Published in: Annals of Intensive Care 1/2017

Open Access 01-12-2017 | Research

A new global and comprehensive model for ICU ventilator performances evaluation

Authors: Nicolas S. Marjanovic, Agathe De Simone, Guillaume Jegou, Erwan L’Her

Published in: Annals of Intensive Care | Issue 1/2017

Login to get access

Abstract

Background

This study aimed to provide a new global and comprehensive evaluation of recent ICU ventilators taking into account both technical performances and ergonomics.

Methods

Six recent ICU ventilators were evaluated. Technical performances were assessed under two FIO2 levels (100%, 50%), three respiratory mechanics combinations (Normal: compliance [C] = 70 mL cmH2O−1/resistance [R] = 5 cmH2O L−1 s−1; Restrictive: C = 30/R = 10; Obstructive: C = 120/R = 20), four exponential levels of leaks (from 0 to 12.5 L min−1) and three levels of inspiratory effort (P0.1 = 2, 4 and 8 cmH2O), using an automated test lung. Ergonomics were evaluated by 20 ICU physicians using a global and comprehensive model involving physiological response to stress measurements (heart rate, respiratory rate, tidal volume variability and eye tracking), psycho-cognitive scales (SUS and NASA-TLX) and objective tasks completion.

Results

Few differences in terms of technical performance were observed between devices. Non-invasive ventilation modes had a huge influence on asynchrony occurrence. Using our global model, either objective tasks completion, psycho-cognitive scales and/or physiological measurements were able to depict significant differences in terms of devices’ usability. The level of failure that was observed with some devices depicted the lack of adaptation of device’s development to end users’ requests.

