Published in: Journal of Medical Case Reports 1/2019

Open Access 01-12-2019 | Pacemaker | Case report

Unusual conservative treatment of a complicated pacemaker pocket infection: a case report

Authors: Wanqiu Kang, Xiaoming Chen, Zicheng Li, Aidong Zhang, Jingwen Liu, Liqiong Yu, Yingzhen Wen

Published in: Journal of Medical Case Reports | Issue 1/2019

Login to get access

Abstract

Background

For patients with complicated generator pocket infection, expert consensuses universally advocate complete device and leads removal followed by delayed replacement on the contralateral side. We cured our patient by partial generator removal and reimplantation of sterilized pulse generator on the ipsilateral side. We also performed a literature review about incomplete removal therapy for the management of cardiac implantable electronic device infection.

Case presentation

An 86-year-old Chinese Han man was diagnosed as having third-degree atrioventricular block and received a permanent double-chamber pacemaker in his left prepectoral area 15 years ago. Nine years later, the entire system was removed because of confirmed infection, and a new device was reimplanted in the contralateral area. He developed skin necrosis around the pacemaker pocket after 1 year, and his generator was renewed without leads extraction at another medical center. He was subsequently admitted several times for surgical tissue debridement at another institution due to extended skin necrosis. At the time of the new admission, he had severe infection, heart failure, and hypoalbuminemia. He was diagnosed as having complicated pacemaker pocket infection. Our preferred treatment strategy was for complete removal of both the generator and transvenous pacing leads, and we intended to implant an epicardial pacemaker in our patient if necessary. However, he rejected the treatment strategy and firmly refused to replace his generator. We had to attempt a novel pacemaker-preserving strategy considering our patient’s severe comorbidities. Finally, we cured him by partial generator removal and reimplantation of the sterilized pulse generator on the ipsilateral side. There was no sign of wound dehiscence or infection during a 6-month follow-up.

