Skip to main content
Top
Published in: European Journal of Medical Research 1/2017

Open Access 01-12-2017 | Reaearch

Improved success rates using videolaryngoscopy in unexperienced users: a randomized crossover study in airway manikins

Authors: Hendrik Eismann, Lion Sieg, Nicola Etti, Lars Friedrich, Christian Schröter, Philipp Mommsen, Christian Krettek, Christian Zeckey

Published in: European Journal of Medical Research | Issue 1/2017

Login to get access

Abstract

Background

Videolaryngoscopy has been proven to be a safe procedure managing difficult airways in the hands of airway specialists. Information about the success rates in unexperienced users of videolaryngoscopy compared to conventional laryngoscopy is sparse. Therefore, we aimed to evaluate if there might be more success in securing an airway if the unexperienced provider is using a videolaryngoscope in simulated airways in a randomized manikin study. Differences between commonly used videolaryngoscopes were elucidated.

Methods

A standardized hands-on workshop prior to the study was performed. For direct laryngoscopy (DL) we used a Macintosh laryngoscope, whereas for videolaryngoscopy (VL) we used the cMac, the dBlade, and a King Vision videolaryngoscope. Endotracheal intubations in three simulated normal and difficult airways were performed. Main outcome parameters were time to view and time to intubation. Cormack and Lehane (C + L) classification and the percentage of glottic opening (POGO) score were evaluated. After every intubation, the participants were asked to review the airway and the device used.

Results

22 participants (14.8 ± 4.0 intubations per year, mostly trauma surgeons) with limited experience in videolaryngoscopy (mean total number of videolaryngoscopy .4 ± .2) were enrolled. We found improved C + L grades with VL in contrast to DL. We saw similar data with respect to the POGO score, where the participants achieved better visibility of the glottis with VL. The hyperangulated blade geometries of videolaryngoscopes provided a better visibility in difficult airways than the standard geometry of the Macintosh-type blade. The subjective performance of the VL devices was better in more difficult airway scenarios.

