Skip to main content
Top
Published in: Critical Care 1/2016

Open Access 01-12-2016 | Research

The usability of ventilators: a comparative evaluation of use safety and user experience

Authors: Plinio P. Morita, Peter B. Weinstein, Christopher J. Flewwelling, Carleene A. Bañez, Tabitha A. Chiu, Mario Iannuzzi, Aastha H. Patel, Ashleigh P. Shier, Joseph A. Cafazzo

Published in: Critical Care | Issue 1/2016

Login to get access

Abstract

Background

The design complexity of critical care ventilators (CCVs) can lead to use errors and patient harm. In this study, we present the results of a comparison of four CCVs from market leaders, using a rigorous methodology for the evaluation of use safety and user experience of medical devices.

Methods

We carried out a comparative usability study of four CCVs: Hamilton G5, Puritan Bennett 980, Maquet SERVO-U, and Dräger Evita V500. Forty-eight critical care respiratory therapists participated in this fully counterbalanced, repeated measures study. Participants completed seven clinical scenarios composed of 16 tasks on each ventilator.
Use safety was measured by percentage of tasks with use errors or close calls (UE/CCs). User experience was measured by system usability and workload metrics, using the Post-Study System Usability Questionnaire (PSSUQ) and the National Aeronautics and Space Administration Task Load Index (NASA-TLX).

Results

Nine of 18 post hoc contrasts between pairs of ventilators were significant after Bonferroni correction, with effect sizes between 0.4 and 1.09 (Cohen’s d). There were significantly fewer UE/CCs with SERVO-U when compared to G5 (p = 0.044) and V500 (p = 0.020). Participants reported higher system usability for G5 when compared to PB980 (p = 0.035) and higher system usability for SERVO-U when compared to G5 (p < 0.001), PB980 (p < 0.001), and V500 (p < 0.001). Participants reported lower workload for G5 when compared to PB980 (p < 0.001) and lower workload for SERVO-U when compared to PB980 (p < 0.001) and V500 (p < 0.001). G5 scored better on two of nine possible comparisons; SERVO-U scored better on seven of nine possible comparisons. Aspects influencing participants’ performance and perception include the low sensitivity of G5’s touchscreen and the positive effect from the quality of SERVO-U’s user interface design.