Conclusions

Despite similar technical performance, some ICU ventilators exhibit low ergonomics performance and a high risk of misusage.
Appendix
Available only for authorised users
Literature
1.
go back to reference The Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. NEJM. 2000;342:1301–8.CrossRef The Acute Respiratory Distress Syndrome Network. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. NEJM. 2000;342:1301–8.CrossRef
2.
go back to reference Mercat A, Richard JC, Vielle B, Jaber S, Osman D, Diehl JL, et al. Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2008;299:646–55.CrossRefPubMed Mercat A, Richard JC, Vielle B, Jaber S, Osman D, Diehl JL, et al. Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2008;299:646–55.CrossRefPubMed
3.
go back to reference Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2000;32:1515–22.CrossRef Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2000;32:1515–22.CrossRef
4.
go back to reference Chatburn RL, El-Khatib M, Mireles-Cabodevila E. A taxonomy for mechanical ventilation: 10 fundamental maxims. Respir Care. 2014;59:1747–63.CrossRefPubMed Chatburn RL, El-Khatib M, Mireles-Cabodevila E. A taxonomy for mechanical ventilation: 10 fundamental maxims. Respir Care. 2014;59:1747–63.CrossRefPubMed
5.
go back to reference Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, et al. Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med. 2002;28:1049–57.CrossRefPubMed Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, et al. Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med. 2002;28:1049–57.CrossRefPubMed
6.
go back to reference Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC. Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med. 2010;36:2074–80.CrossRefPubMed Lyazidi A, Thille AW, Carteaux G, Galia F, Brochard L, Richard JC. Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation. Intensive Care Med. 2010;36:2074–80.CrossRefPubMed
7.
go back to reference Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L. A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med. 2009;35:1368–76.CrossRefPubMedPubMedCentral Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L. A bench study of intensive-care-unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med. 2009;35:1368–76.CrossRefPubMedPubMedCentral
9.
go back to reference Boussen S, Gainnier M, Michelet P. Evaluation of ventilators used during transport of critically Ill patients: a bench study. Respir Care. 2013;58:1911–22.CrossRefPubMed Boussen S, Gainnier M, Michelet P. Evaluation of ventilators used during transport of critically Ill patients: a bench study. Respir Care. 2013;58:1911–22.CrossRefPubMed
10.
go back to reference Institute of Medicine. To err is human: building a safer health system. Washington: The National Academies Press; 1999. Institute of Medicine. To err is human: building a safer health system. Washington: The National Academies Press; 1999.
11.
go back to reference Horsky J, Zhang J, Patel VL. To err is not entirely human: complex technology and user cognition. J Biomed Inform. 2005;38:264–6.CrossRefPubMed Horsky J, Zhang J, Patel VL. To err is not entirely human: complex technology and user cognition. J Biomed Inform. 2005;38:264–6.CrossRefPubMed
12.
go back to reference Richard JC, Kacmarek RM. ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers. Intensive Care Med. 2009;35:1662–3.CrossRefPubMed Richard JC, Kacmarek RM. ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers. Intensive Care Med. 2009;35:1662–3.CrossRefPubMed
13.
go back to reference Templier F, Miroux P, Dolveck F, Descatha A, Goddet NS, Jeleff C, et al. Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators. Respir Care. 2007;52:1701–9.PubMed Templier F, Miroux P, Dolveck F, Descatha A, Goddet NS, Jeleff C, et al. Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators. Respir Care. 2007;52:1701–9.PubMed
14.
go back to reference Uzawa Y, Yamada Y, Suzukawa M. Evaluation of the user interface simplicity in the modern generation of mechanical ventilators. Respir Care. 2008;53:329–37.PubMed Uzawa Y, Yamada Y, Suzukawa M. Evaluation of the user interface simplicity in the modern generation of mechanical ventilators. Respir Care. 2008;53:329–37.PubMed
15.
go back to reference Vignaux L, Tassaux D, Jolliet P. Evaluation of the user-friendliness of seven new generation intensive care ventilators. Intensive Care Med. 2009;35:1687–91.CrossRefPubMed Vignaux L, Tassaux D, Jolliet P. Evaluation of the user-friendliness of seven new generation intensive care ventilators. Intensive Care Med. 2009;35:1687–91.CrossRefPubMed
16.
go back to reference Gonzalez-Bermejo J, Laplanche V, Husseini FE, Duguet A, Derenne JP, Similowski T. Evaluation of the user-friendliness of 11 home mechanical ventilators. Eur Respir J. 2006;27:1236–43.CrossRefPubMed Gonzalez-Bermejo J, Laplanche V, Husseini FE, Duguet A, Derenne JP, Similowski T. Evaluation of the user-friendliness of 11 home mechanical ventilators. Eur Respir J. 2006;27:1236–43.CrossRefPubMed
17.
go back to reference Garnier M, Quesnel C, Fulgencio JP, Degrain M, Carteaux G, Bonnet F, et al. Multifaceted bench comparative evaluation of latest intensive care unit ventilators. Br J Anaesth. 2015;115:89–98.CrossRefPubMed Garnier M, Quesnel C, Fulgencio JP, Degrain M, Carteaux G, Bonnet F, et al. Multifaceted bench comparative evaluation of latest intensive care unit ventilators. Br J Anaesth. 2015;115:89–98.CrossRefPubMed
18.
go back to reference Chao DC, Scheinhorn DJ, Stearn-Hassenpflug M. Patient-ventilator trigger asynchrony in prolonged mechanical ventilation. Chest. 1997;112:1592–9.CrossRefPubMed Chao DC, Scheinhorn DJ, Stearn-Hassenpflug M. Patient-ventilator trigger asynchrony in prolonged mechanical ventilation. Chest. 1997;112:1592–9.CrossRefPubMed
19.
go back to reference Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, et al. Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest. 2004;126:851–9.CrossRefPubMed Vitacca M, Bianchi L, Zanotti E, Vianello A, Barbano L, Porta R, et al. Assessment of physiologic variables and subjective comfort under different levels of pressure support ventilation. Chest. 2004;126:851–9.CrossRefPubMed
20.
go back to reference Brooke J. SUS: A quick and dirty usability scale. In: Jordan PW, Weerdmeester B, Thomas A, McLelland IL, editors. Usability evaluation in industry. London: Taylor and Francis; 1996. Brooke J. SUS: A quick and dirty usability scale. In: Jordan PW, Weerdmeester B, Thomas A, McLelland IL, editors. Usability evaluation in industry. London: Taylor and Francis; 1996.
21.