Conclusions

After a short introduction and hands-on training, a videolaryngoscope seems to be safe and usable by unexperienced providers. We assume a standard geometry laryngoscope is optimal for a patient with normal anatomy, whereas VL device with a hyperangulated blade is ideal for difficult airway situations with limited mouth opening or restricted neck movement.
Literature
1.
go back to reference Lavery GG, McCloskey BV. The difficult airway in adult critical care. Crit Care Med. 2008;36:2163–73.CrossRefPubMed Lavery GG, McCloskey BV. The difficult airway in adult critical care. Crit Care Med. 2008;36:2163–73.CrossRefPubMed
2.
go back to reference Adnet F, Racine SX, Borron SW, Clemessy JL, Fournier JL, Lapostolle F, Cupa M. A survey of tracheal intubation difficulty in the operating room: a prospective observational study. Acta Anaesthesiol Scand. 2001;45:327–32.CrossRefPubMed Adnet F, Racine SX, Borron SW, Clemessy JL, Fournier JL, Lapostolle F, Cupa M. A survey of tracheal intubation difficulty in the operating room: a prospective observational study. Acta Anaesthesiol Scand. 2001;45:327–32.CrossRefPubMed
3.
go back to reference Cobas MA, De la Pena MA, Manning R, Candiotti K, Varon AJ. Prehospital intubations and mortality: a level 1 trauma center perspective. Anesth Analg. 2009;109:489–93.CrossRefPubMed Cobas MA, De la Pena MA, Manning R, Candiotti K, Varon AJ. Prehospital intubations and mortality: a level 1 trauma center perspective. Anesth Analg. 2009;109:489–93.CrossRefPubMed
4.
go back to reference Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesth Analg. 2004;99:607–13.CrossRefPubMed Mort TC. Emergency tracheal intubation: complications associated with repeated laryngoscopic attempts. Anesth Analg. 2004;99:607–13.CrossRefPubMed
5.
go back to reference Koh JC, Lee JS, Lee YW, Chang CH. Comparison of the laryngeal view during intubation using Airtraq and Macintosh laryngoscopes in patients with cervical spine immobilization and mouth opening limitation. Korean J Anesthesiol. 2010;59:314–8.CrossRefPubMedPubMedCentral Koh JC, Lee JS, Lee YW, Chang CH. Comparison of the laryngeal view during intubation using Airtraq and Macintosh laryngoscopes in patients with cervical spine immobilization and mouth opening limitation. Korean J Anesthesiol. 2010;59:314–8.CrossRefPubMedPubMedCentral
6.
go back to reference Lim Y, Yeo SW. A comparison of the GlideScope with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway. Anaesth Intensive Care. 2005;33:243–7.PubMed Lim Y, Yeo SW. A comparison of the GlideScope with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway. Anaesth Intensive Care. 2005;33:243–7.PubMed
7.
go back to reference Malik MA, Maharaj CH, Harte BH, Laffey JG. Comparison of Macintosh, Truview EVO2, Glidescope, and Airwayscope laryngoscope use in patients with cervical spine immobilization. Br J Anaesth. 2008;101:723–30.CrossRefPubMed Malik MA, Maharaj CH, Harte BH, Laffey JG. Comparison of Macintosh, Truview EVO2, Glidescope, and Airwayscope laryngoscope use in patients with cervical spine immobilization. Br J Anaesth. 2008;101:723–30.CrossRefPubMed
8.
go back to reference Serocki G, Bein B, Scholz J, Dorges V. Management of the predicted difficult airway: a comparison of conventional blade laryngoscopy with video-assisted blade laryngoscopy and the GlideScope. Eur J Anaesthesiol. 2010;27:24–30.CrossRefPubMed Serocki G, Bein B, Scholz J, Dorges V. Management of the predicted difficult airway: a comparison of conventional blade laryngoscopy with video-assisted blade laryngoscopy and the GlideScope. Eur J Anaesthesiol. 2010;27:24–30.CrossRefPubMed
9.