Conclusions

This study provides empirical evidence of how four ventilators from market leaders compare and highlights the importance of medical technology design. Within the boundaries of this study, we can infer that SERVO-U demonstrated the highest levels of use safety and user experience, followed by G5. Based on qualitative data, differences in outcomes could be explained by interaction design, quality of hardware components used in manufacturing, and influence of consumer product technology on users’ expectations.
Appendix
Available only for authorised users
Literature
1.
go back to reference Needham DM, Bronskill SE, Calinawan JR, Sibbald WJ, Pronovost PJ, Laupacis A. Projected incidence of mechanical ventilation in Ontario to 2026: preparing for the aging baby boomers. Crit Care Med. 2005;33:574–9.CrossRefPubMed Needham DM, Bronskill SE, Calinawan JR, Sibbald WJ, Pronovost PJ, Laupacis A. Projected incidence of mechanical ventilation in Ontario to 2026: preparing for the aging baby boomers. Crit Care Med. 2005;33:574–9.CrossRefPubMed
2.
go back to reference Edwards JR, Peterson KD, Andrus ML, Tolson JS, Goulding JS, Dudeck MA, et al. National Healthcare Safety Network (NHSN) Report, data summary for 2006, issued June 2007. Am J Infect Control. 2007;35:290–301.CrossRefPubMed Edwards JR, Peterson KD, Andrus ML, Tolson JS, Goulding JS, Dudeck MA, et al. National Healthcare Safety Network (NHSN) Report, data summary for 2006, issued June 2007. Am J Infect Control. 2007;35:290–301.CrossRefPubMed
3.
go back to reference Wunsch H, Wagner J, Herlim M, Chong D, Kramer A, Halpern SD. ICU occupancy and mechanical ventilator use in the United States. Crit Care Med. 2013;41:2712–9.CrossRefPubMed Wunsch H, Wagner J, Herlim M, Chong D, Kramer A, Halpern SD. ICU occupancy and mechanical ventilator use in the United States. Crit Care Med. 2013;41:2712–9.CrossRefPubMed
4.
go back to reference Esteban A, Anzueto A, Alía I, Gordo F, Apezteguía C, Pálizas F, et al. How is mechanical ventilation employed in the intensive care unit? An international utilization review. Am J Respir Crit Care Med. 2000;161:1450–8.CrossRefPubMed Esteban A, Anzueto A, Alía I, Gordo F, Apezteguía C, Pálizas F, et al. How is mechanical ventilation employed in the intensive care unit? An international utilization review. Am J Respir Crit Care Med. 2000;161:1450–8.CrossRefPubMed
5.
go back to reference Metnitz PGH, Metnitz B, Moreno RP, Bauer P, Sorbo LD, Hoermann C, et al. Epidemiology of mechanical ventilation: analysis of the SAPS 3 database. Intensive Care Med. 2009;35:816–25.CrossRefPubMed Metnitz PGH, Metnitz B, Moreno RP, Bauer P, Sorbo LD, Hoermann C, et al. Epidemiology of mechanical ventilation: analysis of the SAPS 3 database. Intensive Care Med. 2009;35:816–25.CrossRefPubMed
8.
go back to reference Gilstrap D, Macintyre N. Patient ventilator interactions: implications for clinical management. Am J Respir Crit Care Med. 2013;188:1058–68.CrossRefPubMed Gilstrap D, Macintyre N. Patient ventilator interactions: implications for clinical management. Am J Respir Crit Care Med. 2013;188:1058–68.CrossRefPubMed
9.
go back to reference Chastre J, Fagon J-Y. Ventilator-associated pneumonia. Am J Respir Crit Care Med. 2002;165:867–903.CrossRefPubMed Chastre J, Fagon J-Y. Ventilator-associated pneumonia. Am J Respir Crit Care Med. 2002;165:867–903.CrossRefPubMed
10.
go back to reference Chatwin M, Heather S, Hanak A, Polkey MI, Simonds AK. Analysis of home support and ventilator malfunction in 1,211 ventilator-dependent patients. Eur Respir J. 2010;35:310–6.CrossRefPubMed Chatwin M, Heather S, Hanak A, Polkey MI, Simonds AK. Analysis of home support and ventilator malfunction in 1,211 ventilator-dependent patients. Eur Respir J. 2010;35:310–6.CrossRefPubMed
11.
