossRef Guler A, Sumer M, Duran I, Sandikci E, Telcioglu N. Resonance frequency analysis of 208 Straumann dental implants during the healing period. J Oral Impl. 2013;39(2):161–7.CrossRef
16.
go back to reference Valderrama P, Oates T, Jones A, Simpson J, Schoolfield J, Cochran D. Evaluation of two different resonance frequency devices to detect implant stability: a clinical trial. J Periodontol. 2007;2:262–72.CrossRef Valderrama P, Oates T, Jones A, Simpson J, Schoolfield J, Cochran D. Evaluation of two different resonance frequency devices to detect implant stability: a clinical trial. J Periodontol. 2007;2:262–72.CrossRef
17.
go back to reference Tözüm T, Bal B, Turkyilmaz I, Gülay G, Tulunoglu I. Which device is more accurate to determine the stability/mobility of dental implants? A human cadaver study. J Oral Rehabil. 2010;37:217–24.CrossRefPubMed Tözüm T, Bal B, Turkyilmaz I, Gülay G, Tulunoglu I. Which device is more accurate to determine the stability/mobility of dental implants? A human cadaver study. J Oral Rehabil. 2010;37:217–24.CrossRefPubMed
18.
go back to reference Pagliani L, Sennersby L, Petersson A, Verrochi D. The relationship between resonance frequency analysis (RFA) and lateral displacement of dental implants: an in vitro study. J of OralRehabilitation. 2013;40:221–7. Pagliani L, Sennersby L, Petersson A, Verrochi D. The relationship between resonance frequency analysis (RFA) and lateral displacement of dental implants: an in vitro study. J of OralRehabilitation. 2013;40:221–7.
19.
go back to reference Haw-Ming H, Ching-Lai C, Ching-Ching Y, Sheng-Yang L. Factors influencing the resonance frequency of dental implants. J Oral Maxillofac Surg. 2003;61:1184–8.CrossRef Haw-Ming H, Ching-Lai C, Ching-Ching Y, Sheng-Yang L. Factors influencing the resonance frequency of dental implants. J Oral Maxillofac Surg. 2003;61:1184–8.CrossRef
Metadata
Title
ISQ calculation evaluation of in vitro laser scanning vibrometry-captured resonance frequency
Authors
Stijn Debruyne
Nicolas Grognard
Gino Verleye
Korneel Van Massenhove
Dimitrios Mavreas
Bart Vande Vannet
Publication date
01-12-2017
Publisher
Springer Berlin Heidelberg
Published in
International Journal of Implant Dentistry / Issue 1/2017
Electronic ISSN: 2198-4034
DOI
https://doi.org/10.1186/s40729-017-0105-3

Other articles of this Issue 1/2017

International Journal of Implant Dentistry 1/2017 Go to the issue