go back to reference Timmermann A, Byhahn C, Wenzel V, Eich C, Piepho T, Bernhard M, Dörges V. Allgemeine Prinzipien der Notfallmedizin. Notfallmedizin up2date. 2012;7:105–20.CrossRef Timmermann A, Byhahn C, Wenzel V, Eich C, Piepho T, Bernhard M, Dörges V. Allgemeine Prinzipien der Notfallmedizin. Notfallmedizin up2date. 2012;7:105–20.CrossRef
10.
go back to reference Gaither JB, Spaite DW, Stolz U, Ennis J, Mosier J, Sakles JJ. Prevalence of difficult airway predictors in cases of failed prehospital endotracheal intubation. J Emerg Med. 2014;47:294–300.CrossRefPubMed Gaither JB, Spaite DW, Stolz U, Ennis J, Mosier J, Sakles JJ. Prevalence of difficult airway predictors in cases of failed prehospital endotracheal intubation. J Emerg Med. 2014;47:294–300.CrossRefPubMed
11.
go back to reference Levitan RM, Pisaturo JT, Kinkle WC, Butler K, Everett WW. Stylet bend angles and tracheal tube passage using a straight-to-cuff shape. Acad Emerg Med. 2006;13:1255–8.CrossRefPubMed Levitan RM, Pisaturo JT, Kinkle WC, Butler K, Everett WW. Stylet bend angles and tracheal tube passage using a straight-to-cuff shape. Acad Emerg Med. 2006;13:1255–8.CrossRefPubMed
12.
go back to reference el-Ganzouri AR, McCarthy RJ, Tuman KJ, Tanck EN, Ivankovich AD. Preoperative airway assessment: predictive value of a multivariate risk index. Anesth Analg. 1996;82:1197–204.PubMed el-Ganzouri AR, McCarthy RJ, Tuman KJ, Tanck EN, Ivankovich AD. Preoperative airway assessment: predictive value of a multivariate risk index. Anesth Analg. 1996;82:1197–204.PubMed
13.
go back to reference Cormack RS, Lehane JR, Adams AP, Carli F. Laryngoscopy grades and percentage glottic opening. Anaesthesia. 2000;55:184.CrossRefPubMed Cormack RS, Lehane JR, Adams AP, Carli F. Laryngoscopy grades and percentage glottic opening. Anaesthesia. 2000;55:184.CrossRefPubMed
14.
go back to reference Ochroch EA, Hollander JE, Kush S, Shofer FS, Levitan RM. Assessment of laryngeal view: percentage of glottic opening score vs Cormack and Lehane grading. Can J Anaesth. 1999;46:987–90.CrossRefPubMed Ochroch EA, Hollander JE, Kush S, Shofer FS, Levitan RM. Assessment of laryngeal view: percentage of glottic opening score vs Cormack and Lehane grading. Can J Anaesth. 1999;46:987–90.CrossRefPubMed
15.
go back to reference Chi JH, Knudson MM, Vassar MJ, McCarthy MC, Shapiro MB, Mallet S, Holcroft JJ, Moncrief H, Noble J, Wisner D, et al. Prehospital hypoxia affects outcome in patients with traumatic brain injury: a prospective multicenter study. J Trauma. 2006;61:1134–41.CrossRefPubMed Chi JH, Knudson MM, Vassar MJ, McCarthy MC, Shapiro MB, Mallet S, Holcroft JJ, Moncrief H, Noble J, Wisner D, et al. Prehospital hypoxia affects outcome in patients with traumatic brain injury: a prospective multicenter study. J Trauma. 2006;61:1134–41.CrossRefPubMed
17.
go back to reference Malik MA, O’Donoghue C, Carney J, Maharaj CH, Harte BH, Laffey JG. Comparison of the Glidescope, the Pentax AWS, and the Truview EVO2 with the Macintosh laryngoscope in experienced anaesthetists: a manikin study. Br J Anaesth. 2009;102:128–34.CrossRefPubMed Malik MA, O’Donoghue C, Carney J, Maharaj CH, Harte BH, Laffey JG. Comparison of the Glidescope, the Pentax AWS, and the Truview EVO2 with the Macintosh laryngoscope in experienced anaesthetists: a manikin study. Br J Anaesth. 2009;102:128–34.CrossRefPubMed
18.
go back to reference Serocki G, Neumann T, Scharf E, Dorges V, Cavus E. Indirect videolaryngoscopy with C-MAC D-Blade and GlideScope: a randomized, controlled comparison in patients with suspected difficult airways. Minerva anestesiologica. 2013;79:121–9.PubMed Serocki G, Neumann T, Scharf E, Dorges V, Cavus E. Indirect videolaryngoscopy with C-MAC D-Blade and GlideScope: a randomized, controlled comparison in patients with suspected difficult airways. Minerva anestesiologica. 2013;79:121–9.PubMed