go back to reference Gravenstein JS. How does human error affect safety in anesthesia? Surg Oncol Clin N Am. 2000;9:81–95.PubMed Gravenstein JS. How does human error affect safety in anesthesia? Surg Oncol Clin N Am. 2000;9:81–95.PubMed
12.
go back to reference Auriant I, Reignier J, Pibarot ML, Bachat S, Tenaillon A, Raphael JC. Critical incidents related to invasive mechanical ventilation in the ICU: preliminary descriptive study. Intensive Care Med. 2002;28:452–8.CrossRefPubMed Auriant I, Reignier J, Pibarot ML, Bachat S, Tenaillon A, Raphael JC. Critical incidents related to invasive mechanical ventilation in the ICU: preliminary descriptive study. Intensive Care Med. 2002;28:452–8.CrossRefPubMed
13.
15.
go back to reference Gajic O, Frutos-Vivar F, Esteban A, Hubmayr RD, Anzueto A. Ventilator settings as a risk factor for acute respiratory distress syndrome in mechanically ventilated patients. Intensive Care Med. 2005;31:922–6.CrossRefPubMed Gajic O, Frutos-Vivar F, Esteban A, Hubmayr RD, Anzueto A. Ventilator settings as a risk factor for acute respiratory distress syndrome in mechanically ventilated patients. Intensive Care Med. 2005;31:922–6.CrossRefPubMed
16.
go back to reference Donchin Y, Seagull FJ. The hostile environment of the intensive care unit. Curr Opin Crit Care. 2002;8:316–20.CrossRefPubMed Donchin Y, Seagull FJ. The hostile environment of the intensive care unit. Curr Opin Crit Care. 2002;8:316–20.CrossRefPubMed
17.
18.
go back to reference Wiklund ME, Kendler J, Strochlic AY. Usability testing of medical devices. Boca Raton: CRC Press; 2011. Wiklund ME, Kendler J, Strochlic AY. Usability testing of medical devices. Boca Raton: CRC Press; 2011.
19.
go back to reference Nielsen J. Usability engineering. Burlington: Morgan Kaufmann; 1993. Nielsen J. Usability engineering. Burlington: Morgan Kaufmann; 1993.
21.
go back to reference Ginsburg G. Human factors engineering: a tool for medical device evaluation in hospital procurement decision-making. J Biomed Inform. 2005;38:213–9.CrossRefPubMed Ginsburg G. Human factors engineering: a tool for medical device evaluation in hospital procurement decision-making. J Biomed Inform. 2005;38:213–9.CrossRefPubMed
22.
go back to reference Rubin J, Chisnell D. Handbook of usability testing: how to plan, design and conduct effective tests. Hoboken: Wiley; 2008. Rubin J, Chisnell D. Handbook of usability testing: how to plan, design and conduct effective tests. Hoboken: Wiley; 2008.
23.
go back to reference Templier F, Miroux P, Dolveck F, Descatha A, Goddet N-S, Jeleff C, et al. Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators. Respir Care. 2007;52:1701–9.PubMed Templier F, Miroux P, Dolveck F, Descatha A, Goddet N-S, Jeleff C, et al. Evaluation of the ventilator-user interface of 2 new advanced compact transport ventilators. Respir Care. 2007;52:1701–9.PubMed
24.
go back to reference McCrory B, Lowndes BR, LaGrange CA, Miller EE, Hallbeck MS. Comparative usability testing of conventional and single incision laparoscopic surgery devices. Hum Factors. 2012;55:619–31.CrossRef McCrory B, Lowndes BR, LaGrange CA, Miller EE, Hallbeck MS. Comparative usability testing of conventional and single incision laparoscopic surgery devices. Hum Factors. 2012;55:619–31.CrossRef
25.
go back to reference Diggory P, Fernandez C, Humphrey A, Jones V, Murphy M. Comparison of elderly people’s technique in using two dry powder inhalers to deliver zanamivir: randomised controlled trial. Br Med J. 2001;322:577–9.CrossRef Diggory P, Fernandez C, Humphrey A, Jones V, Murphy M. Comparison of elderly people’s technique in using two dry powder inhalers to deliver zanamivir: randomised controlled trial. Br Med J. 2001;322:577–9.CrossRef
26.
go back to reference Gonzalez-Bermejo J, Laplanche V, Husseini FE, Duguet A, Derenne J-P, Similowski T. Evaluation of the user-friendliness of 11 home mechanical ventilators. Eur Respir J. 2006;27:1236–43.CrossRefPubMed Gonzalez-Bermejo J, Laplanche V, Husseini FE, Duguet A, Derenne J-P, Similowski T. Evaluation of the user-friendliness of 11 home mechanical ventilators. Eur Respir J. 2006;27:1236–43.CrossRefPubMed