go back to reference Vignaux L, Vargas F, Roeseler J, Tassaux D, Thille AW, Kossowsky MP, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med. 2009;35:840–6.CrossRefPubMed Vignaux L, Vargas F, Roeseler J, Tassaux D, Thille AW, Kossowsky MP, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med. 2009;35:840–6.CrossRefPubMed
22.
go back to reference Vignaux L, Tassaux D, Jolliet P. Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med. 2007;33:1444–51.CrossRefPubMed Vignaux L, Tassaux D, Jolliet P. Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med. 2007;33:1444–51.CrossRefPubMed
23.
go back to reference Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, et al. Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. 2012;142:367–76.CrossRefPubMed Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, et al. Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. 2012;142:367–76.CrossRefPubMed
24.
go back to reference Giraud T, Dhainaut JF, Vaxelaire JF, Joseph T, Journois D. Bleichner G Iatrogenic complications in adult intensive care units: a prospective two-center study. CCM. 1993;21:40–51.CrossRef Giraud T, Dhainaut JF, Vaxelaire JF, Joseph T, Journois D. Bleichner G Iatrogenic complications in adult intensive care units: a prospective two-center study. CCM. 1993;21:40–51.CrossRef
25.
go back to reference Bracco D, Favre JB, Bissonnette B, Wasserfallen JB, Revelly JP, Ravussin P, et al. Human errors in a multidisciplinary intensive care unit: a 1-year prospective study. Intensive Care Med. 2001;27:137–45.CrossRefPubMed Bracco D, Favre JB, Bissonnette B, Wasserfallen JB, Revelly JP, Ravussin P, et al. Human errors in a multidisciplinary intensive care unit: a 1-year prospective study. Intensive Care Med. 2001;27:137–45.CrossRefPubMed
26.
27.
go back to reference Cowan N. The magical number 4 in short-term memory: a reconsideration of mental storage capacity. Behav Brain Sci. 2001;24:87–114 (discussion 114–185).CrossRefPubMed Cowan N. The magical number 4 in short-term memory: a reconsideration of mental storage capacity. Behav Brain Sci. 2001;24:87–114 (discussion 114–185).CrossRefPubMed
28.
go back to reference Morita PP, Weinstein PB, Flewwelling CJ, Banez CA, Chiu TA, Iannuzzi M, et al. The usability of ventilators: a comparative evaluation of use safety and user experience. Crit Care. 2016;20:263.CrossRefPubMedPubMedCentral Morita PP, Weinstein PB, Flewwelling CJ, Banez CA, Chiu TA, Iannuzzi M, et al. The usability of ventilators: a comparative evaluation of use safety and user experience. Crit Care. 2016;20:263.CrossRefPubMedPubMedCentral
29.
go back to reference Hess EH, Polt JM. Pupil size in relation to mental activity during simple problem-solving. Science. 1964;143:1190–2.CrossRefPubMed Hess EH, Polt JM. Pupil size in relation to mental activity during simple problem-solving. Science. 1964;143:1190–2.CrossRefPubMed
30.
go back to reference Beatty J, Kahneman D. Pupillary changes in two memory tasks. Psychon Sci. 2013;5:371–2.CrossRef Beatty J, Kahneman D. Pupillary changes in two memory tasks. Psychon Sci. 2013;5:371–2.CrossRef
31.
go back to reference Chen S, Epps J, Chen F. An investigation of pupil-based cognitive load measurement with low cost infrared webcam under light reflex interference. In: Conference proceedings: annual international conference of the IEEE engineering in medicine and biology society IEEE engineering in medicine and biology society annual conference. 2013. p. 3202–05. Chen S, Epps J, Chen F. An investigation of pupil-based cognitive load measurement with low cost infrared webcam under light reflex interference. In: Conference proceedings: annual international conference of the IEEE engineering in medicine and biology society IEEE engineering in medicine and biology society annual conference. 2013. p. 3202–05.
32.
go back to reference Boiten FA. The effects of emotional behaviour on components of the respiratory cycle. Biol Psychol. 1998;49:29–51.CrossRefPubMed Boiten FA. The effects of emotional behaviour on components of the respiratory cycle. Biol Psychol. 1998;49:29–51.CrossRefPubMed
33.
go back to reference Boiten FA, Frijda NH, Wientjes CJ. Emotions and respiratory patterns: review and critical analysis. Int J Psychophysiol. 1994;17:103–28.CrossRefPubMed Boiten FA, Frijda NH, Wientjes CJ. Emotions and respiratory patterns: review and critical analysis. Int J Psychophysiol. 1994;17:103–28.CrossRefPubMed
34.
go back to reference Anttonen J, Surakka V. Emotions and heart rate while sitting on a chair. In: Anttonen J, Surakka V, editors. Book emotions and heart rate while sitting on a chair. New York: ACM; 2005. p. 491–9. Anttonen J, Surakka V. Emotions and heart rate while sitting on a chair. In: Anttonen J, Surakka V, editors. Book emotions and heart rate while sitting on a chair. New York: ACM; 2005. p. 491–9.
35.
go back to reference Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. In: Peter AH, Najmedin M, editors. Advances in psychology. Amsterdam: North-Holland; 1988. p. 139–83. Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. In: Peter AH, Najmedin M, editors. Advances in psychology. Amsterdam: North-Holland; 1988. p. 139–83.
36.
go back to reference Hart S. Nasa-Task Load Index (Nasa-TLX); 20 years later. In: Hart S (ed) Book Nasa-Task Load Index (Nasa-TLX); 20 years later. City;2006. Hart S. Nasa-Task Load Index (Nasa-TLX); 20 years later. In: Hart S (ed) Book Nasa-Task Load Index (Nasa-TLX); 20 years later. City;2006.
37.
go back to reference McLellan S, Muddimer A, Peres SC. The effect of experience on System Usability Scale ratings. J Usability Stud. 2012;7:56–67. McLellan S, Muddimer A, Peres SC. The effect of experience on System Usability Scale ratings. J Usability Stud. 2012;7:56–67.
38.
go back to reference Martin JL, Norris BJ, Murphy E, Crowe JA. Medical device development: the challenge for ergonomics. Appl Ergon. 2008;39:271–83.CrossRefPubMed Martin JL, Norris BJ, Murphy E, Crowe JA. Medical device development: the challenge for ergonomics. Appl Ergon. 2008;39:271–83.CrossRefPubMed
39.
go back to reference Alessi SM. Fidelity in the design of instructional simulations. J Comput Based Instruct. 1988;15:40–7. Alessi SM. Fidelity in the design of instructional simulations. J Comput Based Instruct. 1988;15:40–7.
Metadata
Title
A new global and comprehensive model for ICU ventilator performances evaluation
Authors
Nicolas S. Marjanovic
Agathe De Simone
Guillaume Jegou
Erwan L’Her
Publication date
01-12-2017
Publisher
Springer International Publishing
Published in
Annals of Intensive Care / Issue 1/2017
Electronic ISSN: 2110-5820
DOI
https://doi.org/10.1186/s13613-017-0285-2

Other articles of this Issue 1/2017

Annals of Intensive Care 1/2017 Go to the issue