19.
go back to reference Szarpak L, Smereka J, Ladny JR. Comparison of Macintosh and Intubrite laryngoscopes for intubation performed by novice physicians in a difficult airway scenario. Am J Emerg Med. 2017;35:796–7.CrossRefPubMed Szarpak L, Smereka J, Ladny JR. Comparison of Macintosh and Intubrite laryngoscopes for intubation performed by novice physicians in a difficult airway scenario. Am J Emerg Med. 2017;35:796–7.CrossRefPubMed
20.
go back to reference Lee RA, van Zundert AA, Maassen RL, Willems RJ, Beeke LP, Schaaper JN, van Dobbelsteen J, Wieringa PA. Forces applied to the maxillary incisors during video-assisted intubation. Anesth Analg. 2009;108:187–91.CrossRefPubMed Lee RA, van Zundert AA, Maassen RL, Willems RJ, Beeke LP, Schaaper JN, van Dobbelsteen J, Wieringa PA. Forces applied to the maxillary incisors during video-assisted intubation. Anesth Analg. 2009;108:187–91.CrossRefPubMed
21.
go back to reference Maassen R, Lee R, van Zundert A, Cooper R. The videolaryngoscope is less traumatic than the classic laryngoscope for a difficult airway in an obese patient. J Anesth. 2009;23:445–8.CrossRefPubMed Maassen R, Lee R, van Zundert A, Cooper R. The videolaryngoscope is less traumatic than the classic laryngoscope for a difficult airway in an obese patient. J Anesth. 2009;23:445–8.CrossRefPubMed
22.
go back to reference van Zundert A, Maassen R, Lee R, Willems R, Timmerman M, Siemonsma M, Buise M, Wiepking M. A Macintosh laryngoscope blade for videolaryngoscopy reduces stylet use in patients with normal airways. Anesth Analg. 2009;109:825–31.CrossRefPubMed van Zundert A, Maassen R, Lee R, Willems R, Timmerman M, Siemonsma M, Buise M, Wiepking M. A Macintosh laryngoscope blade for videolaryngoscopy reduces stylet use in patients with normal airways. Anesth Analg. 2009;109:825–31.CrossRefPubMed
23.
go back to reference Maassen R, Lee R, Hermans B, Marcus M, van Zundert A. A comparison of three videolaryngoscopes: the Macintosh laryngoscope blade reduces, but does not replace, routine stylet use for intubation in morbidly obese patients. Anesth Analg. 2009;109:1560–5.CrossRefPubMed Maassen R, Lee R, Hermans B, Marcus M, van Zundert A. A comparison of three videolaryngoscopes: the Macintosh laryngoscope blade reduces, but does not replace, routine stylet use for intubation in morbidly obese patients. Anesth Analg. 2009;109:1560–5.CrossRefPubMed
24.
go back to reference Raymondos K, Seidel T, Sander B, Gerdes A, Goetz F, Helmstadter V, Panning B, Dieck T. The intubation scoop (i-scoop)-a new type of laryngoscope for difficult and normal airways. Anaesthesia. 2014;69:990–1001.CrossRefPubMed Raymondos K, Seidel T, Sander B, Gerdes A, Goetz F, Helmstadter V, Panning B, Dieck T. The intubation scoop (i-scoop)-a new type of laryngoscope for difficult and normal airways. Anaesthesia. 2014;69:990–1001.CrossRefPubMed
25.
go back to reference McElwain J, Malik MA, Harte BH, Flynn NM, Laffey JG. Comparison of the C-MAC videolaryngoscope with the Macintosh, Glidescope, and Airtraq laryngoscopes in easy and difficult laryngoscopy scenarios in manikins. Anaesthesia. 2010;65:483–9.CrossRefPubMed McElwain J, Malik MA, Harte BH, Flynn NM, Laffey JG. Comparison of the C-MAC videolaryngoscope with the Macintosh, Glidescope, and Airtraq laryngoscopes in easy and difficult laryngoscopy scenarios in manikins. Anaesthesia. 2010;65:483–9.CrossRefPubMed
26.