27.
go back to reference Vignaux L, Tassaux D, Jolliet P. Evaluation of the user-friendliness of seven new generation intensive care ventilators. Intensive Care Med. 2009;35:1687–91.CrossRefPubMed Vignaux L, Tassaux D, Jolliet P. Evaluation of the user-friendliness of seven new generation intensive care ventilators. Intensive Care Med. 2009;35:1687–91.CrossRefPubMed
29.
go back to reference Cassano-Piché A, Cafazzo JA, Chagpar A, Easty AC. How human factors methods can help in the procurement process. Biomed Instrum Technol. 2010;44:49–56.CrossRef Cassano-Piché A, Cafazzo JA, Chagpar A, Easty AC. How human factors methods can help in the procurement process. Biomed Instrum Technol. 2010;44:49–56.CrossRef
30.
go back to reference Kutner MH, Nachtsheim CJ, Neter J. Applied linear statistical models. New York: McGraw-Hill/Irwin; 2005. Kutner MH, Nachtsheim CJ, Neter J. Applied linear statistical models. New York: McGraw-Hill/Irwin; 2005.
31.
go back to reference Gravetter FJ, Wallnau LB. Statistics for the behavioral sciences. 7th ed. Belmont: Wadsworth Publishing; 2006. Gravetter FJ, Wallnau LB. Statistics for the behavioral sciences. 7th ed. Belmont: Wadsworth Publishing; 2006.
32.
go back to reference van Teijlingen ER, Hundley V. The importance of pilot studies. Soc Res Update. 2001;35:1–4. van Teijlingen ER, Hundley V. The importance of pilot studies. Soc Res Update. 2001;35:1–4.
33.
go back to reference Cohen J. Statistical power analysis for the behavioral sciences. London: Routledge; 1988. Cohen J. Statistical power analysis for the behavioral sciences. London: Routledge; 1988.
37.
go back to reference Richard JCM, Richard M, Kacmarek RM. ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers. Intensive Care Med. 2009;35:1662–3.CrossRefPubMed Richard JCM, Richard M, Kacmarek RM. ICU mechanical ventilators, technological advances vs. user friendliness: the right picture is worth a thousand numbers. Intensive Care Med. 2009;35:1662–3.CrossRefPubMed
38.
go back to reference Kamouri AL, Kamouri J, Smith KH. Training by exploration: facilitating the transfer of procedural knowledge through analogical reasoning. Int J Man Mach Stud. 1986;24:171–92.CrossRef Kamouri AL, Kamouri J, Smith KH. Training by exploration: facilitating the transfer of procedural knowledge through analogical reasoning. Int J Man Mach Stud. 1986;24:171–92.CrossRef
39.
go back to reference Carayon P, Wetterneck TB, Hundt AS, Ozkaynak M, Ram P, DeSilvey J, et al. Observing nurse interaction with infusion pump technologies. Adv Pat Saf. 2005;2:349–64. Carayon P, Wetterneck TB, Hundt AS, Ozkaynak M, Ram P, DeSilvey J, et al. Observing nurse interaction with infusion pump technologies. Adv Pat Saf. 2005;2:349–64.
40.
go back to reference Zhang J, Johnson TR, Patel VL, Paige DL, Kubose T. Using usability heuristics to evaluate patient safety of medical devices. J Biomed Inform. 2003;36:23–30.CrossRefPubMed Zhang J, Johnson TR, Patel VL, Paige DL, Kubose T. Using usability heuristics to evaluate patient safety of medical devices. J Biomed Inform. 2003;36:23–30.CrossRefPubMed
41.
go back to reference Middleton B, Bloomrosen M, Dente MA, Hashmat B, Koppel R, Overhage JM, et al. Enhancing patient safety and quality of care by improving the usability of electronic health record systems: recommendations from AMIA. J Am Med Informatics Assoc. 2013;20:e2–8.CrossRef Middleton B, Bloomrosen M, Dente MA, Hashmat B, Koppel R, Overhage JM, et al. Enhancing patient safety and quality of care by improving the usability of electronic health record systems: recommendations from AMIA. J Am Med Informatics Assoc. 2013;20:e2–8.CrossRef
42.
43.