Conclusions

We would posit that partial removal of infected generators combined with conservative treatment may be a proper treatment of complicated generator pocket infection, especially for those who are susceptible to cardiac complications. Reimplantation of a sterilized pulse generator on the ipsilateral side may be an option if patients reject a new device and contralateral vascular condition is not really suitable. Opting for such treatment should be at the consideration of the primary care physician based on the condition of the patient.
Literature
1.
go back to reference Kurtz SM, Ochoa JA, Lau E, et al. Implantation trends and patient profiles for pacemakers and implantable cardioverter defibrillators in the United States:1993-2006. Pacing Clin Electrophysiol. 2010;33(6):705–11.CrossRef Kurtz SM, Ochoa JA, Lau E, et al. Implantation trends and patient profiles for pacemakers and implantable cardioverter defibrillators in the United States:1993-2006. Pacing Clin Electrophysiol. 2010;33(6):705–11.CrossRef
2.
go back to reference Uslan DZ, Tleyjeh IM, Baddour LM, et al. Temporal trends in permanent pacemaker implantation: a population-based study. Am Heart J. 2008;155(5):896–903.CrossRef Uslan DZ, Tleyjeh IM, Baddour LM, et al. Temporal trends in permanent pacemaker implantation: a population-based study. Am Heart J. 2008;155(5):896–903.CrossRef
3.
go back to reference Sridhar AR, Lavu M, Yarlagadda V, et al. Cardiac implantable electronic device-related infection and extraction trends in the U.S. Pacing Clin Electrophysiol. 2017;40(3):286–93.CrossRef Sridhar AR, Lavu M, Yarlagadda V, et al. Cardiac implantable electronic device-related infection and extraction trends in the U.S. Pacing Clin Electrophysiol. 2017;40(3):286–93.CrossRef
4.
go back to reference Sandoe JA, Barlow G, Chambers JB, et al. Guidelines for the diagnosis, prevention and management of implantable cardiac electronic device infection. Report of a joint working party project on behalf of the British Society for Antimicrobial Chemotherapy (BSAC, host organization), British Heart Rhythm Society (BHRS), British Cardiovascular Society (BCS), British Heart Valve Society (BHVS) and British Society for Echocardiography (BSE). J Antimicrob Chemother. 2015;70(2):334–59.CrossRef Sandoe JA, Barlow G, Chambers JB, et al. Guidelines for the diagnosis, prevention and management of implantable cardiac electronic device infection. Report of a joint working party project on behalf of the British Society for Antimicrobial Chemotherapy (BSAC, host organization), British Heart Rhythm Society (BHRS), British Cardiovascular Society (BCS), British Heart Valve Society (BHVS) and British Society for Echocardiography (BSE). J Antimicrob Chemother. 2015;70(2):334–59.CrossRef
5.
go back to reference Baddour LM, Epstein AE, Erickson CC, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121(3):458–77.CrossRef Baddour LM, Epstein AE, Erickson CC, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121(3):458–77.CrossRef
6.
go back to reference Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm. 2009;6(7):1085–104.CrossRef Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm. 2009;6(7):1085–104.CrossRef
7.
go back to reference Kusumoto FM, Schoenfeld MH, Wilkoff BL, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503–51.CrossRef Kusumoto FM, Schoenfeld MH, Wilkoff BL, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503–51.CrossRef
8.
go back to reference Le KY, Sohail MR, Friedman PA, et al. Impact of timing of device removal on mortality in patients with cardiovascular implantable electronic device infections. Heart Rhythm. 2011;8(11):1678–85.CrossRef Le KY, Sohail MR, Friedman PA, et al. Impact of timing of device removal on mortality in patients with cardiovascular implantable electronic device infections. Heart Rhythm. 2011;8(11):1678–85.CrossRef
9.
go back to reference Wazni O, Epstein LM, Carrillo RG, et al. Lead extraction in the contemporary setting: the LExICon study: an observational retrospective study of consecutive laser lead extractions. J Am Coll Cardiol. 2010;55(6):579–86.CrossRef Wazni O, Epstein LM, Carrillo RG, et al. Lead extraction in the contemporary setting: the LExICon study: an observational retrospective study of consecutive laser lead extractions. J Am Coll Cardiol. 2010;55(6):579–86.CrossRef
10.
go back to reference Brunner MP, Yu C, Hussein AA, et al. Nomogram for predicting 30-day all-cause mortality after transvenous pacemaker and defibrillator lead extraction. Heart Rhythm. 2015;12(12):2381–6.CrossRef Brunner MP, Yu C, Hussein AA, et al. Nomogram for predicting 30-day all-cause mortality after transvenous pacemaker and defibrillator lead extraction. Heart Rhythm. 2015;12(12):2381–6.CrossRef
11.
go back to reference Lopez JA. Conservative management of infected pacemaker and implantable defibrillator sites with a closed antimicrobial irrigation system. Europace. 2013;15(4):541–5.CrossRef Lopez JA. Conservative management of infected pacemaker and implantable defibrillator sites with a closed antimicrobial irrigation system. Europace. 2013;15(4):541–5.CrossRef