go back to reference Hagiwara Y, Watase H, Okamoto H, Goto T, Hasegawa K, Japanese Emergency Medicine Network I. Prospective validation of the modified LEMON criteria to predict difficult intubation in the ED. Am J Emerg Med. 2015;33:1492–6.CrossRefPubMed Hagiwara Y, Watase H, Okamoto H, Goto T, Hasegawa K, Japanese Emergency Medicine Network I. Prospective validation of the modified LEMON criteria to predict difficult intubation in the ED. Am J Emerg Med. 2015;33:1492–6.CrossRefPubMed
27.
go back to reference Truszewski Z, Czyzewski L, Smereka J, Krajewski P, Fudalej M, Madziala M, Szarpak L. Ability of paramedics to perform endotracheal intubation during continuous chest compressions: a randomized cadaver study comparing Pentax AWS and Macintosh laryngoscopes. Am J Emerg Med. 2016;34:1835–9.CrossRefPubMed Truszewski Z, Czyzewski L, Smereka J, Krajewski P, Fudalej M, Madziala M, Szarpak L. Ability of paramedics to perform endotracheal intubation during continuous chest compressions: a randomized cadaver study comparing Pentax AWS and Macintosh laryngoscopes. Am J Emerg Med. 2016;34:1835–9.CrossRefPubMed
28.
go back to reference Schebesta K, Hupfl M, Rossler B, Ringl H, Muller MP, Kimberger O. Degrees of reality: airway anatomy of high-fidelity human patient simulators and airway trainers. Anesthesiology. 2012;116:1204–9.CrossRefPubMed Schebesta K, Hupfl M, Rossler B, Ringl H, Muller MP, Kimberger O. Degrees of reality: airway anatomy of high-fidelity human patient simulators and airway trainers. Anesthesiology. 2012;116:1204–9.CrossRefPubMed
29.
go back to reference Benumof JL, Dagg R, Benumof R. Critical hemoglobin desaturation will occur before return to an unparalyzed state following 1 mg/kg intravenous succinylcholine. Anesthesiology. 1997;87:979–82.CrossRefPubMed Benumof JL, Dagg R, Benumof R. Critical hemoglobin desaturation will occur before return to an unparalyzed state following 1 mg/kg intravenous succinylcholine. Anesthesiology. 1997;87:979–82.CrossRefPubMed
30.
go back to reference Trimmel H, Kreutziger J, Fertsak G, Fitzka R, Dittrich M, Voelckel WG. Use of the Airtraq laryngoscope for emergency intubation in the prehospital setting: a randomized control trial. Crit Care Med. 2011;39:489–93.CrossRefPubMed Trimmel H, Kreutziger J, Fertsak G, Fitzka R, Dittrich M, Voelckel WG. Use of the Airtraq laryngoscope for emergency intubation in the prehospital setting: a randomized control trial. Crit Care Med. 2011;39:489–93.CrossRefPubMed
31.
go back to reference Sakles JC, Mosier J, Chiu S, Cosentino M, Kalin L. A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department. Ann Emerg Med. 2012;60:739–48.CrossRefPubMedPubMedCentral Sakles JC, Mosier J, Chiu S, Cosentino M, Kalin L. A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department. Ann Emerg Med. 2012;60:739–48.CrossRefPubMedPubMedCentral
32.
go back to reference Noppens RR, Geimer S, Eisel N, David M, Piepho T. Endotracheal intubation using the C-MAC(R) video laryngoscope or the Macintosh laryngoscope: a prospective, comparative study in the ICU. Crit Care. 2012;16:R103.CrossRefPubMedPubMedCentral Noppens RR, Geimer S, Eisel N, David M, Piepho T. Endotracheal intubation using the C-MAC(R) video laryngoscope or the Macintosh laryngoscope: a prospective, comparative study in the ICU. Crit Care. 2012;16:R103.CrossRefPubMedPubMedCentral
Metadata
Title
Improved success rates using videolaryngoscopy in unexperienced users: a randomized crossover study in airway manikins
Authors
Hendrik Eismann
Lion Sieg
Nicola Etti
Lars Friedrich
Christian Schröter
Philipp Mommsen
Christian Krettek
Christian Zeckey
Publication date
01-12-2017
Publisher
BioMed Central
Published in
European Journal of Medical Research / Issue 1/2017
Electronic ISSN: 2047-783X
DOI
https://doi.org/10.1186/s40001-017-0268-7

Other articles of this Issue 1/2017

European Journal of Medical Research 1/2017 Go to the issue