go back to reference Morita PP, Cafazzo JA. Challenges and paradoxes of human factors in health technology design. JMIR J Hum Factors. 2015;3:1–7. Morita PP, Cafazzo JA. Challenges and paradoxes of human factors in health technology design. JMIR J Hum Factors. 2015;3:1–7.
44.
go back to reference Nielsen J, Levy J. Measuring usability: preference vs. performance. Commun ACM. 1994;37:66–75.CrossRef Nielsen J, Levy J. Measuring usability: preference vs. performance. Commun ACM. 1994;37:66–75.CrossRef
45.
go back to reference Lewis JR. Psychometric evaluation of the PSSUQ using data from five years of usability studies. Int J Hum Comput Interact. 2002;14:463–88.CrossRef Lewis JR. Psychometric evaluation of the PSSUQ using data from five years of usability studies. Int J Hum Comput Interact. 2002;14:463–88.CrossRef
46.
go back to reference Dosani M, Hunc K, Dumont GA, Dunsmuir D, Barralon P, Schwarz SKW, et al. A vibro-tactile display for clinical monitoring. Anesth Analg. 2012;115:588–94.PubMed Dosani M, Hunc K, Dumont GA, Dunsmuir D, Barralon P, Schwarz SKW, et al. A vibro-tactile display for clinical monitoring. Anesth Analg. 2012;115:588–94.PubMed
47.
go back to reference Görges M, Winton P, Koval V, Lim J, Stinson J, Choi PT, et al. An evaluation of an expert system for detecting critical events during anesthesia in a human patient simulator. Anesth Analg. 2013;117:380–91.CrossRefPubMed Görges M, Winton P, Koval V, Lim J, Stinson J, Choi PT, et al. An evaluation of an expert system for detecting critical events during anesthesia in a human patient simulator. Anesth Analg. 2013;117:380–91.CrossRefPubMed
48.
go back to reference Schutte JL, Gales S, Filipponi A, Saptono A, Parmanto B, McCue M. Evaluation of a telerehabilitation system for community-based rehabilitation system. Int J Telerehabilitation. 2012;4:15–24.CrossRef Schutte JL, Gales S, Filipponi A, Saptono A, Parmanto B, McCue M. Evaluation of a telerehabilitation system for community-based rehabilitation system. Int J Telerehabilitation. 2012;4:15–24.CrossRef
49.
go back to reference Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol. 1988;1:139–83.CrossRef Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol. 1988;1:139–83.CrossRef
50.
go back to reference Hart SG. Nasa-Task Load Index (NASA-TLX); 20 years later. Proc Hum Factors Ergon Soc Annu Meet. 2006;50:904–8. SAGE Publications. Hart SG. Nasa-Task Load Index (NASA-TLX); 20 years later. Proc Hum Factors Ergon Soc Annu Meet. 2006;50:904–8. SAGE Publications.
51.
go back to reference Wachter SB, Johnson K, Albert R, Syroid N, Drews F, Westenskow D. The evaluation of a pulmonary display to detect adverse respiratory events using high resolution human simulator. J Am Med Inform Assoc. 2006;13:635–42.CrossRefPubMedPubMedCentral Wachter SB, Johnson K, Albert R, Syroid N, Drews F, Westenskow D. The evaluation of a pulmonary display to detect adverse respiratory events using high resolution human simulator. J Am Med Inform Assoc. 2006;13:635–42.CrossRefPubMedPubMedCentral
52.
go back to reference Görges M, Staggers N. Evaluations of physiological monitoring displays: a systematic review. J Clin Monit Comput. 2008;22:45–66.CrossRefPubMed Görges M, Staggers N. Evaluations of physiological monitoring displays: a systematic review. J Clin Monit Comput. 2008;22:45–66.CrossRefPubMed
54.
go back to reference Sankoh AJ, Huque MF, Dubey SD. Some comments on frequently used multiple endpoint adjustment methods in clinical trials. Stat Med. 1997;16:2529–42.CrossRefPubMed Sankoh AJ, Huque MF, Dubey SD. Some comments on frequently used multiple endpoint adjustment methods in clinical trials. Stat Med. 1997;16:2529–42.CrossRefPubMed
55.
go back to reference Fairbanks RJ, Caplan S. Poor interface design and lack of usability testing facilitate medical error. Jt Comm J Qual Saf. 2004;30:579–84.CrossRefPubMed Fairbanks RJ, Caplan S. Poor interface design and lack of usability testing facilitate medical error. Jt Comm J Qual Saf. 2004;30:579–84.CrossRefPubMed