12.
go back to reference Tan EM, Desimone DC, Sohail MR, et al. Outcomes in patients with cardiovascular implantable electronic device infection managed with chronic antibiotic suppression. Clin Infect Dis. 2017;64(11):1516–21.CrossRef Tan EM, Desimone DC, Sohail MR, et al. Outcomes in patients with cardiovascular implantable electronic device infection managed with chronic antibiotic suppression. Clin Infect Dis. 2017;64(11):1516–21.CrossRef
13.
go back to reference Peacock JE Jr, Stafford JM, Le K, et al. Attempted salvage of infected cardiovascular implantable electronic devices: Are there clinical factors that predict success? Pacing Clin Electrophysiol. 2018;41(5):524–31.CrossRef Peacock JE Jr, Stafford JM, Le K, et al. Attempted salvage of infected cardiovascular implantable electronic devices: Are there clinical factors that predict success? Pacing Clin Electrophysiol. 2018;41(5):524–31.CrossRef
14.
go back to reference Tascini C, Bongiorni MG, Di Cori A, et al. Cardiovascular implantable electronic device endocarditis treated with daptomycin with or without transvenous removal. Heart Lung. 2012;41(6):e24–30.CrossRef Tascini C, Bongiorni MG, Di Cori A, et al. Cardiovascular implantable electronic device endocarditis treated with daptomycin with or without transvenous removal. Heart Lung. 2012;41(6):e24–30.CrossRef
15.
go back to reference Margey R, McCann H, Blake G, et al. Contemporary management of and outcomes from cardiac device related infections. Europace. 2010;12(1):64–70.CrossRef Margey R, McCann H, Blake G, et al. Contemporary management of and outcomes from cardiac device related infections. Europace. 2010;12(1):64–70.CrossRef
16.
go back to reference Pichlmaier M, Knigina L, Kutschka I, et al. Complete removal as a routine treatment for any cardiovascular implantable electronic device-associated infection. J Thorac Cardiovasc Surg. 2011;142(6):1482–90.CrossRef Pichlmaier M, Knigina L, Kutschka I, et al. Complete removal as a routine treatment for any cardiovascular implantable electronic device-associated infection. J Thorac Cardiovasc Surg. 2011;142(6):1482–90.CrossRef
17.
go back to reference Goya M, Nagashima M, Hiroshima K, et al. Lead extractions in patients with cardiac implantable electronic device infections: Single center experience. J Arrhythm. 2016;32(4):308–12.CrossRef Goya M, Nagashima M, Hiroshima K, et al. Lead extractions in patients with cardiac implantable electronic device infections: Single center experience. J Arrhythm. 2016;32(4):308–12.CrossRef
18.
go back to reference Bongiorni MG, Tascini C, Tagliaferri E, et al. Microbiology of cardiac implantable electronic device infections. Europace. 2012;14(9):1334–9.CrossRef Bongiorni MG, Tascini C, Tagliaferri E, et al. Microbiology of cardiac implantable electronic device infections. Europace. 2012;14(9):1334–9.CrossRef
19.
go back to reference Golzio PG, Vinci M, Anselmino M, et al. Accuracy of swabs, tissue specimens, and lead samples in diagnosis of cardiac rhythm management device infections. Pacing Clin Electrophysiol. 2009;32(suppl 1):S76–80.CrossRef Golzio PG, Vinci M, Anselmino M, et al. Accuracy of swabs, tissue specimens, and lead samples in diagnosis of cardiac rhythm management device infections. Pacing Clin Electrophysiol. 2009;32(suppl 1):S76–80.CrossRef
20.
go back to reference Fukunaga M, Goya M, Nagashima M, et al. Identification of causative organism in cardiac implantable electronic device infections. J Cardiol. 2017;70(5):411–5.CrossRef Fukunaga M, Goya M, Nagashima M, et al. Identification of causative organism in cardiac implantable electronic device infections. J Cardiol. 2017;70(5):411–5.CrossRef
21.
go back to reference Deharo JC, Quatre A, Mancini J, et al. Long-term outcomes following infection of cardiac implantable electronic devices: a prospective matched cohort study. Heart. 2012;98(9):724–31.CrossRef Deharo JC, Quatre A, Mancini J, et al. Long-term outcomes following infection of cardiac implantable electronic devices: a prospective matched cohort study. Heart. 2012;98(9):724–31.CrossRef
22.
go back to reference Klug D, Balde M, Pavin D, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007;116(12):1349–55.CrossRef Klug D, Balde M, Pavin D, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007;116(12):1349–55.CrossRef
Metadata
Title
Unusual conservative treatment of a complicated pacemaker pocket infection: a case report
Authors
Wanqiu Kang
Xiaoming Chen
Zicheng Li
Aidong Zhang
Jingwen Liu
Liqiong Yu
Yingzhen Wen
Publication date
01-12-2019
Publisher
BioMed Central
Published in
Journal of Medical Case Reports / Issue 1/2019
Electronic ISSN: 1752-1947
DOI
https://doi.org/10.1186/s13256-019-1987-x

Other articles of this Issue 1/2019

Journal of Medical Case Reports 1/2019 Go to the issue