56.
go back to reference Hodge E, Blunt M, Young P. Capnography activation is improved by better ventilator interface ergonomics. Crit Care. 2011;15:P139.CrossRefPubMedCentral Hodge E, Blunt M, Young P. Capnography activation is improved by better ventilator interface ergonomics. Crit Care. 2011;15:P139.CrossRefPubMedCentral
58.
go back to reference Pickering BW, Litell JM, Herasevich V, Gajic O. Clinical review: the hospital of the future - building intelligent environments to facilitate safe and effective acute care delivery. Crit Care. 2012;16:220.CrossRefPubMedPubMedCentral Pickering BW, Litell JM, Herasevich V, Gajic O. Clinical review: the hospital of the future - building intelligent environments to facilitate safe and effective acute care delivery. Crit Care. 2012;16:220.CrossRefPubMedPubMedCentral
59.
go back to reference Colpaert K, Claus B, Somers A, Vandewoude K, Robays H, Decruyenaere J. Impact of computerized physician order entry on medication prescription errors in the intensive care unit: a controlled cross-sectional trial. Crit Care. 2006;10:R21.CrossRefPubMedPubMedCentral Colpaert K, Claus B, Somers A, Vandewoude K, Robays H, Decruyenaere J. Impact of computerized physician order entry on medication prescription errors in the intensive care unit: a controlled cross-sectional trial. Crit Care. 2006;10:R21.CrossRefPubMedPubMedCentral
60.
go back to reference Lin L, Vicente K, Doyle D. Patient safety, potential adverse drug events, and medical device design: a human factors engineering approach. J Biomed Inform. 2001;34:274–84.CrossRefPubMed Lin L, Vicente K, Doyle D. Patient safety, potential adverse drug events, and medical device design: a human factors engineering approach. J Biomed Inform. 2001;34:274–84.CrossRefPubMed
61.
go back to reference Graham TAD, Kushniruk AW, Bullard MJ, Holroyd BR, Meurer DP, Rowe BH. How usability of a web-based clinical decision support system has the potential to contribute to adverse medical events. AMIA Annu Symp Proc. 2008;257–61. Graham TAD, Kushniruk AW, Bullard MJ, Holroyd BR, Meurer DP, Rowe BH. How usability of a web-based clinical decision support system has the potential to contribute to adverse medical events. AMIA Annu Symp Proc. 2008;257–61.
62.
go back to reference Stadlbauer V, Wallner S, Sojakovic T, Smolle KH. Comparison of three different multi-analyte point-of-care devices during clinical routine on a medical ICU. Crit Care. 2011;433:e1–11. Stadlbauer V, Wallner S, Sojakovic T, Smolle KH. Comparison of three different multi-analyte point-of-care devices during clinical routine on a medical ICU. Crit Care. 2011;433:e1–11.
64.
go back to reference Hugonnet S, Uckay I, Pittet D. Staffing level: a determinant of late-onset ventilator-associated pneumonia. Crit Care. 2007;11:1–7.CrossRef Hugonnet S, Uckay I, Pittet D. Staffing level: a determinant of late-onset ventilator-associated pneumonia. Crit Care. 2007;11:1–7.CrossRef
65.
go back to reference Sevdalis N, Brett SJ. Improving care by understanding the way we work: human factors and behavioural science in the context of intensive care. Crit Care. 2009;13:139.CrossRefPubMedPubMedCentral Sevdalis N, Brett SJ. Improving care by understanding the way we work: human factors and behavioural science in the context of intensive care. Crit Care. 2009;13:139.CrossRefPubMedPubMedCentral
Metadata
Title
The usability of ventilators: a comparative evaluation of use safety and user experience
Authors
Plinio P. Morita
Peter B. Weinstein
Christopher J. Flewwelling
Carleene A. Bañez
Tabitha A. Chiu
Mario Iannuzzi
Aastha H. Patel
Ashleigh P. Shier
Joseph A. Cafazzo
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2016
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-016-1431-1